Abstract

Background

The utility of standard distal bipolar electrograms (sEGMs) for assessing catheter-tissue contact may be obscured by the presence of far-field signals. Microelectrode electrograms (mEGMs) may overcome this limitation.

Methods

We compared 5 mEGM characteristics (amplitude, frequency content, temporal signal variability, presence of injury current, and amplitude differential between bipoles) with the sEGM for determining tissue contact in 20 patients undergoing ablation of typical atrial flutter. Visualization of catheter-tissue contact by intracardiac echocardiography (ICE) served as the gold standard for assessing contact. Correlation between electrograms and ICE-verified contact level was reported as percent concordance.

Results

Three of 5 mEGM characteristics demonstrated significantly better concordance with ICE-verified contact level than the sEGM (52% concordance with ICE): mEGM frequency content (59% concordance with ICE, P < .001 for comparison with sEGM); mEGM amplitude (concordance 59%, P < .001); and mEGM presence of injury current (56% concordance, P = .001). Concordance of amplitude differential between mEGM bipoles with ICE (49%) was not significantly different than the sEGM (P = .638) whereas mEGM temporal variability (39%) was significantly worse than the sEGM. Using a median of all 5 mEGM characteristics provided additive information (concordance with ICE 64%) and was significantly better than all of the individual mEGM characteristics except frequency content (P = .976).

Conclusion

Microelectrode EGMs (in particular frequency content, amplitude, and presence of injury current) can improve real-time assessment of catheter contact compared to the use of standard bipolar EGMs. Broader use of mEGMs may enhance ablation efficacy.

Details

Title
Use of microelectrode near-field signals to determine catheter contact
Author
Levy, Mathew R 1 ; Merchant, Faisal M 1   VIAFID ORCID Logo  ; Langberg, Jonathan J 1 ; Delurgio, David B 1 

 Division of Cardiology, Section of Cardiac Electrophysiology, Emory University School of Medicine, Atlanta, GA, USA 
Pages
23-29
Section
ORIGINAL ARTICLES
Publication year
2018
Publication date
Feb 2018
Publisher
John Wiley & Sons, Inc.
ISSN
1880-4276
e-ISSN
1883-2148
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2329741235
Copyright
© 2018. This work is published under http://creativecommons.org/licenses/by-nc-nd/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.