Abstract
Purpose: To evaluate the effect of selective laser trabeculoplasty (SLT) in normal tension glaucoma (NTG) patients.
Patients and methods: A retrospective review was performed of NTG patients who had undergone SLT at the Duke University Eye Center between 12/2002 and 7/2005. For each eye of each patient at pre-laser and post-laser time points, the IOP measurements were summarized by mean, standard deviation, and range. Then for each of these descriptive statistics, the differences between pre-laser and post-laser values were obtained. Statistical analysis was performed using a random effects model. Main outcome measures: difference in mean IOP, standard deviation of IOP, and range of IOP.
Results: Thirty-one eyes of 18 patients were included for analysis. The average of the mean pre-operative IOP measurements was 14.3 ± 2.6 mmHg compared to 12.2 ± 1.7 mmHg (P < 0.001) post-operatively. The mean pre-operative standard deviation was 1.9 ± 0.9 mmHg compared to 1.0 ± 0.6 mmHg (P = 0.002) post-operatively while the mean IOP range prior to treatment was 4.5 ± 2.5 mmHg compared to 2.5 ± 1.9 mmHg (P = 0.017) after treatment.
Conclusion: In this pilot study, SLT was found to lower mean IOP and intervisit IOP variation in NTG patients. Given the importance of IOP variation and its association with glaucoma progression, measurement of IOP variation following treatment with SLT may be considered.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer