Abstract
Doc number: 37
Abstract
Background: The aim of this study is to determine the test-retest reliability of the measurement of regional myocardial function by cardiovascular magnetic resonance (CMR) tagging using spatial modulation of magnetization.
Methods: Twenty-five participants underwent CMR tagging twice over 12 ± 7 days. To assess the role of slice orientation on strain measurement, two healthy volunteers had a first exam, followed by image acquisition repeated with slices rotated ±15 degrees out of true short axis, followed by a second exam in the true short axis plane. To assess the role of slice location, two healthy volunteers had whole heart tagging. The harmonic phase (HARP) method was used to analyze the tagged images. Peak midwall circumferential strain (Ecc), radial strain (Err), Lambda 1, Lambda 2, and Angle α were determined in basal, mid and apical slices. LV torsion, systolic and early diastolic circumferential strain and torsion rates were also determined.
Results: LV Ecc and torsion had excellent intra-, interobserver, and inter-study intra-class correlation coefficients (ICC range, 0.7 to 0.9). Err, Lambda 1, Lambda 2 and angle had excellent intra- and interobserver ICC than inter-study ICC. Angle had least inter-study reproducibility. Torsion rates had superior intra-, interobserver, and inter-study reproducibility to strain rates. The measurements of LV Ecc were comparable in all three slices with different short axis orientations (standard deviation of mean Ecc was 0.09, 0.18 and 0.16 at basal, mid and apical slices, respectively). The mean difference in LV Ecc between slices was more pronounced in most of the basal slices compared to the rest of the heart.
Conclusions: Intraobserver and interobserver reproducibility of all strain and torsion parameters was excellent. Inter-study reproducibility of CMR tagging by SPAMM varied between different parameters as described in the results above and was superior for Ecc and LV torsion. The variation in LV Ecc measurement due to altered slice orientation is negligible compared to the variation due to slice location.
Trial registration: This trial is registered as NCT00005487 at National Heart, Lung and Blood institute.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer