About the Authors:
Julian Kirchner
Contributed equally to this work with: Julian Kirchner, Lale Umutlu
* E-mail: [email protected]
Affiliation: Department of Diagnostic and Interventional Radiology, University Dusseldorf, Medical Faculty, Dusseldorf, Germany
Lino Morris Sawicki
Affiliation: Department of Diagnostic and Interventional Radiology, University Dusseldorf, Medical Faculty, Dusseldorf, Germany
Saravanabavaan Suntharalingam
Affiliation: Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Johannes Grueneisen
¶‡ These authors also contributed equally to this work
Affiliation: Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Verena Ruhlmann
Affiliation: Department of Nuclear Medicine, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Bahriye Aktas
Affiliation: Department of Gynecology and Obstetrics, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Cornelius Deuschl
¶‡ These authors also contributed equally to this work
Affiliation: Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Ken Herrmann
Affiliation: Department of Nuclear Medicine, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Gerald Antoch
Affiliation: Department of Diagnostic and Interventional Radiology, University Dusseldorf, Medical Faculty, Dusseldorf, Germany
Michael Forsting
Affiliation: Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Lale Umutlu
Contributed equally to this work with: Julian Kirchner, Lale Umutlu
Affiliation: Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany
Introduction
Hybrid imaging in terms of positron emission tomography/computed tomography (PET/CT) has been successfully introduced into oncologic staging and restaging of numerous tumor entities over the last years [1–3]. Due to the combination of metabolic information, derived from the PET-component, and morphologic data, based on whole-body CT, 18F-FDG PET/CT has been shown superior to conventional cross-sectional imaging modalities for detection of tumor lesions, in particular for (recurrent) pelvic malignancies [4–6]. However, apart from its superior diagnostic capability, 18F-FDG PET/CT is affiliated with significantly increased radiation exposure, when compared to conventional imaging [7]. The implementation of integrated positron emission tomography/magnetic resonance imaging (PET/MRI) offers the opportunity to reduce radiation exposure, while preserving high-quality staging of cancer patients. In addition, high soft-tissue contrast of MRI may be beneficial in some pelvic indications. Currently available data demonstrate an equal diagnostic accuracy in staging of oncologic patients [8–10], particularly in recurrent malignancies of the female pelvis [11,12]. Despite initial approaches [13] to shorten the examination time of PET/MRI studies, the markedly prolonged examination time, caused by the acquisition of a variety of MR sequences, still constitutes a major disadvantage of this modality and restricts the implementation into clinical imaging. Hence, careful selection of the needed MRI sequences to a necessary minimum combined with optimizing PET image acquisition protocols as described in previous studies [14] is a promising concept.
Thus, the objective of this study was to evaluate the diagnostic feasibility of an ultra-fast protocol for 18F-FDG PET/MRI, including high-resolution morphological (T1w- and T2w-sequences) and metabolic assessment (PET) compared to 18F-FDG PET/CT and CT for whole-body staging of female patients with suspected recurrence of pelvic malignancies. Additionally, the radiation exposure in the different modalities was investigated and compared.
Materials & methods
Patients
The study was conducted in conformance with the Declaration of Helsinki and approved by the Ethics Commission of the Medical Faculty of the University Duisburg-Essen (study number 11-4822-BO). All patients underwent a clinically indicated whole-body PET/CT and subsequently an additional whole-body PET/MRI after informed written consent was obtained.
Altogether, 81 female patients were enrolled in this trial. All participants were suspect for a recurrence of pelvic malignancy, based on clinical follow-up (pelvic pain, rise in tumor marker in case of ovarian cancer) as well as abnormal findings on imaging follow-up (Ultrasound, CT, MRI). 43 patients (mean age 55 ± 11 years; range 25–79 years) patients met the inclusion criteria of disease recurrence confirmed either by histopathology or corresponding cross-sectional imaging follow-up and were selected for further analysis.
PET/CT
Whole-body 18F-FDG PET / CT was performed on a Biograph mCT 128 scanner (Siemens, Healthcare GmbH, Erlangen, Germany) after intravenous injection of body-weight adapted mean activity of 234 MBq ± 32 MBq (6.32 ± 0.86 mCi) 18F-FDG. To ensure blood glucose levels below 150 mg/dl blood samples were obtained. Whole-body full-dose scans (from skull base to mid thigh) were performed in caudo-cranial scan direction 70 s after 100ml iodinated contrast medium was intravenously administered. The CT component of 18F-FDG PET/CT (CTPET/CT) was acquired using the manufacturer-supplied dose reduction software CareKV and CareDose 4D (Siemens Healthcare GmbH, Erlangen, Germany) with presets of 120 kV, 210 mAs, an increment of 5 mm, a pitch of 1 and a slice thickness of 5 mm.
PET data acquisition was performed in 5–7 bed positions (2 min per bed position, axial field of view (FOV): 21.8 cm, matrix size 256 x 256 and a Gaussian filter of 4mm Full Width at Half Maximum (FWHM)). Iterative reconstruction (3 iterations and 24 subsets) was applied. Attenuation correction was calculated based on obtained CT datasets.
PET/MRI
Whole-body 18F-FDG PET/MRI examinations were performed on an integrated 3 Tesla PET/MRI scanner (Biograph mMR, Siemens Healthcare GmbH, Erlangen, Germany) and obtained with an average delay of 150 ± 47 min after 18F-FDG injection. No additional tracer was injected for the subsequent PET/MRI examinations. Corresponding to PET/CT scan volumes covered skull base to mid-thigh.
PET data acquisition was performed in 4–5 bed positions with a median of 4 min per bed position. PET images were reconstructed using the iterative ordered-subset expectation maximization (OSEM) algorithm, 3 iterations and 21 subsets, a Gaussian filter with 4mm full width at half maximum (FWHM) and a 344 × 344 image matrix. For MR-based attenuation correction a two-point (fat, water) coronal 3D-Dixon-VIBE sequence was performed to generate a four-compartment model (background air, lungs, fat, muscle). MRI data were obtained simultaneously using a 16-channel head-and-neck radiofrequency (RF) coil, a 24-channel spine-array RF coil as well as three-to-five 6-channel flex body-coils, depending on the patient size.
To investigate a (hypothetical) ultra-fast PET/MRI protocol the readers were asked to exclusively read the following sequences (out of a standardized longer protocol) (Table 1):
1. A coronal T1-w 3D-Dixon-VIBE sequence prior contrast administration for attenuation correction only.
2. An axial fat-saturated T2-w HASTE (half Fourier acquisition single shot turbo spin echo)
3. An axial T1-w 3D VIBE (fat-suppressed (fs) volume-interpolated breath-hold examination sequence) after intravenous administration of a gadolinium-based contrast medium (Gadovist; Bayer Healthcare, Germany; 0.1mmol/kg bw).
[Figure omitted. See PDF.]
Table 1. Magnetic Resonance Imaging Parameters.
https://doi.org/10.1371/journal.pone.0172553.t001
Image analysis
The imaging datasets of CT of 18F-FDG PET/CT, 18F-FDG PET/CT and 18F-FDG PET/MRI were evaluated on a dedicated OsiriX Workstation (Pixmeo SARL, Bernex, Switzerland) and independently analyzed by two experienced radiologists in hybrid imaging interpretation. All datasets were evaluated in random order. Both readers were informed regarding the primary cancer of the patients, yet remained blinded to patient identity, clinical findings as well as results of prior or follow-up imaging concerning potential suspicion for tumor recurrence. CTPET/CT, 18F-FDG PET/CT and 18F-FDG PET/MRI were separately assessed in random order with a minimum of two weeks apart to avoid recognition bias.
As previously described [12] the following features were independently evaluated based on each dataset (CTPET/CT, 18F-FDG PET/CT, and 18F-FDG PET/MRI): (a) lesion count, (b) lesion localization, (c) lesion characterization (1 benign; 2 malignant), and (d) diagnostic confidence (1 not confident, 2 rather confident, 3 very confident). On the morphological datasets lesion density (intensity), shape and contrast enhancement were taken into account. Lymph nodes with a short axis diameter exceeding 10 mm, a spherical configuration, or strong contrast enhancement were regarded as suspicious. For lesion characterization on PET, visually increased focal FDG-uptake in comparison to background tissue was considered indicative of malignancy for both 18F-FDG PET/CT and 18F-FDG PET/MRI. In all lesions demonstrating focal 18F-FDG uptake the maximum standardized uptake value (SUVmax) was measured by placing a manually drawn polygonal volume of interest (VOI) over each lesion on attenuation-corrected PET images. Additionally, the maximum diameter of all suspicious lesions was determined on CT and on T1w MR images. Any discrepancies between the two readers were resolved in a subsequent consensus reading.
Reference standard
In accordance with current guidelines, histopathological confirmation of suspected tumor recurrence is not necessarily required for malignancies of the female pelvis. Hence, histopathological verification of disease recurrence was only available in 15 out of 43 patients. As described in previous publications [9,15,16] a modified standard of reference was applied to the remaining cases, comprising previous cross-sectional imaging (mean interval 549 ± 610 days; 18F-FDG PET/CT: n = 10; CT: n = 6; MRI: n = 1) as well as follow-up cross-sectional imaging (mean interval of 270 ± 241 days; 18F-FDG PET/CT: n = 11; CT: n = 13; MRI: n = 1; 18F-FDG PET/MRT: n = 1).
Radiation dose
Effective dose (ED) due to the 18F-FDG PET part of the examination (EDPET) was calculated by using the whole body ED coefficient for females recommended by Andersson et al. [17] based on the International Commission on Radiation Protection publication 2007: EDPET = ADFDG x (1,53 x 10−2); ADFDG is the administered dose in MBq.
ED due to the CT part of the examination (EDCT) was estimated according to a method described by Christner et al. [18] in which the dose-length product and a conversion factor are used.
Statistical analysis
Statistical analysis was performed using IBM SPSS version 22 (IBM Inc, Armonk, NY, USA). Data are presented as mean values ± standard deviation (SD). Sensitivity, specificity, positive predictive value, negative predictive value and diagnostic accuracy for the detection of local tumor recurrences, lymph node and distant metastases were calculated for CTPET/CT, PET/CT and PET/MRI, respectively. The proportion of lesions correctly or falsely rated as malignant or benign or missed on FDG PET/MRI, FDG PET/CT, and CTPET/CT was determined using descriptive statistics.
Results
Out of the 43 enrolled patients 23 (53.5%) were primarily diagnosed with ovarian cancer, 12 (28%) cervical cancer, 4 (9.5%) endometrium cancer, 3 (7%) vulva cancer, and 1 (2%) vaginal cancer.
Patient based analysis
PET/CT and subsequent PET/MRI were completed successfully in all 43 patients. 38 (88%) patients were shown to suffer from tumor recurrence according to the reference standard. PET/CT correctly identified 37 out of 38 patients with tumor relapse (97%), while PET/MRI missed 2 patients (95% correct identification) and CT missed 13 patients (67% correct identification). All modalities (CT, PET/CT, and PET/MRI) classified one patient false positive, indicating recurrent cancer, without confirmation in subsequently performed histopathology.
Lesion-based analysis
In accordance with the reference standard, a total of 154 lesions were included in the final analysis. These comprised 113 (73.4%) malignant and 41 (26.6%) benign lesions based on the reference standard (Table 2).
[Figure omitted. See PDF.]
Table 2. Distribution of benign and malignant lesions.
https://doi.org/10.1371/journal.pone.0172553.t002
Sensitivity, specificity, positive predictive value, negative predictive value and diagnostic accuracy were 98%, 83%, 94%, 94%, and 94% for PET/MRI, respectively. The corresponding results were 97%, 83%, 93%, 94%, and 92% for PET/CT and 50%, 58%, 76%, 31%, and 53% for CT.
Based on PET/MRI readings 2 malignant lesions were falsely rated as benign, comprising a cystic metastasis next to the aorta and spleen, which was incorrectly identified as a seroma as well as an incipient peritoneal carcinosis restricted to the pelvis with low lesion-to-background contrast. Based on PET/CT readings 3 malignant lesions were missed (gluteal muscle metastasis, 2 lymph nodes). In CT 13 lesions were falsely rated as benign, entailing normal-sized lymph nodes without morphologic criteria indicating malignancy. Additionally, 45 lesions were missed, encompassing very small lymph nodes, liver and peritoneal lesions.
Examination time and estimated effective dose of ionizing radiation
Mean scan duration of whole-body CT, PET/CT, and PET/MRI amounted to 3.5 minutes, 18.2 ± 1 minutes, and 18.5 ± 1 minutes (including the scout scan, Dixon sequence for attenuation correction, contrast media administration and breath-hold commands, as well as shimming for PET/MRI), respectively (Table 3).
[Figure omitted. See PDF.]
Table 3. Scan duration of one whole-body CT, PET/CT examination as well as for ultra-fast protocol for whole-body PET/MR imaging.
https://doi.org/10.1371/journal.pone.0172553.t003
The mean effective dose was calculated as described before and amounted to 15.9 ± 8.5 mSv for whole-body CT and 19.5 ± 8.7 mSv for whole-body PET/CT, consisting of the CT component and effective dose of 3.6 ± 0.5 mSv of the administered 18F-FDG (18.5%). Accordingly the mean effective dose for whole-body PET/MRI amounted to 3.6 ± 0.5 mSv (Fig 1).
[Figure omitted. See PDF.]
Fig 1. Estimated Effective Dose (ED) of PET/CT, CT, and PET/MRI.
https://doi.org/10.1371/journal.pone.0172553.g001
Discussion
Our results demonstrate the feasibility and high diagnostic potential of an ultra-fast PET/MRI protocol in females with suspected recurrent pelvic malignancy. On a lesion-based analysis, ultra-fast PET/MRI and PET/CT showed equivalently high diagnostic performance, while being superior to CT (Fig 2). Furthermore, the implementation of an ultra-fast PET/MRI protocol enables restaging of gynecological cancer patients at comparable short scan duration as in PET/CT, while enabling a significant reduction of the radiation exposure when compared to PET/CT and CT.
[Figure omitted. See PDF.]
Fig 2. A 46-year-old patient with recurrent cervical cancer.
Tumor lesion originating from right ovary with contact to the small bowel is hardly detectable in CT (A). The identical lesion shows a slightly higher lesion contrast in the post-contrast VIBE sequence in MRI (C). In fused PET/CT (B) and PET/MR (D) images the lesion can be equally clearly identified as a metastasis.
https://doi.org/10.1371/journal.pone.0172553.g002
PET/CT and PET/MRI have been proven highly valuable in restaging gynecological pelvic cancer due to the combination of metabolic and high-resolution anatomical information [19,20] with previous studies demonstrating the diagnostic equality of PET/MRI compared to PET/CT [12] and diagnostic superiority to MRI alone [11]. Despite excellent diagnostic performance both methods were shown to bear some weak points, namely high radiation exposure in PET/CT and prolonged examination time in PET/MRI. Particularly young patients suffering from tumor entities associated with low mortality rates may be sensitive to the application of repetitive PET/CT examinations with high radiation exposure [21]. The great potential of PET/MRI lies in its multifaceted diagnostic ability. On the one hand multiparametric imaging, requiring dedicated extensive protocols, will be put into focus to provide insights into tumor biology in therapy-naive / therapy-monitored primary tumors [22]. On the other hand, to further enhance the role of PET/MR imaging in clinical routine whole-body imaging, fast efficient protocols are necessary. Initial studies demonstrated the successful implementation of shorter whole-body PET/MRI protocols while preserving the morphologic imaging quality [13,23]. Nevertheless, these studies retained standard PET acquisition times of 4 minutes per bed position, restricting the reduction of the total examination time to a minimum of 26 minutes [13]. Only a few studies investigated the feasibility of truly short PET/MR protocols by utilizing a single sequence, which was originally designed for attenuation-correction, for morphologic correlation [24]. However, this approach degrades the potential of simultaneous PET/MRI to provide the combination of high-resolution morphologic, functional and metabolic information to low-morphologic-metabolic information as in low-dose PET/CT imaging. Hence, the evaluation of ultra-fast PET/MR protocols, entailing the acquisition of high-quality morphologic and metabolic data remains desirable.
