© 2015 Xu et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Purpose

To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes.

Methods

The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was included in this prospective, comparative, observational study. Two separate tests with OQASII were conducted sequentially on the same day by two different examiners. A week later, the first examiner conducted the third measurement. All subjects underwent three consecutive tests during each session. The repeatability and reproducibility of the modulation transfer function cut off frequency (MTF cutoff), the Strehl ratio, the OQAS values (OVs) at contrasts of 100%, 20% and 9%, and the objective scatter index (OSI) were analyzed.

Results

For MTF cutoff, Strehl ratio, OV100%, OV20%, OV9%, and OSI, the mean values were 39.32±9.75cpd, 0.22±0.06, 1.31±0.33, 1.33±0.39, 1.33±0.41, 0.60±0.42, respectively. Repeatability and reproducibility were good with a very low coefficient of variation and high interclass correlation coefficients (>0.88) for all parameters. Bland-Altman plots showed good correlation with 95% limits of agreement ranged from -6.04 to 6.78cpd, -0.05 to 0.05, -0.20 to 0.23, -0.29 to 0.32, -0.40 to 0.42, -0.23 to 0.21 in inter-observer, and -6.56 to 7.42cpd, -0.06 to 0.06, -0.22 to 0.24, -0.30 to 0.32, -0.35 to 0.34, -0.24 to 0.23 in inter-visit, respectively.

Conclusion

The OQASII system yields excellent repeatability and good reproducibility for objective measurements of overall optical quality in clinic.

Details

Title
Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device
Author
Chen-Chen, Xu; Xue, Ting; Qin-Mei, Wang; Yi-Ni, Zhou; Jin-Hai, Huang; A-Yong, Yu
First page
e0117587
Section
Research Article
Publication year
2015
Publication date
Feb 2015
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1650969934
Copyright
© 2015 Xu et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.