Background
This study sought to determine whether several metatarsophalangeal (MTP) fusion techniques require complete immobilization or if some level of weight-bearing could be recommended after surgery. A comparison of synthetic composite to actual bone was included in order to examine the validity of the testing conditions.
Methods
Four MTP fusion modalities were tested in synthetic composite bone models: unlocked plating, locked plating, crossed lag screws, and an unlocked plate with a single lag screw. Stiffness was calculated and then used to find the two most rigid constructs; the load to failure was recorded. Stiffness and load to failure testing for the two more rigid constructs in paired cadaveric bones were followed.
Results
The unlocked plate plus screw and crossed screw constructs were stiffest (p < 0.008). Loads to failure of the unlocked plate plus screw and crossed screws in synthetic bone were 131 and 101 N, respectively and in cadaveric bone were 154 and 94 N, respectively, which are less than the estimated 25% body weight required at the MTP joint. The plate plus screws were statistically more stiff than crossed screws (p = 0.008), but there was no statistical difference between synthetic and cadaveric bone in load to failure (p = 0.296).
Conclusions
The plate plus screw offered the greatest stiffness; the failure test showed that no construct could withstand weight-bearing as tolerated; and, synthetic composite models of the MTP joint did not provide the consistent results in stiffness and failure.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer