PHOTONIC SENSORS / Vol. 6, No. 3, 2016: 214220
Study of a Distributed Feedback Diode Laser Based Hygrometer Combined Herriot-Gas Cell and Waterless Optical Components
Yubin WEI1,2, Jun CHANG1*, Jie LIAN1, Qiang WANG1, and Wei WEI1
1School of Information Science and Engineering and Shandong Provincial Key Laboratory of Laser Technology and
Application, Shandong University, Jinan, 250100, China
2Laser Research Institute of Shandong Sciences Academy, Jinan, 250014, China
*Corresponding author: Jun CHANG E-mail: [email protected]
Abstract: A distributed feedback diode laser (DFB-DL) based hygrometer combined with a long-path-length Herriot gas cell and waterless optical components was proposed and investigated. The main function of this sensor was to simultaneously improve the measurement reliability and resolution. A comparison test between a 10-cm normal transmission-type gas cell and a 3-m Herriot gas cell was carried out to demonstrate the improvement. Reliability improvement was achieved by influence suppression of water vapor inside optical components (WVOC) through combined action of the Herriot gas cell and waterless optical components. The influence of WVOC was suppressed from 726 ppmv to 25 ppmv using the Herriot gas cell. Moreover, combined with waterless optical components, the influence of WVOC was further suppressed to no more than 4 ppmv. Resolution improvement from 11.7 ppmv to 0.32 ppmv was achieved mainly due to the application of the long-path-length Herriot gas cell. The results show that the proposed sensor has a good performance and considerable potential application in gas sensing, especially when probed gas possibly permeates into optical components.
Keywords: Herriot gas cell; hygrometer; wavelength modulation spectroscopy; waterless optical components; water vapor inside optical components
Citation: Yubin WEI, Jun CHANG, Jie LIAN, Qiang WANG, and Wei WEI, Study of a Distributed Feedback Diode Laser Based Hygrometer Combined Herriot-Gas Cell and Waterless Optical Components, Photonic Sensors, 2016, 6(3): 214220.
1. Introduction
Reliable and high-resolution measurement of gases of interest becomes more and more important in the field of scientific research and facility manufacture. Particularly, trace water vapor qualification analysis is indispensable in a wide of applications, ranging from polymer fabrication [1], environment analysis [2] to chemical production [3]. Because of remote detection capability, safety in
hazardous environments, and immunity to electromagnetic radiation, the distributed feedback diode laser (DFB-DL) based hygrometer attracted the widest applications. In the recent years, many techniques were developed to improve the measurement reliability and resolution. For example, the utilization of differential value of two adjacent absorption peaks solved the difficulty of choosing reference point [4]. The utilization of combination of
Received: 19 January 2016 / Revised: 1 June 2016 The Author(s) 2016. This article is published with open access at Springerlink.com DOI: 10.1007/s13320-016-0320-1Article type: Regular
Yubin WEI et al.: Study of a Distributed Feedback Diode Laser Based Hygrometer Combined Herriot Gas Cell and Waterless Optical Components
215
wavelength scanning and intensity modulation significantly improved the detection resolution [5]. The application of wavelength modulation spectroscopy (WMS) could achieve high resolution detection of trace gases at atmospheric pressure [6]. The utilization of the improved WMS technique achieved rapid measurement of water vapor at high temperature [7].
Almost all of DFB-DL based hygrometers consist of several optical components, such as light source (DFB-DL), fiber collimators, and photoelectric detector. However, it was found in [8] that water vapor in these optical components seriously influenced the measurement result not only in resolution but also in reliability. To realize reliable and high-resolution measurement of water vapor, an approach, namely the photo detector (PD) matching, had been reported to achieve preliminary suppression for the influence of water vapor inside optical components (WVOC). However, it is found in practical application the suppression effectiveness of the above approach depends on the match level between the two PDs, which is tedious and not guaranteed.
In this paper, an alternative approach consisting of Herriot cell and waterless optical components is investigated. It can observably improve the reliability and resolution simultaneously. The influence of WVOC can be diluted by introducing a Herriot cell with an effective length of 3 meters. Here we describe the experimental system and the mathematical justification of this approach and demonstrate the validity through exhibiting the improved measurements made by this approach.
2. Experimental details
Figure 1 shows the experimental system. A DFB-DL (with a linewidth of 3 MHz) is used as the light source. The inside temperature is controlled by a chip LTC 1923. The driving current consists of a 20-Hz sawtooth waveform and a 2-kHz sinusoidal waveform, and it is produced by a home-built signal
generator using a chip LPC 1758. The sawtooth waveform ranges from 23 mA to 78 mA. As a result, a sweep range of about 280 pm is given to scan the absorption line of water vapor at 7306.75 cm1. At the same time, the sinusoidal current is used to accomplish wavelength modulation of the DFB-DL. The output of the DFB-DL propagates through a gas cell to interact with water vapor molecules. To introduce a long optical absorption path, a multiple-pass Herriot cell is used. The 34 traversals inside the Herriot cell give a path length of 3 meters. The long optical path length not only improves the effective absorbance during the interaction, which is beneficial to improve the measurement resolution but also dilute the influence of WVOC, which is very important to improve the measurement reliability. After the interaction, the laser beam is detected by an InGaAs PD whose output is fed to RS850 digital lock-in amplifier (LIA). To implement the WMS technique, the second harmonic detection is performed with the LIA. To avoid variation of the phase difference between the WMS derivative signal and reference signal, the reference signal is synchronously generated by the same home-built signal generator. The final output is observed and captured by a computer.
TEC control
Herriott cell
PD
DFB diode laser
LIA
Driving current
Fig. 1 Experimental setup.
3. Analysis of reliability and resolution improvement
The description of laser beam attenuation due to wavelength dependent absorption for a gas absorption line is given by Beer-Lamberts law [9], It(v)=I0(v)exp[(v)CL], where I0(v) and It(v) are the
Photonic Sensors
216
incident and transmitted laser beam intensities, respectively, (v) is the wavelength independent absorption coefficient, C is the mole fraction concentration, and L is the optical path length. For WMS, the simultaneous frequency and amplitude of the modulated laser beam can be represented by
2 0 2 , cos
p p p
I I H
(4)
where vp and I0p donate the frequency and intensity at the absorption peak. To simplify the expression of the relationship between the detected second harmonic amplitude and absorbance, (4) can be rewritten as
I CL
. (5)
The above expression is applicable only when the absorbance is not too high. For the detection of water vapor at 7306.75 cm1 in this research, the above expression is applicable when C is no more than 2000 ppmv. in (5) designates the conversation coefficient, which is relative to the beam intensity at the absorption peak, values of and , and Fourier component H2.
According to the analysis in [8], the water vapor inside the optical components can influence the measurement result. Combined with (5) in the WMS technique, the influenced expression of the water vapor concentration should be
t t
(1)
cos
c
2 p p
I t I i t
(2)
where v(t) is the instantaneous optical frequency, I0(t)
is the laser emission intensity, and is the phase separation [10] between the intensity and optical frequency modulation. vc and I0c donate the slowly-varying central frequency and intensity, as the laser scans across the absorption features. is the sinusoidal modulation frequency, v and i0 are the maximum small-amplitude excursions of the v(t) and I0(t) around vc and I0c.
In the implementation of the WMS technique, the signals at the output can be found using Fourier cosine series expansion of the absorbance line shape function, (v)CL [11]. The result for the primary second harmonic term is shown as
= cos
c
0 0 0
I L
C C
L
L
+ (6)
where Leq and Ceq donate the equivalent absorption length and concentration of the water vapor inside the optical components. An extended optical absorption length can simultaneously improve the measurement reliability and resolution. Firstly, during the practical application, the detected harmonic signal is directly used to invert the wanted water vapor concentration. However, system noises (optical noise and electrical noise) are accumulated and added in the detected harmonic signal. Generally speaking, the system noises have nothing to do with the absorption length. According to (6), for the same optical gas sensor, the measurement resolution can be significantly improved using a longer optical absorption length. Secondly, the influence of the water vapor inside the optical components is relatively certain and invariable. Similarly, the measurement reliability can be improved by weakening the influence of the water
2 eq
eq
p
p
i
0
, cos 2
2
, cos , cos
2
I H
i
I H H
c c
c c c
2 3
(3)
0
0 2 1
where H1, H2, and H3 denote three Fourier components obtained for the absorbance line shape function. designates the phase difference between the detection signal and the laser emission intensity. Although the components of H3 and H1 on the measurement axis of H2 can distort the detected profile of the detected second harmonic term [12], the amplitude of the detected second harmonic signal at the line center is immune to the components of H3 and H1. This is because the amplitudes of H3 and H1 are zero at the absorption peak [11]. Thus the amplitude of the detected second harmonic (i.e. the value of the detected second harmonic signal at the absorption peak) can be expressed as
Yubin WEI et al.: Study of a Distributed Feedback Diode Laser Based Hygrometer Combined Herriot Gas Cell and Waterless Optical Components
217
vapor inside the optical components.
4. Experimental results
4.1 Existence of WVOC
To demonstrate the severity of existence of the water vapor inside the optical components, a simple test was organized. The simple schematic of the test is shown in Fig. 2(a). Laser light from DFB-LD directly couples on the PD. The most distinctive characteristic is there is no gas cell in the test. The process of the detected signal is the same with the introduction above.
PD
DFB diode laser
Reference
(a)
(a)
50
40
30
20
10
Fiber collimators
Stainless steel tube
(a)
Fiber collimators
Dielectric film Quartz lenses
0
-10
-20
-30
-40
-50
0 10 20 30 40 50
Quartz rod
Time (ms)
(b)
Fig. 2 Test of existence of the water vapor inside the optical components: (a) test setup and (b) measured second harmonic signal without the gas cell.
Based on the traditional understanding, there should be no absorption during this setup. However, the detected result shown in Fig. 2 (b) subverts our preliminary cognition. Not only the existence of WVOC badly impacts the measurement result during the practical application of an optical fiber
hygrometer, but also variation of the inside water vapor can further impact the measurement.
4.2 Reliability and resolution improvement
To fully exhibit the influence of the water vapor inside the optical components especially the improvement of reliability and resolution, a comparison research was demonstrated experimentally between the normal transmission-type gas cell with a 10-cm effective path length [8] and Herriot gas cell with a 3-m effective path length. Figure 3(a) shows the structure of the normal gas cell comprising two commercial fiber collimators used in communication. One collimator is used to transfer laser beam into parallel light inside the gas cell, and the parallel light interacts with the probed gas molecules and possesses information of characteristics associated with the probed gas molecules. The other one is used to receive the signal light and couple it into the fiber. The two fiber collimators are fixed on a stainless steel tube to guarantee reliable alignment and achieve low coupling loss (less than 0.5 dB). Holes of the stainless steel tube are used the realize gas exchange.
(b)
Fig. 3 Schematic of gas cell structure: (a) normal transmission-type gas cell with an effective path length of 10 cm and (b) Herriot gas cell with an effective path length of 3 m, the two quartz lenses are coated with the dielectric film.
Photonic Sensors
218
Figure 3(b) shows the structure of the Herriot cell. Two quartz lenses with a diameter of 25.4 mm are fixed with a straight column quartz rod. The two quartz lenses coated with the dielectric film have the same curvature radius of 100 mm and are separated with a distance of about 90 mm. A fiber collimator with a diameter of 1.3 mm is used to introduce light into the Herriot gas cell. After adjusting the incidence angle, 17 reflections exist for each lens. Another identical fiber collimator is used to collect the signal light and couple it into fiber. As a result, the effective optical path length has been extended to about 3 meters, and the insertion loss is lower than
linear fitting process slope of solid line is calculatedto be 0.11 mV/ppmv, thus the intercept of solid lineis 79.9 mV corresponding to a water vapor concentration of 726 ppmv in a 10-cm path length. While the slope of dot line is calculated to be 3.38 mV/ppmv, and the intercept of solid line is 85.1 mV corresponding to a water vapor concentration of 25 ppmv in a 3-m path length. Apparently, the influence of WVOC on the measured probed water vapor concentration has been significantly suppressed.
Fig. 4 Reliability improvement test through WVOC suppression (the lower line is for a 10-cm transmission-type gas cell, and the upper line is for a 3-m Herriot gas cell).
As for the resolution improvement test, the resolution [with signal to noise ratio (SNR) =1] is obtained through deducing the SNR. Given the existence of WVOC, the deducing process is not based on a certain water vapor concentration. On the contrary, the signal amplitude in SNR calculation is derived from two different amplitudes (in Fig. 4) corresponding to two water vapor concentrations, which are measured by S8000. The noise amplitude is derived from the WMS profile corresponding to one of the two water vapor concentrations shown in Fig. 5.
From the above test and process, the resolution of measurement using the normal transmission-type gas cell is calculated to be 11.7 ppmv, and that of measurement using the Herriot gas cell is 0.32 ppmv. Thus reliability can obtain significant improvement using a long path length gas cell.
dB.
During the comparison research, the measurement reliability and resolution were studied. The reliability improvement reflected in influence suppression of WVOC. A linearity test was arranged.
A water vapor generator (DG-70, Shanghai Hema, China) was used to generate sample gases with different water vapor concentrations ranging from20 ppmv to 430 ppmv. To change the inside water vapor concentration, sample gases were introduced into the gas cell at a flow rate of 100 mL/min at
atm and (296 0.5) K. The overflowing sample gas from gas cell was used to monitor the watervapor concentration by a moisture meter (S8000, MICHELL, UK). Precision of S8000 is within 1 ppmv in the range of 1 ppmv 2000 ppmv.
The measurement result is shown in Fig. 4. Amplitude (in mV) is proportional to the water vapor concentration. The solid line named normal transmission-type gas cell is the linearity test result using the 10-cm normal transmission-type gas cell, and the dot line named Herriot gas cell is the linearity test result using the 3-m Herriot gas cell. It can be seen from Fig. 4 that both the lines do not pass through the origin of coordinates (0, 0), and this is caused by the influence of WVOC. That is to say influence of WVOC can be reflected by the intercepts of the measured linearity lines. After the
Yubin WEI et al.: Study of a Distributed Feedback Diode Laser Based Hygrometer Combined Herriot Gas Cell and Waterless Optical Components
219
To obtain the efficiency of water vapor suppression using these waterless optical components, a test was carried out, and the test result is shown in Fig. 6 (a). It is apparent that the intercept of the linearity test has been suppressed to 10.9 mV corresponding to no more than 4 ppmv. Figure 6(b) shows the detected LIA output without the gas cell and a result of fitting procedure based on the Fourier analysis of the Lorentzian profile.
LIA output (mV)
1400
Linearity test using waterless
1200
(a)
1000
Y(mV)=3.19
LIA output (mV)
800
600
400
200
LIA output (mV)
0
0 100 200 300 400 500
Water vapor concentration (ppmv)
(a)
4
Mesured data of a single scan
2f fitting of data
3
2
1
(b)
0
Fig. 5 Detected WMS profiles: (a) corresponds to the normal transmission-type gas cell with a water concentration of 94 ppmv and (b) corresponds to the Herriot gas cell with a water concentration of 92 ppmv.
4.3 Further improvement using waterless optical components
Although a long optical path length gas cell can apparently suppress the influence of WVOC, there still exists an apparent influence, which is shown in Fig. 4. Thus waterless optical components were designed and adopted to further suppress influence of WVOC. A 14-pin butterfly packaged DFB-DL and a PD with the same package were used. The big difference between the two waterless optical components and normal ones is that the inside space is filled of high purity nitrogen. As a result, inside water vapor content could be rather low.
-1
-2
-3
-4
0 10 20 30 40 50
Time (ms)
(b)
Fig. 6 Further suppression results for influence of WVOC using waterless optical components: (a) linearity test and (b) situation of no gas cell.
5. Conclusions
We proposed and demonstrated a technique comprising Herriot Cell waterless optical components. Reliability and resolution of the DFB-DL based hygrometer can be simultaneously improved by this technique. A comparison test between a 10-cm normal transmission-type gas cell
Photonic Sensors
220
and this 3-m Herriot gas cell was carried out. As a result, the influence of WVOC had been suppressed from 726 ppmv to 25 ppmv. Moreover, waterless optical components were used to further suppress the influence of WVOC, and the influence had been further suppressed to be no more than 4 ppmv. In addition, the measurement resolution had been improved from 11.7 ppmv to 0.32 ppmv.
In summary, the reliability improvement had been achieved through suppressing the influence of WVOC using a long path length Herriot gas cell and waterless optical components. The resolution improvement had been achieved through enhancing absorbance using a long path length Herriot gas cell. The proposed sensor has potential applications in the design of the reliable and high-resolution gas sensor, especially when permeation of probed gas into optical components exists.
Acknowledgment
This work was supported by the National Natural Science Foundation of China (60977058 & 61475085), the Science and Technology Development Project of Shandong Province (2014GGX101007), and the Fundamental Research Funds of Shandong University (2014YQ011).
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/ licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Applications to Chemical and Environmental Analysis, Santa Fe, New Mexico United States, 2000. [3] S. Wu, F. Deng, Y. Hao, M. Shima, X. Wang, C. Zheng, et al., Chemical constituents of fine particulate air pollution and pulmonary function in healthy adults: the healthy volunteer natural relocation study, Journal Hazardous Materials, 2013, 260C(6): 183191. [4] Q. Wang, J. Chang, C. G. Zhu, C. Li, F. J. Song, Y. N. Liu, et al., Detection of water vapor concentration based on differential value of two adjacent absorption peaks, Laser Physics Letters, 2012, 9(6): 421425. [5] Q. Wang, J. Chang, Z. Wang, C. Tian, S. Jiang, and G. Lv, Study of an optical fiber water vapor sensor based on a DFB diode laser: combined wavelength scanning and intensity modulation, Journal of Modern Optics, 2014, 61(18): 15381544. [6] A. Upadhyay and A. L. Chakraborty, Residual amplitude modulation method implemented at the phase quadrature frequency of a 1650-nm laser diode for line shape recovery of methane, IEEE Sensors Journal, 2015, 15(2): 11531160.
[7] H. Li, A. Farooq, J. B. Jeffries, and R. K. Hanson, Near-infrared diode laser absorption sensor for rapid measurements of temperature and water vapor in a shock tube, Applied Physics B, 2007, 89(2): 407416.
[8] Q. Wang, J. Chang, F. J. Song, F. P. Wang, C. G. Zhu, Z. Liu, et al., Measurement and analysis of water vapor inside optical components for optical fiber H2O sensing system, Applied Optics, 2013, 52(26): 64456451.
[9] X. Liu, J. B. Jeffries, and R. K. Hanson, Measurements of spectral parameters of water-vapour transitions near 1388 and 1345 nm for accurate simulation of high-pressure absorption spectra, Measurement Science and Technology, 2007, 18(5): 11851194.
[10] X. Zhu and D. T. Cassidy, Modulation spectroscopy with a semiconductor diode laser by injection-current modulation, Journal of the Optical Society of America B, 1997, 14(14): 19451950.
[11] L. C. Philippe and R. K. Hanson, Laser diode wavelength-modulation spectroscopy for simultaneous measurement of temperature, pressure, and velocity in shock-heated oxygen flows, Applied Optics, 1993, 32(30): 60906103.
[12] K. Duffin, A. J. McGettrick, W. Johnstone, G. Stewart, and D. G. Moodie, Tunable diode laser spectroscopy with wavelength modulation: a calibration-free approach to the recovery of absolute gas absorption line shapes, Journal of Lightwave Technology, 2007, 25(10): 3114 3125
.
References
[1] W. Zhang, D. J. Webb, and G. D. Peng, Investigation into time response of polymer fiber Bragg grating based humidity sensors, Journal of Lightwave Technology, 2012, 30(8): 10901096.
[2] J. Goldmeer and D. J. Kane, Real-time 2-D imaging of water vapor in diffusion flames, in Laser
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
UESTC and Springer 2016
Abstract
A distributed feedback diode laser (DFB-DL) based hygrometer combined with a long-path-length Herriot gas cell and waterless optical components was proposed and investigated. The main function of this sensor was to simultaneously improve the measurement reliability and resolution. A comparison test between a 10-cm normal transmission-type gas cell and a 3-m Herriot gas cell was carried out to demonstrate the improvement. Reliability improvement was achieved by influence suppression of water vapor inside optical components (WVOC) through combined action of the Herriot gas cell and waterless optical components. The influence of WVOC was suppressed from 726 ppmv to 25 ppmv using the Herriot gas cell. Moreover, combined with waterless optical components, the influence of WVOC was further suppressed to no more than 4 ppmv. Resolution improvement from 11.7 ppmv to 0.32 ppmv was achieved mainly due to the application of the long-path-length Herriot gas cell. The results show that the proposed sensor has a good performance and considerable potential application in gas sensing, especially when probed gas possibly permeates into optical components.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer