1. Introduction
Spinal Cord Injury (SCI) is a potentially life-changing condition, causing permanent disability, compromising the victim’s physical and psychological well-being and impacting their close environment as well. Brain-computer interfaces (BCIs) are rapidly developing into a field-changing technology for those patients, not only replacing motor functions [1, 2] but even promising to alter the long-term outcome of the condition [3, 4]. Wireless invasive (implantable) neural recording is also an important development [5, 6], especially considering that SCI has become the research target of several assistive technologies [7] including functional electrical stimulation [8] and robotics for neurorehabilitation [9].
On the other hand, noninvasive BCI technology is dominated by the EEG modality, which has great temporal but low spatial resolution. EEG can detect rapid dynamics of the brain but lacks source estimation. Consequently, a major limitation of noninvasive BCI is the low signal to noise ratio (SNR) which can be partially attributed to volume conduction effect [10, 11]. Since the EEG signal measured at the scalp is the superposition of all electrical signals, including those generated by the cortex, discriminating brain activity from artifacts and noise can be technically difficult.
In addition to these neurophysiological limitations, a major drawback of traditional noninvasive BCIs is inconvenience. The subject’s movement is typically constrained by wires connecting the EEG electrode cap to the data acquisition unit, while the entire setup of wet electrodes entails a time-consuming procedure. From the hardware perspective, the main challenges of noninvasive BCI technology are portability and ease-of-use, rendering the wireless dry-electrode cap a promising solution. Lessons could be derived from the—otherwise immobile—MEG modality that boasts contactless sensors and reduced preparation time [13]. In terms of performance, the main challenge for noninvasive BCI technology is to accurately control external devices (multijoint robotic arms, drones, wheelchairs, etc.) that can move along several degrees of freedom (DoFs), a task which is necessary for performing complex operations in the physical world. Such a feat requires highly accurate decoding algorithms, able to discriminate multiple classes of Motor Imagery (MI) under the limitations of low SNR and spatial resolution. It has been demonstrated that the basic operation of a robotic arm which fits the aforementioned description is possible using noninvasive BCI [14].
Continuous advancements in electronics, from solid-state transistors in the 1940s, integration of sensors and powerful digital microprocessors in the 1970s, and other developments in the following decades, such as switching frequency, increased computing power, and programming flexibility, decreased fabrication cost and power consumption and steadily led to lighter and more agile, responsive, and computationally complex robots. The demand in the fields of Medical Rehabilitation and Assistive Technologies (AT) leads to the first combined medical applications for paralysis and stroke patients: robotic sleeves for assisted living, powered orthotics, and even some initial attempts at robotic manipulation arms [15, 16]. This nascent field of Rehabilitation Robotics started taking shape in the 1970s and 1980s, gradually emphasizing novel robotic manipulator designs [17, 18] and attempts to adapt existing industrial robotic manipulators [19].
Unobtrusiveness, portability, and mobility are clear advantages for AT and Medical Robotics. Even though affordable, power-autonomous, full-body robotic exoskeletons are yet to materialize, disabled patients now have a variety of product options, some commercially available: manipulator arms mounted on robotic wheelchairs [20, 21], dexterous robotic prosthetics which can acquire their control signals directly from the patient’s nervous system [1, 22], and adaptations of aforementioned BCI technology specifically for robotic rehabilitation [23–25]. Unfortunately, research in optimal interfaces for Human-Robot Interaction (HRI) is often overlooked in medical robotics development [12], with a potentially significant impact on user acceptance and the validation success rate for new technologies [26]. User perception and overall satisfaction with a robot’s technological interface has equal importance to hardware/software design and quality standards, something that holds particularly true in robotic-assisted rehabilitation. The need for patient’s immersive experience and willingness to collaborate with the robot, physiotherapist, and supervising physician should meaningfully alter the criteria of success of a robot [27]. For instance, acceptance of an external machine as part of one’s own body schema can significantly impact the rehabilitation process and should also be taken into account [15, 28].
In our previous work we have presented our progress towards developing 8-DoF anthropomorphic robotic arms, controlled by wireless off-the-shelf BCI, for AT and rehabilitation applications [12]. We have accounted for development of the robotic arms and electronics, for implementation of the BCI control module, and we have presented pilot experimental applications of the Brain-Robot Interface (BRI) on healthy and disabled individuals [29, 30].
In the remainder of this paper we present an elaborate user-assessment study of our wireless BRI by subjects with SCI and healthy individuals [31]. We focus not only on performance assessment during multiple movements BCI control, but on user perception of the assistive technology as well, analyzed with regard to their neurological condition, independence, imagery capacity, and psychometric evaluation. In Materials and Methods, we briefly present the BRI and we explain the experimental setup and statistical analysis of our collected data, which we then present in Results. In Discussion we attempt to interpret our results, comment on the strengths, and underline the limitations of our approach. We also discuss planned technological development in the direction of robotics, BCI module, signal analysis, and further experiments, as well as the challenges we are yet to meet.
2. Materials and Methods
2.1. The Brain-Robotic Arms Interface
As both the interface’s and robotics’ development have been thoroughly described before [12, 32], we will briefly explain the system’s technical characteristics here, emphasizing wireless capabilities. The Emotiv EPOC is an easily applied wireless 14-saline felt-electrode EEG system (Emotiv, USA), capable of detecting brain activity at a bandwidth of 0.2–43 Hz, employing digital notch filters at 50 Hz and 60 Hz and built-in digital 5th order Sinc filter. Meanwhile, due to being commercial class, it costs significantly less and also is less obtrusive, simple to use, and portable compared to traditional medical EEG devices. Its technical characteristics are also on par with the designated use case, so our team used this device for the development of the BCI modality. The wireless chipset transmits at a proprietary ultra-low energy (ULE) 2.4 Ghz Bluetooth Smart to a USB dongle connected to a dedicated computing laptop. Internal sampling rate of the electrodes is at 2 kHz (sequential sampling and single analog to digital conversion) with 16-bit resolution and signal sampling rate of 128 Hz transmitted. Using the Cognitiv suite a resting state and up to four different mental commands are classified (using proprietary BCI algorithm) and then are appended to key button input through the Emokey software. The input then accesses a control MATLAB script corresponding to movement commands for the robotic arms (Figure 1). Movement coordinates are subsequently transmitted to the robotic arms system through a serial port with a Baud Rate of 9600 bps.
[figure omitted; refer to PDF]
The “Mercury 2.0” robotic arm is a stand-alone electromechanical manipulator system developed by our team, capable of replicating most movements of a physical human arm. The current version of the robot is capable of movement along 8 DoFs [32]. Eight motors are attached in total on each arm, six of them being DC electric motors: two on the robotic shoulder joint, responsible for horizontal (1) and for vertical movement (2); one on the elbow joint, for horizontal movement (3); one on the wrist joint, for horizontal movement (4), one for rotation between the shoulder and elbow (5); and one for rotation between the elbow and wrist (6); finally, two servo motors add the ability of curling movement of the thumb and palm fingers (7) and (8). These 8 DoFs allow for responsive, fluid movement of the robotic arm and provide it with the ability to grab and manipulate small and light objects. Each arm functions with a single attached microcontroller, responsible for motors, connected to the dedicated computer system through a serial port at a Baud Rate of 9600. The microcontroller translates the commands the computer system transmits into positional coordinates for each motor. System responsiveness has been measured to be approximately 0.2 seconds.
2.2. Experimental Setup
2.2.1. Ethical Approval and Recruitment
The institutional bioethics committee approved the experimental protocol [12, 31] and all subjects filled and signed an informed consent form prior to their participation. Criteria for participation included clinical diagnosis and radiological documentation of SCI (evaluated by ASIA Impairment Scale (AIS) [33] and/or neurological examination reporting the condition) or healthy participants [31]. Exclusion criteria were other neurological injury or disease (traumatic brain injury, central nervous system tumors, epilepsy, etc.), recent participation in an interventional study, other grave medical condition that could affect participation or the safety, hearing and visual impairments, illegal drug use, and chronic alcoholism.
Ten subjects with SCI (8 male, 2 female) and ten healthy controls (with an effort for being gender and age matched to the SCI group) were recruited in total. The SCI group had a mean age of 46.0 years (range 28–74, standard deviation (std) 17.64), while the healthy group had a mean age of 46.20 years (range 27–74, std 18.27), as shown in Table 1. The subjects had no prior experience in BCI or robotics.
Table 1
Descriptive statistics of age (mean and standard deviation) for both groups (SCI group and healthy controls).
Group | Age |
SCI | 46.0 (17.64) |
Healthy | 46.2 (18.27) |
2.2.2. Subject Assessment
For both groups demographics and medical history were collected including smoking habits, age, height, weight, Body Mass Index (BMI), and education level. BMI was calculated and the following classification was used: underweight < 18.5, normal weight = 18.5–24.9, overweight = 25–29.9, and obesity ≥ 30. For the SCI group, history data was also collected including age at injury, weight at injury, and cause of injury. Neurological examination was performed by a specialist physician using the International Standards for Neurological Classification of Spinal Cord Injury: severity of injury (classification in AIS), Neurological Level of Injury (NLI), motor scores for upper extremities (UEMS), lower extremities (LEMS), and total (Moto-Total), and sensory scores for light touch (LT), pin prick (PP), and total (Sensory-Total) were recorded [33]. Spasticity, if present, was also recorded using Ashworth [34] and Penn Spasm Frequency Scales [35]. Both groups were assessed for everyday life independence using the Spinal Cord Independence Measure (SCIM-III) [36], translated in Greek (g-SCIM-III) [37]. Total SCIM score and subscores for Self-Care (g-SCIM-III-SC), Respiration and Sphincter Control (g-SCIM-III-RS), and Mobility (g-SCIM-III-M) were recorded.
In the SCI group, 9 out of 10 patients had incomplete injury. Four patients were classified as AIS D (40%), 2 were AIS B (20%), 2 were AIS E (20%), one had complete injury and was classified as AIS A (10%), and one patient was classified as AIS C (10%). Regarding level, 70% were cervical injuries whereas the remaining 30% were thoracic injuries. Moreover, the predominant cause of injury, in 60% (6/10) of patients injury, was motor-vehicle accidents (MVA), while in 30% of participants the injury was induced by fall from heights (Fall) and one patient reported other causes. Table 2 depicts age and gender distributions for both groups and also cause of injury, AIS, and NLI by subject.
Table 2
Subjects by group (SCI and healthy controls): basic demographic data for both groups and cause of injury, ASIA Impairment Scale (AIS), and Neurological Level of Injury.
SCI group | Age | Gender | Cause | AIS | NLI | Healthy group | Age | Gender |
CSI-02-001 | 28 | f | MVA | ASIA C | C4 | CSI-04-001 | 27 | f |
CSI-02-002 | 52 | m | MVA | ASIA D | C4 | CSI-04-007 | 51 | m |
CSI-02-003 | 42 | m | MVA | ASIA D | C8 | CSI-04-003 | 43 | m |
CSI-02-004 | 70 | m | Fall | ASIA D | C5 | CSI-04-006 | 71 | m |
CSI-02-005 | 60 | m | Fall | ASIA E | C6 | CSI-04-009 | 63 | m |
CSI-02-006 | 28 | m | MVA | ASIA D | C5 | CSI-04-004 | 28 | m |
CSI-02-007 | 30 | m | MVA | ASIA E | C5 | CSI-04-005 | 31 | m |
CSI-03-001 | 47 | m | Fall | ASIA A | T7 | CSI-04-008 | 47 | m |
CSI-03-002 | 29 | f | MVA | ASIA B | T4 | CSI-04-002 | 27 | f |
CSI-03-003 | 74 | m | Other | ASIA B | T4 | CSI-04-010 | 74 | m |
|
||||||||
Mean | 46.00 | - | - | - | - | Mean | 46.20 | - |
Std | 17.64 | - | - | - | - | Std | 18.27 | - |
Based on neurological assessment, the 60% of patients that were classified as either AIS D or E showed approximately intact general motor function and excellent UEMS and LEMS (3/10 of patients scored 50 in both categories whereas 2/10 scored 48 in UEMS and 50 in LEMS). The remaining 40% patients were classified as AIS A, B, or C and showed motor deficits, as presented in Table 3. The SCI group was therefore further grouped into positive outcome (60%) and negative outcome (40%) for further analysis as described in the statistical analysis section below. With regard to sensory skills, the patients with good outcome scored as high as the healthy controls in LT and PP sensory examination (Table 3).
Table 3
Neurological evaluation scores of the SCI group.
SCI group | Motor-Total | UEMS | LEMS | Sensory-Total | LT | PP |
CSI-02-001 | 48 | 24 | 24 | 224 | 112 | 112 |
CSI-02-002 | 84 | 39 | 45 | 183 | 94 | 89 |
CSI-02-003 | 98 | 48 | 50 | 224 | 112 | 112 |
CSI-02-004 | 98 | 48 | 50 | 223 | 111 | 112 |
CSI-02-005 | 100 | 50 | 50 | 224 | 112 | 112 |
CSI-02-006 | 100 | 50 | 50 | 222 | 111 | 111 |
CSI-02-007 | 100 | 50 | 50 | 224 | 112 | 112 |
CSI-03-001 | 56 | 50 | 6 | 121 | 58 | 63 |
CSI-03-002 | 50 | 50 | 0 | 156 | 78 | 78 |
CSI-03-003 | 54 | 50 | 4 | 156 | 78 | 78 |
|
||||||
Median | 91.00 | 50.00 | 47.50 | 222.50 | 111.0 | 111.50 |
Interquartile range | 47.00 | 4.30 | 44.50 | 68.00 | 34.00 | 34.00 |
Subject assessment also included subjective reporting of imagery capacity, using Vividness of Visual Imagery Questionnaire (VVIQ-II) [38] with eyes open that was assessed for total score and for each of the four scenarios (VVIQ1-VVIQ4). Psychometric evaluation also entailed answering Beck Depression Inventory (BDI) [39] and Rosenberg Self-Esteem Questionnaire (RSEQ) [40], both translated in Greek [41, 42]. After the participation in the experiment, both subject groups reported on HRI experience using the Godspeed robotics questionnaire [27], also translated in Greek [43]. Godspeed total score (GDSPD-Total) and subscores for Anthropomorphism (GDSPD-Anthr), Animacy (GDSPD-Anim), Likeability (GDSPD-Like), Perceived Intelligence (GDSPD-Int), and Perceived Safety (GDSPD-Safe) were recorded.
2.2.3. Experimental Procedure
As the experimental procedures have been also described in detail [12, 29], we will provide an overview hereby. The subjects were asked to perform Visual Motor Imagery (VMI), meaning they had to mentally rehearse 32 different movements of the upper extremities, while watching a set of arms performing randomly those movements on a computer screen. Each movement was shown 9 times in total. Randomly, walking and an oddball video were also shown. This VMI experience was aimed at preparing participants for the BCI classes they would have to later perform. This VMI procedure was also performed under high-resolution EEG recording, which will not be further elaborated hereby. The subjects had their arms, torso, and legs covered with a black curtain during VMI experience as well as during subsequent BCI training and BCI control to facilitate registration of the projected arms (in the case of VMI) or the robotic arms (in the case of BCI) into the body schema [28]. Also, in the case of VMI the projected arms were gender-matched to the subject.
Following VMI experience, the subjects sat across a computer monitor, between two Mercury 2.0 robotic arms, located in the Thessaloniki Active and Healthy Ageing Living Lab technology showcase room (Thess-AHALL, member of ENoLL) [44–46]. Subjects wore an Emotiv EPOC headset and they began training of the BCI modality using kinesthetic motor imagery (KMI) of movements of the left and the right arms. Action power threshold was set at 50% and all subjects were above threshold during training [29]. Three imagery classes were trained (resting state, left, and right). Five training cycles were allowed for left and right, using continuous mental rehearsal of all movements of left and right arm, respectively (as in the videos that subjects watched) [47]. Training skills (Train-L) (Train-R) classes were recorded. Following BCI training, the subjects were given written commands in the monitor to move a specific DoF towards a specific direction. 32 different commands were given in total and the subjects were asked to use the trained KMI skill in order to control the robotic arms to perform those actions. Each command was projected for 30 sec with a 5 sec rest between them (Figure 2). During the 30 sec period the subject attempted to perform the movement as many times as possible by giving the desired direction of imagery class (left or right). The BCI detected as correct (desired) any activation of the class corresponding to the arm currently targeted and gave the output to move the arm as per the instruction (positive feedback). Incorrect (undesired) activation that was detected resulted in an opposite motion of the target robotic arm (negative feedback). Desired and undesired commands that were detected by the BCI program were recorded through the MATLAB script.
[figures omitted; refer to PDF]
Performance was rated in each different movement with a score from 0 to 5, based on the ratio of desired mental commands to undesired mental commands that were detected and classified (Table 4) [12]. A total score for BCI performance (BCI-total) was then calculated for each subject, adding the scores for each movement (max 160) and the performance of each subject was also converted to a percentage (BCI%). Subscores were calculated for the 16 movements of the left arm (BCI-L), the 16 movements of the right arm (BCI-R), the 16 movements of wrists, fingers, and thumbs (BCI-Distal), and the 16 movements of shoulder and elbow (BCI-Proximal), with max scores of 80. Subscores were also calculated for 24 linear movements and 8 rotational movements but, to allow comparison, an average per movement score was calculated for both categories (BCI-Li/24 and BCI-Ro/8).
Table 4
House-developed performance rating scale for brain-robot interface control, tailored to the needs of the use case of the current experiment.
Performance rating of brain-robot interface control based on ratio of desired to undesired detected & classified mental commands | ||
Score | Rating | Description of rating |
5 | Excellent | Arms were moving towards desired direction for most of the time |
4 | Very Good | 5+ commands were detected, minimal undesired commands |
3 | Good | 4+ commands were detected, only 2-3 undesired commands |
2 | Average | 3+ commands were detected, but also up to +3 undesired commands |
1 | Minimal | 1-2 commands were detected or 3-4 including undesired commands or many commands were detected but most were undesired |
0 | No control | No command or only undesired commands were detected |
2.3. Statistical Analysis
2.3.1. Statistical Tests
Statistical analysis was performed in IBM SPSS Statistics (version 23) and we set a significance level of 0.05 for all statistical tests. The variables were explored for normality assumption using two grouping schemes: (1) SCI and healthy groups as grouping factor for all subjects and (2) positive and negative outcome as grouping factor for SCI subjects.
Normality was explored using visual inspection of histograms, normal Q-Q plots and boxplots, skewness, and kurtosis [48–50] as well as using the normality tests (Shapiro-Wilk test and Kolmogorov-Smirnov Test) [51, 52]. The variable age, in particular, was normally distributed for both groups (skewness: 0.407 (SE = 0.687), kurtosis: −1.418 (SE = 1.334) for healthy group; skewness: 0.651 (SE = 0.687), kurtosis: −0.752 (SE = 1.334) for patient group).
Normality assumption was not met (1) for motor and sensory scores of neurological examination as well as SCIM scores, (2) for VVIQ1 and VVIQ3 scenarios and BDI in SCI and healthy groups, (3) for GDSPD-Safe in either grouping scheme, and (4) for training scores in outcome grouping. All other distributions met normality assumption. Depending on normality assumption group differences were explored using either Independent Samples t-test or Mann–Whitney (U) test. Possible associations between quantitative variables were explored via Pearson correlation coefficient or Spearman’s coefficient depending on normality. Other specific statistical tests or other specific groupings were used as described below. Please also refer to Supplementary Materials for a more detailed report of statistical analysis (available here).
2.3.2. Demographics, Somatometric Data, and Clinical Evaluation
Initially, we planned comparisons of demographic and somatometric data between SCI and healthy groups. Comparisons regarding the education level (basic studies, pregraduate level, graduate level, postgraduate level, and Ph.D. holders) were performed by Mann–Whitney (U) test, since our data are ordinal (Likert-type) [53]. Medical history and neurological data were analyzed via descriptive statistical methods. Moreover, we explored whether the smoking status is independent of the group or not using chi-square test.
2.3.3. Assessment Questionnaires and User Perception
Imagery capacity (VVIQ) and psychometric questionnaires (BDI and RSEQ) were analyzed between SCI and healthy groups, as well as between SCI subgroups of positive and negative outcomes. The scores of Godspeed and its subcategories (GDSPD-Total, GDSPD-Anthr, GDSPD-Anim, GDSPD-Like, GDSPD-Int, and GDSPD-Safe) were analyzed between patient and healthy groups, as well as between patient outcome groups. Finally, possible correlations were explored between Godspeed scores and BCI performance as well as Godspeed scores and VVIQ scores.
2.3.4. BCI Performance
BCI-total, BCI-L, BCI-R, and training skills (Train-L and Train-R) were analyzed between SCI and healthy groups as well as in SCI outcome subgroups. BCI-Distal, BCI-Proximal, BCI-Li, and BCI-Ro were analyzed using descriptive statistics. We further explored differences in BCI performance and training scores between different groups of neurological levels of injury (cervical, thoracic) after testing for normality. Additionally, linear regression analysis was used to model the possible relationship between the independent variable BCI scores and the explanatory variable NLI using linear regression analysis. Possible correlations were further explored across groups (SCI/healthy) between BCI performance and (1) age, (2) imagery capacity (VVIQ), and (3) psychometric questionnaires (BDI, RSEQ) for both groups.
3. Results
3.1. Demographics, Somatometric Data, and Clinical Evaluation
Planned comparisons of age between groups did not reveal any significant difference (Table 1). Regarding education level, no statistically significant difference between groups was shown. Group differences were not revealed in somatometric data either. Most of the injuries (4/10) happened when the participants were between 26.0 and 33.0 years of age. The second more vulnerable age range to injuries (MVAs) seems to be the age range of 50.0–57.0 years. Exploring whether the smoking status is independent of the group or not, we observed a significant association (
Table 5
Statistical tests on demographics, clinical evaluation, and assessment questionnaires.
Demographics | Clinical evaluation | |||||
SCI versus healthy | SCI | Healthy | ||||
Education |
|
|
g-SCIM-III-Total | Median = 94.0, IQR = 51.3 | Median = 100.0, IQR = 0.0 | |
Height |
|
|
g-SCIM-III-SC | Median = 18.0, IQR = 7.3 | Median = 20.0, IQR = 0.0 | |
Weight |
|
|
g-SCIM-III-BS | Median = 40.0, IQR = 19.3 | Median = 40.0, IQR = 0.0 | |
BMI |
|
|
g-SCIM-III-M | Median = 37.5, IQR = 24.3 | Median = 40.0, IQR = 0.0 | |
Smoking |
|
|
||||
|
||||||
Assessment questionnaires | ||||||
SCI | Healthy | SCI versus healthy | Positive versus negative outcome | |||
|
||||||
VVIQ | 65.70 (7.04) | 67.0 (10.92) |
|
|
|
|
VVIQ1 | Median = 16.5, IQR = 4.0 | Median = 18.5, IQR = 4.5 |
|
|
|
|
VVIQ2 | 17.20 (2.25) | 16.50 (2.99) |
|
|
|
|
VVIQ3 | Median = 17.0, IQR = 4.5 | Median = 19.5, IQR = 3.3 |
|
|
|
|
VVIQ4 | 15.10 (3.98) | 14.70 (3.59) |
|
|
|
|
BDI | Median = 14.5, IQR = 12.5 | Median = 3.0, IQR = 5.5 |
|
|
||
RSEQ | 21.40 (3.86) | 24.70 (4.14) |
|
|
3.2. Assessment Questionnaires
VVIQ did not show any significant difference between SCI and healthy groups in total score or any of the scenarios. SCI group scored lower than healthy (Figure 3) but differences did not reach statistical significance. Regarding SCI outcome subgroups, no considerable group differences were revealed in VVIQ scores. Also, even though the SCI group showed increased depressive scores and decreased self-esteem compared to healthy group, group differences were not statistically significant (Figure 4).
[figure omitted; refer to PDF]
[figures omitted; refer to PDF]
3.3. BCI Performance
Planned between-group (SCI/healthy) comparisons of BCI performance revealed a statistically significant difference only in the control of right robotic arm (BCI-R:
Table 6
Statistical tests on BCI performance and training skills.
BCI performance | ||||||
SCI | Healthy | SCI versus healthy | Cervical | Thoracic | ||
|
||||||
BCI-Total | 66.70 (17.49) | 79.50 (18.51) |
|
|
71.57 (14.75) | 55.33 (21.13) |
BCI (%) | 41.69% (10.93%) | 49.65% (11.47%) |
|
|
||
BCI-R | 27.10 (17.97) | 45.80 (14.05) |
|
|
29.86 (20.07) | 20.67 (12.50) |
BCI-L | 39.60 (19.64) | 33.70 (21.32) |
|
|
41.71 (19.20) | 34.67 (24.03) |
BCI-Distal | 33.40 (9.98) | 40.50 (8.45) |
|
|
||
BCI-Proximal | 33.30 (8.82) | 39.00 (11.95) |
|
|
||
BCI-Linear/24 | 2.11 (0.52) | 2.48 (0.60) |
|
|
||
BCI-Rot/8 | 2.25, 1.06 | 2.44, 0.72 |
|
|
||
|
||||||
Positive | Negative | Positive versus negative outcome | Cervical versus thoracic | |||
|
||||||
BCI-Total | 69.50 (15.00) | 62.50 (22.43) |
|
|
|
|
BCI (%) | 43.44% (9.36%) | 39.06% (14.02%) |
|
|
||
BCI-R |
|
|
|
|
||
BCI-L |
|
|
|
|
||
|
||||||
Training skill | ||||||
SCI | Healthy | SCI versus healthy | Cervical | Thoracic | ||
|
||||||
Train-R | 46.30% (18.00%) | 46.40% (10.32%) |
|
|
49.14% (18.05%) | 39.67% (19.66%) |
Train-L | 56.20% (19.61%) | 53.80% (16.67%) |
|
|
53.71% (22.54%) | 62.00% (11.53%) |
|
||||||
Positive | Negative | Positive versus negative outcome | Cervical versus thoracic | |||
|
||||||
Train-R | Median = 48.00%, IQR = 40.25% | Median = 50.50%, IQR = 26.50% |
|
|
|
|
Train-L | Median = 51.00%, IQR =19.00% | Median = 68.00%, IQR = 40.00% |
|
|
|
|
|
||||||
Correlations | ||||||
NLI | Age-SCI | Age-healthy | ||||
|
||||||
BCI-Total |
|
|
|
|
|
|
BCI-R | - | - |
|
|
|
|
BCI-L | - | - |
|
|
|
|
[figures omitted; refer to PDF]
[figure omitted; refer to PDF]
The ability of SCI subjects to control robotic arms did not differentiate depending on the injury location (cervical, thoracic). However, subjects with cervical injuries scored higher than those with thoracic injuries in BCI control. BCI-Total was marginally negatively correlated to NLI in the SCI group
[figures omitted; refer to PDF]
The training skill was not statistically different between SCI and healthy groups neither for the right hand, nor for the left hand, even if SCI subjects showed slightly lower training scores in the right hand and higher in the left hand than healthy participants. Comparing training scores between SCI subjects grouped by outcome (Figure 8), a marginally considerable difference was found only in training scores of left hand (
[figures omitted; refer to PDF]
Total BCI scores were significantly negatively associated with VVIQ total scores (
3.4. User Perception
Based on planned analysis of HRI characteristics of the robotic arms, SCI and healthy groups did not present significant differences in their answers in any subcategory (GDSPD-Anthr:
The scores are illustrated in Figure 9 (GDSPD (/120) (mean (SD)), healthy: 80.80 (14.78); SCI: 78.00 (14.57); GDSPD-Anthr, healthy: 13.60 (4.25); SCI: 10.80 (4.08); GDSPD-Anim, healthy: 18.90 (5.11); SCI: 17.70 (3.53); GDSPD-Like, healthy: 18.50 (3.87); SCI: 18.90 (4.36); GDSPD-Int, healthy: 17.70 (4.60); SCI: 17.80 (4.92); GDSPD-Safe, healthy (median = 12.0, IQR = 5.3); SCI (median = 13.50, IQR = 4.5)).
[figures omitted; refer to PDF]
Godspeed’s scores were further explored after grouping by outcome (positive, negative). Marginally significant difference was found only in Anthropomorphism (
Significant correlation between Godspeed’s and BCI scores was not revealed for any group. GDSPD-Anthr and GDSPD-Int were positively correlated to VVIQ4 and VVIQ1 scenario’s scores, respectively, only in healthy (GDSPD-Int, VVIQ4:
4. Discussion
4.1. User Perception and Performance Assessment
Our results show that healthy controls performed better (49.65%) than the SCI group (41.65%) in BCI control and so did subjects with positive neurological outcome (43.44%) when compared to subjects with negative neurological outcome (39.6%). Both those observations did not reach statistical significance but their interpretation is not as straightforward. Since no patients with complete tetraplegia were included, these findings probably cannot be attributed to some type of “thought extinction process” [54]. Indeed, self-reported imagery capacity (as measured by VVIQ) was found to be nonsignificantly lower in SCI group than in healthy subjects but, on the other hand, subjects with cervical injuries fared better than those subjects with thoracic injuries. Thus the group with injury that could affect upper extremity neural circuits outperformed those subjects that did not have direct injury to those circuits. Moreover this finding was further reinforced by a marginally significant (negative) correlation of BCI performance to neurological level of injury, meaning that the higher the level of injury, the better the subjects performed.
Possible explanations for performance differences could be explored along the lines of depression [55] and motivation [56, 57]. Indeed the SCI group showed increased depressive scores (measured by BDI) and lower self-esteem scores (measured by RSEQ), despite both findings not being statistically significant. Our research protocol did not include any official questionnaire to assess SCI and healthy subject with regard to motivation. Only by anecdotal evidence, during informal debriefing discussions that we held with participants, subjects with thoracic injuries tended to consider somewhat irrelevant this experimental setup to their condition. While reciprocal sensorimotor pathways and sensorimotor cortex representations are affected regardless of the level of SCI, those subjects felt that their clinical condition and independence demands were not addressed by robotic arms technology.
