Abstract

Background

The sweat test (ST) is the gold standard for the diagnosis of cystic fibrosis (CF). However, little is known about sweat induction using different types of currents and waves. In this context, our objective was to develop a device to induce sweat and compare the use of continuous constant current (CCC) and continuous pulsed current (CPC) in individuals with CF and healthy controls.

Methods

A prospective cross-sectional study with experimental intervention. The variables of gender, ethnicity, age, and body mass index (BMI) were considered. The method of Gibson and Cooke was used, and the following markers were evaluated: sweat weight, electrical impedance, sufficient sweat amount, and CF diagnosis. Triangular (TPC) or sinusoidal (SPC) pulsed current was applied to the right arm, and CCC was applied to the left arm.

Results

The study analyzed 260 individuals, 141/213 (54.2%) were female participants, 135/260 (51.9%) were Caucasians. The distribution of individuals by concentration of chloride at the ST was: (CF) 26/260 (10%); (borderlines) 109/260 (41.9%); (healthy) 97/260 (37.3%); (insufficient weight in sweat) 28/260 (10.8%). No association was observed between the sufficient sweat amount to perform the ST when we compared the currents. However, the SPC showed a higher amount of sweat weight. Using Bland and Altman plot considering the agreement between the sweat chloride values achieved from CPC [SPC and TPC] and CCC, there was no proportional bias and mean values are unrelated and only explain less than 8% of the variation. Moreover, TPC presented higher electrical impedance when compared with SPC and CCC. SPC presented lower electrical impedance and higher sweat weight than CCC. Male participants presented lower electrical impedance and higher sweat weight with CCC and TPC, and higher sweat weight with SPC.

Conclusions

The evaluated currents are safe and able to induce and produce sweat in sufficient quantities for the ST. SPC presented lower electrical impedance when compared with other currents. The use of SPC is recommended to induce sweat in patients with sweat problems. Finally, ethnicity, gender, age and BMI did not influence sweat induction at the ST, and no side effect was observed in our study.

Details

Title
Evaluation of continuous constant current and continuous pulsed current in sweat induction for cystic fibrosis diagnosis
Author
Souza Gomez, Carla Cristina; Lima Marson, Fernando Augusto; Maria Fátima Servidoni; Ribeiro, Antônio Fernando; Maria Ângela Gonçalves Oliveira Ribeiro; Veruska Acioli Lopes Gama; Eduardo Tavares Costa; José Dirceu Ribeiro; Francisco Ubaldo Vieira Junior
Publication year
2018
Publication date
2018
Publisher
BioMed Central
e-ISSN
14712466