Abstract

Over the past 60 years, the diamond anvil cell (DAC) has been developed into a widespread high static pressure device. The adaptation of laboratory and synchrotron analytical techniques to DAC enables a detailed exploration in the 100 GPa range. The strain of the anvils under high load explains the 400 GPa limit of the conventional DAC. Here we show a toroidal shape for a diamond anvil tip that enables to extend the DAC use toward the terapascal pressure range. The toroidal-DAC keeps the assets for a complete, reproducible, and accurate characterization of materials, from solids to gases. Raman signal from the diamond anvil or X-ray signal from the rhenium gasket allow measurement of pressure. Here, the equations of state of gold, aluminum, and argon are measured with X-ray diffraction. The data are compared with recent measurements under similar conditions by two other approaches, the double-stage DAC and the dynamic ramp compression.

Details

Title
Toroidal diamond anvil cell for detailed measurements under extreme static pressures
Author
Dewaele, Agnès 1 ; Loubeyre, Paul 1 ; Occelli, Florent 1 ; Olivier, Marie 1 ; Mezouar, Mohamed 2   VIAFID ORCID Logo 

 CEA, DAM, DIF, Arpajon, France 
 ESRF, BP220, Grenoble Cedex, France 
Pages
1-9
Publication year
2018
Publication date
Jul 2018
Publisher
Nature Publishing Group
e-ISSN
20411723
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2076221136
Copyright
© 2018. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.