Introduction
The conformance of Primary Diagnostic Monitors (PDMs) used to make a final interpretation from images generated by radiological devices to the Digital Imaging and Communication in Medicine Grayscale Standard Display Function is increasingly required by several state and city regulators. For example, effective from 1 January 2016, all radiological equipment registrants in New York City (NYC) are required to extend their Quality Assurance (QA) program to their PDMs. This work describes our experience in implementing the PDM performance and conformance with the NYC QA program and characterization of the luminance measurement performance of the internal, built-in photometer of Barco monitors against an externally calibrated luminance meter.
Methods
During the on-site annual testing of PDMs, several luminance characteristics were evaluated, including GSDF conformance, maximum luminance (Lmax), luminance ratio (LR), and luminance uniformity.
Measurements were performed on 79 PDMs: color display system models MDCC-6130 (41), MDCC-6230 (9), MDCC-6330 (1) and Grayscale display system models MDCG 3120-CB (2), MDCG-10130 (2), MFGD 3420 (6) and MFGD 5421 (18) (Barco; Duluth, Georgia) as part of annual QA testing following the NYC regulations. Monitors were located at various hospital sites, radiology physicians’ offices, and radiology reading rooms. All measured PDMs were liquid crystal displays equipped with the manufacturer's built-in photometers (Barco, I-Guard) and connected to Barco MediCal QAWeb service for manual and automatic quality control measurements. PDM combinations included 1, 2, and 4 monitors, depending on the location. TG-18 and Society of Motion Picture and Television Engineers test patterns were used to evaluate monitor performance. An externally calibrated photometer (RaySafe Solo Light; Billdal, Sweden) was used to measure the luminance values.
Measurement and verification of Grayscale standard display function
For each PDM, the luminance response is expressed as the change in luminance between gray levels (dL/L) for the just noticeable difference plotted against the digital driving level (DDL). To validate the results of GSDF calibration that was performed using the built-in photometer, TG18LN1-18 luminance patterns were measured with an external photometer. All measurements were performed under normal reading conditions. NYC QA guidelines and AAPM Task Group 18 require 10% tolerance for GSDF calibration.
Maximum luminance and luminance ratio
The perceived contrast characteristics of an image depend on the ratio of maximum luminance Lmax (all white) to minimum luminance Lmin (all black). Luminance values were measured with an external photometer using TG18-LN01 and TG18-LN18 test patterns. Following NYC regulations and the ACR technical standard, the maximum luminance Lmax of nonmammography PDMs should be at least 350 cd/m2. The LR was calculated using the following equation:
Evaluation of luminance uniformity
Using an external photometer, luminance values were measured at five locations on the PDM: the center and at each quadrant using the TG18-UNL80 test pattern. The luminance uniformity (LU) of a display was calculated using the following equation:
Because the built-in photometer can measure luminance only at a fixed location (the bottom right corner for PDM models MDCG 3120-CB, MFGD 3420, and MFGD 5421 and the top middle portion of the screen for PDM models MDCC-6130, MDCC-6230, MDCC-6330, and MDCG-10130), a LU test is performed only with an external photometer.
Ambient light measurements
A RaySafe Solo Light photometer used in the study was equipped with the clip-on cover that eliminates the effect of ambient light on luminance measurements. All measurements of the luminance test patterns are performed with direct contact of the photometer with the measured PDM. The ambient luminance values were measured as a reflected ambient light from the screen of a turned off monitor (one foot away from the center of the monitor). These measurements were used to adjust the monitor GSDF calibration using the formula:
In addition to luminance measurements, ambient light was measured to confirm that the monitors did not receive outside light flow and they provide reading conditions mentioned in the ACR manual.
Illuminance was measured using the same RaySafe light photometer with the illuminance sensor positioned in the center of the display facing outward while the PDM was turned off. Illuminance values are used to estimate normal reading conditions (20-40 Lux).
Results
Grayscale standard display function
Figure represents the results of GSDF conformance tests performed automatically by the built-in luminance meter (a) and by an external photometer (b). For the built-in luminance meter, the measured luminance values for each DDL were within the 10% tolerance limit for 77 monitors and failed for two monitors (15% and 11.3%). Measurements performed by the external luminance meter showed that 15 out of 79 monitors had failed the 10% tolerance limit within the range of 10.7% and 31.1%. The average deviation of contrast response for all 79 monitors is 8.2% ± 5.6%.
[IMAGE OMITTED. SEE PDF]
Maximum luminance and luminance ratio
Figures and represent the Lmax values measured with an external photometer for PDMs with a target calibration of Lmax = 400 cd/m2 (Fig. ) and Lmax = 500 cd/m2 (Fig. ). Measurement results showed that 10 out of 53 PDMs with a target calibration of Lmax = 400 cd/m2 and 17 out of 26 PDMs with a target calibration of Lmax = 500 cd/m2 did not meet the manufacturer's recommended 10% tolerance limit for the target Lmax calibration. Two out of 79 monitors did not pass the Lmax ≥ 350 cd/m2 NYC QA regulations with values of Lmax = 331 cd/m2 and Lmax = 340 cd/m2. The difference between built-in and external photometer luminance measurements for Lmax was 5.1% ± 4.2% for all tested PDMs.
[IMAGE OMITTED. SEE PDF]
[IMAGE OMITTED. SEE PDF]
Figure shows the results of LR measurements performed by a built-in photometer (a) and an external luminance meter (b). All tested PDMs exceeded the luminance ratio of 250:1 as required by NYC PDM guidelines.
[IMAGE OMITTED. SEE PDF]
Evaluation of luminance uniformity
Figure represents the results of luminance uniformity measurements performed by an external photometer. All PDMs passed the luminance uniformity test, which was 30% according to NYC regulations. Luminance nonuniformity up to 17% was observed in PDMs connected in 2009 through 2016. The Barco MDCC-6130 model connected in 2012 and 2013 had luminance nonuniformity values less than 6%.
[IMAGE OMITTED. SEE PDF]
Ambient light and illuminance measurements
The mean value of illuminance measured with an external meter was 6.2 lux ± 3.6 lux. The mean ambient luminance was 0.19 cd/m2 ± 0.11 cd/m2. These levels are well below the ACR recommended 20 lux for a nonmammography diagnostic monitor radiology reading room.
Discussion
The evaluation of Barco primary diagnostic monitors of various ages and models demonstrated a discrepancy between luminance measurements performed by the manufacturer's built-in photometer when compared with those performed by an externally calibrated luminance meter. This discrepancy was larger for older PDMs, confirming the findings of Silovsky and Marsh. However, the number of PDMs manufactured after 2014 also showed a disagreement in luminance values measured by external and built-in photometers. All tested PDM models had a built-in photometer sensor positioned either at the bottom right corner (models MDCG 3120-CB, MFGD 3420, and MFGD 5421) or at the top middle part of the screen (MDCC-6130, MDCC-6230, MDCC-6330, and MDCG-10130). Figure represents the luminance uniformity for all tested PDM models, sorted by the age of the monitor, where the first monitor is Barco MFGD 5421 (September 2009) and the 79th monitor is Barco MDCC 6330 (January 2016) (See Table ). Barco MDCC-6130 monitors (21–23, 28–33, 36–66, and 75) showed the lowest nonuniformity values, between 2.1% and 6.4% with a mean value of 4.26% and standard deviation of 1.21%. Table summarizes the luminance uniformity and the difference in the maximum luminance measurements between the built-in and external photometers for the Barco PDM model MDCC-6130. There is no direct dependence between PDM uniformity and the difference in luminance measurements for the external and built-in photometers as can be observed.
List of all measured Barco PDM models included in the study and sorted by the year of first use
# | Model | In use since | # | Model | In use since | # | Model | In use since |
1 | MFGD 5421 | 09/2009 | 28 | MDCC-6130 | 03/2012 | 55 | MDCC-6130 | 01/2014 |
2 | MFGD 5421 | 09/2009 | 29 | MDCC-6130 | 05/2012 | 56 | MDCC-6130 | 01/2014 |
3 | MFGD 5421 | 10/2009 | 30 | MDCC-6130 | 05/2012 | 57 | MDCC-6130 | 01/2014 |
4 | MFGD 5421 | 10/2009 | 31 | MDCC-6130 | 05/2012 | 58 | MDCC-6130 | 02/2014 |
5 | MFGD 5421 | 10/2009 | 32 | MDCC-6130 | 05/2012 | 59 | MDCC-6130 | 03/2014 |
6 | MFGD 5421 | 10/2009 | 33 | MDCC-6130 | 05/2012 | 60 | MDCC-6130 | 03/2014 |
7 | MDCG 3120-CB | 11/2009 | 34 | MFGD 3420 | 06/2012 | 61 | MDCC-6130 | 03/2014 |
8 | MDCG 3120-CB | 11/2009 | 35 | MFGD 3420 | 06/2012 | 62 | MDCC-6130 | 03/2014 |
9 | MFGD 3420 | 11/2009 | 36 | MDCC-6130 | 09/2012 | 63 | MDCC-6130 | 03/2014 |
10 | MFGD 3420 | 11/2009 | 37 | MDCC-6130 | 10/2012 | 64 | MDCC-6130 | 03/2014 |
11 | MFGD 5421 | 11/2009 | 38 | MDCC-6130 | 11/2012 | 65 | MDCC-6130 | 03/2014 |
12 | MFGD 5421 | 11/2009 | 39 | MDCC-6130 | 11/2012 | 66 | MDCC-6130 | 03/2014 |
13 | MFGD 5421 | 11/2009 | 40 | MDCC-6130 | 08/2013 | 67 | MDCG-10130 | 05/2014 |
14 | MFGD 5421 | 11/2009 | 41 | MDCC-6130 | 08/2013 | 68 | MDCG-10130 | 05/2014 |
15 | MFGD 5421 | 11/2009 | 42 | MDCC-6130 | 08/2013 | 69 | MDCC-6230 | 06/2014 |
16 | MFGD 5421 | 11/2009 | 43 | MDCC-6130 | 08/2013 | 70 | MDCC-6230 | 09/2014 |
17 | MFGD 5421 | 11/2009 | 44 | MDCC-6130 | 08/2013 | 71 | MDCC-6230 | 09/2014 |
18 | MFGD 5421 | 11/2009 | 45 | MDCC-6130 | 08/2013 | 72 | MDCC-6230 | 11/2014 |
19 | MFGD 3420 | 06/2011 | 46 | MDCC-6130 | 08/2013 | 73 | MDCC-6230 | 11/2014 |
20 | MFGD 3420 | 06/2011 | 47 | MDCC-6130 | 08/2013 | 74 | MDCC-6230 | 11/2014 |
21 | MDCC-6130 | 03/2012 | 48 | MDCC-6130 | 08/2013 | 75 | MDCC-6130 | 12/2014 |
22 | MDCC-6130 | 03/2012 | 49 | MDCC-6130 | 12/2013 | 76 | MDCC-6230 | 06/2015 |
23 | MDCC-6130 | 03/2012 | 50 | MDCC-6130 | 12/2013 | 77 | MDCC-6230 | 06/2015 |
24 | MFGD 5421 | 03/2012 | 51 | MDCC-6130 | 01/2014 | 78 | MDCC-6230 | 06/2015 |
25 | MFGD 5421 | 03/2012 | 52 | MDCC-6130 | 01/2014 | 79 | MDCC-6330 | 01/2016 |
26 | MFGD 5421 | 03/2012 | 53 | MDCC-6130 | 01/2014 | |||
27 | MFGD 5421 | 03/2012 | 54 | MDCC-6130 | 01/2014 |
Luminance uniformity and maximum luminance values for the PDM Barco MDCC 6130 measured by external and built-in photometers
# | Lmax external meter | Lmax built-in meter | Nonuniformity | Difference for Lmax | # | Lmax external meter | Lmax built-in meter | Nonuniformity | Difference for Lmax |
(cd/m2) | (cd/m2) | (%) | (%) | (cd/m2) | (cd/m2) | (%) | (%) | ||
1 | 383.2 | 400.6 | 3.4 | 4.3 | 22 | 372.5 | 401.0 | 4.7 | 7.1 |
2 | 413.3 | 400.9 | 2.3 | 3.1 | 23 | 391.9 | 400.9 | 5.0 | 2.2 |
3 | 395.5 | 400.4 | 4.8 | 1.2 | 24 | 401.5 | 400.7 | 5.8 | 0.2 |
4 | 409.3 | 400.3 | 3.3 | 2.2 | 25 | 384.2 | 400.7 | 5.5 | 4.1 |
5 | 382.3 | 400.0 | 4.2 | 4.4 | 26 | 397.0 | 400.6 | 3.7 | 0.9 |
6 | 398.7 | 400.5 | 4.2 | 0.5 | 27 | 403.7 | 400.7 | 5.0 | 0.8 |
7 | 331.4 | 400.3 | 3.3 | 17.2 | 28 | 407.3 | 400.4 | 3.1 | 1.7 |
8 | 417.8 | 400.4 | 5.9 | 4.4 | 29 | 391.4 | 400.4 | 3.3 | 2.2 |
9 | 391.3 | 400.6 | 4.1 | 2.3 | 30 | 395.0 | 400.6 | 4.5 | 1.4 |
10 | 399.8 | 400.5 | 2.6 | 0.2 | 31 | 360.8 | 400.6 | 2.1 | 9.9 |
11 | 370.5 | 400.8 | 4.8 | 7.6 | 32 | 396.9 | 400.5 | 4.9 | 0.9 |
12 | 340.2 | 400.3 | 3.9 | 15.0 | 33 | 388.8 | 400.7 | 5.8 | 3.0 |
13 | 368.8 | 400.5 | 6.4 | 7.9 | 34 | 400.4 | 400.7 | 3.8 | 0.1 |
14 | 382.0 | 400.6 | 4.3 | 4.6 | 35 | 419.9 | 400.7 | 2.4 | 4.8 |
15 | 380.1 | 400.7 | 5.4 | 5.1 | 36 | 382.0 | 400.3 | 5.8 | 4.6 |
16 | 368.0 | 399.8 | 4.9 | 8.0 | 37 | 385.6 | 400.6 | 3.1 | 3.8 |
17 | 379.8 | 400.6 | 2.5 | 5.2 | 38 | 385.8 | 400.4 | 5.4 | 3.7 |
18 | 381.0 | 400.6 | 3.8 | 4.9 | 39 | 394.3 | 400.8 | 3.7 | 1.6 |
19 | 388.5 | 400.2 | 5.6 | 2.9 | 40 | 387.3 | 400.7 | 6.4 | 3.3 |
20 | 370.5 | 400.7 | 3.7 | 7.5 | 41 | 350.3 | 400.6 | 2.2 | 12.6 |
21 | 393.3 | 400.3 | 5.1 | 1.8 |
However, as indicated in Table , built-in photometer measurements appear to be remarkably consistent across a large number of monitors even though the external photometer measurements vary significantly. Because PDMs are automatically calibrated by the system software, this software uses an internal photometer calibration value of 400 cd/m2. Therefore, in the event that the luminance value of a PDM fluctuates from the calibration value, the internal photometer will always read luminance values close to 400 cd/m2 because that is the calibration point. To test this, additional tests were run where the luminance value of the same monitor using four different external photometers (RaySafe Solo Light, RaySafe Xi, Konica–Minolta CA-210, and Konica-Minolta CS-100A) was measured. The results showed that luminance values varied 0.8% between these photometers (the results are not presented here). In light of this test, we speculate that external measurements are closest to an actual luminance value, whereas internal measurement is biased to the PDM calibration level.
One area of future work is to examine the effect of monitor warm-up time on the maximum luminance measurement by external and built-in photometers. As a small pilot study, 5 Barco MDCC-6130 PDMs located in the same radiology reading room were measured under the same conditions. The mean change in maximum luminance measurements during the first 30 min after the monitor was turned on (following a 12-hour turned off period), was 12.7 cd/m2 ± 7.6 cd/m2 for the measurements performed by an external photometer and 0.9 cd/m2 ± 0.5 cd/m2 for those performed by the built-in photometer. This shows the sensitivity of an external photometer to PDM warm-up period and confirms the importance of performing the PDM QA measurements after at least a 30-minute warm-up time for the monitor. We plan to examine the possible effect of the PDM's temperature during GSDF calibration on its performance and luminance characteristics.
Conclusion
We have investigated internal and external photometer performance of PDMs and have found that there are many more monitors that fail external photometer luminance measurements compared with those measured with an internal photometer when tested according to NYC PDM guidelines.
The discrepancy between the luminance measurement by external and built-in photometers should be followed up with the hospital IT department and manufacturer shortly after PDM testing.
Acknowledgments
None.
Conflict of interest
None.
NYC. Quality Assurance Guidelines for Primary Diagnostic Monitors, 2015. New York, NY: New York City Health Department's Office of Radiological Health (ORH); 2015.
Barten P. Physical model for the contrast sensitivity of the human eye. SPIE Human Vision, Visual Processing and Digital Display III, Vol. 57. 1992.
Digital Imaging and Communications in Medicine (DICOM). Part 14: Grayscale Standard Display Function. Report No.: PS 3.14‐2004. Rosslyn, VA: National Electrical Manufacturers Association; 2004, 55 pp.
Samei E, Badano A, Chakraborty D, et al. Assessment of display performance for medical imaging systems [Internet]. College Park, MD: American Association of Physicists in Medicine; 2005. 155. Available from: https://www.aapm.org/pubs/reports/OR_03.pdf
American College of Radiology (ACR), The American Association of Physicists in Medicine (AAPM), The Society for Imaging Informatics in Medicine (SIIM); ACR‐AAPM‐SIIM Technical Standard for Electronic Practice of Medical Imaging. 2014 [Published 2007. Amended 2014 (Resolution39)]; 2014, 18 pp. Available from: http://www.acr.org/~/media/AF1480B0F95842E7B163F09F1CE00977.pdf
Silosky M, Marsh RM. Constancy of built‐in luminance meter measurements in diagnostic displays. Med Phys. 2013;40:121902.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright John Wiley & Sons, Inc. 2017
Abstract
The purpose of this work was to perform the initial evaluation of primary diagnostic monitor (
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer