© 2019 Ceresnak et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

In previous pilot work we demonstrated that a novel automated signal analysis tool could accurately identify successful ablation sites during Wolff-Parkinson-White (WPW) ablation at a single center.

Objective

We sought to validate and refine this signal analysis tool in a larger multi-center cohort of children with WPW.

Methods

A retrospective review was performed of signal data from children with WPW who underwent ablation at two pediatric arrhythmia centers from 2008–2015. All patients with WPW ≤ 21 years who underwent invasive electrophysiology study and ablation with ablation signals available for review were included. Signals were excluded if temperature or power delivery was inadequate or lesion time was < 5 seconds. Ablation lesions were reviewed for each patient. Signals were classified as successful if there was loss of antegrade and retrograde accessory pathway (AP) conduction or unsuccessful if ablation did not eliminate AP conduction. Custom signal analysis software analyzed intracardiac electrograms for amplitudes, high and low frequency components, integrated area, and signal timing components to create a signal score. We validated the previously published signal score threshold 3.1 in this larger, more diverse cohort and explored additional scoring options. Logistic regression with lasso regularization using Youden’s index criterion and a cost-benefit criterion to identify thresholds was considered as a refinement to this score.

Results

347 signals (141 successful, 206 unsuccessful) in 144 pts were analyzed [mean age 13.2 ± 3.9 years, 96 (67%) male, 66 (45%) left sided APs]. The software correctly identified the signals as successful or unsuccessful in 276/347 (80%) at a threshold of 3.1. The performance of other thresholds did not significantly improve the predictive ability. A signal score threshold of 3.1 provided the following diagnostic accuracy for distinguishing a successful from unsuccessful signal: sensitivity 83%, specificity 77%, PPV 71%, NPV 87%.

Conclusions

An automated signal analysis software tool reliably distinguished successful versus unsuccessful ablation electrograms in children with WPW when validated in a large, diverse cohort. Refining the tools using an alternative threshold and statistical method did not improve the original signal score at a threshold of 3.1. This software was effective across two centers and multiple operators and may be an effective tool for ablation of WPW.

Details

Title
Validation of a novel automated signal analysis tool for ablation of Wolff-Parkinson-White Syndrome
Author
Ceresnak, Scott R; Pass, Robert H; Dubin, Anne M; Yang, Lingyao; Motonaga, Kara S; Hedlin, Haley; Avasarala, Kishor; Trela, Anthony; McElhinney, Doff B; Janson, Christopher; Nappo, Lynn; Ling, Xuefeng B; Gates, Gregory J
First page
e0217282
Section
Research Article
Publication year
2019
Publication date
Jun 2019
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2247705696
Copyright
© 2019 Ceresnak et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.