J. Sens. Sens. Syst., 5, 137145, 2016 www.j-sens-sens-syst.net/5/137/2016/ doi:10.5194/jsss-5-137-2016 Author(s) 2016. CC Attribution 3.0 License.
Christophe Coillot1, Rahima Sidiboulenouar1, Eric Nativel2, Michel Zanca1,4, Eric Alibert1, Maida Cardoso1, Guillaume Saintmartin1,3, Harun Noristani3, Nicolas Lonjon3,4, Marine Lecorre3,4, Florence Perrin3, and Christophe Goze-Bac1
1Laboratoire Charles Coulomb (L2C-UMR5221), BioNanoNMRI group, University of Montpellier, Place Eugene Bataillon, 34095 Montpellier, France
2Institut dElectronique et des Systmes (IES-UMR5214), University of Montpellier,
Campus Saint-Priest, 34095 Montpellier, France
3Institut des Neurosciences de Montpellier (INSERM U1051), University of Montpellier, 34095 Montpellier, France
4Nuclear medicine, CMC Gui de Chauliac, University Hospital Montpellier, 34095 Montpellier, France
Correspondence to: Christophe Coillot ([email protected])
Received: 23 November 2015 Revised: 7 March 2016 Accepted: 23 March 2016 Published: 6 April 2016
Abstract. Nuclear magnetic resonance imaging (NMRI) is a powerful tool for biological investigations. Nevertheless, the imaging resolution performance results in the combination of the magnetic eld (B0) and the antenna efciency. This latter one results in a compromise between the size of the sample, the location of the region of interest and the homogeneity requirement. In the context of spinal cord imaging on mice, a ribbon solenoid coil is used to enhance the efciency of the MRI experiment. This paper details the calculation of the local magnetization contribution to the induced voltage of MRI coils. The modeling is illustrated on ribbon solenoid antennas used in emitterreceiver mode for the study. The analytical model, which takes into account the emitting mode, the receiving step and the imaging sequence, is compared to the measurement performed on a 9.4 T VARIAN MRI apparatus. The efciency of the antenna, in terms of signal to noise ratio, is signicantly enhanced with respect to a commercial quadrature volumic antenna, given a signicant advantage for the study of spinal cord injuries.
1 Context of the study: the spinal cord injuries
Spinal cord injuries (SCIs) are devastating neuropathologies that affect over 2.5 million patients worldwide, yield major handicaps and represent high costs to our society (from aboutUSD 1 million to up than 4 million per patient, National SCIStatistical Care, Sekhon and Fehlings, 2001). Neurological difculty depends on the spinal level and lesion severity.Unfortunately, there is no effective treatment for any symptoms associated with SCI. MRI is indeed well-established as the most commonly used imaging approach to diagnose and follow-up spinal cord injury patients. In the context of spinal cord injury studies in animals, MRI allows the localization of the region of the lesion and its evolution in order to
Signal modeling of an MRI ribbon solenoid coil dedicated to spinal cord injury investigations
understand the fundamental biological mechanisms and the perspective of translation to clinics, to evaluate the effect of therapeutical trials. Even if it is preferentially used for in vivo studies, in vitro imaging of tissue has the advantage of an enhanced resolution because of the acquisition time which is less constrained (Nor et al., 2015).
The aim of the ex vivo MRI study is to deepen the in vivo analysis of altered tissues by means of higher MRI spatial resolution and to evaluate putative correlation with histology. Nevertheless, the imaging resolution performance results from the combination of the magnetic eld (B0), the acquisition time and the antenna efciency in terms of signal to noise ratio (SNR). By a literal shortcut, MRI experimenters usually dene the SNR as the ratio between the
Published by Copernicus Publications on behalf of the AMA Association for Sensor Technology.
138 C. Coillot et al.: Modeling of an MRI ribbon solenoid coil
Figure 1. Spinal cord tissue of a CX3CR1 mouse: length is typically about 2535 mm and diameter is about 23 mm.
