Introduction
Bladder cancer is a cause of considerable morbidity and mortality worldwide. The intensive care and treatment of bladder cancer represents a considerable burden to patients and society (Botteman et al., 2003). Intravesical therapy has been used to maximize the exposure of bladder tumors to the drug, while limiting systemic effects and toxicity. Mitomycin C (MMC) is the most widely used drug for intravesical irrigation in low and intermediate risk tumors (Bouffioux et al., 1995; Sylvester, Oosterlinck & Van der Meijden, 2004), while Bacillus Calmette-Guérin (BCG) is recommended for high-risk tumors (Sylvester, Van der Meijden & Lamm, 2002; Shelley et al., 2004; Böhle & Bock, 2004). Intravesical chemotherapy with MMC is used immediately after transurethral resection of tumor (TURBT) to treat circulating tumor cells that can reimplant at other sites of the bladder, or as an ablative effect on residual tumour cells at the resection site (Pan et al., 1989; Brocks, Büttner & Böhle, 2005). MMC is also used in series of 8-weekly intravesical irrigations for patients with high risk of recurrence (Tolley et al., 1996; Sylvester, Oosterlinck & Witjes, 2008).
MMC migrates poorly deeper into the lamina propria, which may explain MMC’s poor efficacy treating T1 tumors invading this layer of the mucosa. Electrokinetic forces may accelerate and increase drug delivery across the urothelium. Electromotive Drug Administration (EMDA) is a method that combines intravesical chemotherapy with local electric field application. Di Stasi et al. (1999) suggested in an ex vivo model that concentrations of MMC may be increased in all the layers of the bladder wall, when the drug was delivered with EMDA. Clinical studies where EMDA/MMC was applied showed lower recurrence rates than MMC alone, and equivalent efficacy for the treatment of high-risk bladder tumors when compared to BCG (Di Stasi et al., 2003; Di Stasi et al., 2006) with few side effects noted. Thus EMDA method seems safe and effective.
It has been proposed that EMDA leads to the combination of several electro-molecular interactions; iontophoresis, electroosmosis/electrophoresis, and electroporation (Di Stasi et al., 1997). All these phenomena may potentially be responsible for electromotive transport of drug molecules through biological membranes and into the underlying tissue when the electric field is applied. EMDA method is being used experimentally in relatively few centers, possibly due to the fact that the mechanisms of action are poorly understood. However, the actual worldwide shortage of BCG has led to the use of alternative methods and increased use of EMDA/MMC. Consequently there is a need to further elucidate the mechanisms of action of this treatment, which led to the present study.
The objective of this study is to perform calculations of the electric field present in the bladder wall during EMDA treatments, and in that way estimate what biological mechanisms may be at play when the tissue is exposed to this particular electric field.
Materials and Methods
The main goal of the following steps was to build an in-silico representation of a human body during the delivery of an EMDA treatment. This was done by reconstructing the anatomical shape of the body and the relevant tissues from CT images and then using a computational approach to determine the relevant electric quantities in the tissues: electric field and current density. These quantities were then analyzed to get an insight into the possible mechanisms of EMDA.
Delivery setup
EMDA is administered by a battery powered current generator (Physionizer®) that delivers a controlled electric current of 0–30 mA/0–55 V DC, which is passed between electrodes: (1) the active intravesical electrode is a silver spiral integrated to the tip of a specifically designed transurethral catheter, (2) two dispersive ground electrodes, placed in the lower abdominal skin. The operator sets the active electrode’s polarity and current intensity on the generator.
When applying intravesical EMDA with MMC for non-muscle invasive bladder cancer, the catheter is inserted into the bladder followed by instillation of 40 mg of MMC in 50 ml 0.9% NaCl solution. The catheter and the ground electrodes are connected to the generator, and the active electrode is set to positive polarity. The generator gradually increases the current to a predetermined level (20–25 mA) at a rise rate of 50 mA/s. The total treatment time is 30 min. This setup is shown in Fig. 1.
