Abstract

Electron dosimetry can be performed using cylindrical chambers, plane‐parallel chambers, and diode detectors. The finite volume of these detectors results in a displacement effect which is taken into account using an effective point of measurement (EPOM). Dosimetry protocols have recommended a shift of 0.5 rcav for cylindrical chambers; however, various studies have shown that the optimal shift may deviate from this recommended value. This study investigated the effect that the selection of EPOM shift for cylindrical chamber has on percentage depth dose (PDD) curves. Depth dose curves were measured in a water phantom for electron beams with energies ranging from 6 to 18 MeV. The detectors investigated were of three different types: diodes (Diode‐E PTW 60017 and SFD IBA), cylindrical (Semiflex PTW 31010, PinPoint PTW 31015, and A12 Exradin), and parallel plate ionization chambers (Advanced Markus PTW 34045 and Markus PTW 23343). Depth dose curves measured with Diode‐E and Advanced Markus agreed within 0.2 mm at R50 except for 18 MeV and extremely large field size. The PDDs measured with the Semiflex chamber and Exradin A12 were about 1.1 mm (with respect to the Advanced Markus chamber) shallower than those measured with the other detectors using a 0.5 rcav shift. The difference between the PDDs decreased when a Pinpoint chamber, with a smaller cavity radius, was used. Agreement improved at lower energies, with the use of previously published EPOM corrections (0.3 rcav). Therefore, the use of 0.5 rcav as an EPOM may result in a systematic shift of the therapeutic portion of the PDD (distances < R90). Our results suggest that a 0.1 rcav shift is more appropriate for one chamber model (Semiflex PTW 31010).

Details

Title
The use of 0.5r cav as an effective point of measurement for cylindrical chambers may result in a systematic shift of electron percentage depth doses
Author
Anusionwu, Princess C 1 ; Alpuche Aviles, Jorge E 2 ; Pistorius, Stephen 2 

 CancerCare Manitoba, Winnipeg, Canada; Department of Physics & Astronomy, University of Manitoba, Winnipeg, Canada 
 CancerCare Manitoba, Winnipeg, Canada; Department of Physics & Astronomy, University of Manitoba, Winnipeg, Canada; Department of Radiology, University of Manitoba, Winnipeg, Canada 
Pages
117-126
Section
RADIATION MEASUREMENTS
Publication year
2020
Publication date
Jan 2020
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2338983448
Copyright
© 2020. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.