Full text

Turn on search term navigation

© 2018 Hojreh et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

Modern CT scanners provide automatic dose adjustment systems, which are promising options for reducing radiation dose in pediatric CT scans. Their impact on patient dose, however, has not been investigated sufficiently thus far.

Objective

To evaluate automated tube voltage selection (ATVS) in combination with automated tube current modulation (ATCM) in non-contrast pediatric chest CT, with regard to the diagnostic image quality.

Materials and methods

There were 160 non-contrast pediatric chest CT scans (8.7±5.4 years) analyzed retrospectively without and with ATVS. Correlations of volume CT Dose Index (CTDIvol) and effective diameter, with and without ATVS, were compared using Fisher’s z-transformation. Image quality was assessed by mean signal-difference-to-noise ratios (SDNR) in the aorta and in the left main bronchus using the independent samples t-test. Two pediatric radiologists and a general radiologist rated overall subjective Image quality. Readers’ agreement was assessed using weighted kappa coefficients. A p value <0.05 was considered significant.

Results

CTDIvol correlation with the effective diameter was r = 0.62 without and r = 0.80 with ATVS (CI: -0.04 to -0.60; p = 0.025). Mean SDNR was 10.88 without and 10.03 with ATVS (p = 0.0089). Readers’ agreement improved with ATVS (weighted kappa between pediatric radiologists from 0.1 (0.03–0.16) to 0.27 (0.09–0.45) with ATVS; between general and each pediatric radiologist from 0.1 (0.06–0.14) to 0.12 (0.05–0.20), and from 0.22 (0.11–0.34) to 0.36 (0.24–0.49)).

Conclusion

ATVS, combined with ATCM, results in a radiation dose reduction for pediatric non-contrast chest CT without a loss of diagnostic image quality and prevents errors in manual tube voltage setting, and thus protecting larger children against an unnecessarily high radiation exposure.

Details

Title
Automated tube voltage selection in pediatric non-contrast chest CT
Author
Hojreh, Azadeh; ⨯ Peter Homolka; Gamper, Jutta; Unterhumer, Sylvia; ⨯ Daniela Kienzl-Palma; Balassy, Csilla; ⨯ Thomas Wrba; ⨯ Helmut Prosch
First page
e0204794
Section
Research Article
Publication year
2018
Publication date
Oct 2018
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2117861908
Copyright
© 2018 Hojreh et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.