Abstract

Purpose

There remain uncertainties due to inter‐ and intraobserver variability in soft‐tissue‐based patient positioning even with the use of image‐guided radiation therapy (IGRT). This study aimed to reveal observer uncertainties of soft‐tissue‐based patient positioning on cone‐beam computed tomography (CBCT) images for prostate cancer IGRT.

Methods

Twenty‐six patients (7–8 fractions/patient, total number of 204 fractions) who underwent IGRT for prostate cancer were selected. Six radiation therapists retrospectively measured prostate cancer location errors (PCLEs) of soft‐tissue‐based patient positioning between planning CT (pCT) and pretreatment CBCT (pre‐CBCT) images after automatic bone‐based registration. Observer uncertainties were evaluated based on residual errors, which denoted the differences between soft‐tissue and reference positioning errors. Reference positioning errors were obtained as PCLEs of contour‐based patient positioning between pCT and pre‐CBCT images. Intraobserver variations were obtained from the difference between the first and second soft‐tissue‐based patient positioning repeated by the same observer for each fraction. Systematic and random errors of inter‐ and intraobserver variations were calculated in anterior–posterior (AP), superior–inferior (SI), and left–right (LR) directions. Finally, clinical target volume (CTV)‐to‐planning target volume (PTV) margins were obtained from systematic and random errors of inter‐ and intraobserver variations in AP, SI, and LR directions.

Results

Interobserver variations in AP, SI, and LR directions were 0.9, 0.9, and 0.5 mm, respectively, for the systematic error, and 1.8, 2.2, and 1.1 mm, respectively, for random error. Intraobserver variations were <0.2 mm in all directions. CTV‐to‐PTV margins in AP, SI, and LR directions were 3.5, 3.8, and 2.1 mm, respectively.

Conclusion

Intraobserver variability was sufficiently small and would be negligible. However, uncertainties due to interobserver variability for soft‐tissue‐based patient positioning using CBCT images should be considered in CTV‐to‐PTV margins.

Details

Title
Observer uncertainties of soft tissue‐based patient positioning in IGRT
Author
Taka‐aki Hirose 1 ; Arimura, Hidetaka 2 ; Jun‐ichi Fukunaga 1 ; Ohga, Saiji 3 ; Yoshitake, Tadamasa 3 ; Shioyama, Yoshiyuki 3 

 Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan 
 Faculty of Medical Sciences, Kyushu University, Fukuoka, Japan 
 Department of Clinical Oncology, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan 
Pages
73-81
Section
RADIATION ONCOLOGY PHYSICS
Publication year
2020
Publication date
Feb 2020
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2354742536
Copyright
© 2020. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.