Figure 1: Schematic illustration of the EMDA application system. (A) Numbers indicate the electrodes: 1, helical bladder catheter electrode; 2, abdominal patch electrodes. (B) The catheter is inserted in the urinary bladder and connected to the current generator. DOI: 10.7717/peerj.2309/fig-1
Segmentation of medical images
Different tissues have different electrical properties. To be able to accurately compute the electrical quantities in the body we need to differentiate between these tissues and assign them their appropriate properties. Abdominal CT images of a patient treated with EMDA were segmented into the following tissues (bladder wall, bladder lumen, small intestine, prostate, muscle, bone, fat, and internal air) using ITK-Snap version 2.4 (Yushkevich et al., 2006). The patient was a participant in a clinical trial where EMDA was tested, and the Ethics Committee of the Capital Region of Denmark approved this trial (study number H-1-2012-050). As the current study is not part of the clinical trial, permission for the use of the CT images was obtained from the patient orally.
A radiologist verified the segmentation. The pixel size in the segmentation was 0.781 mm, while the slice separation was 3 mm. In total 72 slices were segmented, corresponding to 14 slices below the bladder and 3 slices above the top edge of the external electrodes.
3-D model building
A 3-D model of the patient’s anatomy was constructed in COMSOL Multiphysics version 5.2 (Comsol AB, Stockholm, Sweden). The outside of the body was constructed using a planar contour method using an algorithm written in Matlab (Mathworks, Natick, MA, USA) (Sel, Lebar & Miklavcic, 2007). The tissues of the interior of the body were taken into account by building a spatial conductivity matrix in Matlab by assigning conductivity values based on the segmentation (Aström et al., 2009). The tissue properties were taken from the literature (Gabriel, Lau & Gabriel, 1996a; Gabriel, Lau & Gabriel, 1996b) and the actual values used are reported in Table 1. The conductivity of the saline used for bladder irrigation was also taken from the literature (Stogryn, 1971). Electrodes were added by CAD modelling directly in COMSOL based on the dimensions of the actual electrodes. The balloon of the catheter (12 mm diameter) was placed right above the bladder neck. The catheter electrode consists of a helical wire with a diameter of 0.4 mm and 20 turns. The major radius of the helix was 1.25 mm, while the turn pitch was 1.6 mm; therefore, the total combined length of the electrode is 32 mm. This electrode shape limits the intensity of the electric field and current density around the electrode. The outer electrodes were modelled as a patch with outer dimensions of 5 × 7 cm. The resulting model is shown in Fig. 2. The catheter balloon was modelled as a perfect insulator, so it was excluded from the computation.
Tissue | Conductivity [S/m] | Reference |
---|---|---|
Bladder wall | 0.2 | Gabriel, Lau & Gabriel (1996a), Gabriel, Lau & Gabriel (1996b) |
Bladder lumen | 1.5 | Stogryn (1971) |
Prostate | 0.41 | Gabriel, Lau & Gabriel (1996a), Gabriel, Lau & Gabriel (1996b) |
Fat | 0.012 | Gabriel, Lau & Gabriel (1996a), Gabriel, Lau & Gabriel (1996b) |
Muscle | 0.2 | Gabriel, Lau & Gabriel (1996a), Gabriel, Lau & Gabriel (1996b) |
Intestines | 0.4 | Gabriel, Lau & Gabriel (1996a), Gabriel, Lau & Gabriel (1996b) |
Bone | 0.02 | Gabriel, Lau & Gabriel (1996a), Gabriel, Lau & Gabriel (1996b) |
DOI: 10.7717/peerj.2309/table-1
Figure 2: 3-D model with electrodes and different tissues. (A–C) sagittal, coronal and axial views. (D) rotated 3D view. The abdominal and bladder electrodes are indicated. In all panels, the bladder, intestine, prostate and internal air are shown, but clipped to only half of the model for clarity. DOI: 10.7717/peerj.2309/fig-2
Computational setup
We used the stationary finite element method solvers in COMSOL Multiphysics to solve the Laplace equation for electric potential in the 3-D model of the patient’s anatomy. The model was discretized using a free tetrahedral mesh, which consisted of 2,779,405 elements. We used the conductive media DC interface in COMSOL Multiphysics to run a static simulation of the applied current density and electric fields inside the body. The electrodes were driven with a current-source terminal boundary condition, while the skin was taken into account only at the skin-electrode interface, by using a built-in thin resistive layer boundary condition in COMSOL. The results were extracted on a grid with data points positioned at the centers of the image elements in the original anatomical CT scans.
Results Global quantities
Since the electrodes were driven by a current source (20 mA), it is possible to compute the resulting electrode voltages with the data from the manufacturer’s specifications. The voltage between the two electrodes was 13.1 V, which is within the generator specification and corresponds well to the expected voltage. Unfortunately, the generator does not specify the output voltage at the time of the treatment so the exact value used in the treatment is not available for the validation of the model.
Applied fields in the body
Electric fields in the whole body were extracted. The highest magnitude of electric field in the simulations was 271 V/m, in the subcutaneous fat tissue near the surface electrodes. The maximum values in the bladder wall were 71.9 V/m. The mean and SD of electric field magnitude in the bladder wall were 6.0 V/m and 7.1 V/m, respectively.
The largest current density was found around the catheter electrode: 134 A/m2. The largest current density in the bladder wall was 14.4 A/m2. Mean current density in the bladder wall was 1.2 A/m2, with the standard deviation being 1.4 A/m2. The highest current density in all other tissues was 35.6 A/m2, however this value was found in a single voxel and may be due to numerical error. The current density is mostly perpendicular to the bladder, although the current density is much higher at the anterior part than at the posterior as would be expected based on electrode positioning Fig. 3.
Figure 3: Visualization of the current in the whole body and the current density in the bladder wall. (A) Current lines from the bladder electrodes to the abdominal patch electrodes. (B) Surface plot of the current density on the bladder wall in the sagittal view (from the left side). (C) Surface plot of the current density in a coronal view (from the front). The current density is markedly higher in the upper anterior bladder wall in comparison to the posterior and lower bladder walls. DOI: 10.7717/peerj.2309/fig-3
Figure 4: Electric field and current density in the bladder wall. The figure shows an axial slice at the vertical center of the bladder wall. Both panels show the same slice. The panels show: (A) Electric field, (B) Current density. DOI: 10.7717/peerj.2309/fig-4
Discussion
Based on the values of electric fields, we can conclude that electroporation is highly unlikely to be the predominant mechanism of action in EMDA. Typically, electroporation occurs at electric field strengths that are above 40,000 V/m (Cemazar et al., 1998; Miklavcic et al., 1998; Gehl et al., 1999; Pucihar et al., 2011) also on urothelial cells (JL Vásquez, 2013, unpublished data), while the electric fields reached established by EMDA are well below even 100 V/m. Although electroporation thresholds decrease with increasing duration of electric field application, the lowest reported in the literature is above 10,000 V/m for 100 ms pulse duration (Pucihar et al., 2011). The highest amplitude of the induced transmembrane voltage (ITV) according to Schwann’s equation (Schwan, 1957) is determined by the equation ITVMax = 1.5ER, where E is the electric field strength to which the cell of radius R is exposed. Based on the size of the cells used by Pucihar et al. in their experiments, the highest transmembrane voltage at 100 ms pulses was approximately 150 mV. The superficial layer of the bladder contains polyhedral cells of up to 250 µm in diameter (Khandelwal, Abraham & Apodaca, 2009). Given the value of the electric field in the bladder, the highest expected ITV during the treatment is 14 mV or lower, which is well within physiological values of transmembrane potential. The other layers of the bladder contain much smaller cells, which would result in even lower ITV.
Typically, iontophoresis setups for trans-dermal drug delivery use current densities in the range of 0.1–0.5 mA/cm2 (Kalia et al., 2004), which is equal to 1–5 A/m2 and therefore very close to the values produced by our model. The values of current density in the bladder wall support the hypothesis, that iontophoresis is involved in EMDA, and that is likely to be the contributing factor for reducing recurrence, due to the increase of the drug concentration near the point of excision.
Due to the placement of the abdominal electrodes on the front part of the abdominal wall, the predominant direction of the current flow is towards the front, as can be seen in Figs. 3 and 4. This means that the current density of the posterior bladder wall could be below the current density required for efficient iontophoretic flow, i.e., the treatment would be more successful for tumors located on the front half of the bladder than for the back half.
The present study is limited by the fact that it is based on a single patient, and that the exact positioning of the intravesical electrode during the treatment is unknown. Additionally, the variability between patients would affect the values of electrical quantities in the tissue. However, even with these variations, the difference in the electric fields and current density would not be so great to change our findings.
Our study shows that electroporation is not the mechanism of action in EMDA. However, electroporation combined with chemotherapy (electrochemotherapy) is a novel and interesting treatment modality in use for the treatment of cutaneous tumors and in clinical trials for tumors in internal organs (Campana et al., 2009; Matthiessen et al., 2011; Edhemovic et al., 2011; Miklavčič et al., 2014). Also for bladder cancer, there are interesting perspectives for electrochemotherapy, both using MMC and cisplatin (Vásquez, Gehl & Hermann, 2012; Vásquez et al., 2015). These data on cisplatin used in electrochemotherapy experiments may give inspiration for new advances in iontophoretic drug delivery.
Conclusion
Based on numerical computational modelling based on realistically dimensioned models of EMDA treatment, we suggest that iontophoretic forces are predominant in this treatment. Electrode placements could mean that the anterior wall of the bladder receives a higher treatment field than other parts of the bladder. Also, due to the presence of a small air pocket in the bladder, the location directly above the air pocket receives a reduced treatment. Based on the numerical results alternate electrode positions should be considered if the original location of the excised tumor is not on the anterior bladder wall.
Additional Information and Declarations
Competing Interests
The authors declare there are no competing interests.
Author Contributions
Bor Kos conceived and designed the experiments, performed the experiments, analyzed the data, wrote the paper, prepared figures and/or tables, reviewed drafts of the paper.
Juan Luis Vásquez conceived and designed the experiments, analyzed the data, wrote the paper, prepared figures and/or tables, reviewed drafts of the paper.
Damijan Miklavčič wrote the paper, reviewed drafts of the paper.
Gregers G.G. Hermann analyzed the data, wrote the paper, reviewed drafts of the paper.
Julie Gehl conceived and designed the experiments, wrote the paper, reviewed drafts of the paper.
Human Ethics
The following information was supplied relating to ethical approvals (i.e., approving body and any reference numbers):
The patient was a participant in a clinical trial where EMDA was tested and the Ethics Committee of the Capital Region of Denmark approved this trial (study number H-1-2012-050).
As the current study is not part of the clinical trial, permission for the use of the CT images was obtained from the patient orally.
Data Availability
The following information was supplied regarding data availability:
Kos, Bor; Luis Vásquez, Juan; Miklavcic, Damijan; G Hermann, Gregers; Gehl, Julie (2016): Investigation of the mechanisms of action behind Electromotive Drug Administration (EMDA). Figshare https://dx.doi.org/10.6084/m9.figshare.3118759.v1.
Funding
The authors received no funding for this work.
Aström M, Zrinzo LU, Tisch S, Tripoliti E, Hariz MI, Wårdell K. 2009. Method for patient-specific finite element modeling and simulation of deep brain stimulation. Medical and Biological Engineering and Computing 47:21-28
Böhle A, Bock PR. 2004. Intravesical bacille Calmette-Guérin versus mitomycin C in superficial bladder cancer: formal meta-analysis of comparative studies on tumor progression. Urology 63:682-686
Botteman MF, Pashos CL, Redaelli A, Laskin B, Hauser R. 2003. The health economics of bladder cancer: a comprehensive review of the published literature. PharmacoEconomics 21:1315-1330
Bouffioux C, Kurth KH, Bono A, Oosterlinck W, Kruger CB, De Pauw M, Sylvester R. 1995. Intravesical adjuvant chemotherapy for superficial transitional cell bladder carcinoma: results of 2 European Organization for Research and Treatment of Cancer randomized trials with mitomycin C and doxorubicin comparing early versus delayed instillations and short-term versus long-term treatment. European Organization for Research and Treatment of Cancer Genitourinary Group. The Journal of Urology 153:934-941
Brocks CP, Büttner H, Böhle A. 2005. Inhibition of tumor implantation by intravesical gemcitabine in a murine model of superficial bladder cancer. The Journal of Urology 174:1115-1118
Campana L, Mocellin S, Basso M, Puccetti O, De Salvo G, Chiarion-Sileni V, Vecchiato A, Corti L, Rossi C, Nitti D. 2009. Bleomycin-based electrochemotherapy: clinical outcome from a single institution’s experience with 52 patients. Annals of Surgical Oncology 16:191-199
Cemazar M, Jarm T, Miklavcic D, Lebar AM, Ihan A, Kopitar NA, Sersa G. 1998. Effect of electric-field intensity on electropermeabilization and electrosensitivity of various tumor-cell lines in vitro. Electro- and Magnetobiology 17:263-272
Di Stasi SM, Giannantoni A, Giurioli A, Valenti M, Zampa G, Storti L, Attisani F, De Carolis A, Capelli G, Vespasiani G, Stephen RL+1 more. 2006. Sequential BCG and electromotive mitomycin versus BCG alone for high-risk superficial bladder cancer: a randomised controlled trial. The Lancet Oncology 7:43-51
Di Stasi SM, Giannantoni A, Massoud R, Dolci S, Navarra P, Vespasiani G, Stephen RL. 1999. Electromotive versus passive diffusion of mitomycin C into human bladder wall: concentration-depth profiles studies. Cancer Research 59:4912-4918
Di Stasi SM, Giannantoni A, Stephen RL, Capelli G, Navarra P, Massoud R, Vespasiani G. 2003. Intravesical electromotive mitomycin C versus passive transport mitomycin C for high risk superficial bladder cancer: a prospective randomized study. The Journal of Urology 170:777-782
Di Stasi SM, Vespasiani G, Giannantoni A, Massoud R, Dolci S, Micali F. 1997. Electromotive delivery of mitomycin C into human bladder wall. Cancer Research 57:875-880
Edhemovic I, Gadzijev EM, Brecelj E, Miklavcic D, Kos B, Zupanic A, Mali B, Jarm T, Pavliha D, Marcan M, Gasljevic G, Gorjup V, Music M, Vavpotic TP, Cemazar M, Snoj M, Sersa G+7 more. 2011. Electrochemotherapy: a new technological approach in treatment of metastases in the liver. Technology in Cancer Research and Treatment 10:475-485
Gabriel S, Lau R, Gabriel C. 1996a. The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz–20 GHz. Physics in Medicine and Biology 41:2251-2269
Gabriel S, Lau R, Gabriel C. 1996b. The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues. Physics in Medicine and Biology 41:2271-2293
Gehl J, Sorensen TH, Nielsen K, Raskmark P, Nielsen SL, Skovsgaard T, Mir LM. 1999. In vivo electroporation of skeletal muscle: threshold, efficacy and relation to electric field distribution. Biochimica Et Biophysica Acta-General Subjects 1428:233-240
Kalia YN, Naik A, Garrison J, Guy RH. 2004. Iontophoretic drug delivery. Advanced Drug Delivery Reviews 56:619-658
Khandelwal P, Abraham SN, Apodaca G. 2009. Cell biology and physiology of the uroepithelium. American Journal of Physiology. Renal Physiology 297:F1477-F1501
Matthiessen LW, Chalmers RL, Sainsbury DCG, Veeramani S, Kessell G, Humphreys AC, Bond JE, Muir T, Gehl J. 2011. Management of cutaneous metastases using electrochemotherapy. Acta Oncologica 50:621-629
Miklavcic D, Beravs K, Semrov D, Cemazar M, Demsar F, Sersa G. 1998. The importance of electric field distribution for effective in vivo electroporation of tissues. Biophysical Journal 74:2152-2158
Miklavčič D, Mali B, Kos B, Heller R, Serša G. 2014. Electrochemotherapy: from the drawing board into medical practice. Biomedical Engineering Online 13:29
Pan JS, Slocum HK, Rustum YM, Greco WR, Gaeta JF, Huben RP. 1989. Inhibition of implantation of murine bladder tumor by thiotepa in cauterized bladder. The Journal of Urology 142:1589-1593
Pucihar G, Krmelj J, Reberšek M, Napotnik TB, Miklavčıč D. 2011. Equivalent pulse parameters for electroporation. IEEE Transactions on Bio-Medical Engineering 58:3279-3288
Schwan HP. 1957. Electrical properties of tissue and cell suspensions. Advances in Biological and Medical Physics 5:147-209
Sel D, Lebar A, Miklavcic D. 2007. Feasibility of employing model-based optimization of pulse amplitude and electrode distance for effective tumor electropermeabilization. IEEE Transactions on Biomedical Engineering 54:773-781
Shelley MD, Wilt TJ, Court J, Coles B, Kynaston H, Mason MD. 2004. Intravesical bacillus Calmette-Guérin is superior to mitomycin C in reducing tumour recurrence in high-risk superficial bladder cancer: a meta-analysis of randomized trials. BJU International 93:485-490
Stogryn A. 1971. Equations for calculating the dielectric constant of saline water (correspondence) IEEE Transactions on Microwave Theory and Techniques 19:733-736
Sylvester RJ, Oosterlinck W, Van der Meijden A. 2004. A single immediate postoperative instillation of chemotherapy decreases the risk of recurrence in patients with stage Ta T1 bladder cancer: a meta-analysis of published results of randomized clinical trials. The Journal of Urology 171:2186-2190 quiz 2435
Sylvester RJ, Oosterlinck W, Witjes JA. 2008. The schedule and duration of intravesical chemotherapy in patients with non-muscle-invasive bladder cancer: a systematic review of the published results of randomized clinical trials. European Urology 53:709-719
Sylvester RJ, Van der Meijden APM, Lamm DL. 2002. Intravesical bacillus Calmette-Guerin reduces the risk of progression in patients with superficial bladder cancer: a meta-analysis of the published results of randomized clinical trials. The Journal of Urology 168:1964-1970
Tolley DA, Parmar MK, Grigor KM, Lallemand G, Benyon LL, Fellows J, Freedman LS, Grigor KM, Hall RR, Hargreave TB, Munson K, Newling DW, Richards B, Robinson MR, Rose MB, Smith PH, Williams JL, Whelan P+8 more. 1996. The effect of intravesical mitomycin C on recurrence of newly diagnosed superficial bladder cancer: a further report with 7 years of follow up. The Journal of Urology 155:1233-1238
Vásquez JL, Gehl J, Hermann GG. 2012. Electroporation enhances mitomycin C cytotoxicity on T24 bladder cancer cell line: a potential improvement of intravesical chemotherapy in bladder cancer. Bioelectrochemistry 88:127-133
Vásquez JL, Ibsen P, Lindberg H, Gehl J. 2015. In vitro and in vivo experiments on electrochemotherapy for bladder cancer. The Journal of Urology 193:1009-1015
Yushkevich PA, Piven J, Hazlett HC, Smith RG, Ho S, Gee JC, Gerig G. 2006. User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. NeuroImage 31:1116-1128
Bor Kos1, Juan Luis Vásquez2, Damijan Miklavčič1, Gregers G.G. Hermann3, Julie Gehl2 1 Faculty of Electrical Engineering, University of Ljubljana, Ljubljana, Slovenia 2 Department of Oncology, Copenhagen University Hospital Herlev, Copenhagen, Denmark 3 Department of Urology, Copenhagen University Hospital Frederiksberg, Copenhagen, Denmark
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2016 Kos et al. This is an open access article distributed under the terms of the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, reproduction and adaptation in any medium and for any purpose provided that it is properly attributed. For attribution, the original author(s), title, publication source (PeerJ) and either DOI or URL of the article must be cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Objective
Bladder cancer is a cause of considerable morbidity worldwide. Electromotive Drug Administration is a method that combines intravesical chemotherapy with local electric field application. Electroporation has been suggested among other mechanisms as having a possible role in the therapy, so the goal of the present study was to investigate the electric fields present in the bladder wall during the treatment to determine which mechanisms might be involved.
Material and Methods
Electromotive Drug Administration involves applying intravesical mitomycin C with direct current of 20 mA delivered through a catheter electrode for 30 min. For numerical electric field computation we built a 3-D nonhomogeneous patient specific model based on CT images and used finite element method simulations to determine the electric fields in the whole body.
Results
Results indicate that highest electric field in the bladder wall was 37.7 V/m. The mean electric field magnitude in the bladder wall was 3.03 V/m. The mean magnitude of the current density in the bladder wall was 0.61 A/m2.
Conclusions
The present study shows that electroporation is not the mechanism of action in EMDA. A more likely explanation of the mechanism of action is iontophoretic forces increasing the mitomycin C concentration in the bladder wall.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer