© 2019 Weiss et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Introduction

In cases with residual-hearing (RH) loss after cochlear implantation, a safe method is needed to provide full spectral resolution and as much auditory information as possible without implant replacement. Aim of this study was to prove the feasibility of accessing a partially inserted cochlear-implant-electrode for complete insertion to its maximum length through the external ear canal using a transcanal approach.

Methods

Two CI electrodes were customized with 18 stimulating channels. The electrode design enables the use of 12 active channels available for electrical stimulation inside the cochlea both after partial and full insertion. 10 CI electrodes were implanted in 10 fresh human cadaveric temporal bones. After initial partial insertion by posterior tympanotomy, the electrode was inserted to its maximum length via a transcanal approach. Radiographs and CT scans were performed to confirm the electrode position. The electrodes were investigated via x-ray after removal.

Results

X-ray and CT-scans confirmed the electrode prototypes covering an angular insertion depth between 236° to 307° after initial insertion. Accessing the electrode in the middle ear space was feasible and insertion to its full length was successful. Post-insertion CT confirmed insertion of the 28mm and 31.5mm electrode arrays covering an angular insertion depth between 360° and 540° respectively. No tip foldovers were detected.

Conclusion

This study confirms the feasibility of extending the electrode insertion to its maximum insertion length using a transcanal approach in temporal bone specimens. This constitutes a second stage procedure on demand in EAS-surgery. This may be beneficial for EAS-patients providing electrical stimulation beyond the basal turn of the cochlea once the functional residual hearing is lost, without replacing the entire CI.

Details

Title
Surgical approach for complete cochlear coverage in EAS-patients after residual hearing loss
Author
Weiss, Nora M; Anandhan Dhanasingh; Schraven, Sebastian P; Schulze, Marko; Langner, Soenke; Mlynski, Robert
First page
e0223121
Section
Research Article
Publication year
2019
Publication date
Sep 2019
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2298559017
Copyright
© 2019 Weiss et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.