Abstract

This paper discusses a procedure for measuring the absolute spectral responsivity of optical-fiber power meters and computation of the calibration uncertainty. The procedure reconciles measurement results associated with a monochromator-based measurement system with those obtained with laser sources coupled with optical fiber. Relative expanded uncertainties based on the methods from the Guide to the Expression of Uncertainty in Measurement and from Supplement 1 to the "Guide to the Expression of Uncertainty in Measurement"-Propagation of Distributions using a Monte Carlo Method are derived and compared. An example is used to illustrate the procedures and calculation of uncertainties. [PUBLICATION ABSTRACT]

Details

Title
Uncertainty Calculation for Spectral-Responsivity Measurements
Author
Lehman, John H; Wang, C M; Dowell, Marla L; Hadler, Joshua A
Pages
287-291
Publication year
2009
Publication date
Sep/Oct 2009
Publisher
Superintendent of Documents
ISSN
1044677X
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
214785807
Copyright
Copyright Superintendent of Documents Sep/Oct 2009