Full Text

Turn on search term navigation

© 2019 Wang et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Purpose

Central foveal thickness (CFT) measurements from optical coherence tomography (OCT) scans provide a precise measure of severity of pathologic changes in the fovea, progress of disease and response to treatment. Although these measures are additionally valuable to assess foveal development in infants, their reproducibility is not known. The goal of this retrospective study is to evaluate the variation and reproducibility of CFT measurements using handheld spectral-domain OCT (hh-SDOCT) in supine infants compared to conventional adult tabletop imaging.

Methods

Imaging sessions with multiple macular, volume scans in one eye were selected for analysis from two participant groups: Group 1, 25 imaging sessions from 21 preterm infants without macular edema imaged supine in the nursery using hh-SDOCT (Leica/Bioptigen Envisu C2300, RTP, NC); Group 2, 25 imaging sessions from 25 adults imaged using tabletop Bioptigen SDOCT. For each imaging session, three macular OCT volumes with acceptable image quality were selected for analysis. CFTs were measured using a customized script for automatic segmentation. An expert grader and a typical grader corrected the segmentation lines for the central foveal frame. Coefficient of variations (CV) and intraclass correlation coefficients (ICC) were calculated for graders and systems and compared to the previous literature on OCT reproducibility.

Results

CFT measurements were repeatable and reproducible for both handheld and tabletop SDOCT systems. For handheld, grader ICC (CI) and mean CV were 0.94 (0.90–0.97) and 3.8 (typical) and 0.98 (0.96–0.99) and 2.9 (expert), and for tabletop were 0.91(0.83–0.96) and 2.1 (typical) and 0.92 (0.86–0.96) and 1.9 (expert). Intergrader reproducibility of handheld and tabletop SDOCT systems were ICC(CI) 0.97 (0.95–0.98) and 0.93 (0.89–0.96) respectively, and both are comparable to previously reported reproducibility of tabletop systems.

Conclusion

Handheld SDOCT is a reproducible instrument to measure foveal thicknesses in supine infants. It can be used in clinical research to evaluate foveal changes during retinal development and pathological conditions.

Details

Title
Understanding the variability of handheld spectral-domain optical coherence tomography measurements in supine infants
Author
Wang, Kira L; Chen, Xi; Stinnett, Sandra; Tai, Vincent; Winter, Katrina P; Tran-Viet, Du; Toth, Cynthia A
First page
e0225960
Section
Research Article
Publication year
2019
Publication date
Dec 2019
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2324896075
Copyright
© 2019 Wang et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.