Considering recent data demonstrating the questionable added value of diffusion-weighted imaging to PET/MRI as well as a potential reduction of the PET scan time to 2 minutes per bed position (mpb) in PET/MR imaging [14], the aim of this study was to evaluate the diagnostic potential of a hypothetical ultra-fast protocol with preselected sequences from a longer PET/MRI protocol, enabling a significant reduction of the scan time, while preserving high morphological and metabolic data acquisition. Hartung-Knemeyer et al [14] investigated the potential difference in diagnostic quality of PET (derived from PET/MRI), comparing 2 to 8 minutes per bed position PET acquisitions. While the subjective image quality significantly declined from 8 to 2 mpb, neither lesion detectability, nor objective measurements, including signal-to-noise and contrast-to-noise ratio, or SUVmean and SUVmax significantly differed among the acquisition times. Accordingly, we showed the implementation of an ultra-fast PET/MRI protocol, comprising a pre-contrast T1w and T2w sequence as well as a post-contrast (post-PET) T1w sequence within a 2 mpb PET scan time, seems feasible and may lead to a reasonable reduction of the overall scan time to 18.5 ± 1 minutes (including scout scan, shimming, breath hold commands, contrast media application) which would be comparable to PET/CT (comprising 6–7 bed position of 2 mpb).
Considering the results of our study, our hypothesis in demonstrating the diagnostic comparability of ultra-fast-PET/MRI to PET/CT and superiority towards CT may be proven valid. While all three imaging modalities falsely rated one patient with perforated sigmoid diverticulitis and reactive FDG-accumulation as indicative for recurrent cancer, both hybrid imaging techniques showed their significant superiority over CT and comparable diagnostic competence among each other. PET/MRI utilizing an ultra-fast protocol demonstrated its diagnostic equivalence to PET/CT, providing slightly inferior sensitivity rates based on patient-based analysis and superior rates based on lesion-based-analysis (Fig 3). Two misinterpreted lesions in PET/MRI leading to misclassifications may be due to the study design, involving subsequent data-acquisition. Due to ethical standards all PET/MRI datasets were obtained after clinically indicated PET/CT examination, resulting in an additional delay of the PET acquisition for PET/MRI. As previously published data demonstrate this delay leads to an increased and altered distribution of FDG accumulation in several parenchymatous organs and background in PET/MRI [25,26]. Accordingly, an increasing FDG accumulation in surrounding non-cancerous tissues and background during the time of prolonged tracer uptake might lead to a decrease of lesion-to-background ratio and false interpretation. Nevertheless, previous publications also reported on increasing FDG accumulation in tumor lesions due to prolonged timing after tracer injection. [27]. However, we did not encounter this in our study. Three missed lesions on PET/CT included a soft tissue metastasis in right gluteal muscle and two inguinal lymph nodes. The soft tissue metastasis was misinterpreted as increased muscle activity because of lacking morphological correlation in CT. Lymph node metastases were not detected with PET/CT because of artifacts due to a hip prosthesis.
[Figure omitted. See PDF.]
Fig 3. Images of a 58-year-old patient with recurrence of ovarian cancer.
CT image (A) shows a peritoneal mass next to the liver with pathological FDG metabolism in fused PET/CT images (B). The same lesion is visible in post-contrast media VIBE sequence in MRI (C) and in PET/MRI (D). No FDG metabolism can be seen in a benign liver cyst in PET/CT or PET/MRI (A-D).
https://doi.org/10.1371/journal.pone.0172553.g003
Our study is not without some limitations. Investigating a hypothetical ultra-fast-PET/MR protocol, the analyzed sequences were selected from prolonged examination protocols, comprising a median of 4 minute PET acquisition time per bed position. While the investigation of a “true” ultra-fast-protocol would have been desirable, our primary responsibility lies in proper patient care and diagnostics, hence performing a non-evaluated, potentially insufficient ultra-fast-protocol would not have merited our ethical standards. Hence, our study results should be considered as a platform to utilize future studies and evaluate true ultra-fast PET/MR imaging. Secondly, due to our demanding standards regarding long follow-up periods and / or histopathological confirmation for reference standard, the enrolled and analyzed patient cohort is rather small. Hence, our results should be regarded as preliminary and future studies with larger patient cohorts should be performed.
In conclusion, our study results demonstrate the high diagnostic potential of an ultra-fast PET/MR protocol, shortening the examination time to PET/CT acquisition times while preserving high quality MR morphological imaging and equivalent diagnostic performance when compared to PET/CT. This may be of particular interest in case of younger patients, for therapy monitoring as well as long-term aftercare of patients suffering from tumor entities that are associated to low mortality rates, such as early stage cervical cancer or lymphoma. In addition, radiation exposure can be markedly reduced with PET/MRI compared to full-dose PET/CT and CT. With regard to patient comfort and an improved implementation of PET/MRI into routine clinical imaging, ultra-fast PET/MRI may play a crucial role for the diagnostic work-up of cancer patients.
Author Contributions
1. Conceptualization: JK CD LU JG BA.
2. Formal analysis: JK LS KH GA VR SS.
3. Funding acquisition: MF GA KH LU.
4. Investigation: JK CD JG LU BA.
5. Methodology: JK VR CD BA JG MF LS LU.
6. Project administration: JK JG LU CD VR.
7. Resources: GA MF KH BA.
8. Supervision: GA MF KH LU.
9. Validation: VR JG CD MF KH.
10. Visualization: JK.
11. Writing – original draft: JK LU.
12. Writing – review & editing: LS SS JG VR BA CD KH GA MF.
Citation: Kirchner J, Sawicki LM, Suntharalingam S, Grueneisen J, Ruhlmann V, Aktas B, et al. (2017) Whole-body staging of female patients with recurrent pelvic malignancies: Ultra-fast 18F-FDG PET/MRI compared to 18F-FDG PET/CT and CT. PLoS ONE 12(2): e0172553. https://doi.org/10.1371/journal.pone.0172553
1. Czernin J, Allen-Auerbach M, Schelbert HR (2007) Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006. J Nucl Med 48 Suppl 1: 78S–88S.
2. Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. (2000) A combined PET/CT scanner for clinical oncology. J Nucl Med 41: 1369–1379. pmid:10945530
3. Antoch G, Vogt FM, Freudenberg LS, Nazaradeh F, Goehde SC, Barkhausen J, et al. (2003) Whole-body dual-modality PET/CT and whole-body MRI for tumor staging in oncology. JAMA 290: 3199–3206. pmid:14693872
4. Kitajima K, Murakami K, Yamasaki E, Domeki Y, Kaji Y, Morita S, et al. (2009) Performance of integrated FDG-PET/contrast-enhanced CT in the diagnosis of recurrent uterine cancer: comparison with PET and enhanced CT. Eur J Nucl Med Mol Imaging 36: 362–372. pmid:18931841
5. Kitajima K, Murakami K, Yamasaki E, Domeki Y, Kaji Y, Fukasawa I, et al. (2008) Performance of integrated FDG-PET/contrast-enhanced CT in the diagnosis of recurrent ovarian cancer: comparison with integrated FDG-PET/non-contrast-enhanced CT and enhanced CT. Eur J Nucl Med Mol Imaging 35: 1439–1448. pmid:18418592
6. Gu P, Pan LL, Wu SQ, Sun L, Huang G (2009) CA 125, PET alone, PET-CT, CT and MRI in diagnosing recurrent ovarian carcinoma: a systematic review and meta-analysis. Eur J Radiol 71: 164–174. pmid:18378417
7. Brix G, Nosske D, Lechel U (2014) Radiation exposure of patients undergoing whole-body FDG-PET/CT examinations: an update pursuant to the new ICRP recommendations. Nuklearmedizin 53: 217–220. pmid:24919708
8. Spick C, Herrmann K, Czernin J (2016) 18F-FDG PET/CT and PET/MRI Perform Equally Well in Cancer: Evidence from Studies on More Than 2,300 Patients. J Nucl Med 57: 420–430. pmid:26742709
9. Sawicki LM, Grueneisen J, Schaarschmidt BM, Buchbender C, Nagarajah J, Umutlu L, et al. (2016) Evaluation of (1)(8)F-FDG PET/MRI, (1)(8)F-FDG PET/CT, MRI, and CT in whole-body staging of recurrent breast cancer. Eur J Radiol 85: 459–465. pmid:26781152
10. Heusch P, Nensa F, Schaarschmidt B, Sivanesapillai R, Beiderwellen K, Gomez B, et al. (2015) Diagnostic accuracy of whole-body PET/MRI and whole-body PET/CT for TNM staging in oncology. Eur J Nucl Med Mol Imaging 42: 42–48. pmid:25112399
11. Grueneisen J, Beiderwellen K, Heusch P, Gratz M, Schulze-Hagen A, Heubner M, et al. (2014) Simultaneous positron emission tomography/magnetic resonance imaging for whole-body staging in patients with recurrent gynecological malignancies of the pelvis: a comparison to whole-body magnetic resonance imaging alone. Invest Radiol 49: 808–815. pmid:25010207
12. Beiderwellen K, Grueneisen J, Ruhlmann V, Buderath P, Aktas B, Heusch P, et al. (2015) [(18)F]FDG PET/MRI vs. PET/CT for whole-body staging in patients with recurrent malignancies of the female pelvis: initial results. Eur J Nucl Med Mol Imaging 42: 56–65. pmid:25223420
13. Grueneisen J, Schaarschmidt BM, Heubner M, Suntharalingam S, Milk I, Kinner S, et al. (2015) Implementation of FAST-PET/MRI for whole-body staging of female patients with recurrent pelvic malignancies: A comparison to PET/CT. Eur J Radiol 84: 2097–2102. pmid:26321491
14. Hartung-Knemeyer V, Beiderwellen KJ, Buchbender C, Kuehl H, Lauenstein TC, Bockisch A, et al. (2013) Optimizing positron emission tomography image acquisition protocols in integrated positron emission tomography/magnetic resonance imaging. Invest Radiol 48: 290–294. pmid:23399811
15. Buchbender C, Hartung-Knemeyer V, Beiderwellen K, Heusch P, Kuhl H, Lauenstein TC, et al. (2013) Diffusion-weighted imaging as part of hybrid PET/MRI protocols for whole-body cancer staging: does it benefit lesion detection? Eur J Radiol 82: 877–882. pmid:23428414
16. Beiderwellen K, Gomez B, Buchbender C, Hartung V, Poeppel TD, Nensa F, et al. (2013) Depiction and characterization of liver lesions in whole body [(1)(8)F]-FDG PET/MRI. Eur J Radiol 82: e669–675. pmid:24011443
17. Andersson M, Johansson L, Minarik D, Leide-Svegborn S, Mattsson S (2014) Effective dose to adult patients from 338 radiopharmaceuticals estimated using ICRP biokinetic data, ICRP/ICRU computational reference phantoms and ICRP 2007 tissue weighting factors. EJNMMI Phys 1: 9. pmid:26501451
18. Christner JA, Kofler JM, McCollough CH (2010) Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection publication 103 or dual-energy scanning. AJR Am J Roentgenol 194: 881–889. pmid:20308486
19. Kim SK, Choi HJ, Park SY, Lee HY, Seo SS, Yoo CW, et al. (2009) Additional value of MR/PET fusion compared with PET/CT in the detection of lymph node metastases in cervical cancer patients. Eur J Cancer 45: 2103–2109. pmid:19403303
20. Caobelli F, Alongi P, Evangelista L, Picchio M, Saladini G, Rensi M, et al. (2016) Predictive value of (18)F-FDG PET/CT in restaging patients affected by ovarian carcinoma: a multicentre study. Eur J Nucl Med Mol Imaging 43: 404–413. pmid:26381775
21. Torre LA, Bray F, Siegel RL, Ferlay J, Lortet-Tieulent J, Jemal A (2015) Global cancer statistics, 2012. CA Cancer J Clin 65: 87–108. pmid:25651787
22. Pinker K, Andrzejewski P, Baltzer P, Polanec SH, Sturdza A, Georg D, et al. (2016) Multiparametric [18F]Fluorodeoxyglucose/ [18F]Fluoromisonidazole Positron Emission Tomography/ Magnetic Resonance Imaging of Locally Advanced Cervical Cancer for the Non-Invasive Detection of Tumor Heterogeneity: A Pilot Study. PLoS One 11: e0155333. pmid:27167829
23. Grueneisen J, Sawicki LM, Schaarschmidt BM, Suntharalingam S, von der Ropp S, Wetter A, et al. (2016) Evaluation of a Fast Protocol for Staging Lymphoma Patients with Integrated PET/MRI. PLoS One 11: e0157880. pmid:27327617
24. Kohan AA, Kolthammer JA, Vercher-Conejero JL, Rubbert C, Partovi S, Jones R, et al. (2013) N staging of lung cancer patients with PET/MRI using a three-segment model attenuation correction algorithm: initial experience. Eur Radiol 23: 3161–3169. pmid:23765261
25. Kershah S, Partovi S, Traughber BJ, Muzic RF Jr., Schluchter MD, O'Donnell JK, et al. (2013) Comparison of standardized uptake values in normal structures between PET/CT and PET/MRI in an oncology patient population. Mol Imaging Biol 15: 776–785. pmid:23632951
26. Drzezga A, Souvatzoglou M, Eiber M, Beer AJ, Furst S, Martinez-Moller A, et al. (2012) First clinical experience with integrated whole-body PET/MR: comparison to PET/CT in patients with oncologic diagnoses. J Nucl Med 53: 845–855. pmid:22534830
27. Law WP, Maggacis N, Jeavons SJ, Miles KA (2016) Concordance of 18F-FDG PET Uptake in Tumor and Normal Tissues on PET/MRI and PET/CT. Clin Nucl Med.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2017 Kirchner et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Objectives
To evaluate the diagnostic feasibility of an ultra-fast 18F-FDG PET/MRI protocol, including T2-w and contrast-enhanced T1-w imaging as well as metabolic assessment (PET) in comparison to 18F-FDG PET/CT and CT for whole-body staging of female patients with suspected recurrence of pelvic malignancies.
Methods
43 female patients with suspected tumor recurrence were included in this study. Suspicion was based on clinical follow-up and abnormal findings on imaging follow-up. All patients underwent a PET/CT and a subsequent PET/MRI examination. Two readers were asked to evaluate ultra-fast PET/MRI, PET/CT as well as CT datasets of PET/CT separately for suspect lesions regarding lesion count, lesion localization and lesion characterization. Statistical analyses were performed both, on a per-patient and a per-lesion basis.
Results
Tumor relapse was present in 38 of the 43 patients. Based on CT readings 25/38 tumor relapses were correctly identified. PET/CT enabled correct identification of 37/38 patients, PET/MRI correctly identified 36 of the 38 patients with recurrent cancer. On a lesion-based analysis PET/MRI enabled the correct detection of more lesions, comprising a lesion-based sensitivity, specificity, positive predictive value, negative predictive value and diagnostic accuracy of 50%, 58%, 76%, 31%, and 53% for CT, 97%, 83%, 93%, 94%, and 92% for PET/CT and 98%, 83%, 94%, 94%, and 94% for PET/MRI, respectively. Mean scan duration of ultra-fast PET/MRI, PET/CT and whole-body CT amounted to 18.5 ± 1 minutes, 18.2 ± 1 minutes and 3.5 minutes, respectively.
Conclusion
Ultra-fast PET/MRI provides equivalent diagnostic performance and examination time when compared to PET/CT and superior diagnostic performance to CT in restaging female patients suspected to have recurrent pelvic cancer.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer