A Disposable paper breathalyzer with an alcohol sensing organic electrochemical transistor
OPEN
R A
P
Bihar,,,*, Yingxin Deng,*, Takeo Miyake,, Mohamed Saadaoui, George G. Malliaras & Marco Rolandi,
Breathalyzers estimate Blood Alcohol Content (BAC) from the concentration of ethanol in the breath. Breathalyzers are easy to use but are limited either by their high price and by environmental concerns, or by a short lifetime and the need for continuous recalibration. Here, we demonstrate a proof-of- ethylenedioxythiophene):poly(styrenesulfonate) (PEDOT:PSS), and is printed on paper. ADH and its cofactor nicotinamide adenine dinucleotide (NAD+) are immobilized onto the OECT with an electrolyte gel. When the OECT-breathalyzer is exposed to ethanol vapor, the enzymatic reaction of ADH and ethanol transforms NAD+ into NADH, which causes a decrease in the OECT source drain current. In disposable, ecofriendly, and integrated with wearable devices for real-time BAC monitoring.
The euphoria from drinking alcoholic beverages makes them popular worldwide. Abuse in alcohol (ethanol) consumption leads to dependence, behavioral problems, and fatal accidents1. In 2013, 10,076 people lost their lives in alcohol-related-driving accidents in the United States alone, accounting for nearly 31% of all traffic related deaths1. Driving under the inuence is illegal and the maximum allowed blood alcohol concentration (BAC) is 0.050.08%, in most countries2. A breathalyzer measures the concentration of ethanol in the breath to estimate the BAC of an individual. The rst generation of breathalyzers uses a liquid dye sensitive to ethanol exposure, potassium dichromate, and a photodetector3. Reliability of these detectors is a challenge and potassium dichro-mate is environmentally toxic3. A new generation of breathalyzers uses the ethanol in the breath to power a fuel cell whose output is proportional to the ethanol concentration4. These breathalyzers are connected directly to smart phones to test ones alcohol level before attempting to drive. However, these breathalyzers are still impractical because they require constant recalibration4. To-date the most reliable BAC tests and the only one that is admissible in court is the blood test, which is difficult to administer on site or for preventative purposes. A breath-alyzer as easy to use as an inexpensive and disposable glucose paper-strip sensor would greatly simplify BAC testing5.
Organic electrochemical transistors (OECTs) are excellent candidates for disposable biosensors because they are inexpensive, they can be made on exible substrates, and they can be printed on paper6,7. OECTs are typically made with poly(3,4-ethylenedioxythiophene) doped with poly(styrenesulfonate) PEDOT:PSS. PEDOT:PSS is a p-type organic semiconductor with several applications in bioelectronics8,9. Coupled with enzymes in the electrolyte, PEDOT:PSS OECTs are able to detect micro molar glucose concentration in human blood and sweat10. OECTs sensors are amenable to screen-printing11 and inkjet printing12 for rapid and inexpensive manufacturing. Here, we demonstrate an early stage proof-of-concept OECT-breathalyzer on paper by integrating a PEDOT:PSS OECT with the enzyme alcohol dehydrogenase. This proof-of-concept OECT-breathalyzer may
Department of Bioelectronics, Ecole
Department of Materials Science and Engineering, University of Washington, Seattle, WA, *These
SCIENTIFIC REPORTS
1
www.nature.com/scientificreports/
Figure 1. Concept of the OECT-breathalyzer. (A) Simply breathing on the printed PEDOT:PSS OECT allows for alcohol detection. (B) The alcohol dehydrogenase (ADH) enzyme and the OECT are the key components of the sensor. The OECT is printed on paper, and comprises a channel, source (S), drain (D), and gate (G) electrodes made of PEDOT:PSS. The enzyme electrolyte gel is deposited to the OECT bridging the channel and gate. (C) Id response of the OECT upon exposure to ethanol.
aid the development of a breathalyzer that is easy-to-use, inexpensive, easily calibrated, and can be coupled with a cell-phone or a smart watch for BAC self-testing to reduce alcohol related traffic accidents.
Results
The OECT is printed on a paper substrate that is approximately 1.5cm 1cm and it is easy to handle. Breathing onto the device enables detection of ethanol concentration in the breath (Fig.1A). The printed OECT on paper has planar geometry with channel, source (S), drain (D), and gate (G) electrodes made of PEDOT:PSS (Fig.1B), a structure that is compatible with rapid, one-step fabrication of the device. We modify commercially available PEDOT:PSS to make it compatible with inkjet printing on paper (see Suppl. Info). A key challenge for the fabrication of an enzyme-based OECT sensor is the immobilization of the enzyme and its cofactor. In this work, the enzyme alcohol dehydrogenase (ADH) and its cofactor nicotinamide adenine dinucleotide (NAD+) are trapped in a collagen-based gel deposited onto the channel of the OECT (Fig.1B). Exposing the OECT-breathalyzer to ethanol from the breath causes a marked decrease of the source-drain current, Id, which is used as the output signal for detection (Fig.1C).
The decrease in Id upon ethanol exposure arises from a series of chemical reactions that oxidize ethanol into acetaldehyde and produce electrons as byproducts, which in turn decrease the conductivity of the OECT channel (Fig.2A).
SCIENTIFIC REPORTS
2
www.nature.com/scientificreports/
Figure 2. Alcohol detection in solution. (A) Enzymatic reaction of ethanol and ADH in electrolyte solution. Ethanol is oxidized to acetaldehyde, and NAD+ is reduced to NADH. (B) Step change in Id when adding 0.0004%, 0.004%, and 0.04% ethanol solutions, with Vg=0.5V, Vd=0.7V. (C) Variation of Id is plotted at dierent alcohol concentrations in solution. Id=(I0Id), where I0 is Id before exposure to ethanol. The dash line is a t to equation(3). n=3.
First the reaction:
3 2 3
yields the reduced form of nicotinamide adenine dinucleotide (NADH), which itself oxidizes according to13:
+ +
+ +
e
NADH NAD H 2 (2)
The electrons produced by the NADH oxidation are collected from the gate electrode of the OECT-breathalyzer (Fig.2A), and cause a shi of the applied gate potential to the channel/electrolyte interface, leading to a decrease in Id14. Id is written as14:
^
d m
where gm is the transconductance of the OECT and C is the concentration of the molecule contributing electrons to the gate (NADH in this case) (Figs S1 and S2). Since the concentration of NADH is directly related to the concentration of ethanol in the breath1517, this mechanism leads to quantitative ethanol detection. This was
ADH
+ +
CH CH OH NAD CH CHO NADH H (1)
+ + +
I g kT
2e ln [C] (3)
SCIENTIFIC REPORTS
3
www.nature.com/scientificreports/
Figure 3. Alcohol detection in breath. (A) Step change in Id when breathing 0%, 0.01%, 0.06%, 0.2% ethanol vapor on to the OECT, with Vg=0.5V, Vd=0.7V. Breath alcohol concentration is tested 30minutes aer drinking one glass of wine (0.01%) and two glasses of wine (0.06%). 0.2% breath ethanol vapor is from mouthwash, which contains 21.6% alcohol. The BAC is calibrated with a commercially available breathalyzer before testing on the device. (B) Normalized response (NR) of Id is plotted at dierent breath alcohol concentrations (BAC). NR=(I0Id)/I0. n=3.
conrmed by exposing the OECT-breathalyzer to a series of phosphate-buered saline (PBS) solutions containing ADH, NAD+, and dierent amounts of ethanol. The results, shown in Fig.2B, demonstrate that ethanol detection is achieved for concentrations as low as 0.0004%, and that higher ethanol concentration causes a bigger drop in Id. The Id drop occurs in seconds with an immediate enzyme response and steady state is reached in seconds. As a control, in absence of ADH, there is no response from the device upon addition of dierent ethanol concentrations (Fig. S3). We calibrate the sensor to dierent concentrations (Fig.2C). Higher ethanol concentration corresponds to higher Id as observed with NADH (Fig. S1). The response of the sensor is logarithmic with ethanol concentration, as expected from Eq.(3).
Finally, we demonstrate that the OECT-breathalyzer detects ethanol content in the breath of human subjects. For this demonstration, we use bovine gelatin to integrate the ADH and NAD+ onto the OECT, as breathing onto a liquid electrolyte causes excessive noise and makes packaging of the OECT-breathalyzer challenging. Ethanol in the breath is related to BAC by a factor of approximately 1/2100, which tends to vary with each individual18.
Seven volunteers participate in the breath alcohol test for BAC detection. One volunteer serves as the control, while other volunteers consume dierent amounts (120 ml and 240 ml) of red wine (Les 3 lles, 2014, Merlot, 13% alcohol content). The experiment has been reproduced for each amount of wine with 3 volunteers. Thirty minutes aer wine consumption, the volunteers are subject to a breath test with a commercial breathalyzer (BreathometerTM) for calibration and then are subject to the same test with the OECT-breathalyzer. Subsequently, the volunteer who serves as the control rinses their mouth with mouthwash (Listerine, 21.6% ethanol content)
and immediately takes another set of breath tests. The results from the two tests are compared for accuracy. The test from the volunteer serving as control (0% BAC) results in no response from the OECT-breathalyzer. In contrast, when the volunteers who have consumed wine take the test aer one glass (0.01% BAC) and two glasses (0.06% BAC), an immediate response in Id of the OECT-breathalyzer is observed (Fig.3A). This response scales with the amount of alcohol consumed. The test corresponding to mouthwash registers an apparent BAC of 0.2%, consistent with the high alcohol concentration in this solution. (Fig.3B). Similarly to commercial breathalyzers, the OECT-breathalyzer does not seem to be aected by variation in breathing time from dierent volunteers as indicated by the small spread in the data for the NR (Fig.3B). We suggest that as long as a volunteer breathes on the OECT-breathalyzer long enough, the ethanol concentration on the device will equilibrate with the ethanol concentration in the breath. This proof-of-concept demonstrates that this simple OECT-breathalyzer is able to detect BAC in human subjects with performance comparable with a commercial breathalyzer.
Discussion and Conclusions
We demonstrate the rst alcohol sensor made with an organic electrochemical transistor integrated with the enzyme alcohol dehydrogenase and its cofactor. This OECT-breathalyzer is easy to fabricate using printing techniques, requires no metal deposition, and is made on an inexpensive, disposable, and biodegradable paper support. The use of a gel to immobilize the enzyme and its cofactor makes the device robust and easy to use. We show that the OECT-breathalyzer detects ethanol in both solution and vapor (such as breath). We conduct preliminary tests with a limited pool of human volunteers and compare the performance of the OECT-breathalyzer with the performance of a commercial breathalyzer. The OECT-breathalyzer is able to detect the consumption of just one glass of red wine. For further optimization, studies with a larger pool of human volunteers are required. This work may help develop alcohol sensors that are easy to integrate with portable/wearable electronics such as smartphones and smart watches. These devices could, in turn, prevent a vehicle to start if they detect the driver to be under the inuence and therefore reduce alcohol-associated traffic accidents.
SCIENTIFIC REPORTS
4
www.nature.com/scientificreports/
Materials and Methods
PEDOT:PSS ink. The PEDOT:PSS ink consists of the commercially available PEDOT:PSS (Heraeus, Clevios PH1000) dispersion with 20wt% ethylene glycol (Sigma Aldrich) and a combination of organic solvents. We add 0.8wt% glycidoxypropyltrimethoxysilane (GOPS, Sigma Aldrich) to the ink to prevent delamination, and 0.3% surfactants to match the surface tension of the ink with the substrate.
Ink-jet printing. We use a Dimatix DMP-2800 inkjet printer to print the OECT onto a coated paper (Arjo Wiggins, Inc.). In this study, two layers of PEDOT:PSS are deposited for a total thickness of 190nm. The printed device is cured in a conventional oven at 160 C for 30 min. The dimensions of the channel are 1 5 mm2 and 25mm2 for the gate (Fig.1). For the alcohol solution measurements, 9mM NAD+ (Sigma Aldrich) are mixed in 0.1M standard phosphate buer solution (PBS, Sigma Aldrich). For each measurement, 1.5mg/mL ADH (Sigma Aldrich) in PBS is added to the in NAD+ and PBS mix at a 1:10 ratio of the total volume. The pH of the electrolyte is adjusted to 8.2 and measured with a pH meter to meet the requirements of the enzyme. A PDMS well is attached to the OECT to conne the electrolyte, dening an active device area of 44mm2. The well is lled with 20 L electrolyte. For the alcohol vapor measurements, we formulate a 2wt% bovine gelatin (Sigma Aldrich) in PBS, containing the same proportion of ADH and NAD+ as in the liquid electrolyte above. 15L gel solution is drop-casted onto the device and cured at 4C for 30min to form gel.
OECT Electrical Characterization. Electrical characterization is conducted with an Agilent 4155C semiconductor parameter analyzer. During the experiment, Vg varies from 0 to 0.7V and the Vd from 0.8 to 0V. For the measurements, we apply constant Vg= 0.5V and Vd=0.7V respectively. So carbon electrodes connect the OECT contacts and the Agilent. We prepare dierent concentrations of ethanol solution in DI water: 0.0004wt%, 0.004 wt%, 0.008 wt%, 0.01 wt%, 0.03 wt%, 0.04 wt%, 0.08 wt%, 0.1 wt%, 0.12 wt%, 0.15 wt% and measure the response of the device for each concentration three times. During the alcohol solution tests (Fig.2), we rst wait for 120 s until the source drain current (Id) stabilizes aer the application of bias on the OECT, and then add 2L of ethanol solution to the electrolyte at a 1:10 ratio. For the BAC tests (Fig.3), we replace the electrolyte solution with gel on top of the OECT. Volunteers consuming dierent quantities of alcoholic beverages rst have their BAC tested with a commercial breathalyzer: The Original Breathometer (Breathometer, Inc.), and then breathe onto the OECT-breathalyzer for a certain duration. The experiment has been tested three times for each concentration with volunteers. The methods were carried out in accordance with the approved guidelines, and all experimental protocols were approved by the direction of research of the Ecole des Mines de St. Etienne. All volunteers provided informed signed consent to participate in the study.
References
1. Park, K. K. et al. Capacitive micromachined ultrasonic transducers for chemical detection in nitrogen. Appl. Phys. Lett. 91, 094102 (2007).
2. Steven, B. & Karch, M. D. F. Drug Abuse Handbook, Second Edition. 1129 (CRC Press, 2006).3. U.S. Environmental Protection Agency, Potassium Dichromate Listing Background Document For The Inorganic Chemical Listing Determin Ation. Date of access: 07/09/2015. Link: http://www3.epa.gov/epawaste/hazard/wastetypes/wasteid/inorchem/docs/potdich.pdf.
4. Breathometer Inc., Calibration Service, Date of access: 07/09/2015. Link: https://www.breathometer.com/calibration.5. Cui, G. et al. Disposable amperometric glucose sensor electrode with enzyme-immobilized nitrocellulose strip. Talanta 54, 11051111 (2001).
6. Strakosas, X., Bongo, M. & Owens, R. M. The organic electrochemical transistor for biological applications. J. Appl. Polym. Sci. 132, 41735 (2015).
7. Lin, P., Yan, F., Yu, J., Chan, H. L. & Yang, M. The Application of Organic Electrochemical Transistors in Cell-Based Biosensors. Adv. Mater. 22, 36553660 (2010).
8. Rivnay, J., Owens, R. M. & Malliaras, G. G. The rise of organic bioelectronics. Chem. Mater. 26, 679685 (2013).9. Berggren, M. & Richter-Dahlfors, A. Organic bioelectronics. Adv. Mater. 19, 32013213 (2007).10. Scheiblin, G. et al. Screen-printed organic electrochemical transistors for metabolite sensing. MRS Communications 5, 507511 (2015).
11. Andersson Ersman, P., Nilsson, D., Kawahara, J., Gustafsson, G. & Berggren, M. Fast-switching all-printed organic electrochemical transistors. Org. Electron. 14, 12761280 (2013).
12. Mannerbro, R., Ranlf, M., Robinson, N. & Forchheimer, R. Inkjet printed electrochemical organic electronics. Synth. Met. 158, 556560 (2008).
13. Kitagawa, Y., Kitabatake, K., Kubo, I., Tamiya, E. & Karube, I. Alcohol Sensor Based on Membrane-Bound Alcohol-Dehydrogenase. Anal Chim Acta 218, 6168 (1989).
14. Bernards, D. A. et al. Enzymatic sensing with organic electrochemical transistors. J. Mater. Chem. 18, 116120 (2008).15. Shan, C. et al. Electrochemical determination of NADH and ethanol based on ionic liquid-functionalized graphene. Biosens. Bioelectron. 25, 15041508 (2010).
16. Raj, C. R. & Behera, S. Mediatorless voltammetric oxidation of NADH and sensing of ethanol. Biosens. Bioelectron. 21, 949956 (2005).
17. Ramesh, P., Sivakumar, P. & Sampath, S. Renewable surface electrodes based on dopamine functionalized exfoliated graphite: NADH oxidation and ethanol biosensing. J.Electroanal. Chem. 528, 8292 (2002).
18. Alobaidi, T. A. A., Hill, D. W. & Payne, J. P. Signicance of Variations in Blood-Breath Partition-Coefficient of Alcohol. Brit Med J.
2, 14791481 (1976).
Acknowledgements
We are grateful to Scott Keene, Zahra Hemmatian, Eloise Loen, Esma Ismailova, Yi Zhang, Thomas Lonjaret, Paschalis Gkoupidenis, and MicroVitae Technologies for help with some of the experiments and background research for this work. The work is supported by the National Science Foundation CAREER Award # DMR-1150630, the PolyMed project within the FP7-PEOPLE-2013-IRSES framework, PIRSES-GA-2013-612538, and a fellowship from the Japan Society for the Promotion of Science (T.M.).
SCIENTIFIC REPORTS
5
www.nature.com/scientificreports/
Author Contributions
M.R. and G.G.M. coordinated the research. E.B., Y.D. and T.M. designed and performed the experiment. M.S. provided materials. E.B. fabricated the device. E.B., Y.D. and T.M. analyzed the data. Y.D. and E.B. and Y.D. wrote the manuscript with help from all the authors. All the authors revised the manuscript.
Additional Information
Supplementary information accompanies this paper at http://www.nature.com/srep
Competing nancial interests: The authors declare no competing nancial interests.
How to cite this article: Bihar, E. et al. A Disposable paper breathalyzer with an alcohol sensing organic electrochemical transistor. Sci. Rep. 6, 27582; doi: 10.1038/srep27582 (2016).
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the articles Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
SCIENTIFIC REPORTS
6
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Nature Publishing Group Jun 2016
Abstract
Breathalyzers estimate Blood Alcohol Content (BAC) from the concentration of ethanol in the breath. Breathalyzers are easy to use but are limited either by their high price and by environmental concerns, or by a short lifetime and the need for continuous recalibration. Here, we demonstrate a proof-of-concept disposable breathalyzer using an organic electrochemical transistor (OECT) modified with alcohol dehydrogenase (ADH) as the sensor. The OECT is made with the conducting polymer poly(3,4-ethylenedioxythiophene):poly(styrenesulfonate) (PEDOT:PSS), and is printed on paper. ADH and its cofactor nicotinamide adenine dinucleotide (NAD+ ) are immobilized onto the OECT with an electrolyte gel. When the OECT-breathalyzer is exposed to ethanol vapor, the enzymatic reaction of ADH and ethanol transforms NAD+ into NADH, which causes a decrease in the OECT source drain current. In this fashion, the OECT-breathalyzer easily detects ethanol in the breath equivalent to BAC from 0.01% to 0.2%. The use of a printed OECT may contribute to the development of breathalyzers that are disposable, ecofriendly, and integrated with wearable devices for real-time BAC monitoring.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer




