1. Introduction
Catheter ablation (CA) is more effective than antiarrhythmic drugs to achieve rhythm control for symptomatic atrial fibrillation (AF), with additional benefits of reducing arrhythmic burden and improving quality of life [1–3]. However, the role of AF ablation in long-term stroke prevention has not been well established [4]. In patients with increased risk of stroke and bleeding, left atrial appendage closure (LAAC) has been proven a safe and effective alternative to long-term anticoagulation [5]. Recently, combined procedures of CA for AF and LAAC for stroke prevention have raised increasing attention in selected patients with high risk of stroke and bleeding. Observational studies without controls have reported the feasibility and safety of performing combined procedures [6–12]. However, clinical outcomes on procedural safety, AF recurrence, and adverse events of the combined procedure compared with those of the single procedures have not been reported. Thus, the aim of this study was to compare the procedural and long-term outcomes of the combined procedure with isolated CA and LAAC.
2. Methods
2.1. Study Population
This single-center retrospective study enrolled consecutive patients with nonvalvular AF between April 2017 and September 2018. All participants were included based on the following criteria: (1) age >18 years undergoing the first CA and/or LAAC procedure, (2) indications for CA were symptomatic nonvalvular AF refractory to antiarrhythmic drugs, and (3) all patients who underwent LAAC met at least one of the following indications: (i) CHA2DS2-VASc scores ≥2 and HAS-BLED scores ≥3; (ii) intolerance to chronic oral anticoagulation (OAC); and (iii) stroke/TIA or thromboembolism (TE) even under OAC treatment [4, 13, 14]. A total of 76 patients underwent combined procedures who constituted the combined group. During the same period of time, isolated CA or isolated LAAC procedures were performed in 549 and 142 patients, respectively. Propensity score matching was used to select controls to minimize potential confounding bias. Patients of the combined group were matched 1 : 1 to those who underwent isolated CA (CA-only group) and to those who underwent isolated LAAC (LAAC-only group), respectively. A total of 228 patients in three groups were included in the final analysis (Figure 1). This retrospective study was approved by the Ethics Committee of Xinhua Hospital affiliated to Shanghai Jiao Tong University School of Medicine and complied with the Declaration of Helsinki. Written informed consent was obtained from each patient.
[figure omitted; refer to PDF]2.2. Preprocedural Assessment
All the procedures were performed in a high-volume AF center (>1000 cases of AF intervention per year) and undertaken by experienced operators who had passed the learning curves of either CA or LAAC. Left atrial appendage (LAA) thrombus exclusion and size measurement were conducted by transesophageal echocardiography (TEE) before procedures. A cardiac computed tomography (CT) scan and 3-dimensional reconstruction of the left atrium were performed before procedure in 82% (187/228) of patients to assist CA and/or LAAC.
2.3. Catheter Ablation Procedure
Under conscious sedation, a decapolar catheter was positioned in the coronary sinus, and two transseptal accesses were obtained through the right femoral vein. Mapping and ablation were performed under the guidance of CARTO (Biosense Webster, Diamond Bar, CA, USA) or EnSite (St. Jude Medical, St Paul, MN, USA) 3-dimensional electroanatomic mapping systems in addition to standard fluoroscopy. For patients with paroxysmal AF, standard pulmonary vein isolation (PVI) was performed, and for those with persistent AF, additional linear and/or complex fragmented atrial electrogram (CFAE) ablations were performed according to the physician’s discretion. Sinus rhythm was restored by either ablation or electric cardioversion.
2.4. Left Atrial Appendage Closure Procedure
LAAC procedure was performed under local anesthesia, and TEE was introduced under deep sedation after device deployment to reconfirm the position of the device before release. The procedure was performed as described previously [15]. In brief, the ostial width and depth of the LAA were measured by LAA angiography or cardiac CT images. A mean left atrial pressure above 10 mmHg was obtained before measurement. Only WATCHMANTM (Boston Scientific, MA, USA) devices were used. The device with appropriate size was chosen when the depth was allowed, generally 3–6 mm oversizing based on the maximum diameter [16]. LAAC implantation was performed after AF ablation was completed in the combined group. We preferred a relatively larger device in the combined procedure since pulmonary vein ridge edema may occur after ablation and in case of the novel leak after the procedure. The device was then advanced into the delivery sheath and deployed by sheath retraction guided by fluoroscopy. Preliminary assessment was performed by angiography and TUG test under fluoroscopy to check the device position and stability. TEE was then performed to reconfirm the position with minimal (<5 mm) to no residual peridevice leaks and a proper compression ratio under deep sedation. The device was released if it was verified by the assessment of “PASS” criteria.
2.5. Postprocedural Anticoagulation
In each group, patients received OAC therapy for at least 3 months following the procedure, unless there were contraindications. In the CA-only group, life-long OAC was recommended for those with a CHA2DS2-VASc score ≥2, and OAC was discontinued for those with a CHA2DS2-VASc score <2 if no AF/AT recurrence was observed 6 months following the procedure [4]. For patients in the combined group and the LAAC-only group, dual antiplatelet therapy was recommended for another 3 months, and then life-long aspirin was prescribed if follow-up TEE showed either complete closure of the LAA or limited residual peridevice flow (jet <5 mm in width) [14].
2.6. Follow-Up
After discharge, office or transtelephonic visits were scheduled at the 3rd month, 6th month, and 12th month following the procedure and once half a year thereafter. ECG or 24 h Holter monitoring was performed at each office visit for patients of the combined group and the CA-only group. Antiarrhythmic drug therapy was discontinued after 3 months if no clinical or documented AF recurrences were identified. In the combined group and the LAAC-only group, TEE was performed to assess the device occlusion safety and efficiency at 45 days and 6 months of follow-up time points according to the percutaneous LAA occlusion Munich Consensus Document [14]. New peridevice leaks were defined as newly detected leaks during the TEE follow-up that were not observed immediately after the device was implanted. Persistent leaks indicated unresolved peridevice leaks from implantation till reassessment.
2.7. Statistical Analysis
Propensity score matching was used to select controls to minimize potential confounding bias. The propensity scores were constructed using the following variables: age, sex, AF type, hypertension, heart failure, diabetes mellitus, coronary artery disease, and previous stroke/TIA/TE. We conducted propensity score matching using nearest neighbor matching, and no matching samples were removed. Patients of the CA-only group and the LAAC-only group were matched to the combined group at a 1 : 1 ratio with a caliper width of 0.10 and 0.18 standard deviations of the propensity score, respectively.
Continuous variables are described as mean ± standard deviation (median (interquartile range) for nonnormal data) and are compared using Student’s t-test (Mann–Whitney U test if normality not satisfied). Categorical variables are presented as percentages and are analyzed using chi-square test or Fisher exact test where appropriate. Variables of the combined group were compared with those of the CA-only and the LAAC-only group, respectively. All analyses were performed using SPSS version 22.0 (IBM Software Inc., Armonk, NY). Two-sided
3. Results
3.1. Baseline Characteristics
Baseline characteristics before propensity score matching are shown in Table 1. Compared with the combined group, patients in the CA-only group were younger and more likely to have paroxysmal AF and a smaller left atrium but less likely to have heart failure and previous stroke/TIA/TE. The CHA2DS2-VASc and HAS-BLED scores of the CA-only group were both lower than those of the combined group. Compared with the combined group, patients in the LAAC-only group were older and more likely to have persistent AF and a larger left atrium.
Table 1
Baseline characteristics of the study population before propensity score matching.
Combined (n = 76) | CA-only (n = 549) | LAAC-only (n = 142) | |||
Female | 37 (48.7) | 285 (51.9) | 0.598 | 67 (47.2) | 0.833 |
Age (years) | 69.9 ± 7.9 | 65.2 ± 7.8 | <0.001 | 72.7 ± 8.4 | 0.019 |
BMI (kg/m2) | 24.6 ± 3.2 | 24.9 ± 3.9 | 0.551 | 24.1 ± 3.0 | 0.274 |
Paroxysmal AF | 37 (48.7) | 344 (62.7) | 0.019 | 30 (21.1) | <0.001 |
Persistent AF | 39 (51.3) | 205 (37.3) | 0.019 | 112 (78.9) | <0.001 |
Coronary artery disease | 16 (21.1) | 74 (13.5) | 0.246 | 37 (26.1) | 0.204 |
Hypertension | 56 (73.7) | 358 (65.2) | 0.143 | 118 (83.1) | 0.099 |
Heart failure | 23 (30.3) | 68 (12.4) | <0.001 | 36 (25.4) | 0.437 |
Diabetes mellitus | 14 (18.4) | 87 (15.8) | 0.568 | 37 (26.1) | 0.204 |
Previous stroke/TIA/TE | 23 (30.3) | 63 (11.5) | <0.001 | 35 (24.6) | 0.371 |
Left atrial diameter (mm) | 42.7 ± 5.7 | 38.4 ± 5.5 | <0.001 | 44.7 ± 6.0 | 0.023 |
LVEF (%) | 63.9 ± 6.3 | 64.8 ± 6.8 | 0.271 | 62.2 ± 7.4 | 0.104 |
CHA2DS2-VASc score | 3.6 ± 1.3 | 2.5 ± 1.5 | <0.001 | 3.9 ± 1.3 | 0.122 |
HAS-BLED score | 3.3 ± 1.1 | 1.7 ± 0.9 | <0.001 | 3.5 ± 1.1 | 0.105 |
After propensity score matching, three groups with each of 76 subjects were matched (Table 2). There were no significant differences in age, AF type, left atrium size, CHA2DS2-VASc score, and comorbidities with heart failure and previous stroke/TIA/TE between the CA-only group and combined group after matching. The HAS-BLED score of the CA-only group was still lower than that of the combined group (2.6 ± 0.9 vs. 3.3 ± 1.1,
Table 2
Baseline characteristics of the study population after propensity score matching.
Combined (n = 76) | CA-only (n = 76) | LAAC-only (n = 76) | |||
Female | 37 (48.7) | 40 (52.6) | 0.626 | 38 (50.0) | 0.871 |
Age (years) | 69.9 ± 7.9 | 69.5 ± 7.8 | 0.703 | 71.36 ± 8.9 | 0.300 |
BMI (kg/m2) | 24.6 ± 3.2 | 24.9 ± 3.7 | 0.543 | 24.3 ± 3.3 | 0.644 |
Paroxysmal AF | 37 (48.7) | 38 (50.0) | 0.871 | 27 (35.5) | 0.100 |
Persistent AF | 39 (51.3) | 38 (50.0) | 0.871 | 49 (64.5) | 0.100 |
Coronary artery disease | 16 (21.1) | 15 (19.7) | 0.840 | 18 (23.7) | 0.426 |
Hypertension | 56 (73.7) | 54 (71.1) | 0.717 | 57 (75.0) | 0.853 |
Heart failure | 23 (30.3) | 21 (27.6) | 0.721 | 19 (25.0) | 0.468 |
Diabetes mellitus | 14 (18.4) | 15 (19.7) | 0.836 | 15 (19.7) | 0.836 |
Previous stroke/TIA/TE | 23 (30.3) | 15 (19.7) | 0.134 | 20 (26.3) | 0.589 |
Left atrial diameter (mm) | 42.7 ± 5.7 | 41.7 ± 4.9 | 0.226 | 44.1 ± 5.9 | 0.163 |
LVEF (%) | 63.9 ± 6.3 | 64.2 ± 5.4 | 0.746 | 62.7 ± 6.9 | 0.273 |
CHA2DS2-VASc score | 3.6 ± 1.3 | 3.4 ± 1.4 | 0.484 | 3.7 ± 1.4 | 0.719 |
HAS-BLED score | 3.3 ± 1.1 | 2.6 ± 0.9 | <0.001 | 3.4 ± 1.1 | 0.413 |
3.2. Catheter Ablation and Arrhythmia Follow-Up
The periprocedural outcomes of catheter ablation are given in Table 3. A total of 39 patients (51.3%) of the combined group and 41 patients (53.9%) of the CA-only group underwent PVI only (
Table 3
Catheter ablation and arrhythmia follow-up.
Combined (n = 76) | CA-only (n = 76) | ||
Mapping system | |||
CARTO | 33 (43.4) | 37 (48.7) | 0.515 |
EnSite | 43 (56.6) | 39 (51.3) | 0.515 |
PVI only | 39 (51.3) | 41 (53.9) | 0.745 |
PVI plus linear/CFAE ablation | 37 (48.7) | 35 (46.1) | 0.745 |
Cardioversion | 36 (47.4) | 32 (42.1) | 0.514 |
Procedure time | 129 ± 24.5 | 133.1 ± 26.4 | 0.388 |
Fluoroscopy time | 5.4 ± 2.4 | 5.9 ± 2.6 | 0.336 |
Complications | 3 (3.9) | 2 (2.6) | 0.650 |
Pericardial effusion | 1 (1.3) | 1 (1.3) | 1.000 |
Stroke | 0 (0.0) | 0 (0.0) | — |
Major bleeding events | 0 (0.0) | 0 (0.0) | — |
Death | 0 (0.0) | 0 (0.0) | — |
Complications of vascular access | 2 (2.6) | 1 (1.3) | 0.561 |
Average follow-up (months) | 24.0 ± 5.2 | 23.7 ± 4.9 | 0.717 |
AF-free follow-up | |||
Overall | 51 (67.1) | 53 (69.7) | 0.727 |
Paroxysmal AF | 26 (70.3) | 29 (76.3) | 0.554 |
Persistent AF | 25 (64.1) | 24 (63.2) | 0.931 |
Redo ablation | 12 (15.8) | 9 (11.8) | 0.481 |
Follow-up events | 3 (3.9) | 4 (5.3) | 0.699 |
Ischemic stroke events | 1 (1.3) | 1 (1.3) | 1.000 |
Bleeding events | 2 (2.6) | 3 (3.9) | 0.650 |
After a mean of 24 months follow-up, 51 patients (67.1%) in the combined group and 53 patients (69.7%) in the CA-only group were AF-free of antiarrhythmic drugs (
3.3. LAAC and TEE Follow-Up
The morphology of the LAA was equally distributed between the combined group and the LAAC-only group (Table 4). The device size used in the combined group was larger according to the ostium width of the LAA, which resulted in obvious larger (but not significantly) device compression (19.6 ± 4.5% vs. 18.4 ± 4.3%,
Table 4
Left atrial appendage closure and device follow-up.
Combined (n = 76) | LAAC-only (n = 76) | ||
Morphology of LAA | |||
Cauliflower | 48 (63.2) | 52 (68.4) | 0.494 |
Chicken wing | 16 (21.1) | 8 (10.5) | 0.075 |
Cactus | 6 (7.9) | 9 (11.8) | 0.415 |
Windsock | 6 (7.9) | 7 (9.2) | 0.772 |
Procedure time | 26.1 ± 6.4 | 28.0 ± 8.7 | 0.149 |
Fluoroscopy time | 4.9 ± 1.9 | 4.5 ± 2.4 | 0.273 |
LAA ostium width (mm) | 22.9 ± 3.1 | 23.5 ± 3.4 | 0.220 |
Device size (mm) | 29.0 ± 3.0 | 29.1 ± 3.2 | 0.876 |
Number of device size changes per patient | |||
0 | 73 (96.1) | 71 (93.4) | 0.468 |
1 | 3 (3.9) | 5 (6.6) | 0.468 |
2 | 0 (0.0) | 0 (0.0) | — |
Device compression (%) | 19.6 ± 4.5 | 18.4 ± 4.3 | 0.085 |
Successful implantation | 76 (100) | 76 (100) | — |
Peridevice leak at implantation | |||
Complete occlusion of LAA | 72 (94.7) | 70 (92.1) | 0.513 |
Leak ≤5 mm | 4 (5.3) | 6 (7.9) | 0.513 |
Leak >5 mm | 0 (0.0) | 0 (0.0) | — |
Complications | 3 (3.9) | 2 (2.6) | 0.650 |
Pericardial effusion | 1 (1.3) | 0 (0.0) | 0.317 |
Stroke | 0 (0.0) | 0 (0.0) | — |
Coronary air embolism | 0 (0.0) | 1 (1.3) | 0.317 |
Bleeding | 0 (0.0) | 0 (0.0) | — |
Death | 0 (0.0) | 0 (0.0) | — |
Complications of vascular access | 2 (2.6) | 1 (1.3) | 0.561 |
Peridevice leak at 45 days | |||
Complete occlusion of LAA | 63 (82.9) | 65 (85.5) | 0.656 |
New peridevice leaks | 10 (13.2) | 7 (9.2) | 0.440 |
Persistent peridevice leaks | 3 (3.9) | 4 (5.3) | 0.699 |
Device-associated thrombosis | 0 (0.0) | 1 (1.3) | 0.317 |
Average follow-up (months) | 24.0 ± 5.2 | 24.1 ± 4.9 | 0.961 |
Follow-up events | 3 (3.9) | 2 (2.6) | 0.650 |
Ischemic stroke events | 1 (1.3) | 0 (0.0) | 0.317 |
Bleeding events | 2 (2.6) | 2 (2.6) | 1.000 |
Data on TEE imaging at least 45 days after the procedure were available in 74 patients (97.3%) of the combined group and 73 patients (96.1%) of the LAAC-only group. Five patients were evaluated by CT imaging. Proper seal (residual leak ≤5 mm) was found in all the patients, and device-related thrombosis (DRT) was detected on TEE in one patient (1.3%) of the LAAC-only group resolving without clinical sequelae on continued oral anticoagulation. Patients with complete occlusion (82.9% vs. 85.5%,
3.4. Antithrombotic Therapy and Clinical Outcomes
After the procedure, 100.0% of the CA-only group, 96.0% of the combined group, and 97.4% of the LAAC-only group were prescribed an OAC, while the rest of the patients were given antiplatelet therapy. During the latest follow-up, 2.6% of the combined group remained on OAC, while antiplatelet was prescribed in 88.2% (80.3% single and 7.9% dual), and the remainder 9.2% received no therapy (Figure 2). The distribution of antithrombotic therapy in the LAAC-only group was similar to the combined group (Figure 2). The rates of antithrombotic therapy in the CA-only group were as follows: 61.8% OAC, 26.3% antiplatelet therapy (17.1% single and 9.2% dual), and 11.8% on no therapy.
[figure omitted; refer to PDF]In the combined group, a total of two patients had bleeding events (one pulmonary and one epistaxis), and one had ischemic stroke during follow-up. The stroke happened in the seventh month after implantation, and the patient was recommenced to change from antiplatelet therapy back to OAC thereafter. There was no evidence for device thrombus in this patient. The event rate for the CA-only group and the LAAC-only group was 5.3% (3 bleeding and 1 ischemic stroke) and 2.6% (2 bleeding and 0 ischemic stroke), respectively (Tables 3 and 4). The event rates of the CA-only group and the LAAC-only group did not differ significantly from those of the combined group (
4. Discussion
The current study presents a comparison between combined procedures and the isolated CA or LAAC procedures. By comparing the single procedures, the combined therapy (i) does not introduce additional periprocedural complications; (ii) has comparable rate of AF recurrence and successful LAA occlusion, and (iii) provides satisfactory outcomes of stroke prevention and bleeding events during a mean of 2 years of follow-up. The results support and supplement the safety and efficacy of the combined procedure of AF ablation and LAAC, which were previously published in noncontrol studies [8–12].
Catheter ablation for AF has been proven to improve symptoms and quality of life, but no randomized clinical trial has shown a reduction in long-term ischemic stroke. [1] On the contrary, percutaneous LAAC, especially with WATCHMAN device, has been demonstrated in randomized trials to reduce stroke and therefore can be an alternative to warfarin therapy for stroke prevention [17, 18]. The combination of LAAC with CA is an effective way to ameliorate the symptoms of AF, while at the same time, reducing the risk of stroke, which was first reported by Swaans et al. in a small observational study of 30 patients [6]. Since then, a number of studies, from single-center experiences to multicenter registry results, support and strengthen that the combined therapy can be feasible, safe, and successful [7–12]. However, all these observational studies lack control groups. In this study, we selected a CA-alone group and a LAAC-alone group by the propensity matching score as control groups for combined therapy. Our results further confirmed favorable outcomes of the combined CA and LAAC therapy in the aspect of safety, AF recurrence, and adverse events.
4.1. Periprocedural Safety
Although AF ablation or LAAC alone has been shown to be safe and effective, concomitant intervention of these two procedures remains to be concerned about whether there are additional periprocedural risks. In this study, the procedure-related complications included one pericardial effusion and two groin haematomas but no death. Compared with CA alone (2.6%) and LAAC alone (2.6%), combined therapy (3.9%) did not significantly increase procedure-related complications. The procedural-related complication/serious adverse event rate of combined therapy reported by two multicenter studies was 6.2% and 2.1% [9, 12], which was consistent with the results of this study. A Chinese multicenter registry with the same population of our study documented a similar rate of 4.1% [11]. Our results support good periprocedural safety of the combined procedure.
4.2. Assessment of AF Recurrence
The impact of LAAC on the rhythm outcomes of AF ablation is still not clear. A small randomized trial suggested that the combination of LAAC with CA did no influence on the long-term AF recurrences [19]. In this case-control study, the AF-free rate of the combined group was comparable with that of the CA-only group after nearly 2 years of follow-up. No differences of recurrence were also observed in subgroups of paroxysmal and persistent AF either. Of note, the presence of the WATCHMAN device did not appear to interfere when doing a repeat ablation. Larger randomized trials are needed to assess whether the use of LAA occlusion devices will impact the outcomes of AF ablation.
4.3. Assessment of LAA Occlusion
The complete occlusion rate of the combined group decreased from 94.7% at implant to 82.9% at 45 days follow-up. Combined procedural data on CA and LAAC of the evolution and wasp studies published by Phillips et al. showed reduction in complete sealing from 97.1% to 61.0% at the first follow-up [8]. Wintgens et al. reported a similar reduction in a multicenter study [9]. Pulmonary vein ridge edema, which may lead to underestimation of the true LAA diameter, was the main concern in combined therapy. In our experience, a larger device size or measurement of the LAA diameter by preprocedure CT may reduce this risk. However, a similar reduction in complete occlusion was observed in the LAAC-only group. The PROTECT AF trial with LAAC procedure alone also demonstrated a peridevice leak rate of 40.9% at 45 days follow-up [20]. Explanations and clinical consequences of the new peridevice leaks require further investigation.
4.4. Antithrombotic Therapy and Clinical Follow-Up
LAAC with WATCHMAN device has been demonstrated superior to warfarin in randomized trials to reduce stroke and bleeding risk [17]. In our study, more than 95% of patients in the combined group and the LAAC-only group discontinued OAC, while more than 60% of the CA-only group continued on OAC. The stroke and bleeding event rates of the combined and LAAC-only groups were similar to those of the CA-only group, although patients of the prior groups had higher bleeding risk and discontinued OAC after 3 months of the procedure. The efficacy of combined therapy in reduction of stroke and bleeding compared with the calculated risk based on CHA2DS2-VASc and HAS-BLED scores was also reported by the two multicenter studies [9, 12].
4.5. Limitations
This was a single-center retrospective study with a moderate sample size. Because the cohort consists of three different nonrandomized groups, a selection bias may exist. We attempted to minimize the selection bias by using a propensity score to select patients to be included into the CA-only and LAAC-only groups. Future larger randomized trials are required to compare combined therapy with CA alone or LAAC alone. The follow-up method of ECG and 24 h Holter recordings for detecting AF recurrence is another limitation. Asymptomatic arrhythmias or nondocumented symptomatic episodes might have been missed.
5. Conclusions
The combination of AF-CA and LAAC is safe and efficacious compared with single procedures alone. Randomized control trials are needed to further verify the benefit of combined therapy in selected patients with AF.
Acknowledgments
This work was supported by grants from the State Major Research Program of the National Natural Science Foundation of China (no. 91949128 to Q.-S.W.), the National Natural Science Foundation of China (no. 81873485 to Q.-S.W.), and the Medical-Engineering Cross Foundation of Shanghai Jiaotong University (no. YG2019QNA47 to B.-F.M.).
[1] D. L. Packer, D. B. Mark, R. A. Robb, "Effect of catheter ablation vs antiarrhythmic drug therapy on mortality, stroke, bleeding, and cardiac arrest among patients with atrial fibrillation: the CABANA randomized clinical trial," Jama, vol. 321 no. 13, pp. 1261-1274, 2019.
[2] C. Blomström-Lundqvist, S. Gizurarson, J. Schwieler, "Effect of catheter ablation vs antiarrhythmic medication on quality of life in patients with atrial fibrillation," Jama, vol. 321 no. 11, pp. 1059-1068, DOI: 10.1001/jama.2019.0335, 2019.
[3] H. Calkins, G. Hindricks, R. Cappato, "HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation," Europace, vol. 20, pp. e1-e160, 2017.
[4] P. Kirchhof, S. Benussi, D. Kotecha, "2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS," Europace, vol. 18 no. 11, pp. 1609-1678, DOI: 10.1093/europace/euw295, 2016.
[5] D. R. Holmes, S. K. Doshi, S. Kar, "Left atrial appendage closure as an alternative to warfarin for stroke prevention in atrial fibrillation," Journal of the American College of Cardiology, vol. 65 no. 24, pp. 2614-2623, DOI: 10.1016/j.jacc.2015.04.025, 2015.
[6] M. J. Swaans, M. C. Post, B. J. W. M. Rensing, L. V. A. Boersma, "Ablation for atrial fibrillation in combination with left atrial appendage closure: first results of a feasibility study," Journal of the American Heart Association, vol. 1 no. 5,DOI: 10.1161/jaha.112.002212, 2012.
[7] N. Calvo, N. Salterain, H. Arguedas, "Combined catheter ablation and left atrial appendage closure as a hybrid procedure for the treatment of atrial fibrillation," Europace, vol. 17 no. 10, pp. 1533-1540, DOI: 10.1093/europace/euv070, 2015.
[8] K. P. Phillips, E. Pokushalov, A. Romanov, "Combining Watchman left atrial appendage closure and catheter ablation for atrial fibrillation: multicentre registry results of feasibility and safety during implant and 30 days follow-up," EP Europace, vol. 20 no. 6, pp. 949-955, DOI: 10.1093/europace/eux183, 2018.
[9] L. Wintgens, A. Romanov, K. Phillips, "Combined atrial fibrillation ablation and left atrial appendage closure: long-term follow-up from a large multicentre registry," EP Europace, vol. 20 no. 11, pp. 1783-1789, DOI: 10.1093/europace/euy025, 2018.
[10] K. P. Phillips, D. T. Walker, J. A. Humphries, "Combined catheter ablation for atrial fibrillation and Watchman left atrial appendage occlusion procedures: five-year experience," Journal of Arrhythmia, vol. 32 no. 2, pp. 119-126, DOI: 10.1016/j.joa.2015.11.001, 2016.
[11] X. Du, H. Chu, P. Ye, "Combination of left atrial appendage closure and catheter ablation in a single procedure for patients with atrial fibrillation: multicenter experience," Journal of the Formosan Medical Association, vol. 118 no. 5, pp. 891-897, DOI: 10.1016/j.jfma.2018.10.006, 2019.
[12] K. P. Phillips, A. Romanov, S. Artemenko, "Combining left atrial appendage closure and catheter ablation for atrial fibrillation: 2-year outcomes from a multinational registry," Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology, vol. 22 no. 2, pp. 225-231, 2020.
[13] M. Glikson, R. Wolff, G. Hindricks, "EHRA/EAPCI expert consensus statement on catheter-based left atrial appendage occlusion-an update," EuroIntervention, vol. 15 no. 13, pp. 1133-1180, DOI: 10.4244/eijy19m08_01, 2020.
[14] A. Tzikas, D. R. Holmes, S. Gafoor, "Percutaneous left atrial appendage occlusion: the Munich consensus document on definitions, endpoints, and data collection requirements for clinical studies," Europace, vol. 19 no. 1,DOI: 10.1093/europace/euw141, 2017.
[15] B. F. Mo, Y. Wan, A. Alimu, "Image fusion of integrating fluoroscopy into 3D computed tomography in guidance of left atrial appendage closure," European Heart Journal-Cardiovascular Imaging,DOI: 10.1093/ehjci/jez286, 2019.
[16] K. Korsholm, S. Berti, X. Iriart, "Expert recommendations on cardiac computed tomography for planning transcatheter left atrial appendage occlusion," JACC: Cardiovascular Interventions, vol. 13 no. 3, pp. 277-292, DOI: 10.1016/j.jcin.2019.08.054, 2020.
[17] V. Y. Reddy, H. Sievert, J. Halperin, "Percutaneous left atrial appendage closure vs warfarin for atrial fibrillation," Jama, vol. 312 no. 19, pp. 1988-1998, DOI: 10.1001/jama.2014.15192, 2014.
[18] D. R. Holmes, S. Kar, M. J. Price, "Prospective randomized evaluation of the watchman left atrial appendage closure device in patients with atrial fibrillation versus long-term warfarin therapy," Journal of the American College of Cardiology, vol. 64 no. 1,DOI: 10.1016/j.jacc.2014.04.029, 2014.
[19] A. Romanov, E. Pokushalov, S. Artemenko, "Does left atrial appendage closure improve the success of pulmonary vein isolation? Results of a randomized clinical trial," Journal of Interventional Cardiac Electrophysiology, vol. 44 no. 1,DOI: 10.1007/s10840-015-0030-4, 2015.
[20] J. F. Viles-Gonzalez, S. Kar, P. Douglas, "The clinical impact of incomplete left atrial appendage closure with the watchman device in patients with atrial fibrillation," Journal of the American College of Cardiology, vol. 59 no. 10, pp. 923-929, DOI: 10.1016/j.jacc.2011.11.028, 2012.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2020 Bin-Feng Mo et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. http://creativecommons.org/licenses/by/4.0/
Abstract
Aim. The feasibility and safety of performing the combined procedure of catheter ablation (CA) and left atrial appendage closure (LAAC) for atrial fibrillation (AF) have been reported by observational studies without controls. The aim of this study was to compare the procedural and long-term outcomes of combined procedures with isolated CA or LAAC. Methods and Results. This study included patients who underwent combined CA and LAAC (combined group), CA alone (CA-only group), or LAAC alone (LAAC-only group). Propensity score matching was used to select controls from the CA-only and LAAC-only groups. Each group contained 76 subjects. The procedures were successfully performed in all the patients. Procedure-related complications of the combined group included one pericardial effusion and two groin haematomas, which did not differ significantly with those of the CA-only group (3.9% vs. 2.6%,
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer