© 2016 Lee, Ahn. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Purpose

To investigate whether corneal thickness parameters measured by optical coherence tomography (OCT), such as central corneal thickness (CCT), central corneal stromal thickness (CCST), and central corneal epithelial thickness (CCET), influence the intraocular pressure (IOP) difference measured by Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT).

Methods

In total, 50 eyes from 50 subjects without glaucomatous defects were included in this retrospective, cross-sectional study. We measured IOP using GAT and NCT and calculated the difference between the two methods. CCT was measured by a Cirrus HD-OCT device using anterior segment imaging. The basement membrane of the epithelium, which was seen as a high-reflection line in the OCT image, was taken as a reference line to measure CCST and CCET.

Results

The mean IOP measured by GAT and NCT was 16.7 ± 3.0 and 18.1 ± 3.8 mmHg, respectively. The mean IOP difference was 1.5 ± 1.7 mmHg, and the IOP measured by NCT was 8.4% ± 11.3% higher than that measured by GAT. The CCET and CCST were 57.9 ± 5.6 and 501.7 ± 33.8 μm, respectively. CCT showed a positive correlation with both GAT IOP (r = 0.648, P < 0.001) and NCT IOP (r = 0.676, P < 0.001). Although CCST showed a significant correlation with GAT IOP and NCT IOP, CCET did not. The difference between GAT IOP and NCT IOP increased with CCT (r = 0.333, P = 0.018), and CCET was positively correlated with the IOP difference between GAT and NCT (r = 0.435, P = 0.002).

Conclusions

IOP increased with greater CCT, and CCST seemed to have a more important role than CCET. CCET also increased with greater CCT, and this may be a possible explanation for the increasing difference in IOP between GAT and NCT with increasing CCT.

Details

Title
Effects of Central Corneal Stromal Thickness and Epithelial Thickness on Intraocular Pressure Using Goldmann Applanation and Non-Contact Tonometers
Author
Lee, Marvin; Ahn, Jaehong
First page
e0151868
Section
Research Article
Publication year
2016
Publication date
Mar 2016
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1774764263
Copyright
© 2016 Lee, Ahn. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.