Abstract
Background
Changes in oxygen saturation (SpO2) exposure have been shown to have a marked impact on neonatal outcomes and therefore careful titration of inspired oxygen is essential. In routine use, however, the frequency of SpO2 alarms not requiring intervention results in alarm fatigue and its corresponding risk. SpO2 control systems that automate oxygen adjustments (Auto-FiO2) have been shown to be safe and effective. We speculated that when using Auto-FiO2, alarm settings could be refined to reduce unnecessary alarms, without compromising safety.
Methods
An unblinded randomized crossover study was conducted in a single NICU among infants routinely managed with Auto-FiO2. During the first 6 days of respiratory support a tight and a loose alarm strategy were switched each 24 h. A balanced block randomization was used. The tight strategy set the alarms at the prescribed SpO2 target range, with a 30-s delay. The loose strategy set the alarms 2 wider, with a 90-s delay. The effectiveness outcome was the frequency of SpO2 alarms, and the safety outcomes were time at SpO2 extremes (< 80, > 98%). We hypothesized that the loose strategy would result in a marked decrease in the frequency of SpO2 alarms, and no increases at SpO2 extremes with 20 subjects. Within subject differences between alarm strategies for the primary outcomes were evaluated with Wilcoxon signed-rank test.
Results
During a 13-month period 26 neonates were randomized. The analysis included 21 subjects with 49 days of both tight and loose intervention. The loose alarm strategy resulted in a reduction in the median rate of SpO2 alarms from 5.2 to 1.6 per hour (p < 0.001, 95%-CI difference 1.6–3.7). The incidence of hypoxemia and hyperoxemia were very low (less than 0.1%-time) with no difference associated with the alarm strategy (95%-CI difference less than 0.0–0.2%).
Conclusions
In this group of infants we found a marked advantage of the looser alarm strategy. We conclude that the paradigms of alarm strategies used for manual titration of oxygen need to be reconsidered when using Auto-FiO2. We speculate that with optimal settings false positive SpO2 alarms can be minimized, with better vigilance of clinically relevant alarms.
Trial registration
Retrospectively registered 15 May 2018 at ISRCTN (49239883).
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer