© 2019 Chung et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

VISION Max (Ortho-Clinical Diagnostics, Raritan, NJ, USA) is a newly introduced automated blood bank system. Cross-matching (XM) is an important test confirming safety by simulating reaction between packed Red Blood Cells (RBCs) and patient blood in vitro before transfusion. We assessed the benefits of VISION Max automated XM (A-XM) in comparison with those of manual XM (M-XM) by using multidimensional analysis (cost-effectiveness and quality improvement).

Materials and methods

In a total of 327 tests (130 patients), results from A-XM and M-XM were compared. We assessed the concordance rate, risk priority number (RPN), turnaround time, hands-on time, and the costs of both methods. We further simulated their annual effects based on 37,937 XM tests in 2018.

Results

The concordance rate between A-XM and M-XM was 97.9% (320/327, kappa = 0.83), and the seven discordant results were incompatible for transfusion in A-XM, while compatible for transfusion in M-XM. None of the results was incompatible for transfusion in A-XM, while compatible for transfusion in M-XM, meaning A-XM detect agglutination more sensitively and consequently provides a more safe result than M-XM. A-XM was estimated to have a 6.3-fold lower risk (229 vs. 1,435 RPN), shorter turnaround time (19.1 vs. 23.3 min, P < 0.0001), shorter hands-on time (1.1 vs. 5.3 min, P < 0.0001), and lower costs per single test than M-XM (1.44 vs. 2.70 USD). A-XM permitted annual savings of 46 million RPN, 15.1 months of daytime workers’ labor, and 47,042 USD compared with M-XM.

Conclusion

This is the first attempt to implement A-XM using VISION Max. VISION Max A-XM appears to be a safe, practical, and reliable alternative for pre-transfusion workflow with the potential to improve quality and cost-effectiveness in the blood bank.

Details

Title
Benefits of VISION Max automated cross-matching in comparison with manual cross-matching: A multidimensional analysis
Author
Hee-Jung, Chung; Hur, Mina; Choi, Sang Gyeu; Hyun-Kyung, Lee; Lee, Seungho; Kim, Hanah; Hee-Won Moon; Yeo-Min, Yun
First page
e0226477
Section
Research Article
Publication year
2019
Publication date
Dec 2019
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2330057958
Copyright
© 2019 Chung et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.