The HRI characteristics of the robot were measured (after the participation) by Godspeed robotics questionnaire. It seems that user perception did not differ significantly between SCI and healthy subjects. While in Figure 9 we show absolute values for the total questionnaire score and its subscores, since no standard for direct comparison of those can be identified, we feel that intergroup comparisons are more useful. As such, we believe that this is a positive finding, interpreting it as users with SCI appreciating the robot more or less the same as their healthy counterparts. Among SCI subgroups, those with negative outcome gave higher scores to Anthropomorphism of the robot (mean 13.75 out of max 25) than those with positive outcome (mean 8.83 out of max 25), a finding that can be considered marginally significant (see Figure 9(b)). Indirectly, since user perception did not correlate with BCI performance and did not vary between SCI and healthy subjects in all other comparisons, perhaps this could be attributed to psychometric attributes of the participants and it should be further investigated in an SCI group with wider participation.
As we mentioned, no statistically significant negative correlations were found between BCI performance and the age of the subjects, a finding that also needs further study in order to be validated. As much of the BCI research field endeavors have been shaped around the issue of providing affordable, acceptable, and useful assistive systems for the benefit of the disabled, the particular characteristics of elderly patients should be the focus of investigation [58, 59].
Finally, BCI performance was significantly negatively associated
Some correlation was also revealed between reported imagery capacity and Godspeed questionnaire. SCI subjects that had higher scores in VVIQ rated Perceived Safety higher
Our work adds onto an envisaged puzzle of BCI technology and uses case scenarios with an emphasis on affordability (off-the-shelf component system) and realism at a reasonable accuracy. Multi DoF robotics control by users with SCI has been proven to be feasible even by commercial-class wireless BCI. Nonetheless, repeated training in several BCI sessions is probably needed to improve performance by the end-users. Although users with complete lesions by definition would benefit more from ongoing research in robotics and BCIs, the use of brain controlled assistive technologies holds special importance for users with incomplete injuries too [62]. As a result of preserved reciprocal neural pathway communication, robotic-assisted training and brain controlled functional stimulation have been shown to improve arm motor function of those patients [63, 64] but expectations from assistive technologies and motivation to use them should be among key targets for further user experience investigations.
4.2. Further Technological Development
Our main development directions include (1) improving accuracy, speed, responsiveness, fluidity, and efficiency; (2) enhancing the integration of the system into the operator’s perceived body mental image [17]. The pipeline that we overview below is tested offline, since several aspects, such as the solution of the inverse problem, entail high computational cost and are planned to be realized with dedicated processing units and use of artificial intelligence (AI) [12]. We are currently developing and testing a novel BCI decoding algorithm. Our goal is to increase discriminative ability to multiple MI classes projecting sensor data to cortical source domain [65, 66]. The COLIN27 anatomy is used [67], but tests are also made on individual head models. Our approach is based on studies indicating that the features extracted from the source domain enhance the discrimination [68–70], since the higher dimensional space of sources incorporates anatomical and physiological information. Spatial features are extracted in predefined Regions of Interest (ROIs) [12] and Common Spatial Pattern (CSP) filters [71] are calculated on each ROI. Tuning and feature selection using a linear classifier (LDA) are a crucial part in order to determine the most valuable ROIs and improve BCI performance. The use of time-frequency and connectivity features in conjunction will also be explored.
4.3. Limitations
As we have already mentioned [12], multiple MI class discrimination is a challenge that state-of-the-art noninvasive BCI technology has yet to meet with success. Depending on a proprietary BCI algorithm, such as in our case, further underlines this problem, as it was not tailored to our specific needs. While the commercial-class device meets the use case needs and will be retained in future implementations, the proprietary BCI algorithm will be substituted with the one currently under development, in order to explore and reach practically usable performance of the entire system. This setup will be used in future experiments, including tests on the EEG data that have been already recorded. In general, off-the-shelf BCI technology seems to be able to meet the demands set by the field for unobtrusiveness, portability, and simplicity but it is possible that multi-DoF control may be not feasible or impractical without extensive use of supportive AI and advances in electronics charged with decision-making. Finally, from the investigation’s perspective, although our study is one among those featuring adequate sized sample of disabled end-users for BCI technology [72, 73], our approach would benefit from wider subject participation for validation and extraction of statistically significant—and relevant to the advancement of the research field—results.
5. Conclusions
Much of the BCI research endeavor has been shaped around the issue of providing affordable, accurate, and real-time assistive systems for the benefit of the disabled. This piece of research adds onto that envisaged puzzle, but tackling affordability (off-the-shelf component system) of realistic scenarios at a reasonable accuracy. Healthy controls, SCI subjects with positive neurological outcome, and subjects with cervical injuries performed better in BCI control. Depression and motivation could play significant roles in BCI and robotics control. Reported imagery capacity was nonsignificantly lower for SCI subjects. User perception of the robot did not differ between SCI and healthy subjects, but, among SCI participants, those with negative outcome rated anthropomorphic characteristics higher. More BCI sessions are expected to improve performance of SCI and healthy subjects. The herein results demonstrate that by developing BCI decoding algorithms capable of true control of multiple DoFs and addressing the computational cost of online implementation of such an approach, it could be within short-term reach. While maintaining requirements for wireless, unobtrusive solutions constitutes challenges yet to be met, the exploitability of such a system by real patients at a reasonable accuracy cannot be understated. Improving electronics, fluidity, and accuracy of the robotic system, enhancing HRI experience, and implementing a source-based BCI algorithm for multiple class control, as well as further investigations with disabled users, are among our next steps.
Ethical Approval
This study was conducted in accordance with the Declaration of Helsinki (1964) and its following amendments. The Bioethics & Ethics Committee of Faculty of Medicine, Aristotle University of Thessaloniki, approved the study.
Consent
All experiments were conducted with the subjects’ understanding and written informed consent.
Conflicts of Interest
The authors declare that there are no conflicts of interest regarding the publication of this article.
[1] L. R. Hochberg, D. Bacher, B. Jarosiewicz, N. Y. Masse, J. D. Simeral, J. Vogel, S. Haddadin, J. Liu, S. S. Cash, P. Van Der Smagt, J. P. Donoghue, "Reach and grasp by people with tetraplegia using a neurally controlled robotic arm," Nature, vol. 485 no. 7398, pp. 372-375, DOI: 10.1038/nature11076, 2012.
[2] J. L. Collinger, B. Wodlinger, J. E. Downey, W. Wang, E. C. Tyler-Kabara, D. J. Weber, A. J. C. McMorland, M. Velliste, M. L. Boninger, A. B. Schwartz, "High-performance neuroprosthetic control by an individual with tetraplegia," The Lancet, vol. 381 no. 9866, pp. 557-564, DOI: 10.1016/S0140-6736(12)61816-9, 2013.
[3] A. R. C. Donati, S. Shokur, E. Morya, D. S. F. Campos, R. C. Moioli, C. M. Gitti, P. B. Augusto, S. Tripodi, C. G. Pires, G. A. Pereira, F. L. Brasil, S. Gallo, A. A. Lin, A. K. Takigami, M. A. Aratanha, S. Joshi, H. Bleuler, G. Cheng, A. Rudolph, M. A. L. Nicolelis, "Long-term training with a brain-machine interface-based gait protocol induces partial neurological recovery in paraplegic patients," Scientific Reports, vol. 6 no. 1,DOI: 10.1038/srep30383, 2016.
[4] M. Alam, W. Rodrigues, B. N. Pham, N. V. Thakor, "Brain-machine interface facilitated neurorehabilitation via spinal stimulation after spinal cord injury: Recent progress and future perspectives," Brain Research, vol. 1646, pp. 25-33, DOI: 10.1016/j.brainres.2016.05.039, 2016.
[5] R. R. Harrison, R. J. Kier, C. A. Chestek, V. Gilja, P. Nuyujukian, S. Ryu, B. Greger, F. Solzbacher, K. V. Shenoy, "Wireless neural recording with single low-power integrated circuit," IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 17 no. 4, pp. 322-329, DOI: 10.1109/TNSRE.2009.2023298, 2009.
[6] A. V. Nurmikko, J. P. Donoghue, L. R. Hochberg, W. R. Patterson, Y.-K. Song, C. W. Bull, D. A. Borton, F. Laiwalla, S. Park, Y. Ming, J. Aceros, "Listening to brain microcircuits for interfacing with external world—progress in wireless implantable microelectronic neuroengineering devices," Proceedings of the IEEE, vol. 98 no. 3, pp. 375-388, DOI: 10.1109/JPROC.2009.2038949, 2010.
[7] J. L. Collinger, S. Foldes, T. M. Bruns, B. Wodlinger, R. Gaunt, D. J. Weber, "Neuroprosthetic technology for individuals with spinal cord injury," The Journal of Spinal Cord Medicine, vol. 36 no. 4, pp. 258-272, DOI: 10.1179/2045772313Y.0000000128, 2013.
[8] P. H. Peckham, J. S. Knutson, "Functional electrical stimulation for neuromuscular applications," Annual Review of Biomedical Engineering, vol. 7, pp. 327-360, DOI: 10.1146/annurev.bioeng.6.040803.140103, 2005.
[9] L. Pignolo, "Robotics in neuro-rehabilitation," Journal of Rehabilitation Medicine, vol. 41 no. 12, pp. 955-960, DOI: 10.2340/16501977-0434, 2009.
[10] B. He, Modeling, Imaging of Bioelectrical Activity, 2005.
[11] P. L. Nunez, R. Srinivasan, "Electric Fields of the Brain: The neurophysics of EEG," Electric Fields of the Brain: The neurophysics of EEG,DOI: 10.1093/acprof:oso/9780195050387.001.0001, 2009.
[12] A. Athanasiou, I. Xygonakis, N. Pandria, P. Kartsidis, G. Arfaras, K. R. Kavazidi, N. Foroglou, A. Astaras, P. D. Bamidis, "Towards Rehabilitation Robotics: Off-the-Shelf BCI Control of Anthropomorphic Robotic Arms," BioMed Research International, vol. 2017,DOI: 10.1155/2017/5708937, 2017.
[13] S. Baillet, "Magnetoencephalography for brain electrophysiology and imaging," Nature Neuroscience, vol. 20 no. 3, pp. 327-339, DOI: 10.1038/nn.4504, 2017.
[14] J. Meng, S. Zhang, A. Bekyo, J. Olsoe, B. Baxter, B. He, "Noninvasive electroencephalogram based control of a robotic arm for reach and grasp tasks," Scientific Reports, vol. 6 no. 1, article 38565,DOI: 10.1038/srep38565, 2016.
[15] Y. Kim, A. M. Cook, "Manipulation and Mobility Aids," Electronic Devices for Rehabilitation, 1985.
[16] M. Hillman, "2 Rehabilitation Robotics from Past to Present A Historical Perspective," Advances in Rehabilitation Robotics, pp. 25-44, 2006.
[17] W. Seamone, G. Schmeisser, Evaluation of the APL/JHU Robot Arm Work Station, 1986.
[18] C. P. Mason, E. Peizer, "Medical Manipulator for Quadriplegics, Colloques IRIA," Proceedings of the in International Conference on Telemanipulators for the Physically Handicapped, pp. 309-312, .
[19] H. H. Kwee, M. Tramblay, R. Barbier, M. Dupeyroux, M. F. Vinceneux, P. Semoulin, S. Pannier, "First experimentation of the spartacus telethesis in a clinical environment," Paraplegia, vol. 21 no. 5, pp. 275-286, DOI: 10.1038/sc.1983.47, 1983.
[20] Y. Wakita, W.-K. Yoon, N. Yamanobe, "User evaluation to apply the robotic arm RAPUDA for an upper-limb disabilities Patient's Daily Life," Proceedings of the 2012 IEEE International Conference on Robotics and Biomimetics, ROBIO 2012, pp. 1482-1487, DOI: 10.1109/ROBIO.2012.6491178, .
[21] V. Maheu, P. S. Archambault, J. Frappier, F. Routhier, "Evaluation of the JACO robotic arm: Clinico-economic study for powered wheelchair users with upper-extremity disabilities," Proceedings of the Rehab Week Zurich 2011 - 2011 IEEE International Conference on Rehabilitation Robotics, ICORR 2011,DOI: 10.1109/ICORR.2011.5975397, .
[22] M. S. Johannes, J. D. Bigelow, J. M. Burck, S. D. Harshbarger, M. V. Kozlowski, T. Van Doren, "An overview of the developmental process for the modular prosthetic limb," Johns Hopkins APL Technical Digest, vol. 30 no. 3, pp. 207-216, 2011.
[23] J. J. Daly, J. R. Wolpaw, "Brain-computer interfaces in neurological rehabilitation," The Lancet Neurology, vol. 7 no. 11, pp. 1032-1043, DOI: 10.1016/S1474-4422(08)70223-0, 2008.
[24] M. Gomez-Rodriguez, M. Grosse-Wentrup, J. Hill, A. Gharabaghi, B. Scholkopf, J. Peters, "Towards brain-robot interfaces in stroke rehabilitation," Proceedings of the IEEE International Conference on Rehabilitative Robotics, vol. 2011, .
[25] K. K. Ang, C. Guan, K. S. Chua, B. T. Ang, C. Kuah, C. Wang, K. S. Phua, Z. Y. Chin, H. Zhang, "A clinical study of motor imagery-based brain-computer interface for upper limb robotic rehabilitation," Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC '09), pp. 5981-5984, DOI: 10.1109/IEMBS.2009.5335381, .
[26] K. Dautenhahn, "Methodology & themes of human-robot interaction: a growing research field," International Journal of Advanced Robotic Systems, vol. 4 no. 1, pp. 103-108, 2007.
[27] C. Bartneck, D. Kulić, E. Croft, S. Zoghbi, "Measurement instruments for the anthropomorphism, animacy, likeability, perceived intelligence, and perceived safety of robots," International Journal of Social Robotics, vol. 1 no. 1, pp. 71-81, DOI: 10.1007/s12369-008-0001-3, 2009.
[28] O. Christ, M. Reiner, "Perspectives and possible applications of the rubber hand and virtual hand illusion in non-invasive rehabilitation: Technological improvements and their consequences," Neuroscience & Biobehavioral Reviews, vol. 44, pp. 33-44, DOI: 10.1016/j.neubiorev.2014.02.013, 2014.
[29] G. Arfaras, A. Athanasiou, P. Niki, R. K. Kyriaki, P. Kartsidis, A. Astaras, D. B. Panagiotis, "Visual Versus Kinesthetic Motor Imagery for BCI Control of Robotic Arms (Mercury 2.0)," Proceedings of the 2017 IEEE 30th International Symposium on Computer-Based Medical Systems (CBMS), pp. 440-445, DOI: 10.1109/CBMS.2017.34, .
[30] A. Athanasiou, G. Arfaras, I. Xygonakis, P. Kartsidis, N. Pandria, K. R. Kavazidi, A. Astaras, N. Foroglou, K. Polyzoidis, P. D. Bamidis, "Commercial BCI Control and Functional Brain Networks in Spinal Cord Injury: A Proof-of-Concept," Proceedings of the 2017 IEEE 30th International Symposium on Computer-Based Medical Systems (CBMS), pp. 262-267, DOI: 10.1109/CBMS.2017.35, .
[31] "Brainwave control of a wearable robotic arm for rehabilitation and neurophysiological study in cervical spine injury (CSI:Brainwave)," . https://clinicaltrials.gov/ct2/show/NCT02443558
[32] N. Moustakas, P. Kartsidis, A. Athanasiou, A. Astaras, P. D. Bamidis, "Development of MERCURY version 2.0 robotic arms for rehabilitation applications," Proceedings of the 8th ACM International Conference on PErvasive Technologies Related to Assistive Environments (PETRA '15),DOI: 10.1145/2769493.2769558, .
[33] S. C. Kirshblum, S. P. Burns, F. Biering-Sorensen, W. Donovan, D. E. Graves, A. Jha, M. Johansen, L. Jones, A. Krassioukov, M. J. Mulcahey, M. Schmidt-Read, W. Waring, "International standards for neurological classification of spinal cord injury (revised 2011)," The Journal of Spinal Cord Medicine, vol. 34 no. 6, pp. 535-546, DOI: 10.1179/204577211X13207446293695, 2011.
[34] R. W. Bohannon, M. B. Smith, "Interrater reliability of a modified Ashworth scale of muscle spasticity," Physical Therapy in Sport, vol. 67 no. 2, pp. 206-207, DOI: 10.1093/ptj/67.2.206, 1987.
[35] J. T. C. Hsieh, D. L. Wolfe, W. C. Miller, A. Curt, "Spasticity outcome measures in spinal cord injury: Psychometric properties and clinical utility," Spinal Cord, vol. 46 no. 2, pp. 86-95, DOI: 10.1038/sj.sc.3102125, 2008.
[36] M. Itzkovich, I. Gelernter, F. Biering-Sorensen, "The Spinal Cord Independence Measure (SCIM) version III: Reliability and validity in a multi-center international study," Disability and Rehabilitation, vol. 29 no. 24, pp. 1926-1933, DOI: 10.1080/09638280601046302, 2007.
[37] A. Athanasiou, A. Alexandrou, E. Paraskevopoulos, N. Foroglou, A. Prassas, P. D. Bamidis, "Towards a Greek adaptation of the Spinal Cord Independence Measure (SCIM)," in proceedings of the 15th European Congress of Neurosurgery (EANS 14), pp. 181-184, .
[38] D. F. Marks, "Directions in Mental-Imagery Research," Journal of Mental Imagery, vol. 19 no. 3-4, pp. 153-167, 1995.
[39] A. T. Beck, R. A. Steer, M. G. Garbin, "Psychometric properties of the Beck Depression Inventory: twenty-five years of evaluation," Clinical Psychology Review, vol. 8 no. 1, pp. 77-100, DOI: 10.1016/0272-7358(88)90050-5, 1988.
[40] W. Petersen, "Society and the Adolescent Self-Image. Morris Rosenberg. Princeton University Press, Princeton, N.J., 1965. xii + 326 pp. $6.50," Science, vol. 148 no. 3671,DOI: 10.1126/science.148.3671.804, 1965.
[41] M. Giannakou, P. Roussi, M.-E. Kosmides, G. Kiosseoglou, A. Adamopoulou, G. Garyfallos, "Adaptation of the beck depression inventory-II to greek population," Hellenic Journal of Psychology, vol. 10 no. 2, pp. 120-146, 2013.
[42] C. Galanou, M. Galanakis, E. Alexopoulos, C. Darviri, "Rosenberg Self-Esteem Scale Greek Validation on Student Sample," Psychology, vol. 05 no. 08, pp. 819-827, DOI: 10.4236/psych.2014.58093, 2014.
[43] A. Astaras, A. Athanasiou, A. Alexandrou, P. Kartsidis, N. Moustakas, P. D. Bamidis, "Double-blind greek translation and online implementation of the Godspeed robotics questionnaire," Proceedings of the in 6th Panhellenic Conference on Biomedical Technology Conference, .
[44] E. I. Konstantinidis, A. Billis, C. Bratsas, A. Siountas, P. D. Bamidis, "Thessaloniki active and healthy ageing living lab: The roadmap from a specific project to a living lab towards openness," Proceedings of the 9th ACM International Conference on PErvasive Technologies Related to Assistive Environments (PETRA '2016),DOI: 10.1145/2910674.2935846, .
[45] E. I. Konstantinidis, P. D. Bamidis, "Density based clustering on indoor kinect location tracking: A new way to exploit active and healthy aging living lab datasets," Proceedings of the 2015 IEEE 15th International Conference on Bioinformatics and Bioengineering (BIBE),DOI: 10.1109/BIBE.2015.7367714, .
[46] E. I. Konstantinidis, P. E. Antoniou, G. Bamparopoulos, P. D. Bamidis, "A lightweight framework for transparent cross platform communication of controller data in ambient assisted living environments," Information Sciences, vol. 300 no. 1, pp. 124-139, DOI: 10.1016/j.ins.2014.10.070, 2015.
[47] T. Sollfrank, D. Hart, R. Goodsell, J. Foster, T. Tan, "3D visualization of movements can amplify motor cortex activation during subsequent motor imagery," Frontiers in Human Neuroscience, vol. 9, article 463,DOI: 10.3389/fnhum.2015.00463, 2015.
[48] D. Cramer, "Fundamental statistics for social research," Step-by-step calculations and computer techniques using SPSS for Windows,DOI: 10.4324/9780203360613, 1988.
[49] D. Cramer, D. Howitt, The SAGE Dictionary of Statistics,DOI: 10.4135/9780857020123, 2004.
[50] D. P. Doane, L. E. Seward, "Measuring skewness: a forgotten statistic?," Journal of Statistics Education, vol. 19 no. 2, 2011.
[51] N. M. Razali, Y. B. Wah, M. Sciences, "Power comparisons of Shapiro-Wilk, Kolmogorov-Smirnov," Lilliefors and Anderson-Darling tests, vol. 2 no. 1, pp. 21-33, 2011.
[52] S. S. Shapiro, M. B. Wilk, "An analysis of variance test for normality: Complete samples," Biometrika, vol. 52, pp. 591-611, DOI: 10.1093/biomet/52.3-4.591, 1965.
[53] N. Mogey, "So You Want to Use a Likert Scale?," Edinburgh: 2 Evaluation Cookbook Learning Technology Dissemination Initiative, 1998.
[54] N. Birbaumer, P. Sauseng, "Brain–Computer Interface in Neurorehabilitation," 2010.
[55] A. Vuckovic, J. A. Pineda, K. LaMarca, D. Gupta, C. Guger, "Interaction of BCI with the underlying neurological conditions in patients: pros and cons," Frontiers in Neuroengineering, vol. 7,DOI: 10.3389/fneng.2014.00042, 2014.
[56] F. Nijboer, N. Birbaumer, A. Kübler, "The influence of psychological state and motivation on brain-computer interface performance in patients with amyotrophic lateral sclerosis - a longitudinal study," Frontiers in Neuroscience, vol. 4, article no. 55,DOI: 10.3389/fnins.2010.00055, 2010.
[57] E. M. Hammer, S. Halder, B. Blankertz, C. Sannelli, T. Dickhaus, S. Kleih, K.-R. Müller, A. Kübler, "Psychological predictors of SMR-BCI performance," Biological Psychology, vol. 89 no. 1, pp. 80-86, DOI: 10.1016/j.biopsycho.2011.09.006, 2012.
[58] C. Papadelis, C. Kourtidou-Papadeli, P. Bamidis, M. Albani, "Effects of imagery training on cognitive performance and use of physiological measures as an assessment tool of mental effort," Brain and Cognition, vol. 64 no. 1, pp. 74-85, DOI: 10.1016/j.bandc.2007.01.001, 2007.
[59] F. González-Palau, M. Franco, P. D. Bamidis, R. Losada, E. Parra, S. G. Papageorgiou, A. B. Vivas, "The effects of a computer-based cognitive and physical training program in a healthy and mildly cognitive impaired aging sample," Aging & Mental Health, vol. 18 no. 7, pp. 838-846, DOI: 10.1080/13607863.2014.899972, 2014.
[60] S. M. Roldan, "Object Recognition in Mental Representations: Directions for Exploring Diagnostic Features through Visual Mental Imagery," Frontiers in Psychology, vol. 8,DOI: 10.3389/fpsyg.2017.00833, 2017.
[61] E. Broadbent, "Interactions with Robots: The Truths We Reveal about Ourselves," Annual Review of Psychology, vol. 68, pp. 627-652, DOI: 10.1146/annurev-psych-010416-043958, 2017.
[62] E. López-Larraz, F. Trincado-Alonso, V. Rajasekaran, S. Pérez-Nombela, A. J. del-Ama, J. Aranda, J. Minguez, A. Gil-Agudo, L. Montesano, "Control of an ambulatory exoskeleton with a brain-machine interface for spinal cord injury gait rehabilitation," Frontiers in Neuroscience, vol. 10,DOI: 10.3389/fnins.2016.00359, 2016.
[63] G. E. Francisco, N. Yozbatiran, J. Berliner, M. K. O’Malley, A. U. Pehlivan, Z. Kadivar, K. Fitle, C. Boake, "Robot-Assisted Training of Arm and Hand Movement Shows Functional Improvements for Incomplete Cervical Spinal Cord Injury," American Journal of Physical Medicine & Rehabilitation, vol. 96, pp. S171-S177, DOI: 10.1097/PHM.0000000000000815, 2017.
[64] A. Jackson, J. B. Zimmermann, "Neural interfaces for the brain and spinal cord - Restoring motor function," Nature Reviews Neurology, vol. 8 no. 12, pp. 690-699, DOI: 10.1038/nrneurol.2012.219, 2012.
[65] R. Grech, T. Cassar, J. Muscat, K. P. Camilleri, S. G. Fabri, M. Zervakis, P. Xanthopoulos, V. Sakkalis, B. Vanrumste, "Review on solving the inverse problem in EEG source analysis," Journal of NeuroEngineering and Rehabilitation, vol. 5, article 25,DOI: 10.1186/1743-0003-5-25, 2008.
[66] S. Baillet, J. C. Mosher, R. M. Leahy, "Electromagnetic brain mapping," IEEE Signal Processing Magazine, vol. 18 no. 6, pp. 14-30, DOI: 10.1109/79.962275, 2001.
[67] C. J. Holmes, R. Hoge, L. Collins, R. Woods, A. W. Toga, A. C. Evans, "Enhancement of MR images using registration for signal averaging," Journal of Computer Assisted Tomography, vol. 22 no. 2, pp. 324-333, DOI: 10.1097/00004728-199803000-00032, 1998.
[68] B. J. Edelman, B. Baxter, B. He, "EEG source imaging enhances the decoding of complex right-hand motor imagery tasks," IEEE Transactions on Biomedical Engineering, vol. 63 no. 1,DOI: 10.1109/TBME.2015.2467312, 2016.
[69] H. Yuan, T. Liu, R. Szarkowski, C. Rios, J. Ashe, B. He, "Negative covariation between task-related responses in alpha/beta-band activity and BOLD in human sensorimotor cortex: an EEG and fMRI study of motor imagery and movements," NeuroImage, vol. 49 no. 3, pp. 2596-2606, DOI: 10.1016/j.neuroimage.2009.10.028, 2010.
[70] B. Kamousi, A. N. Amini, B. He, "Classification of motor imagery by means of cortical current density estimation and Von Neumann entropy," Journal of Neural Engineering, vol. 4 no. 2, article no. 002, pp. 17-25, DOI: 10.1088/1741-2560/4/2/002, 2007.
[71] H. Ramoser, J. Müller-Gerking, G. Pfurtscheller, "Optimal spatial filtering of single trial EEG during imagined hand movement," IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 8 no. 4, pp. 441-446, DOI: 10.1109/86.895946, 2000.
[72] G. R. Müller-Putz, I. Daly, V. Kaiser, "Motor imagery-induced EEG patterns in individuals with spinal cord injury and their impact on brain-computer interface accuracy," Journal of Neural Engineering, vol. 11 no. 3,DOI: 10.1088/1741-2560/11/3/035011, 2014.
[73] G. Onose, C. Grozea, A. Anghelescu, C. Daia, C. J. Sinescu, A. V. Ciurea, T. Spircu, A. Mirea, I. Andone, A. Spânu, C. Popescu, A.-S. Mihǎescu, S. Fazli, M. Danóczy, F. Popescu, "On the feasibility of using motor imagery EEG-based brain-computer interface in chronic tetraplegics for assistive robotic arm control: A clinical test and long-term post-trial follow-up," Spinal Cord, vol. 50 no. 8, pp. 599-608, DOI: 10.1038/sc.2012.14, 2012.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2017 Alkinoos Athanasiou et al. This work is licensed under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Patients suffering from life-changing disability due to Spinal Cord Injury (SCI) increasingly benefit from assistive robotics technology. The field of brain-computer interfaces (BCIs) has started to develop mature assistive applications for those patients. Nonetheless, noninvasive BCIs still lack accurate control of external devices along several degrees of freedom (DoFs). Unobtrusiveness, portability, and simplicity should not be sacrificed in favor of complex performance and user acceptance should be a key aim among future technological directions. In our study 10 subjects with SCI (one complete) and 10 healthy controls were recruited. In a single session they operated two anthropomorphic 8-DoF robotic arms via wireless commercial BCI, using kinesthetic motor imagery to perform 32 different upper extremity movements. Training skill and BCI control performance were analyzed with regard to demographics, neurological condition, independence, imagery capacity, psychometric evaluation, and user perception. Healthy controls, SCI subgroup with positive neurological outcome, and SCI subgroup with cervical injuries performed better in BCI control. User perception of the robot did not differ between SCI and healthy groups. SCI subgroup with negative outcome rated Anthropomorphism higher. Multi-DoF robotics control is possible by patients through commercial wireless BCI. Multiple sessions and tailored BCI algorithms are needed to improve performance.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details



1 Biomedical Electronics Robotics & Devices (BERD) Group, Lab of Medical Physics, School of Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki (AUTH), 54124 Thessaloniki, Greece; 1st Department of Neurosurgery, “AHEPA” University General Hospital, Aristotle University of Thessaloniki (AUTH), 54636 Thessaloniki, Greece
2 Biomedical Electronics Robotics & Devices (BERD) Group, Lab of Medical Physics, School of Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki (AUTH), 54124 Thessaloniki, Greece
3 1st Department of Neurosurgery, “AHEPA” University General Hospital, Aristotle University of Thessaloniki (AUTH), 54636 Thessaloniki, Greece
4 Biomedical Electronics Robotics & Devices (BERD) Group, Lab of Medical Physics, School of Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki (AUTH), 54124 Thessaloniki, Greece; Robotics Laboratory, Computer Science Department, American College of Thessaloniki (ACT), 55535 Thessaloniki, Greece