© 2017 de Beaufort et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Objectives

Thoracic endovascular aortic repair (TEVAR) has been shown to lead to increased aortic stiffness. The aim of this study was to investigate the effect of stent graft type and stent graft length on aortic stiffness in a controlled, experimental setting.

Methods

Twenty porcine thoracic aortas were connected to a pulsatile mock loop system. Intraluminal pressure was recorded at two sites in order to measure pulse wave velocity (PWV) for each aorta: before stent graft deployment (t1); after deployment of a 100-mm long stent graft (t2); and after distal extension through deployment of a second 100-mm long stent graft (t3). Four different types of stent grafts (Conformable Gore® TAG® Device, Bolton Relay® Device, Cook Zenith Alpha™, and Medtronic Valiant®) were evaluated.

Results

For the total cohort of 20 aortas, PWV increased by a mean 0.6 m/s or 8.9% of baseline PWV after deployment of a 100-mm proximal stent graft (P<0.001), and by a mean 1.4 m/s or 23.0% of baseline PWV after distal extension of the stent graft (P<0.001). Univariable regression analysis showed a significant correlation between aortic PWV and extent of stent graft coverage, (P<0.001), but no significant effect of baseline aortic length, baseline aortic PWV, or stent graft type on the percentual increase in PWV at t2 or at t3.

Conclusions

In this experimental set-up, aortic stiffness increased significantly after stent graft deployment with each of the four types of stent graft, with the increase in aortic stiffness depending on the extent of stent graft coverage.

Details

Title
Changes in aortic pulse wave velocity of four thoracic aortic stent grafts in an ex vivo porcine model
Author
Hector W L de Beaufort; Coda, Margherita; Conti, Michele; Theodorus M J van Bakel; Nauta, Foeke J H; Lanzarone, Ettore; Moll, Frans L; van Herwaarden, Joost A; Auricchio, Ferdinando; Trimarchi, Santi
First page
e0186080
Section
Research Article
Publication year
2017
Publication date
Oct 2017
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1991850735
Copyright
© 2017 de Beaufort et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.