Full text

Turn on search term navigation

Copyright Superintendent of Documents Mar/Apr 2008

Abstract

NIST developed the alanine dosimetry system in the early 1990s to replace radiochromic dye film dosimeters. Later in the decade the alanine system was firmly established as a transfer service for high-dose radiation dosimetry and an integral part of the internal calibration scheme supporting these services. Over the course of the last decade, routine monitoring of the system revealed a small but significant observation that, after examination, led to the characterization of a previously unknown absorbed-dose-dependent, dose-rate effect for the alanine system. Though the potential impact of this effect is anticipated to be extremely limited for NIST's customer-based transfer dosimetry service, much greater implications may be realized for international measurement comparisons between National Measurement Institutes. [PUBLICATION ABSTRACT]

Details

Title
An Absorbed-Dose/Dose-Rate Dependence for the Alanine-EPR Dosimetry System and Its Implications in High-Dose Ionizing Radiation Metrology
Author
Desrosiers, M F; Puhl, J M; Cooper, S L
Pages
79-83,85-95
Publication year
2008
Publication date
Mar/Apr 2008
Publisher
Superintendent of Documents
ISSN
1044677X
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
214790842
Copyright
Copyright Superintendent of Documents Mar/Apr 2008