mean voxel intensity for a given sample (which is related to the magnetization quantity and to the coil sensitivity (in V /T )) divided by the voxel intensity in a region outside the sample (which is closely related to the coil plus preamplier stochastic noise contribution). In the following, we will consider the SNR of MRI experimenters. In order to enhance the imaging performances a dedicated antenna must be designed. The MRI coil design will result in a compromise between the size of the sample, the location of the region of interest and the homogeneity requirement. The requirement homogeneity on the MRI coil is dened (cf. Mispelter et al., 2006, p. 309) as a [notdef]5 % magnetic eld variation. Since we
will focus on the nal induced voltage, this criterion is excluded. Thus, we dene the homogeneity zone as the region where the induced voltage variation remains within 10 % of its maximum. The MRI coils can be used either in emitter receiver mode or solely in one of the two modes. In case of the separation between emitter and receiver mode an active or passive decoupling is mandatory. Next, an impressive variety of MRI coils have been invented and used (Mispelter et al., 2006): solenoid, saddle coil, loop coil, loop gap, scroll coil and bird cage. The conguration of the sample (its size) and the requirements of the experiment (in terms of SNR and homogeneity) could dictate the choice. In the context of spinal cord tissue (as shown in Fig. 1) the choice of the coil is restricted to the solenoid coil, the scroll coil or the loop gap. The solenoid coil appears to be a relevant choice for simplicity of manufacturing and signal to noise ratio efciency reasons (Hidalgo et al., 2009) even if the scroll coils seems to be competitive (Grant et al., 2010; Mem et al., 2013). The use of ribbon wire instead of round wire is guided by homogeneity considerations over the sample volume.
The homogeneity of the image remains however an important issue for all MRI experiments. MRI coil designers usually anticipate it through a magnetic eld intensity mapping (Mispelter et al., 2006; Hidalgo et al., 2009; Mem et al., 2013). Neglecting the MRI pulse sequence dependency a contrario, some authors have deduced the mapping of the radio-frequency coil using sequence dependency (Akoka et al., 1993; Insko and Bolinger, 1993) to correct it a posteriori. The purpose of this work is to derive a simple analytical model of the induced voltage for the solenoid coil used in emittingreceiving mode, which anticipates the effect of the MRI pulse sequence. We believe this analytic model could
offer a useful tool to guide the MRI coil designer by evaluating the signal homogeneity in the longitudinal direction of the solenoid prior to its realization. The method could be applied to other coils and combined with magnetic eld numerical simulation, to get it in the whole sample volume.
2 The nuclear magnetic resonance (NMR)-induced voltage: from global to local
When the magnetic eld (B0) is applied to paramagnetic matter a macroscopic nuclear moment (M0) arises while precessing at the Larmor frequency !0 (Bloch, 1946):
!0 = B0. (1)
The intensity of the net magnetic moment depends on the intensity of the polarization magnetic eld B0 (assumed in z direction following Fig. 3). Then, a varying magnetic eld at Larmor frequency (B1) is used to rotate the magnetization transverse to the polarizing magnetic eld (cf. Fig. 4). The ip angle of the magnetization ( ) will depend on B1 magnitude and duration ():
= B1. (2)
After application of B1 magnetic eld, the spins precess transversally to B0 and are associated with an electromagnetic eld whose magnetic component is classically measured by means of a coil.
The pioneer work on NMR antenna from Hoult and Richards (1976) invokes the Lorentzs reciprocity theorem to give a formulation of the induction law suited to NMR experiments:
=
[integraldisplay]
Sample
B1I M
[parenrightBig]
@ @t dVS, (3)
where is the electromotive force, B1 is the varying magnetic eld, I is the electrical current, M is the magnetization of the sample and VS is the sample volume. This formula is a well-known basis for the NMR coil SNR formulation. SNR is one of the most important parameter featuring the antenna efciency, the other one being the homogeneity of the radio-frequency magnetic eld over the sample. However, the equation derived by Hoult and Richard (namely Eq. 3) hides the dependency of the detected signal to the location of the spins while it is the quintessence of NMRI-induced voltage.
For this reason, the formulation of the induced voltage due to local elementary magnetization proposed by Pimmel (1990) in his PhD work (which is unfortunately in French but has been reported in the book of Mispelter et al., 2006) is a well suited approach to describe the NMRI signal dependency on the magnetization location r = (x,y,z):
e(t) =
@
@t ( m(t) [notdef] (B1(r)/I)), (4)
J. Sens. Sens. Syst., 5, 137145, 2016 www.j-sens-sens-syst.net/5/137/2016/
C. Coillot et al.: Modeling of an MRI ribbon solenoid coil 139
where e(t) is the contribution to the total electromotive force of the elementary magnetization m(t) when a magnetic eld by unit current B1(x,y,z) is applied also at point (x,y,z).
This formulation assumes the elementary magnetization at point (x,y,z) is constant (equal to M0 Ve, where Ve is the volume element), but the effect of the pulse sequence on the local magnetization is not implicit. This 3-D formulation of the local contribution is elegant but confusing since the precessing component of the magnetization only appears in the plane perpendicular to B0. For this reason, the local contribution to the induced voltage of the NMR signal proposed by Jacquinot and Sakellariou (2011) gives a more precise indication of the problem:
e(r) =
@
@t [ m?(r) [notdef] (B1?(r)/I)], (5)
where B1?(r) and m?(r) are the components in the plane
perpendicular to B0.
A generalization of the induced voltage, taking into account the propagative phenomena in the sample, is proposed by Insko et al. (1998). Thus Eq. (3) is generalized to
e(r) =
Figure 2. Illustration of the local elementary magnetization ( m) position with respect to the coil turn for the vector potential calculation over the turn.
where r is the distance vector from the elementary magnetization location to the point where the vector potential is computed (cf. Fig. 2) and f (r) = eikr(1ikr) summarizes the
contribution of the near and far eld (Insko et al., 1998).
Next, from the relation between the magnetic eld (B) and the vector potential (A),
B =
h[notdef] A. (9) From Stokes theorem, for the magnetic ux through the surface, (S)
=
[integraldisplay][integraldisplay]
(S)
BdS =
@
@t [ m?(r) [notdef] (B[prime]1?(r)/I)], (6)
where B[prime]1?(r) is the generalized magnetic eld retarded po
tential form:
B[prime]1?(r) =
[notdef]0 4
[contintegraldisplay]
[integraldisplay][integraldisplay]
(S)
h[notdef] AdS =
[contintegraldisplay]
Adl. (10)
By substituting the vector potential by its mathematical equation (as given by Eq. 8),
=
[contintegraldisplay]
[notdef]0 4
eikr(1 ikr)
dl [notdef] r [bardbl] r[bardbl]3
, (7)
where k = !p"r"0[notdef]r[notdef]0 is the wave number, "r is the relative
permittivity of the sample and [notdef]r its relative permeability. As emphasized by Insko et al. (1998), the eikr(1 ikr) term
can be omitted in the near-eld approximation (for kr 1 it
follows eikr(1 ikr) ! 1).
When we try to feel the nature of the induced voltage by means of Eq. (3), we have to face some inconsistencies. First, at the time where the induced voltage is measured the magnetic eld B1 and the radio-frequency current I are both null, and consequently the term B1/I is undened. Second, the scalar product between the magnetization vector and a magnetic eld is usually associated with the Zeeman energy, which is confusing. So, even if it is remarkably true from a mathematical point of view, the magnicent intuition of Hoult and Richard, which have gave birth to their famous formula, leads to misunderstanding for beginners. For these reasons, we derive below another way to write the NMRI-induced voltage.
We start from the mathematical form of the vector potential (A) associated with the magnetic dipole moment ( m) corresponding to the magnetization of a small volume ( m =
M0 V ):
A =
[notdef]0 4
m(r,t) [notdef] r [bardbl] r[bardbl]3
f (r)dl, (11)
the total ux [Phi1], which represents the summation over N turns, is
[Phi1] = N
[contintegraldisplay]
[notdef]0 4
m(r,t) [notdef] r [bardbl] r[bardbl]3
f (r)dl, (12)
which allows one to derive the local magnetization contribution to the induced voltage:
e(r,t) = N
@ @t
[contintegraldisplay]
[notdef]0 4
m(r,t) [notdef] r [bardbl] r[bardbl]3
f (r)dl. (13)
By decomposing the components as m = mxx + myy,
we can write
e(r,t) =
N[notdef]0 4
[contintegraldisplay]
f (r)x [notdef] r [bardbl] r[bardbl]3
dl @ mx(r,t)
@t (14)
N[notdef]0 4
[contintegraldisplay]
f (r)y [notdef] r [bardbl] r[bardbl]3
dl @ my(r,t)
@t .
Finally, we can write the local magnetization contribution to the induced voltage in the standard way of writing a signal
www.j-sens-sens-syst.net/5/137/2016/ J. Sens. Sens. Syst., 5, 137145, 2016
m [notdef] r [bardbl] r[bardbl]3
f (r), (8)
140 C. Coillot et al.: Modeling of an MRI ribbon solenoid coil
coming from a sensor which involves the physical quantity to be measured (here the magnetization) multiplied by the sensor sensitivity (a coefcient or a function S):
e(r,t) =
Sx(r)@ mx(r,t)
@t + Sy(r) @ my(r,t) @t
, (15)
where e(r,t) represents the induced voltage, mx(r,t) and my(r,t) are respectively the x and y component of the magnetization vector m, while Sx(r) and Sy(r) are the local coil sensitivities in x and y directions related to the coil geometry, dened as
Sx(r) =
[notdef]0N
4
Figure 3. Illustration of the ribbon solenoid coil.how the signal is detected by the coil when it is used in receiver mode.
3.1 Emitter mode: the magnetization tilt
The magnetic eld component generated by a solenoid coil (Fig. 3) on y axis (B1(y,t)) is given by Biot and Savarts law.
This one can be formulated using sensitivity equations:
B1(y,t) = Sy(y)I(t), (20)
where Sy(y) is well known for a solenoid while it could be calculated using Eq. (19). It follows that
Sy(y) = (21)
[notdef]0N
2L
[contintegraldisplay]
f (r)x [notdef] r [bardbl] r[bardbl]3
dl, (16)
Sy(r) =
[notdef]0N
4
[contintegraldisplay]
f (r)y [notdef] r [bardbl] r[bardbl]3
dl, (17)
where x and y are the unit vectors along x and y axis.
By virtue of the scalar triple product, the sensors sensitivity coefcient can also be expressed:
Sx(r) =
[notdef]0N
4 x [notdef]
[contintegraldisplay]
f (r)dl [notdef] r [bardbl] r[bardbl]3
, (18)
Sy(r) =
[notdef]0N
4 y [notdef]
[contintegraldisplay]
f (r)dl [notdef] r [bardbl] r[bardbl]3
, (19)
where we denitively recognize the BiotSavart law at a sign nearby (or the classical B/I term multiplied by the units vector). The sign difference comes from the reverse r direction convention between the usual form of Biot and Savart law with respect to the vector potential writing of Eq. (8) (Insko et al., 1998).
3 1-D NMRI signal modeling of a ribbon solenoid coil
In this section, we will detail the calculation of the induced voltage. We will perform the calculation on N turns of ribbon solenoid of length L and radius R as the one represented in Fig. 3. The helicity of the antenna will be neglected. We assume a perfect homogeneity in the transverse plane (xz), which is a valid hypothesis for solenoid where the sample is not too close to the coils wire, as reported in Hidalgo et al. (2009). Next, we assume a homogeneous current distribution owing through the conductor. Moreover the time propagative phenomenon in the coil can be neglected since we will assume that total wire length will remain much smaller than /2. Then, as discussed in Hoult (2009), we can neglect the far-eld contribution (and consequently f (r) is assumed close to 1) even if, according to Insko et al. (1998), it seems to be a rough assumption since kr value at 400 MHz is close to 1. Lastly, the elementary magnetization will be designated as M in the following for simplicitys sake.
The NMRI coil is supposed to be used both in emitter and receiver mode. We will discuss in the following how the magnetization is tilted when the coil is used in emitter mode and
L/2 y
pR2 + (L/2 y)2 [vextendsingle][vextendsingle][vextendsingle][vextendsingle][vextendsingle]
.
In the following, the time dependence of both B1 and I will be omitted.
Next, by expanding Eq. (2), the distribution of the angle magnetization along the y axis ( (y)) will be directly related to the magnetic eld distribution:
(y) = B1(y). (22)
Practically, the tilt angle magnetization distribution will be related to the calibration pulse sequence conditions. In this study we assume a calibration pulse performed on a small thickness slice at the center of the antenna (i.e., y = 0). In
case of a different pulse condition (for instance /2 pulse obtained over the whole sample volume) the modeling of the magnetization tilt would differ.
Thus, under the hypothesis of a centered pulse calibration, the /2 magnetization angle is expressed: 2 = B1(0)0, (23)
where B1(0) is the magnetic eld at the center of the antenna and 0 is the pulse duration.
By combining Eqs. (22) and (23), it appears that tilt angle is proportional to magnetic eld (B1(y)) independently to the
J. Sens. Sens. Syst., 5, 137145, 2016 www.j-sens-sens-syst.net/5/137/2016/
[vextendsingle][vextendsingle][vextendsingle][vextendsingle][vextendsingle]
L/2 + y
pR2 + (L/2 + y)2 +
C. Coillot et al.: Modeling of an MRI ribbon solenoid coil 141
Figure 4. Magnetization vector representation in the static frame. hardware strategy (adjustment of magnetic eld magnitude or pulse duration):
(y) =
2B1(0)0
where ERNST is the pulse duration required to tilt the magnetization at Ernst angle at the center of the sample.
Thus, the magnetization in xy plane (component Mxy(y), cf. Eq. 25) will be
Mxy(y) = M0
sin( (y))
1 e TR T1
[parenrightbigg]
e TET 2 , (28)
where (y) is the magnetization angle distribution given by Eq. (27).
3.1.3 Spin echo sequence: the magnetization tilt
For the spin echo sequence, the magnetization will be tilted by a /2 pulse followed by a pulse. We use the signal dependency given by Akoka et al. (1993) and Insko and Bolinger (1993) for spin echo sequence: sin3( ). Thus the magnetization in xy plane (Mxy( (y))) will be given by
Mxy( (y)) = M0sin3( (y))[parenleftbigg]
1 e
1 cos( (y))e
TR
T1
B1(y). (24)
Finally, the components of the magnetization vector M = [Mx;My;Mz] represented in Fig. 4 are expressed:
0
@
Mxy(y)sin(!0t) Mxy(y)cos(!0t)
Mz(y)
TR
T1
A, (25)
where Mxy(y) is the magnetization magnitude in the xy plane.
3.1.1 Single pulse sequence: the magnetization tilt
In the context of NMR single pulse sequence, the magnetization components will take the following form:
0
B
B
B
@
M0 sin( (y))e TET 2 (1 e
e
TET2 , (29)
where (y) is the magnetization angle distribution given by Eq. (24) and (T2) is the true transverse relaxation time.
3.2 Receiver mode: the induced voltage
Once the magnetization ip is determined, we can establish the induced voltage associated with the magnetization precession. The induced voltage created by the elementary magnetization at location y can be simply expressed from Eq. (15) by considering only the sensitivity along y axis:
e(t,y) = Sy(y)
dMy(y)
1
1
C
C
C
A
dt , (30)
where Sy(y) is deduced from Eq. (21).
Since My(y) = Mxy(y)cos(!0t), the induced voltage in
harmonic regime will be
|e(y)[notdef] = [notdef]!0Sy(y)Mxy(y)[notdef], (31)
where Mxy( (y)) is given either by Eqs. (28) or (29) depending of the running pulse sequence.
4 Design of the NMRI ribbon solenoid coil
To design the NMRI ribbon solenoid coil, the rst point is to determine the total length of the wire (Lw). In order to neglect the propagative phenomenon into the coil, a length of the solenoid coil about /6 is classically used (Mispelter
et al., 2006), while is determined by the nuclear frequency of interest into the magnet. In our study, we performed 1 H measurement on a 9.4 T Varian MRI. Thus, the gyromagnetic
www.j-sens-sens-syst.net/5/137/2016/ J. Sens. Sens. Syst., 5, 137145, 2016
TR
T1 )sin(!0t)
M0 sin( (y))e TET 2 (1 e
TR
T1 )cos(!0t)
M0 cos( (y))(1 e
TR
T1 )
, (26)
where (T1) is the longitudinal relaxation time, (TR) is the repetition time, (T 2) is the transverse relaxation time in a heterogeneous magnetic eld and (TE) is the echo time.
Let us now determine the ip angle distribution for two important MRI pulse sequences, namely gradient echo and spin echo.
3.1.2 Gradient echo sequence: the magnetization tilt
For a gradient echo sequence it is a desirable condition to perform it at the Ernst angle ( ERNST). Following the same reasoning that has led to Eq. (24), the ip angle distribution ( (y)) for a gradient echo sequence will follow
(y) = ERNST
B1(y)
B1(0)ERNST , (27)
142 C. Coillot et al.: Modeling of an MRI ribbon solenoid coil
Table 1. Solenoid coil design parameter summary.
N D (mm) w (mm) t (mm)
3 8 10 2
frequency (f0) is about 400 MHz, and it follows that the total length of the solenoid coil should be limited to 12.5 cm:
Lw =
NDcos( ) /6, (32)
where is the pitch angle. The turn number is then deduced from the size of the sample (or its mechanical support). In our study the tissue length is 40 mm while the tube diameter
(D) is equal to 8 mm. The copper ribbon has 10 mm width (w) and 50 m thickness (tw). From Eq. (32) it follows that
N = 3. Choosing a space between turns (t) equal to 2 mm
results in an average length of the solenoid L 36 mm (the
design parameters are summarized in Table 1).
4.1 NMRI coil electrical model
Basically, the coil can be represented by the electromotive force (given by Eq. 15) in series with an inductance (L1) and a resistance which takes into account the occurrence of the skin effect (R1AC). When considering a single ribbon, the current density will tend to ow at the ends of the ribbon: this effect is known as lateral skin effect (Belevitch, 1971). When considering a multiple-turn solenoid (ribbon or round wire), the current density between neighbor conductors will be strengthened especially at the extremities of the coil: this effect is known as the proximity skin effect (Butterworth, 1925). The analytic modeling of these phenomena is beyond the scope of this paper, but they can be efciently approached by electromagnetic numerical simulations. Finally these effects will dictate the current distribution at high frequencies and thus the homogeneity. In practice, even if it increases the coils resistance and thus the noise, the use of ribbon wire tends to improve the homogeneity (Grant et al., 2010; Mem et al., 2013). In the case of ribbon solenoid the spacing between turns should be minimized in order to preserve homogeneity on one side but should be sufcient to avoid to strengthen the proximity effect on the other side. The different skins effects are illustrated in Fig. 5.
Lastly, the occurrence of the coils resonance at a frequency (f0) where the wavelength ( ) is about twice the wire length (namely = c/f0 [similarequal] 2Lc where c is the vacuum light
velocity), will imply f0 [similarequal] c/(2Lc)) can be interpreted by a
capacitance (C1) in parallel (cf. Fig. 6) with the previous components (Knight, 2013b).
4.2 Tuningmatching circuit
The intrinsic self-resonance of the coil is much higher than the one to observe; moreover the electrical resonance fre-
Figure 5. Illustration of the different skin effect regimes in the section of one, two and ve turns: (a) usual skin effect, (b) proximity skin effect and (c) lateral skin effect.
Figure 6. Electrical circuit of the coil plus the tuningmatching circuit.
quency will be affected by the sample dielectric properties.
For these reasons, a variable capacitance is usually added in parallel to the coil terminal to adjust the resonance frequency (the tuning circuit is represented by capacitances CT and CpreT in Fig. 6). On the other side, the coil must be connected to the radio-frequency power amplier and impedance of the coil must be matched to the standard 50[Omega1] at the frequency of use. For this purpose the capacitances CM, CpreM1
and CpreM2 are used in series (the electrical component values are summarized in Table 2).
5 Experimental results
The experiments have been performed on a 9.4 T magnet from AGILENT. The ribbon solenoid coil has been wound on a glass tube. A small printed circuit board (PCB) is used to realize the tuningmatching circuit. Variable non-magnetic capacitances from VOLTRONICS (Ref. NMKJ10HVE from0.5 up to 9 pF) are used. Copper foil connected to the ground BNC cable has been used on the reverse face of the mechanical structure to perform an electromagnetic shielding and to prevent the sensibility of the circuit during the manual adjustment. Special care has been given to the connection distance between the ground and the copper foil to prevent occurrence of resonance in the frequency range of interest. A mechanical structure to maintain the glass tube and the PCB, represented in Fig. 7, has been realized using a 3-D printer with polylactic acid material.
J. Sens. Sens. Syst., 5, 137145, 2016 www.j-sens-sens-syst.net/5/137/2016/
C. Coillot et al.: Modeling of an MRI ribbon solenoid coil 143
Table 2. Electrical parameter summary.
R1AC L1 C1 (pF) CpreT0 CpreM1 CpreM2 Q
0.3([Omega1]) 15 nH 0.55 pF 6.8 pF 0.4 pF 0.4 pF 135
Figure 8. Ex vivo MR images (multi-echo multi-slice sequence: TR = 1155 ms; TE = 14 ms; NE = 1; FOV = 10 mm [notdef] 10 mm;
60 slices; thickness = 0.6 mm; resolution = 256 [notdef] 256) of spinal
cords from adult mice. Hypersignal (buttery shape) represents the grey matter whereas the surrounding hyposignal corresponds to the white matter: (a) image obtained with RAPID Biomedical 43 mm volumic quadrature coil in 14 h 0 min; (b) image obtained with the ribbon solenoid coil of Fig. 7 in 1 h 30 min.
5.1 Gradient echo sequence: method and experimental results
According to Eq. (27), is proportional to B1(y)/y, which is equivalent to the sensitivity term Sy(y), resulting in (y) /
Sy(y). Then, we normalize the tilt angle distribution to the Ernst angle:
(y) = ERNST(y)
Sy(y)
Figure 7. Photograph of the ribbon solenoid coil dedicated to spinal cord injury.
The pulse power needed to tilt the magnetization at /2 (on a 100 mm length and 4 mm diameter glass tube lled with potable water) in a centered thin slice is about 4.8 dB (the SNR is 1600/45, obtained using a gradient echo sequence with the following parameters: FOV = 10[notdef]10: TR = 250 ms,
TE = 4.32 ms, ip angle = 80 , average = 2, resolution 128[notdef]
128, 20 slices of thickness = 1 mm). For comparison, the
pulse power, on the same water sample, needed by a commercial volumetric quadrature antenna with 43 mm inner diameter (from RAPID Biomedical) in the same conditions is about 20 dB (the SNR on the image in the same conditions is 290/45). The enhancement of the SNR between the ribbon solenoid coil and the commercial antenna is 5.5, which
is well correlated with the pulse power attenuation. This increase of the SNR allows one either to perform faster acquisition ( 25 times) for a given resolution or to enhance the res
olution for a given acquisition time. In the context of spinal cord injury studies, this improvement in the SNR (as demonstrated on a T2-weighted spin echo sequence in Fig. 8) was crucial. The time acquisition has been divided by 10 while the image quality has been signicantly enhanced allowing one to combine high-resolution T2-weighted acquisition and diffusion MRI imaging to investigate accurately the lesion site of the spinal cord.
Sy(0) , (33)
where Sy(y) is given by Eq. (21). Combining Eqs. (28) and (33) into the MRI-induced voltage (Eq. 31) leads to
e(y) / (y)
sin( (y))
(1 cos( (y))e
e TET 2 . (34)
For a given set of experimental conditions parameters (TR, T1, ERNST) and given solenoid coil size parameters, the induced voltage e(y) can be plotted. The model (values are normalized) is compared to the normalized experimental data (cf. Fig. 10).
5.2 Spin echo sequence: method and experimental results
For a spin echo sequence, the distribution of the magnetization angle tilt will be
(y) =
TR
T1 )
Sy(y)
Sy(0) , (35)
www.j-sens-sens-syst.net/5/137/2016/ J. Sens. Sens. Syst., 5, 137145, 2016
144 C. Coillot et al.: Modeling of an MRI ribbon solenoid coil
Figure 9. Comparisons of longitudinal homogeneity proles for different solenoid coil aspect ratios.
where Sy(y) is given by Eq. (21). Combining Eqs. (29) and (35) into the MRI-induced voltage (Eq. 31) leads to
e(y) / (y)sin3( (y))(1 e
TR
T1 )e
TE
T2 . (36)
The model can be used to anticipate the longitudinal homogeneity range for different aspect ratios (i.e., L/(2R) ratios) of solenoid coils in the context of spin echo sequence. It highlights how to use the model to guide the designer. For instance, it highlights the inefciency of the short solenoid coil where less than 50 % of the length provides enough signal (considering the homogeneity criterion as 10 % variation of the signal intensity).
Finally, the model so obtained is compared to experimental data in Fig. 10. The measured data have been obtained by inserting a tube of water (100 mm length) into the ribbon solenoid coil. The measurements have been done through the VNMRJ software interface using circular region of interest over the whole diameter.
The comparison between gradient echo and spin echo (Fig. 10a) allows one to illustrate the y axis homogeneity dependency on the pulse sequence. The measurement also shows differences between the two type of sequences, the echo gradient sequence exhibiting a signicantly wider homogeneity range. For both sequences we can notice a signicant smooth decrease of the signal far from the coil in practice while the decrease is predicted as more abrupt by the model. It is certainly related to the far-eld contribution predicted by Insko et al. (1998) conversely to the conclusion given by Hoult (2009).
The real ribbon solenoid coil signal exhibits some magnitude uctuations which are attributed to the spacing between turns. The occurrence of maxima at the ends of the solenoid coil is attributed to the high-frequency current density distribution discussed above, where the different skin effects tend to distribute the current density at its ends increasing both B1 and the sensitivity. Finally, the modeling allows the anticipation of the y axis homogeneity tendency at an early stage of the coil design.
Figure 10. Normalized induced signal comparisons: (a) gradient echo sequence (GEMS) versus spin echo sequence (MEMS); (b) gradient echo sequence measured (GEMSMEAS) versus model (GEMS) (TR = 688 ms; TE = 4.5 ms;
FOV = 10 mm [notdef] 10 mm; 80 slices; thickness = 1 mm; resolu
tion = 128 [notdef] 128); (c) spin echo sequence measured (MEMS
MEAS) versus model (MEMS) (TR = 10 s; TE = 10 ms;
FOV = 10 mm [notdef] 10 mm; 80 slices; thickness = 1 mm; resolu
tion = 128 [notdef] 128).
6 Conclusions
The step-by-step modeling presented in this paper enables the estimation of the longitudinal signal homogeneity of a solenoid coil depending on the running imaging sequence.The formulation of the local contribution of the elementary magnetization to the induced voltage, using sensitivities coefcients, derived in this paper is well suited for MRI coil designers while it avoids the use of confusing notations. We believe this type of modeling could be applied to other coil
J. Sens. Sens. Syst., 5, 137145, 2016 www.j-sens-sens-syst.net/5/137/2016/
C. Coillot et al.: Modeling of an MRI ribbon solenoid coil 145
shapes in order to guide MRI coil designers in their choice of coils. Two important phenomena reduce the analytic modeling validity, namely, the current density distribution at high frequency and the far-eld contribution, which will be more signicant for higher frequencies and higher coil and sample sizes. Quantitative modeling of the homogeneity over the volume could be attained by combining classical numerical computations of magnetic eld with the MRI pulse sequence conditions.
Finally, the customized antennas are a relevant and cheap way of enhancing the performance of the MRI studies with respect to the commercial antennas.
Acknowledgements. The authors would like to thank Association Verticale, who funded the 9.4 T MRI dedicated to SCI studies, and also thank the Labex NUMEV, who funded the electronic material needed to perform this study.
Edited by: I. BrsonyReviewed by: two anonymous referees
References
Akoka, S., Franconi, F., Seguin, F., and Le Pape, A.: Radiofrequency map of an NMR coil by imaging, Magnetic Resonance Imaging, 11, 437441, 1993.
Belevitch, V.: The lateral skin effect in a at conductor, Philips techRev., 32, 221231, 1971.
Bloch, F.: Nuclear Induction, Physical Review, 70, 460474, 1946. Butterworth, S.: On the alternating current resistance of solenoidal coils, Proc. R. Soc. Lon. Ser.-A, 107, 693715, 1925.
Grant, C. V., Wu, C. H., and Opella, S. J.: Probes for high eld solid-state NMR of lossy biological samples, J. Magn. Reson., 204, 180188, 2010.
Hidalgo, S. S., Jirak, D., Solis, S. E., and Rodriguez, A. O.: Solenoid coil for mouse-model MRI with a clinical 3-Tesla imager: body imaging, Rev. Mex. Fis., 55, 140144, 2009.
Hoult, D. I.: The Origins and Present Status of the Radio Wave Controversy in NMR, Concept. Magn. Reson. A, 34, 193216, 2009.
Hoult, D. I. and Richards R. E.: The signal-to-noise ratio of the nuclear magnetic resonance experiment, J. Magn. Reson., 24, 71 85, 1976.
Insko, E. K. and Bolinger, L.: Mapping of the Radiofrequency eld,J. Magn. Reson., 103, 8285, 1993.
Insko, E. K., Elliott, M. A., Schotland, J. C., and Leigh, J. S.: Generalized Reciprocity, J. Magn. Reson., 131, 111117, 1998. Jacquinot, J.-F. and Sakellariou, D.: NMR Signal Detection using
Inductive Coupling: Applications to Rotating Microcoils, Concept. Magn. Reson. A, 38, 3351, 2011.
Knight, D. W.: Solenoid Inductance Calculation, version of Sept. 2015, available at: http://www.g3ynh.info/zdocs/magnetics/Solenoids.pdf
Web End =http://www.g3ynh.info/zdocs/magnetics/ http://www.g3ynh.info/zdocs/magnetics/Solenoids.pdf
Web End =Solenoids.pdf , 2013a.
Knight, D. W.: The self-resonance and self-capacitance of solenoid coils, version of Sept. 2015, available at: http://www.g3ynh.info/
Web End =http://www.g3ynh.info/ , 2013b.
Meme, S., Joudiou, N., and Szmereta, F.: In vivo magnetic resonance microscopy of Drosophilae at 9.4T, Magn. Reson. Imaging, 31, 109119, 2013.
Mispelter, J., Lupu, M., and Briguet, A.: NMR probehads for biophysical and biomedical experiments: theoretical principles and practical guidelines, Imperial College Press, 2006.
Noristani, H. N., Lonjon, N., Cardoso, M., Le Corre, M., Chan-
Seng, E., Captier, G., Privat, A., Coillot, C., Goze-Bac, C. and Perrin, F. E.: Correlation of in vivo and ex vivo 1H-MRI with histology in two severities of mouse spinal cord injury, Front. Neuroanat., 9, 24, doi:http://dx.doi.org/10.3389/fnana.2015.00024
Web End =10.3389/fnana.2015.00024 http://dx.doi.org/10.3389/fnana.2015.00024
Web End = , 2015. Pimmel, P.: Les Antennes en Resonance Magntique Nuclaire:
Fonctionnement et Ralisation, PhD Thesis, Univ. Claude Bernard, Lyon, 1990.
Sekhon, L. H. and Fehlings, M. G.: Epidemiology, demographics, and pathophysiology of acute spinal cord injury, Spine, 26, S2 12, 2001.
Webb, A. G.: Radiofrequency microcoils for magnetic resonance imaging and spectroscopy, J. Magn. Reson., 31, 5566, 2013. Wheeler, H. A.: Simple Inductance Formulas for Radio Coils, P.
IRE, 16, 13981400, 1928.
www.j-sens-sens-syst.net/5/137/2016/ J. Sens. Sens. Syst., 5, 137145, 2016
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Copernicus GmbH 2016
Abstract
Nuclear magnetic resonance imaging (NMRI) is a powerful tool for biological investigations. Nevertheless, the imaging resolution performance results in the combination of the magnetic field (<i>B</i><sub>0</sub>) and the antenna efficiency. This latter one results in a compromise between the size of the sample, the location of the region of interest and the homogeneity requirement. In the context of spinal cord imaging on mice, a ribbon solenoid coil is used to enhance the efficiency of the MRI experiment. This paper details the calculation of the local magnetization contribution to the induced voltage of MRI coils. The modeling is illustrated on ribbon solenoid antennas used in emitter-receiver mode for the study. The analytical model, which takes into account the emitting mode, the receiving step and the imaging sequence, is compared to the measurement performed on a 9.4-T VARIAN MRI apparatus. The efficiency of the antenna, in terms of signal to noise ratio, is significantly enhanced with respect to a commercial quadrature volumic antenna, given a significant advantage for the study of spinal cord injuries.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer