ARTICLE
Received 11 Feb 2014 | Accepted 6 Mar 2014 | Published 7 Apr 2014
Yusuke Yoshioka1,2,3, Nobuyoshi Kosaka1, Yuki Konishi1,4, Hideki Ohta5, Hiroyuki Okamoto5, Hikaru Sonoda5, Ryoji Nonaka6, Hirofumi Yamamoto6, Hideshi Ishii7, Masaki Mori6, Koh Furuta8, Takeshi Nakajima9,Hiroshi Hayashi4, Hajime Sugisaki4, Hiroko Higashimoto4, Takashi Kato2, Fumitaka Takeshita1 & Takahiro Ochiya1
Cancer cells secrete small membranous extracellular vesicles (EVs) into their micro-environment and circulation. Although their potential as cancer biomarkers has been promising, the identication and quantication of EVs in clinical samples remains challenging. Here we describe a sensitive and rapid analytical technique for proling circulating EVs directly from blood samples of patients with colorectal cancer. EVs are captured by two types of antibodies and are detected by photosensitizer-beads, which enables us to detect cancer-derived EVs without a purication step. We also show that circulating EVs can be used for detection of colorectal cancer using the antigen CD147, which is embedded in cancer-linked EVs. This work describes a new liquid biopsy technique to sensitively detect disease-specic circulating EVs and provides perspectives in translational medicine from the standpoint of diagnosis and therapy.
1 Division of Molecular and Cellular Medicine, National Cancer Center Research Institute, Chuo-ku, Tokyo 104-0045, Japan. 2 Integrative Bioscience and Biomedical Engineering, Graduate School of Advanced Science and Engineering, Waseda University, Shinjuku, Tokyo 162-8480, Japan. 3 Research Fellow of the Japan Society for the Promotion of Science (JSPS), Chiyoda-Ku, Tokyo 102-0083, Japan. 4 R&D Department, SRL Inc., Hino-shi, Tokyo 191-0002, Japan.
5 Diagnostics Division, Shionogi & Co., LTD., Settsu-shi, Osaka 566-0022, Japan. 6 Department of Gastroenterological Surgery, Graduated School of Medicine, Osaka University, Suita, Osaka 565-0871, Japan. 7 Department of Frontier Science for Cancer and Chemotherapy, Osaka University, Graduate School of Medicine, Suita, Osaka 565-0871, Japan. 8 Division of Clinical Laboratories, National Cancer Center Hospital, Chuo-ku Tokyo, 104-0045, Japan. 9 Endoscopy Division, National Cancer Center Hospital, Chuo-ku, Tokyo 104-0045, Japan. Correspondence and requests for materials should be addressed to T.O. (email: mailto:[email protected]
Web End [email protected] ).
NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 1
& 2014 Macmillan Publishers Limited. All rights reserved.
DOI: 10.1038/ncomms4591 OPEN
Ultra-sensitive liquid biopsy of circulating extracellular vesicles using ExoScreen
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4591
Cancer cells secrete various types of humoral factors into their microenvironment that are biomarkers for disease diagnosis and prognosis, including cytokines, chemokines
and nucleic acids. Extracellular vesicles (EVs), including exosome and microvesicles from cancer cells, have also been found in the blood of cancer patients17 and therefore provide a novel type of biomarker for various patient scenarios.
EVs are small membranous vesicles that differ in their cellular origin, abundance and biogenesis8, and are naturally secreted by almost all cell types to transport bioactive molecules intercellularly. EVs are positive for tetraspanin family proteins, such as CD63, CD81 and CD9 (refs 911), and contain cell surface proteins as well as both mRNA and microRNA12. Conventional methods of analyzing EVs generally require large quantities of EVs to be concentrated and processed via time-consuming immunoblotting or enzyme-linked immunosorbent assay (ELISA) assays; these methods are impractical in most clinical settings. In this study, we establish a highly sensitive and rapid analytical technique for proling surface proteins in EVs
from patient blood that can be used to identify biomarkers of colorectal cancer, named ExoScreen. ExoScreen could monitor circulating EVs in serum without the need for purication step. In addition, we show that ExoScreen is superior for the detection of EVs to conventional methods, immunoblotting and ELISA. Furthermore, we nd that ExoScreen enables to detect CD147 and CD9 double-positive EVs, which is abundantly secreted from colorectal cancer cells, in serum from colorectal cancer patients. Our results demonstrate that ExoScreen can be a tool for detection of EVs from as little as 5 ml of cancer patients serum to detect circulating cancer-derived EVs.
ResultsEstablishment of ExoScreen to detect EVs in serum. To realize the usage of EVs in clinical situation, we develop methods that specically detect circulating EVs in the serum based on an amplied luminescent proximity homogeneous assay using photosensitizer-beads13 without a purication step of EVs
Clinical samples
Blood (serum)
ExoScreen method
Total time: 1.5 h Total time: ~12 h
Sample number 68
Sample(for example, serum)
Conventional method
1O2
Extracellular vesicle
Donor beads (~100 nm [p10]) Acceptor beads
Purification (Ultracentrifugation: 3 h)
Purified sample
Sample preparation (Quantification: 1 h)
Sample analysis: 58 h
Immunoblotting
Data analysis
Step 1
Step 2
Step 3
Step 4
+
+
Sample 5 l(for example, serum)
Biotinylated antibodies
Acceptor bead antibodies
Donor beads
Step 1 Add samples and antibodies
+
High throughput: > 96 samples (96-well plate)
ELISA
Step 2
Data analysis
Figure 1 | Schematic overview depicting the method for detecting circulating EVs via conventional methods and ExoScreen. In the case of conventional methods, nearly 12 h are needed to detect the expression of certain protein in circulating EVs. In addition, excessive volumes of serum are required. Conversely, ExoScreen is completed within 2 h and requires only 5 ml of serum. In this system, streptavidin-coated donor beads capture an analyte-specic biotinylated antibody and are used in conjunction with acceptor beads conjugated to a second antibody. The streptavidin-coated donor beads are excited with a laser at 680 nm, resulting in the release of singlet oxygen, which excites an amplied uorescent signal in the acceptor bead that emits at 615 nm when the beads are within 200 nm of the captured analyte.
2 NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4591 ARTICLE
Heathy donor (n=3)
Particle size (nm)
CD9/CD9 ExoScreen
Colorectal cancer (n=3) CD63/CD63
97 nm
con
/ml
E6
con
/ml
E6
95 nm 7.06.05.04.03.02.01.0 0
0 50 100 150 200 250
100
80
60
40
20
0
Signal intensity (104)
300
Extracelluar
vesicles CD63
recombinant
protein
(ng)
0 100 200 300 400 500 0 100 200 300 400 500
Particle size (nm)
CD63/CD63
ExoScreen
CD63/CD9 ExoScreen
Signal intensity (104)
100 50
40
30
20
10
0 0 50 100
80
60
40
20
R2 = 0.974 R2 = 0.999 R2 = 0.997
bCD9 only aCD9 only bCD9/aCD9
bCD63 only
aCD63 only
bCD63/aCD63
Signal intensity (104)
Signal intensity (104)
300 150 200 250 300Extracellular vesicles (ng) Extracellular vesicles (ng) Extracellular vesicles (ng)
0 50 100 150 200 250 300
0 0 50 100 150 200 250
20
15
10
5
0 No treatment
Triton0.05%
CD9/CD9 CD9/CD9
Signal intensity (104)
Signal intensity (104)
20
16
12
8
4
Triton0.5%
0 Prok() Prok(+)
Figure 2 | Establishment of ExoScreen to detect the EVs. (a) Analysis of the size distribution in the serum of healthy donors (n 3) and colorectal
cancer patients (n 3) by the NanoSight nanoparticle tracking system. (b) Detection of EVs or monomeric recombinant CD63 protein by ExoScreen using
a CD63 antibody. EV protein concentration was measured via the Qubit system. The concentration of recombinant CD63 was adjusted with that of protein in EVs puried from HCT116 CM. Error bars are s.e.m. (n 3 for each condition). (c) Correlation between ExoScreen measurements for CD9 positive EVs,
CD63 positive EVs or CD63/CD9 double-positive EVs and EV protein concentration in a dilution series. EV protein concentration was measured via the Qubit system. EVs were puried from HCT116 cell CM. The addition of biotinylated CD9 or CD63 antibodies without acceptor beads conjugated to antibodies is denoted bCD9 only or bCD63 only, while aCD9 only or aCD63 only means addition of only acceptor beads conjugated to CD9 or CD63 antibodies without biotinylated antibodies. The addition of biotinylated antibodies and acceptor beads conjugated antibodies is denoted bCD9/aCD9 or bCD63/aCD63. Right panel shows the addition of biotinylated CD63 antibodies and acceptor beads conjugated CD9 antibodies. Error bars are s.e.m. (n 3 for each condition). (d) Evaluation of ExoScreen specicity against puried EVs from HCT116 cell treated with or without 0.05% and 0.5% Triton
X-100. Two hundred fty ng of EVs were detected by ExoScreen using CD9 antibodies. Error bars are s.e.m. (n 3 for each condition). (e) Evaluation
of ExoScreen specicity against EVs from HCT116 cells treated with (Prok( )) or without (Prok( )) Proteinase K. Two hundred fty ng of EVs were
detected by ExoScreen using CD9 antibodies. Error bars are s.e.m. (n 3 for each condition). Data are representative of at least three independent
experiments each.
(Fig. 1). This system utilizes streptavidin-coated donor beads to capture an analyte-specic biotinylated antibody, and acceptor beads conjugated to a second antibody that recognizes an epitope of the analyte. The donor beads are excited with a laser at 680 nm, resulting in the release of singlet oxygen, which excites an amplied uorescent signal in the acceptor beads. As a result, the acceptor beads emit light at 615 nm, but only if they are within 200 nm of the analyte captured by both antibodies. As shown in Fig. 2a, the size of EVs measured by the Nanosight particle tracking system was approximately 100 nm, which prevented the detection of larger vesicles, such as apoptotic bodies, shedding vesicles or protein complexes. In addition, we could not obtain signals from CD63 recombinant protein by ExoScreen, indicating that this assay does not detect antigen monomers (Fig. 2b). We call this assay ExoScreen because the target of the assay is EVs
and because it has a possibility to screen for biomarker of various diseases.
To conrm the reliability for detecting EVs by ExoScreen, we selected CD9 and CD63, which are abundant on the surface of EVs and are expressed in numerous cells, to detect EVs. Conditioned medium (CM) of prostate cancer, prostate epithelial, breast cancer and colorectal cancer cell lines were processed to obtain puried EVs. ExoScreen was able to quantify the amount of EVs present in cell culture supernatants with CD9 and CD63 positive EVs detectable in a dose-dependent manner (Fig. 2c and Supplementary Fig. 1). The negative controls, represented by only the biotinylated antibody or acceptor bead-conjugated antibody, resulted in a minimal uorescent signal (Fig. 2c). In addition, the signal was decreased after detergent treatment (Fig. 2d and Supplementary Fig. 2) or Proteinase K treatment (Fig. 2e and
NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 3
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4591
CD63
15.6 31.3 62.5 125 250
(ng)
75 kDa
50 kDa
25 kDa
CD9
15.6 31.3 62.5 125 250
(ng)
30
25
15
20
10
5
CD9/CD9 CD9/CD9
ELISA
1.40
1.20
1.00
Absorbance
2.0
CD63/CD63 CD63/CD916
Signal intensity (104)
Signal intensity (104)
R2 = 0.998 R
2 = 0.752
R2 = 0.990 R2 = 0.995
1.6
1.2
0.8
0.4
Signal intensity (104)
12
8
4
0.80
0.60
0.40
0.20
0.00
0 0 1 2 3 4 6
Conditioned medium (l)
0 0 1 2 3 4 5
5 6
Conditioned medium (l)
0 0 1 2 3 4 6
Conditioned medium (l)
0 50 100
150
200
250
300
5
Extracellular vesicles (ng)
Figure 3 | Comparison of ExoScreen and conventional methods. (a) Immunoblotting analysis of CD63 (upper panels) or CD9 (lower panels)against the EVs isolated from HCT116 cells. EV protein concentration were measured via the Qubit system. EVs were puried from HCT116 cell CM. (b) Correlation between ExoScreen measurements for CD9 positive, CD63 positive or CD63/CD9 double-positive EVs and HCT116 cell CM in a dilution series. CM was prepared for 5 ml and diluted as indicated. Error bars are s.e.m. (n 3 for each condition). (c) Correlation between ELISA measurements
for CD9 positive EVs and EV protein concentration in a dilution series. EV protein concentration were measured via the Qubit system. EVs were puried from HCT116 cell CM. Error bars are s.e.m. (n 3 for each condition). Data are representative of at least three independent experiments each.
Table 1 | Comparison of ExoScreen and ELISA.
ExoScreen ELISAIncubation time 1.53 h 36 h coating time
Steps 2 More than 5 Washes No Yes Throughput High Low Sample volume Less than 5 ml* 50200 mlw
The volume of antibodies Low HighDynamic range 34 logs 2 logsAnalytic rangez 3 orders of magnitude 2 orders of magnitude
Sensitivity High High Plate format 96-well or 384-well 96-well
ELISA, enzyme-linked immunosorbent assay.
*Total assay volume 50 ml.
wTotal assay volume 200 ml. zCalculated by CD9/CD9 analysis of EVs derived from HCT116 cells
Supplementary Fig. 3), indicating that ExoScreen detected complexes of membranous vesicle and transmembrane proteins. Immunoblotting of the same puried EVs preparations conrmed the data obtained by ExoScreen. In fact, CD9 and CD63 proteins were detectable via immunoblotting (Fig. 3a and Supplementary Fig. 4). As shown in Fig. 3a, approximately 32 ng of EV proteins were needed to properly detect CD63 by immunoblotting, while ExoScreen could detect 15.6 ng of puried EVs (Fig. 2c). Furthermore, EVs from only 1 ml of culture medium are enough to detect by ExoScreen (Fig. 3b and Supplementary Fig. 5). In addition, ExoScreen has a wide working range compared with ELISA (Figs 2c and 3c). Moreover, because ExoScreen is a mix-and-read assay, these conventional methods require many steps and substantial time compared with
ExoScreen (Fig. 1). Thus, the ExoScreen assay increases throughput while substantially decreasing hands-on. Taken together, these results indicate that ExoScreen is superior for the detection of EVs to conventional immunoblotting and ELISA (Table 1). The results of EVs detection in culture supernatant without purication (Fig. 3b, Supplementary Figs 5 and 6) prompted us to investigate whether ExoScreen could detect and characterize EVs in human serum. To develop ExoScreen as a diagnostic tool for clinical use, we optimized the method to detect EVs in serum without purication because the protocol exhibited non-linearity of ExoScreen signals against serum samples (Fig. 4a). This non-linearity is most likely a result of the aggregation of condensed proteins in serum. Indeed, we added dextran-500 to suppress serum protein aggregation, and this
4 NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4591 ARTICLE
2.0 0.7
0.6
0.5
0.4
0.3
0.2
0.1
0 0 1 2 3 4 5 6
CD9/CD9
CD63/CD63
Signal intensity (x104)
Signal intensity (x104)
1.81.61.41.21.00.80.60.40.2 0
0.80.70.60.50.40.30.20.1 0
Signal intensity (104)Signal intensity (104)
Healthy donor 1 ()
Healthy donor 1 (+)
Healthy donor 2 ()
Healthy donor 2 (+)
Healthy donor 1 ()
Healthy donor 1 (+)
Healthy donor 2 ()
Healthy donor 2 (+)
Healthy donor 1
Healthy donor 2
Healthy donor 3
Healthy donor 4
0 1 2 3 4
Serum volume (l) Serum volume (l)
5 6
CD9/CD9 CD63/CD63
Healthy donor 1
Healthy donor 2
Healthy donor 3
Healthy donor 4
0.80.70.60.50.40.30.20.1 0
0 1 2 3 4Concentration of antibodies
and beads
5 6 0 1 2 3 4 5 6
Concentration of antibodies
and beads
Figure 4 | Detection of circulating EVs in healthy donor sera. (a) Correlation between ExoScreen measurements for CD9 or CD63 and serum volume in a dilution series with ( ) or without ( ) Dextran-500. The nal concentration of Dextran-500 was 1 mg ml 1. (b) Concentration of 1 means the
original concentration of donor and acceptor beads which we used in this study (see Methods section). In addition to original concentration, increased (twofold and vefold) and decreased (0.5-fold and 0.25-fold) amount of donor and acceptor beads were evaluated by ExoScreen using serum from four healthy donors. Data are representative of at least three independent experiments each.
50
CD147/CD9
Extracellular vesicles (ng)
CCD-18Co
HCT116
HCT15
HT29
WiDr
COLO201
COLO205
SW1116
HCT15
Signal intensity (104)
40
HCT116
30
HT29
COLO201
75 kDa
20
COLO205
CD147
50 kDa
WiDr
10
SW1116
25 kDa
CCD-18Co
CD9
0 0 50 100 150 200 250 300
Figure 5 | Analysis of the amount of CD147 in EVs derived from various colon cancer cell lines and a normal colon broblast cell line.(a) Immunoblotting analysis of CD147 or CD9 against puried EVs isolated from CCD-18Co cells, HCT116 cells, HCT15 cells, HT29 cells, WiDr cells, COLO201 cells, COLO205 cells and SW1116 cells. EV proteins (250 ng) was used for the detection of CD147 and CD9. (b) Correlation between ExoScreen detection of CD147/CD9 double-positive EVs and EV protein concentration in a dilution series. EVs protein concentration was measured via the Qubit system. EVs were puried from CCD-18Co, HCT116, HCT15, HT29, WiDr, COLO201, COLO205 or SW1116 CM. Error bars are s.e.m.(n 3 for each condition). Data are representative of at least three independent experiments each.
treatment eliminated the disruption of signals by protein aggregation (Fig. 4a). As shown in Fig. 4a, ExoScreen revealed that serum EVs were captured and expressed both CD9 and CD63 without purication. Further, these signals were detectable in a dose-dependent manner (Fig. 4a). In addition, we assessed whether the concentration of beads, which we employed in this study, was appropriate for the detection of circulating EVs in serum by checking the various concentrations of beads via ExoScreen, and found that the concentration of beads we employed in this study was adequate (Fig. 4b). Taken together, these results indicated that ExoScreen could monitor circulating EVs in serum without the need for a purication process.
Enrichment of CD147 on EVs from colorectal cancer cell lines. Because EVs are known to represent an important and specic
route of intercellular communication14, we reasoned that tumour-derived EVs may differ from circulating EVs in normal physiological conditions. Previous reports showed that the protein components of EVs from cancer cells were different from normal cells15,16. Indeed, it has been recently reported that for patients with stage III melanoma, the amount of specic protein in EVs was signicantly increased in individuals who eventually developed metastatic disease, indicating that EVs might have great potential for cancer diagnosis6. To identify cancer-derived EVs in cancer patients, EVs derived from the colorectal cancer cell line HCT116 cells and a normal colon broblast cell line CCD-18Co cells were subjected to proteomic analysis (Supplementary Table 1). When EVs isolated from CCD-18Co cells were compared with HCT116 cells, the amount of CD147, which is the immunoglobulin superfamily member, was found to be signicantly high in the EVs of HCT116 cells, whereas the expression could not be observed in CCD-18Co cells.
NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 5
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4591
CD147/CD9 in serum
CD147/CD9
in serum
CD147/CD9 in serum
P<0.01
12 7.0
P<0.005
Signal intensity (104 )
10
8 6
Signal intensity (104 )
6.05.04.03.02.01.0
4 2 0
8.07.06.05.04.03.02.01.0 0
Healthy donor
(n=191)
Cancer patient
(n=194)
Pre operation Post operation
CD147/CD9
in purified extracellular vesicles
Signal intensity (104 )
Signal intensity (104 )
1.61.41.21.00.80.60.40.2 0
Healthy
donor
(n=10)
Healthy
donor
(n=10)
Stage I
(n=8)
Stage II
(n=10)
Stage III
(n=9)
Stage IV
(n=8)
0
Stage I
(n=8)
Stage II
(n=10)
Stage III
(n=9)
Stage IV
(n=8)
1.0
0.8
0.6
CEA
CA199
CD147/CD9
Sensitivity
Sensitivity
1.0
0.4
0.8
0.6
0.4
0.2
0
0.2
AUC CEA: 0.669 CA19-9: 0.622
AUC: 0.820
0 0 0.6 0.8 1.0
1-specificity
0.4
0.2
0.4
0.6 0.8 1.0
0
0.2
1-specificity
Figure 6 | Analysis of circulating CD147 and CD9 double-positive EVs in healthy donors or colorectal cancer patient sera. (a) Serum levels of CD147/ CD9 double-positive EVs in colorectal cancer patients without any purication. The panel shows a scatter plot for healthy donors (n 191) and colorectal
cancer patients (n 194). The P-value was calculated by using Wilcoxon rank-sum test. (b) Changes in serum levels of CD147/CD9 double-positive EVs in
colorectal cancer patients (stage I or II: n 15) before (preoperation) and after (postoperation after 734 days) surgical removal of the tumor. Box lengths
represent the interquartile range (rst to third quartiles). The line in the center of the boxes represents the median value. Data represented by the asterisks are extreme values (greater than three times the interquartile range over the third quartile). The P-value was calculated by using Wilcoxon signed-rank test. (c) The results of ExoScreen detection of circulating EVs (left panel) and puried circulating EVs (right panel) in sera from healthy donors (n 10) and
colorectal cancer patients (n 35) using CD147 and CD9 antibodies. The panels show the signal intensities from each samples measured for CD147/CD9
double-positive EVs using ExoScreen. (d) Receiver operating characteristic curves between healthy donors and colorectal cancer patients assesing by CD147/CD9 double-positive EVs (left panel), CEA and CA19-9 (right panel). left panel, CD147/CD9 double-positive EVs (healthy donors versus colorectal cancer patients; AUC: 0.820); right panel, CEA (AUC: 0.669); CA19-9 (AUC: 0.622) Data are representative of at least three independent experiments each. AUC, area under the curve.
