Abstract

Doc number: 77

Abstract

Background: Implantation of implantable cardioverter-defibrillators (ICD) from the left pectoral region is the standard therapeutical method. Increasing numbers of system revisions due to lead dysfunction and infections will consecutively increase the numbers of right-sided implantations. The reliability of devices implanted on the right pectoral side remains controversially discussed, and the question of testing these devices remains unanswered.

Methods: In a prospectively designed study all 870 patients (60.0±14 years, 689 male) who were treated with a first ICD from July 2005 until May 2012 and tested intraoperatively according to the testing protocol were analyzed. The indication for implantation was primary prophylactic in 71.5%. Underlying diseases included ischemic cardiomyopathy (50%), dilative cardiomyopathy (37%), and others (13%). Mean ejection faction was 27±12%. Implantation site was right in 4.5% and left in 95.5%.

Results: Five patients supplied with right-sided ICD (13%, p = 0.02 as compared to left-sided) failed initial intraoperative testing with 21 J. 3 patients were male. The age of the patients failing intraoperative testing with right-sided devices appeared higher than of patients with left-sided devices (p = 0.07). The ejection fraction was 28±8%. All patients reached a sufficient DFT ≤ 21 J after corrective procedures.

Conclusion: Implantation of ICDs on the right side results in significantly higher failure rate of successful termination of intraoperatively induced ventricular fibrillation. The data of our study suggest the necessity of intraoperative ICD testing in right-sided implanted ICDs.

Details

Title
Significance of intraoperative testing in right-sided implantable cardioverter-defibrillators
Author
Keyser, Andreas; Hilker, Michael K; Ücer, Ekrem; Wittmann, Sigrid; Schmid, Christof; Diez, Claudius
Pages
77
Publication year
2013
Publication date
2013
Publisher
BioMed Central
e-ISSN
1749-8090
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1347013787
Copyright
© 2013 Keyser et al.; licensee BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.