Notably, CD147 is plasma membrane protein and this is suitable for applying to the ExoScreen. We observed that CD147 is expressed on all of the colorectal cancer cell lines, but their expression levels are not uniform (Fig. 5a,b and Supplementary Figs 7 and 8). In addition, CD147 in EVs from CCD-18Co cells was hardly detectable. Several reports have shown that CD147 is expressed in the majority of human tumour types including colorectal cancer17,18, although CD147 is expressed in a variety of embryonic and adult tissues, such as spermatocytes, neuronal cells, erythrocyte and so on19. In addition, CD147 functions in lactate transporter, which is an important feature of cancer cell, because of the excessive anaerobic glycolysis phenomenon in cancer cells referred to as the Warburg effect20. Indeed, associations between high expression of CD147 and poor prognosis have previously been shown in colorectal cancer21, thus representing a potential marker for ex vivo analysis of tumour-derived EVs.
CD147 and CD9 double-positive EVs in clinical samples. Next, we used ExoScreen to detect cancer-derived EVs in human clinical samples (Fig. 6a). As shown in Fig. 6a, we found that
CD147 and CD9 double-positive EVs were signicantly higher in serum from cancer patients (n 194) than in serum from healthy
donors (n 191). Most importantly, most of CD147 in cancer
patient sera reduced after surgery (Fig. 6b), suggesting that the reduced signal of CD147 obtained from ExoScreen is originated from cancer-derived EVs, even the variety of cells expressed CD147. To conrm whether by ExoScreen really reects the protein prole of EVs in circulation, EVs were puried from the sera of tumour patients (n 35) and healthy donors (n 10)
and analysed for expression of CD147 via immunoblotting (Supplementary Fig. 9). We also performed ExoScreen against the same serum samples obtained by ultracentrifugation (Fig. 6c, right panel). As depicted in Supplementary Fig. 9, the expression of CD147 in EVs isolated from the sera of cancer patients correlated clearly with the results obtained from the ExoScreen assay (Fig. 6c), indicating that the accuracy of ExoScreen was conrmed and that it can be used to monitor EVs in circulation without any purication. Taken together, these results demonstrate that ExoScreen can be a tool for detection of EVs from as little as 5 ml of cancer patients serum to detect circulating cancer-derived EVs.
6 NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4591 ARTICLE
DiscussionIn summary, we propose a rapid, highly sensitive and widely usable detection method based on the amplied luminescent proximity homogeneous assay using photosensitizer-beads for cancer cell-derived EVs. Notably, different antibodies can be conjugated to capture different analytes, such as CD147, thus various types of cancer can be targetd. There are various colorectal cancer screening tests. For example, the fecal occult blood test has been recommended widely as a screening test for colorectal cancer; however, the fecal occult blood displays low sensitivity and specicity for detecting colorectal cancer22. Moreover, carcinoembrionic antigen (CEA) and carbohydrate antigen 19-9 (CA19-9) are the most commonly used tumour-associated antigens in the management of patients with colorectal cancer, although those biomarkers are not sensitive enough for early colorectal cancer23,24. In fact, the high levels of CD147 detected in patient sera showed the normal value range of CEA and CA19-9 in stage I patients (Supplementary Tables 2 and 3). In addition, the receiver operating characteristic curve indicates a diagnostic advantage of CD147/CD9 double-positive EVs in comparison with CEA and CA19-9 (Fig. 6d). From these aspects and the result shown in Fig. 6b, ExoScreen detecting CD147/CD9 double-positive EVs might be used for monitoring the status of cancer after the surgery and during chemotherapy, resulting in increase in QOL of the patients and providing doctor for the proper assessment of patient status. Further studies are needed to know whether our ExoScreen reduces colorectal cancer mortality as a screening test. It should be noted that CD147/CD9 double-positive EVs were also detected in samples with early stage colorectal cancer that invade into submucosal layer (T1 stage according to UICC classication) (Supplementary Table 2) by the ExoScreen assay. These results also indicate that ExoScreen can be used to detect biomarkers for diseases that are currently difcult to diagnose and monitor not only cancer, but autoimmune disease and degenerative disease of the brain. Thus, our data suggest that ExoScreen, in addition to being a novel liquid biopsy platform for the detection of circulating EVs, may aid variety of disease diagnosis and help to identify companion biomarkers that are important for new drug development.
Methods
Cell cultures. Human colorectal cancer cell lines (HCT116 cells, HCT15 cells, HT29 cells, COLO201 cells, COLO205 cells, WiDr cells and SW1116 cells) and normal colon broblast cell line CCD-18Co cells were purchased from American Type Culture Collection. HCT116 and HT29 cells were cultured in McCoys 5A medium supplemented with 10% heat-inactivated fetal bovine serum (FBS) and an antibioticantimycotic solution (Invitrogen) at 37 C in 5% CO2. WiDr cells
CCD-18Co cells were cultured in minimal essential medium (MEM) containing 2 mM L-gulutamine, an antibioticantimycotic solution, nonessential amino acids and 10% FBS at 37 C in 5% CO2. HCT15 cells, COLO201 cells and COLO205 cells were cultured in RPMI 1640 medium supplemented with 10% heat-inactivated FBS and an antibioticantimycotic solution at 37 C in 5% CO2. SW1116 cells were cultured in Leibovitz L15 medium supplemented with 10% heat-inactivated FBS and an antibioticantimycotic solution at 37 C in without CO2. The following additional cell lines were used: PNT2 cells, an immortalized normal adult prostatic epithelial cell line (DS Pharma Biomedical Co., Ltd. Osaka, Japan); PC3 cells, a human prostate cancer cell line initiated from a bone metastasis of a grade IV prostatic adenocarcinoma (American Type Culture Collection); MDA-MB-231-luc-D3H2LN cells (MDA-MB-231LN), a highly metastatic human breast cancer cell line (Xenogen); and MCF7 cells, a human breast cancer cell line which expresses oestrogen receptor (American Type Culture Collection). The above cells were cultured in RPMI 1640 medium supplemented with 10% FBS and an antibioticantimycotic solution at 37 C in 5% CO2.
Patient serum samples. Collection and usage of human serum from corolectal cancer patients (n 194) and healthy donor (n 94) were approved by Osaka
university Institutional Review Board (No.11343). Serum was aliquoted and kept at
80 C until used, and freeze-thawing was avoided as much as possible after that. Some part of the serum samples (n 97) from healthy donor shown in Fig. 6a were
purchased from BizCom Japan (Tokyo, Japan). Serum samples containing red
blood cells were excluded from the analysis. Informed consent was obtained from all patients.
Preparation of conditioned media and EVs. The cells were washed with phosphate-buffered saline (PBS), and the culture medium was replaced with advanced Dulbeccos Modied Eagle Medium for HCT116 cells, WiDr cells, SW1116 cells, HT29 cells and CCD-18Co cells, or advanced RPMI medium for the other cell lines, containing an antibioticantimycotic and 2 mM L-glutamine (but not containing FBS). After incubation for 48 h, the CM was collected and centrifuged at 2,000 g for 10 min at 4 C. To thoroughly remove cellular debris, the supernatant was ltered through a 0.22 mm lter (Millipore). The CM was then used for EV isolation. To prepare EVs, CM or the sera from colorectal patients and healthy donors were ultracentrifuged at 110,000 g for 70 min at 4 C. The pellets were washed with 11 ml of PBS, ultracentrifuged at 110,000 g for 70 min at 4 C and resuspended in PBS. The putative EVs fraction was measured for its protein content using a Quant-iT Protein Assay with Qubit2.0 Fluorometer (Invitrogen).
Reagents. The following antibodies were used for immunoblotting: mouse monoclonal anti-human CD63 antibody (clone H5C6, dilution 1:200) from BD Biosciences, mouse monoclonal anti-human CD9 antibody (clone ALB 6, dilution 1:200) from SantaCruz Biotechnology, mouse monoclonal anti-human CD147 antibody (clone MEM-M6/1, dilution 1:1,000) from Novus Biologicals and mouse monoclonal anti-Actin (clone C4, dilution 1:1,000) from Millipore. The secondary antibody (horseradish peroxidase-labeled sheep anti-mouse) were purchased from GE HealthCare.
The following antibodies used for ExoScreen and ELISA were developed in Shionogi & Co., LTD.: mouse monoclonal anti-human CD63 antibody (clone 8A12) and mouse monoclonal anti-human CD9 antibody (clone 12A12). Mouse monoclonal anti-human CD147 antibody (clone MEM-M6/1) was purchased from Novus Biologicals. Antibodies were used to modify either acceptor bead or biotin following the manufacturers protocol.
AlphaLISA reagents (Perkin Elmer, Inc., Waltham, MA 02451, USA) consisted of AlphaScreen Streptavidin-coated donor beads (6760002), AlphaLISA Unconjugated-acceptor beads (6062011) and AlphaLISA Universal buffer (AL001F). AlphaLISA assays were performed in 96-well half-area white plates (6005560) and read in an EnSpire Alpha 2300 Multilabel Plate reader (Perkin Elmer, Inc.).
ExoScreen assay. A 96-well half-area white plate was lled with 5 ml of sample, 5 nM biotinylated antibodies and 50 mg ml 1 AlphaLISA acceptor beads conjugated antibodies in the universal buffer. The volume of each reagent was10 ml. The plate was then incubated for 13 h at room temperature. Without a washing step, 25 ml of 80 mg ml 1 AlphaScreen streptavidin-coated donor beads were added. The reaction mixture was incubated in the dark for another 30 min at room temperature and the plate was then read on the EnSpire Alpha 2300 Multilabel Plate reader using an excitation wavelength of 680 nm and emission detection set at 615 nm. Background signals obtained from PBS were subtracted from the measured signals.
ELISA. Ninety-six well-plates (Nunc) were coated with 2.5 mg ml 1 anti-human CD9 or -CD63 antibodies in a volume of 50 ml per well of carbonate buffer (pH 9.6) and incubated for 4 h at room temperature. After 2 washes with 0.01% Tween-20 in PBS, 100 ml per well of Blocking One solution (Nacalai Tesque) was added at room temperature for 1 h. Following 3 washes in PBS, EVs puried from cell culture supernatants were added in a nal volume of 50 ml and incubated for 1 h at room temperature. After 3 washes with PBS, 50 ml of biotinylated anti-human CD9or -CD63 antibodies diluted to 1 mg ml 1 were added and incubated for 1 h at room temperature. After 3 washes with PBS, the plate was incubated with 100 ml of HRP-conjugated streptavidin (Cell Signalling Technology) diluted 1:2,000 in Blocking One solution for 1 h at room temperature. After the nal 3 washes with PBS, the reaction was developed with Peroxidase (TMB One Component HRP Microwell Substrate, SurModics). The reaction was arrested with 450 nm Stop Reagent for TMB Microwell Substrates (SurModics) and optical densities were recorded at 450 nm.
Immunoblotting. Equal amounts of EVs or whole-cell lysates were loaded onto 415% Mini-PROTEAN TGX gels (Bio-Rad, Munich, Germany). Following electrophoresis (100 V, 30 mA), the proteins were transferred to a polyvinylidene diuoride membrane. The membranes were blocked with Blocking One solution and then incubated with primary antibodies. After washing, the membranes were incubated with horseradish peroxidase-conjugated sheep anti-mouse IgG and then subjected to enhanced chemiluminescence using ImmunoStar LD (Wako). CD63, CD9 and CD147 were detected under non-reducing conditions. Original scans of the cropped images in the main gures (Figs 3a and 5a) are presented in Supplementary Fig. 10.
NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 7
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4591
Measurement of size distribution by NTA. Nanoparticle tracking analysis (NTA) was carried out using the Nanosight system (NanoSight) on sera diluted 1000-fold with PBS for analysis. The system focuses a laser beam through a suspension of the particles of interest. These are visualized by light scattering using a conventional optical microscope aligned perpendicularly to the beam axis, which collects light scattered from every particle in the eld of view. A 60 s video recorded all events for further analysis by NTA software. The Brownian motion of each particle was tracked between frames to calculate its size using the StokesEinstein equation.
References
1. Iero, M. et al. Tumour-released exosomes and their implications in cancer immunity. Cell Death Differ. 15, 8088 (2008).
2. Webber, J., Steadman, R., Mason, M. D., Tabi, Z. & Clayton, A. Cancer exosomes trigger broblast to myobroblast differentiation. Cancer Res. 70, 96219630 (2010).
3. Al-Nedawi, K. et al. Intercellular transfer of the oncogenic receptor EGFRvIII by microvesicles derived from tumour cells. Nat. Cell Biol. 10, 619624 (2008).
4. Skog, J. et al. Glioblastoma microvesicles transport RNA and proteins that promote tumour growth and provide diagnostic biomarkers. Nat. Cell Biol. 10, 14701476 (2008).
5. Taylor, D. D. & Gercel-Taylor, C. MicroRNA signatures of tumor-derived exosomes as diagnostic biomarkers of ovarian cancer. Gynecol. Oncol. 110, 1321 (2008).
6. Peinado, H. et al. Melanoma exosomes educate bone marrow progenitor cells toward a pro-metastatic phenotype through MET. Nat. Med. 18, 883891 (2012).
7. Kosaka, N., Iguchi, H. & Ochiya, T. Circulating microRNA in body uid:a new potential biomarker for cancer diagnosis and prognosis. Cancer. Sci. 101, 20872092 (2010).
8. Raposo, G. & Stoorvogel, W. Extracellular vesicles: exosomes, microvesicles, and friends. J. Cell Biol. 200, 373383 (2013).
9. Mathivanan, S. & Simpson, R. J. ExoCarta: a compendium of exosomal proteins and RNA. Proteomics 9, 49975000 (2009).
10. Bobrie, A. et al. Rab27a supports exosome-dependent and -independent mechanisms that modify the tumor microenvironment and can promote tumor progression. Cancer Res. 72, 49204930 (2012).
11. Yoshioka, Y. et al. Comparative marker analysis of extracellular vesicles in different human cancer types. J. Extracell. Vesicles 2, doi:http://dx.doi.org/10.3402/jev.v2i0.20424
Web End =10.3402/jev.v2i0.20424 (2013).
12. Valadi, H. et al. Exosome-mediated transfer of mRNAs and microRNAs is a novel mechanism of genetic exchange between cells. Nat. Cell Biol. 9, 654659 (2007).
13. Eglen, R. M. et al. The use of AlphaScreen technology in HTS: current status. Curr. Chem. Genomics 25, 210 (2008).
14. Kosaka, N. et al. Neutral sphingomyelinase 2 (nSMase2)-dependent exosomal transfer of angiogenic microRNAs regulate cancer cell metastasis. J. Biol. Chem. 288, 1084910859 (2013).
15. Xiao, D. et al. Identifying mRNA, microRNA and protein proles of melanoma exosomes. PLoS One 7, e46874 (2012).
16. Choi, D. S. et al. Quantitative proteomics of extracellular vesicles derived from human primary and metastatic colorectal cancer cells. J. Extracell. Vesicles 1, doi:http://dx.doi.org/10.3402/jev.v1i0.18704
Web End =10.3402/jev.v1i0.18704 (2012).
17. Riethdorf, S. et al. High incidence of EMMPRIN expression in human tumors. Int. J. Cancer. 119, 18001810 (2006).
18. Li, Y. et al. HAb18G (CD147), a cancer-associated biomarker and its role in cancer detection. Histopathology 54, 677687 (2009).
19. Muramatsu, T. & Miyauchi, T. Basigin (CD147): a multifunctional transmembrane protein involved in reproduction, neural function, inammation and tumor invasion. Histol. Histopathol. 18, 981987 (2003).
20. Weidle, U. H., Scheuer, W., Eggle, D., Klostermann, S. & Stockinger, H. Cancer-related issues of CD147. Cancer. Genomics. Proteomics. 7, 157169 (2010).
21. Boye, K. et al. EMMPRIN is associated with S100A4 and predicts patient outcome in colorectal cancer. Br. J. Cancer. 107, 667674 (2012).
22. Ahlquist, D. A. et al. Accuracy of fecal occult blood screening for colorectal neoplasia. A prospective study using Hemoccult and HemoQuant tests. JAMA 269, 12621267 (1993).
23. Chapman, M. A., Buckley, D., Henson, D. B. & Armitage, N. C. Preoperative carcinoembryonic antigen is related to tumour stage and long-term survival in colorectal cancer. Br. J. Cancer. 78, 13461349 (1998).
24. Kuusela, P. et al. Comparison of CA 19-9 and carcinoembryonic antigen (CEA) levels in the serum of patients with colorectal diseases. Br. J. Cancer. 49, 135139 (1984).
Acknowledgements
This work was supported in part by a Grant-in-Aid for the Third-Term Comprehensive 10-Year Strategy for Cancer Control, a Grant-in-Aid for Scientic Research on Priority Areas Cancer from the Ministry of Education, Culture, Sports, Science and Technology, and the Program for Promotion of Fundamental Studies in Health Sciences of the National Institute of Biomedical Innovation (NiBio), and the Japan Society for the Promotion of Science (JSPS) through the Funding Program for World-Leading Innovative R&D on Science and Technology (FIRST Program) initiated by the Council for Science and Technology Policy (CSTP), and a Grant-in-aid for Project for Development of Innovative Research on Cancer Therapeutics (P-Direct), and Grant-in-Aid for Scientic Research on Innovative Areas (functional machinery for non-coding RNAs) from the Japanese Ministry of Education, Culture, Sports, Science, and Technology, and Comprehensive Research and Development of a Surgical Instrument for Early Detection and Rapid Curing of Cancer Project (P10003) of the New Energy and IndustrialTechnology Development Organization (NEDO), a research program of the Project for Development of Innovative Research on Cancer Therapeutics (P-Direct), Ministry of Education, Culture, Sports, Science and Technology of Japan. We thank Ms. Ayako Irie at Quantum design Japan for supporting the Nanosight tracking analysis. We thank Ms. Ayako Inoue for excellent technical assistance. We thank Dr Roger Bosse and Dr Nami Kamura at Perkin Elmer for advising protocol for AlphaLisa system.
Author contributions
T.O. originated the concept. Y.Y., N.K. and T.O. carried out the project design. H.Oh.,H.Ok. and H.S. developed the antibodies. R.N., H.Y., H.I., M.M., K.F. and T.N provided the serum samples. H.Ha., H.S., H.Hi., F.T. and T.K. assisted with data interpretation. Y.K. and Y.Y. performed the experiments and T.O. supervised the project; Y.Y., N.K. and T.O. contributed to the writing of the manuscript.
Additional information
Supplementary Information accompanies this paper at http://www.nature.com/nature
Web End =http://www.nature.com/nature communications
Competing nancial interests: The authors declare no competing nancial interests.
Reprints and permission information is available online at http://npg.nature.com/reprintsand
Web End =http://npg.nature.com/ http://npg.nature.com/reprintsand
Web End =reprintsand permissions/
How to cite this article: Yoshioka, Y. et al. Ultra-sensitive liquid biopsy of circulating extracellular vesicles using ExoScreen. Nat. Commun. 5:3591 doi: 10.1038/ncomms4591 (2014).
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivs 3.0 Unported License. The images or other third party material in this article are included in the articles Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by-nc-nd/3.0/
Web End =http://creativecommon http://creativecommons.org/licenses/by-nc-nd/3.0/
Web End =s.org/licenses/by-nc-nd/3.0/
8 NATURE COMMUNICATIONS | 5:3591 | DOI: 10.1038/ncomms4591 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Nature Publishing Group Apr 2014
Abstract
Cancer cells secrete small membranous extracellular vesicles (EVs) into their microenvironment and circulation. Although their potential as cancer biomarkers has been promising, the identification and quantification of EVs in clinical samples remains challenging. Here we describe a sensitive and rapid analytical technique for profiling circulating EVs directly from blood samples of patients with colorectal cancer. EVs are captured by two types of antibodies and are detected by photosensitizer-beads, which enables us to detect cancer-derived EVs without a purification step. We also show that circulating EVs can be used for detection of colorectal cancer using the antigen CD147, which is embedded in cancer-linked EVs. This work describes a new liquid biopsy technique to sensitively detect disease-specific circulating EVs and provides perspectives in translational medicine from the standpoint of diagnosis and therapy.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer