1 Introduction
Ultrafine aerosol particles ( ) might cause severe effects on human health and impact the global climate through the aerosol indirect effect . In the atmosphere, aerosol particles have primary and secondary sources. Secondary particles form when gaseous precursors oxidize to low-volatility compounds, which at high enough abundances can form molecular clusters. Clusters grow by condensation into the nucleation- and Aitken-mode size range and eventually reach sizes at which they can act as cloud condensation nuclei (CCN). A detailed understanding of the mechanisms leading to aerosol formation and growth, i.e., new particle formation (NPF; ), requires the careful measurement of particles in the cluster–particle transition and nucleation-mode size regime. Quantitative measurements of aerosol particles in the size range 1–10 remain a major challenge for the understanding of the mechanisms responsible for NPF.
Condensation particle counters (CPCs), which optically detect nanometer-sized particles after condensational growth, are state-of-the art instruments for particle measurements in the sub-10 nm size regime. Since the introduction of Aitken's “dust counter” towards the end of the 19th century , particle counters capable of detecting smaller and smaller aerosol particles have been developed . Thus, CPCs are used in many different fields, including atmospheric studies , the characterization of combustion processes , monitoring occupational health hazards and clean-room monitoring systems .
CPCs are based on two fundamental processes: heterogeneous nucleation and subsequent condensational growth. The seed particles are mixed with a condensable vapor, followed by a rapid temperature change, leading to vapor supersaturation, particle activation by heterogeneous nucleation and subsequent condensational growth. The seed particles grow to sizes large enough that they can be detected optically. This process can be achieved by different working fluids such as -butanol, water or diethylene glycol (DEG), which are commonly used in commercial instruments. Previous research has shown that the molecular weight and the surface tension of the working fluid have an impact on particle activation . Higher molecular weights and higher surface tensions allow for the activation of smaller particles.
The detection performance of CPCs is commonly quantified by the 50 % cutoff diameter (). The total detection efficiency of a CPC, , is given by the following relation :
1 Here, is the sampling efficiency accounting for the particle losses inside the instrument. corresponds to the activation efficiency of the seed particles by heterogeneous nucleation, and is the detection efficiency of the particles in the optical system.
Heterogeneous nucleation processes are influenced by the physicochemical properties of the seed particles. If the seed particle is entirely soluble in the condensing liquid, Köhler theory can describe the activation behavior. If the seed is insoluble, heterogeneous nucleation theory is applicable. Other relevant seed particle properties for their activation are size, shape, wettability and charging state . Previous studies have shown that the charging state of the seed particles might also have an influence on their respective detection efficiencies . Chemical processes on the molecular level mediate particle activation, including interactions related to the chemical composition of the seed particles and the molecules of the working fluids . Previous research has identified potential molecular characteristics that act as docking stations to the seed particle. These docking stations are polar groups, like – . The distribution of polar groups, in general, defines the chemical polarity of a vapor molecule. A table summarizing the relevant physical properties of the working fluids can be found in the Supplement (Table S1). Furthermore, it has been shown that the activation probability is influenced by the supersaturation occurring in the condenser of the CPC and the working fluid used . As a result the combination of CPCs based on different working fluids might even provide information concerning the chemical composition of the seed particles by showing differing detection efficiencies . Additionally, previous studies have shown that the detection efficiency of CPCs for sub-2 nm particles can be improved by tuning the instrument settings .
Here we present the results of studying the effect of the working fluid on the activation efficiencies with respect to various seed particle compositions. The broad array of investigated particle detectors was composed of -butanol-based CPCs (TSI 3772, TSI 3776 and Airmodus A20), water-based CPCs (TSI 3788 and V-WCPC 3789) and a DEG-based booster stage (TSI 3777). The experiments have been performed under reproducible laboratory conditions and provide further insights into particle activation in CPCs.
2 Experimental methodsA schematic of the experimental setup is displayed in Fig. . The counting efficiencies of the CPCs were measured using four different types of seed particles generated from sodium chloride (, pro analysi, Merck KGaA, Darmstadt, Germany), silver (, wool for elemental analysis, CAS no.: 7440-22-4, Merck KGaA, Darmstadt, Germany), ammonium sulfate (, pro analysi, Merck KGaA, Darmstadt, Germany) and -caryophyllene (, BCY, %, CAS no.: 87-44-5, Sigma Aldrich, St. Louis, USA). To avoid a possible influence of the relative humidity of the carrier gas on the detection efficiency , synthetic air (ALPHAGAZ 1 Air, 99.999 % (5.0), ppm mol, Air Liquide) was used as a carrier gas and the aerosol was generated in dry conditions in all measurements. Sodium chloride, silver and ammonium sulfate particles were produced in a tube furnace manufactured by Carbolite Gero GmbH & Co. KG, Germany. The particle material is put onto a crucible and inserted into the cylindrical furnace. The material evaporates in the heated furnace, and rapid cooling downstream of the heated section leads to particle formation. Table S2 contains the approximate furnace temperatures for different seed particle materials.
Figure 1
Schematic of the experimental setup: the figure shows a schematic of the experimental setup that was used to measure the detection efficiency of the ultrafine particle counters. A Faraday cup electrometer was used as a reference. seeds were generated using BCY (BC), an ozone generator (OG) and a flow tube. The curved arrows mark the sheath air cycle. Panel (a) shows the setup for the measurements involving just CPCs. Due to the higher aerosol inlet flow rate of the TSI 3777 the setup was modified (b).
[Figure omitted. See PDF]
Subsequent to the aerosol generator, a dilution flow of synthetic air is joined with the aerosol flow. A soft X-ray charger (TSI 3088 Advanced Aerosol Neutralizer) is used to achieve a steady-state charge distribution for the analyzed aerosol flow. In the next step, the aerosol enters a custom-made Vienna-type differential mobility analyzer (DMA, as presented in , and referred to as nano-DMA; see the Supplement), where particles are selected according to their electrical mobility. The electrical mobility diameters of the selected particles range between 1 and 25 . According to , the flow rate ratio defines the limiting resolution in the absence of diffusion when the DMA flows are balanced and is given by the sum of the aerosol inlet flow and the sample outlet flow divided by the sum of the sheath flow rate and the excess flow rate : 2 The limiting resolution throughout our experiments is therefore approximated by 0.15. It is important to note that in the 1–2 nm size range diffusional broadening can degrade the DMA resolution . The selected aerosol might be polydisperse and the selected particle diameter must be seen as an average diameter with an envelope (see Fig. S1). In our study this envelope is given by based on a maximum uncertainty approximation by taking into account the results of . The diffusional broadening of the transfer function of the nano-DMA between 1 and 5 was also presented by . In their study the authors calculated a geometric standard deviation of approximately 1.07 for a size-classified output aerosol composed of particles with a mean diameter of 1 . By applying positive voltage to the nano-DMA, particles with negative charge were selected. Negatively charged particles of the same composition are assumed to achieve higher activation efficiencies at small sizes independently of the CPCs used and are thus preferred for applications in size distribution measurements down to particles as small as 1. Previous studies on the counting efficiency of CPCs involving sub-2 nm particles were conducted using high-resolution DMAs . In order to assess the suitability of the nano-DMA used for seed particles with diameters smaller than 2 , previous measurements of negatively charged and seeds using a tuned TSI 3776, the TSI 3777 and a high-resolution DMA (UDMA) were used. Parts of the results and the experimental setup used were published by . The results of the comparison are displayed in the Supplement.
A flow unit, including a total particle filter and a silica gel dryer, maintains a sheath airflow rate of 19.5 L min (see Fig. a) or 33.0 L min (see Fig. b), respectively, in the DMA. The silica gel was exchanged frequently in order to maintain a dry sheath air cycle (closed-loop system). Downstream of the DMA the aerosol is evenly distributed among a CPC and a Faraday cup electrometer (TSI 3068B Aerosol Electrometer) using a T junction. The flow rates were adjusted and monitored using a TSI 4140 mass flowmeter. For the measurements with the DEG-based booster stage, a different flow setup was used due to the higher inlet flow rate of the instrument (see Fig. b). Accordingly, the tube furnace was operated with 2.5 L min of synthetic air and the aerosol inlet flow of the nano-DMA was increased to 5.0 L min. Consequently, the setup used for the generation of seed particles based on , and bears a close resemblance to the setup described in , wherein the authors also present mass spectra of furnace-generated clusters. The study by can be used to infer the chemical composition of some of the seed particles used in this study.
Particles consisting of oxidized -caryophyllene (, ) were produced in a flow tube . The seeds were produced by evaporating BCY into a clean airstream and subsequently mixing it with ozone, allowing the ozonolysis reaction to take place inside the flow tube. Ozone was generated using a custom-made ultraviolet (UV) lamp with adjustable intensity. An intensity–ozone calibration was performed prior to the experiments with an ozone monitor (ThermoFisher Scientific, model i49), suggesting that the ozone concentrations were in the range of 100–500 . The mode diameter of the resulting size distribution can be shifted by varying the length of the reaction path of the flow tube using an adjustable piston and thus the reaction time of the organic compounds. The diameter of the flow tube used was 0.05 , and the length of the reaction path was varied between 0.1 and 0.4 . The temperature of the BCY evaporator was kept at 283.15 .
The CPCs used and the booster stage are presented in Table . All particle counters are based on the laminar-flow principle and have 50 % cutoff diameters below 10 according to the manufacturers. Here, it should be noted that the TSI 3789 was operated in the 2 nm mode for all experiments. The 2 nm mode is the default setting by the manufacturer for low-cutoff measurements based on calibrations with sucrose. The -butanol- and DEG-based instruments implement a typical architecture consisting of a heated saturator and a cooled condenser. In the case of the TSI 3772 and the Airmodus A20 the entire aerosol flow reaches the heated saturator. The TSI 3776 and TSI 3777 are additionally equipped with a capillary regulating the aerosol flow . This capillary–sheath layout helps to keep the aerosol flow centered and enhances the detection efficiency by achieving a higher and sharper supersaturation profile within the condenser and reducing diffusional losses . In the case of the water-based TSI 3788 a cooler conditioner is followed by a heated growth tube . Here, the aerosol flow is first cooled and then enters a heated region of supersaturated vapor. The temperatures of the sections have to be switched due to the higher diffusion coefficient of water compared to heat transfer. Lastly, the TSI 3789 (V-WCPC 3789) is based on a three-step principle: a cool conditioner, a warm initiator and a cooler moderator . The moderator is necessary for removing water vapor and heat while maintaining supersaturated conditions . In contrast to the TSI 3788, the TSI 3789 is not based on a capillary–sheath layout.
Table 1Instrumentation used: the table shows the particle counters used, including their model numbers, the working fluids used and the 50 % cutoff diameter as stated by the manufacturers.
Manufacturer | Model | Working fluid | |
---|---|---|---|
(nm) | |||
TSI | 3776 | -butanol | 2.5 |
TSI | 3772 | -butanol | 10.0 |
TSI | 3788 | water | 2.5 |
TSI | 3789 | water | 2.2 |
TSI | 3777 | DEG | 1.4 |
Airmodus | A20 | -butanol | 7.0 |
By changing the internal settings of three selected CPCs, the 50 % cutoff diameters were pushed to smaller sizes using particles. In the case of the TSI 3777 and the TSI 3789, the boosting was achieved by changing the temperature settings. In the case of the tuned TSI 3776 (denoted as TSI 3776), the internal flows were also adjusted. The valve regulating the sheath airflow is adjusted such that the sample flow rate of the CPC increases to 2.5 L min .
The detection efficiency is determined by comparing the particle number concentration measured by a CPC, , to the particle number concentration of the Faraday cup electrometer (FCE), : 3 After the tube furnace had reached a stable state, the number concentrations were simultaneously measured for 45 s. An uncertainty analysis of measurement data was performed according to the rules of Gaussian error propagation.
3 Results and discussion3.1 Composition-dependent counting efficiencies
Figure shows the detection efficiencies measured with two CPCs (TSI 3776 and TSI 3789) and the CPC–conditioner combination (TSI 3777 and TSI 3772). The abscissa is kept between 1 and 10 in order to set focus on the differences between the seed particles used. The detection efficiency shows a clear dependence on the seed particle composition and the working fluid. The smallest 50 % cutoff diameters using the DEG-based TSI 3777 were accomplished by using and seeds. Both seed particles showed a 50 % cutoff diameter of . In the case of the TSI 3776, seeds were found to have the smallest 50 % cutoff diameter of . Lastly, and seeds exhibited the smallest 50 % cutoff diameter using the TSI 3789 ( ). Interestingly, the TSI 3777 shows a significant difference between the activation of ionically bound (and polar) salts ( and ) and less polar seed particles ( and ): the polar compounds are activated at smaller diameters compared to the nonpolar ones. In the case of the TSI 3776 the detection efficiency curves are very similar for all seed particles, except for . The related detection efficiency curve is shifted towards larger diameters. The water-based CPC shows a much smaller 50 % cutoff diameter for and ammonium sulfate compared to the butanol-based counter ( vs. for ). Overall, the difference between the smallest and the largest 50 % cutoff diameter is approximately 1 for all three CPCs.
Figure 2
Detection efficiencies and working fluids: the figure shows the detection efficiencies for different seed particle materials as a function of the electrical mobility equivalent diameter. Different colors correspond to different seed particles, and every plot is related to a different working fluid: -butanol (a), DEG (b) and water (c).
[Figure omitted. See PDF]
The measured detection efficiency curves show the following two features: (i) the slopes corresponding to the DEG-based TSI 3777 are steeper compared to the -butanol-based TSI 3776 and the water-based TSI 3789. (ii) The counting efficiency of the TSI 3776 and the TSI 3777 reaches 1 around 10 in both cases. In contrast, the detection efficiency of the TSI 3789 reaches 1 at about 20 (see Figs. S2 and S3).
3.2 Comparison of the water-based CPCsSubsequently, the two investigated water-based CPCs were directly compared to each other. Figure shows a direct comparison of the detection efficiency curves for the two different water-based CPCs (TSI 3788 and TSI 3789). The TSI 3788 shows almost identical curves for and ammonium sulfate, with a 50 % cutoff diameter of for ammonium sulfate, while the curves for the less polar compounds ( and ) are shifted by approximately 1 (see Fig. a). The TSI 3789, on the other hand, shows a different kind of compound dependence, since the detection efficiency curves for polar and less polar seeds are not separated as clearly. All 50 % cutoff diameters range from 2.3 to 3.4 and are similar to the cutoff diameters of the TSI 3788, with having the lowest and having a rather high . The curves measured with the TSI 3788 are steeper compared to the TSI 3789. As shown in Fig. c, the lower detection efficiencies for the TSI 3789 below 10 are due to higher internal losses compared to the TSI 3788. This behavior is most probably linked to the fact that the TSI 3789 is not based on a capillary–sheath layout. The upper curve shows the detection efficiency corrected for diffusional losses in the inlet and conditioner according to for seeds. The corrected detection efficiency curves of the other seeds are presented in the Supplement (see Fig. S5).
Figure 3
Detection efficiencies of the water-based CPCs: the figure shows the detection efficiencies as a function of the electrical mobility equivalent diameter. Panel (a) corresponds to the TSI 3788, and panel (b) shows the data measured with the TSI 3789. Panel (c) displays the data set for that has been corrected for diffusion losses; panels (a, b) show uncorrected data.
[Figure omitted. See PDF]
3.3Comparison of the -butanol-based CPCs
Figure shows the direct comparison of the butanol-based CPCs. Three out of the four butanol-based CPCs, i.e., the TSI 3772, the TSI 3776 and the Airmodus A20, show the expected activation pattern: and seeds are activated at smaller diameters compared to and seeds. The Airmodus A20 reaches its plateau at detection efficiency values of about 0.9 at 20 .
The TSI 3776 barely shows a composition dependence of the 50 % cutoff diameters, thereby confirming the results of for seeds. There is no significant difference between the smallest and the largest 50 % cutoff diameters, which are centered around 2.1 . The detection efficiency curves are steep and already reach plateau levels of 1 at 4.5 .
Figure 4
Detection efficiencies of the butanol-based CPCs: the figure shows the detection efficiencies as a function of the electrical mobility equivalent diameter. Every plot corresponds to a different butanol-based CPC: TSI 3772 (a), TSI 3776 (b), Airmodus A20 (c) and TSI 3776 (d).
[Figure omitted. See PDF]
3.450 % cutoff diameters
Figure summarizes the 50 % cutoff diameters measured with the aforementioned instruments operated at standard settings. The discussed dependence of the detection efficiencies on the working fluids and the seed particle compositions can be seen clearly. The DEG-based TSI 3777 and the two water-based CPCs (TSI 3788 and TSI 3789) show the smallest 50 % cutoff diameters for polar and ammonium sulfate seeds. The -butanol-based CPCs activate less polar seeds at the smallest 50 % cutoff diameters. Additionally, our results clearly show that the dependence of the detection efficiency on the seed particle material also influences CPCs with 50 % cutoff diameters larger than 5 as the absolute shifts in the diameters get larger for these models. This applies for the Airmodus A20 and the TSI 3772.
Figure 5
The 50 % cutoff diameters for different working fluids: the figure shows the 50 % cutoff diameters measured with the TSI 3776, Airmodus A20, TSI 3772, TSI 3777, TSI 3788 and TSI 3789. Different colors correspond to different seed particle compositions.
[Figure omitted. See PDF]
3.5 Effect of instrument boostingAs the internal temperature settings in all CPCs used in this work can be adjusted by the customer, a set of experiments was performed to test the instruments with different temperature settings (see Fig. ). The tests were done using seed particles, and the tested CPCs were the butanol-based TSI 3776, the DEG-based TSI 3777 and the water-based TSI 3789. In the case of the TSI 3777 and the TSI 3789, only the temperatures were adjusted, and for the TSI 3776 the flows were also changed (TSI 3776; see Table for tuned settings). The adjustment of the temperatures did not increase the background count rate of the instruments. The results are shown in Fig. . For all tested CPCs, the 50 % cutoff diameters could be reduced (see Table ). The shapes of the curves did not show any significant change except for the TSI 3776, which is due to a change in the internal flow rates, leading to reduced diffusion losses at the instrument's inlet. As a result, the overall counting efficiency is enhanced and exhibits a steeper slope. The 50 % cutoff diameter could be lowered by approximately 35 % in the case of the TSI 3776, TSI 3777 and TSI 3789. Figure a shows the effect of changing the temperature settings and the inlet flow of the TSI 3776. The effect of just changing the temperature settings is shown in Fig. b and c (TSI 3777 and TSI 3789).
Table 2
Temperature settings: the table summarizes the standard temperature settings according to the user manuals of the particle counters and the empirically derived tuned temperature settings (see also ).
Instrument | ||||||
---|---|---|---|---|---|---|
Standard settings (C) | ||||||
TSI 3776 | 10.0 | 39.0 | 40.0 | – | – | – |
TSI 3789 | – | – | 40.0 | 7.0 | 90.0 | 15.0 |
TSI 3777 | 12.0 | 62.0 | – | – | – | – |
Tuned settings (C) | ||||||
TSI 3776 | 1.1 | 33.1 | 34.1 | – | – | – |
TSI 3789 | – | – | 40.0 | 2.0 | 95.0 | 23.0 |
TSI 3777 | 12.0 | 68.0 | – | – | – | – |
Summarized 50 % cutoff diameters: the table shows the measured 50 % cutoff diameters for the four different seed particle materials and for every particle counter in use.
Seed | Working | Instrument | ||
---|---|---|---|---|
fluid | () | () | ||
-butanol | TSI 3776 | 4.0 | 0.4 | |
TSI 3776 | 2.2 | 0.2 | ||
TSI 3772 | 10.0 | 1.0 | ||
A20 | 7.6 | 0.8 | ||
water | TSI 3788 | 2.1 | 0.1 | |
TSI 3789 | 2.3 | 0.1 | ||
DEG | TSI 3777 | 1.6 | 0.3 | |
-butanol | TSI 3776 | 3.0 | 0.3 | |
TSI 3776 | 2.0 | 0.2 | ||
TSI 3772 | 8.8 | 1.0 | ||
A20 | 6.2 | 0.6 | ||
water | TSI 3788 | 4.2 | 0.2 | |
TSI 3789 | 2.7 | 0.1 | ||
DEG | TSI 3777 | 2.5 | 0.3 | |
-butanol | TSI 3776 | 3.1 | 0.4 | |
TSI 3776 | 2.0 | 0.2 | ||
TSI 3772 | 9.9 | 1.2 | ||
A20 | 7.1 | 0.7 | ||
water | TSI 3788 | 2.2 | 0.1 | |
TSI 3789 | 2.3 | 0.2 | ||
DEG | TSI 3777 | 1.6 | 0.3 | |
-butanol | TSI 3776 | 2.8 | 0.3 | |
TSI 3776 | 2.1 | 0.2 | ||
TSI 3772 | 6.7 | 0.7 | ||
A20 | 4.7 | 0.5 | ||
water | TSI 3788 | 3.9 | 0.2 | |
TSI 3789 | 3.4 | 0.2 | ||
TSI 3789 | 3.1 | 0.2 | ||
DEG | TSI 3777 | 2.7 | 0.3 | |
TSI 3777 | 2.3 | 0.3 |
Figure 6
Detection efficiencies of the tuned instruments: the figure shows the detection efficiencies for seeds as a function of the electrical mobility equivalent diameter. The green lines correspond to the standard settings, and the orange lines correspond to the tuned settings. Every plot shows data from a different instrument: TSI 3776 (a), TSI 3777 (b) and TSI 3789 (c).
[Figure omitted. See PDF]
3.6 Discussion of the measurement resultsThe results of the performed measurements show that chemical similarities between the seed particle material and the working fluids influence the detection efficiencies of CPCs. As a rule of thumb, compounds of similar chemical structure dissolve more easily. -butanol is a rather nonpolar fluid and interacts more strongly with particles of nonpolar substances. In contrast, water and DEG are highly polar fluids, and ammonium sulfate and are ionic compounds. Therefore, it is expected that ionic compounds such as and ammonium sulfate easily dissolve in the polar working fluids and that the less polar and particles interact more strongly with the nonpolar -butanol. These chemical similarities of the seed and working fluid are indeed reflected in the counting efficiencies: the 50 % cutoff diameters of the ionic seeds are smaller when measured with the TSI 3789 or TSI 3777. On the other hand, the 50 % cutoff diameters of and seeds are smaller when measured with a butanol-based CPC (see Figs. –). This different activation behavior also requires different theoretical approaches: while particle activation of two soluble compounds can be described by Köhler theory , the Fletcher theory of heterogeneous nucleation needs to be applied when the seed particle is not entirely soluble in the working fluid .
Additionally, the detection efficiency of the instruments is influenced by technical factors. The TSI 3776 and the TSI 3777 are based on a capillary–sheath layout. As a result, particle losses due to diffusion are smaller and the detection efficiencies reach higher values at smaller diameters compared to the TSI 3789 (see Fig. ). Additionally, the detection efficiency curves become steeper.
By changing the temperature settings, higher peak saturation ratios are achieved, leading to smaller 50 % cutoff diameters (see Fig. ). The chosen settings increase the temperature difference between the cooled and heated regions inside the instruments, while simultaneously avoiding significant homogeneous nucleation. Homogeneous nucleation inside the instrument is tested by connecting a total particle filter to the inlet and verifying that no signal is detected. As shown by , lowering the operating temperatures of a TSI 3776 by keeping the temperature difference between the saturator and condenser constant leads to an elevated saturation ratio profile in the condenser tube.
Furthermore, the interpretation of the results of seed particles with diameters is restricted to the knowledge of the exact composition of the seed particle material and cannot be extended to the composition of the produced aerosol particles. Previous research has shown that the use of tube furnaces and aerosol chargers based on radioactivity causes impurities in the produced aerosol particles .
presented results of detection efficiency measurements from the TSI 3776 using and seeds. Our results for and are in perfect agreement with the results for a dry aerosol flow of . As highlight, there is no composition dependence of the detection efficiencies of the TSI 3776. The 50 % cutoff diameters for negatively charged seeds presented by the authors are also confirmed by our measurements. We report the same 50 % cutoff diameters for the TSI 3776 and as reported by , who also used a tuned TSI 3776. We confirm that the discussed trend of smaller 50 % cutoff diameters when using CPCs based on water and DEG (with DEG being linked to the smallest one) is backed by our results. performed a laboratory characterization of the TSI 3788 and compared the results to the TSI 3776. The seed particles used also include negatively charged seeds. The 50 % cutoff diameters for positively charged measured with the TSI 3788 are in good agreement, but we report distinctly larger 50 % cutoff diameters for positively charged measured with the TSI 3776. In their publication about DEG-based particle counters, present 50 % cutoff diameters for negatively charged seeds. The results of our study also support these measurements.
3.7 Supersaturation profiles
To confirm our measurement results regarding the cutoff diameters with theoretical calculations, we simulated the maximum supersaturations for the three butanol-based CPCs (TSI 3776, TSI 3776 and TSI 3772). The supersaturations in the Airmodus A20 were provided by the manufacturer. Measurements of neutral seeds performed with a size analyzing nuclei counter (SANC; ) were used to correctly simulate the maximum supersaturation that seed particles were exposed to . Therefore, we evaluated the heat and mass transfer for a fully developed laminar flow with no mixing according to . The original geometry was transformed to a circular tube to solve the equations with constant boundary conditions (constant temperatures; ). For the flow a parabolic velocity flow profile was considered and used to simulate the convective–diffusive heat and mass transfer for the simulation domain following . The results are presented in Table , showing the calculated corresponding minimum diameters that can be activated. The highest supersaturations (=4.61) are reached in the TSI 3776, and particles are exposed to the lowest supersaturations in the TSI 3772 (), as expected according to our calculations.
Table 4
Calculated supersaturation ratios: the table shows the results of the modeling of the supersaturation profiles of four different particle counters. The maximum supersaturation ratio is related to the smallest particle diameter necessary for the activation of neutral silver seeds.
Instrument | ||
---|---|---|
(nm) | ||
TSI 3776 | 3.43 | 2.20 |
TSI 3776 | 4.61 | 1.80 |
TSI 3772 | 1.68 | 5.40 |
Airmodus A20 | 1.93 | 4.20 |
We correlated the measured 5 % cutoff diameter with the diameters that can be theoretically activated (). The comparison was done for silver particles. In the theoretical calculations neutral particles were used, whereas in the case of the measurements we used negatively charged silver particles. Figure shows a remarkable correlation between the theory and the measurements, resulting in an of 0.98. In the case of the smallest diameter, the calculated diameter is 1.8 and the measured one is . This deviation is most likely due to the charge effect . We are therefore able to correctly predict the onset of detection in CPCs from simulated supersaturation profiles; this implies that we can infer the maximum supersaturation inside the condenser for specific CPCs, which was generally unknown for most CPCs so far. Calculated supersaturation profiles can be found in the Supplement (see Figs. S6 and S7).
Figure 7
Theoretical diameters: the figure shows the theoretically minimal diameters necessary for particle activation of neutral silver seeds as a function of the measured 5 % cutoff diameter for negatively charged silver seeds. Every color corresponds to a different particle counter. Data points on the dashed line stand for the equality of the theoretical and measured diameters.
[Figure omitted. See PDF]
4 ConclusionsThe effect of the chemical composition of seed particles on the detection efficiency in commonly used CPCs based on different working fluids was investigated. These characterizations included organic seed particles that were generated in a controlled way. We present the first characterization measurements of the newly developed TSI 3789 using seed particles with diameters between 1 and 25 . Furthermore, the onset of detection in CPCs was successfully and correctly predicted based on simulated supersaturation profiles.
Chemical similarities between the seed particle material and the working fluids used have been found to have an impact on the detection efficiency of CPCs. It has been shown that shifts in the detection efficiencies also occur for CPCs with 50 % cutoff diameters larger than 5 . Additionally, it was confirmed that the detection efficiency can be improved by changing the temperature and flow rate settings of the instruments. Remarkably, the TSI 3776 also did not show any shifts in the detection efficiency for , and seeds when operated using the tuned settings. This behavior is linked to the higher saturation ratio after tuning the CPC. The saturation ratio might be high enough that differences in the detection efficiencies of different seed particles no longer occur.
Consequently, we conclude that the 50 % cutoff diameter as the sole parameter is not sufficient to characterize the detection efficiency of a CPC. Shifts of the detection efficiency and curve shapes are influenced by the aforementioned interactions. We recommend that CPC characterizations be performed using various polar and less polar seed particles, including seeds, in order to correctly present the detection efficiency of a CPC. The chemical composition of the measured aerosol particles should be considered when instruments are calibrated. The authors recommend following the calibration standard based on seeds, introduced by , when a variety of seed particle materials is not at hand. In the future, measurements of the activation efficiency with could then be used to infer the achieved supersaturation in the CPC. Subsequent comparisons to further SANC measurements with other seed materials (e.g., ; ) then allow researchers to infer the corresponding counting efficiencies. Future research needs to focus on the extension and verification of the presented results for sub-2 nm particles with known composition. Hence, the presented study suggests a novel and improved approach to determining the counting efficiencies of CPCs when a calibration with a variety of seed particles is not available.
Data availability
Raw data are available upon request from the authors.
The supplement related to this article is available online at:
Author contributions
DW, SHS and PMW presented the idea. PJW, DS, SB and DW designed the setup and performed the measurements. CT calculated the supersaturation profiles for the TSI CPCs. PJW, DS, CT and DW analyzed the data. PJW, DS, CT, SB, SHS, PMW and DW were involved in the scientific interpretation of the results. PJW, DS, SB and DW wrote the paper. All authors participated in reviewing the paper.
Competing interests
The authors declare that they have no conflict of interest. This study was independently performed and was not co-funded by TSI Inc. and Airmodus Ltd.
Acknowledgements
The authors thank TSI Inc. for kindly providing a TSI 3789 for the measurements. Furthermore, the authors acknowledge Axel Zerrath for providing technical support concerning the newly developed TSI 3789. Joonas Enroth is acknowledged for calculating the supersaturation profile of the Airmodus A20 and providing it to the authors.
Financial support
This research has been supported by the Austrian Science Fund (FWF) (project no. J3951-N36) and the European Research Council (grant no. 616075).
Review statement
This paper was edited by Pierre Herckes and reviewed by three anonymous referees.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2020. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Condensation particle counters (CPCs) are crucial instruments for detecting sub-10 nm aerosol particles. Understanding the detection performance of a CPC requires thorough characterization under well-controlled laboratory conditions. Besides the size of the seed particles, chemical interactions between the working fluid and the seed particles also influence the activation efficiencies. However, common seed particle materials used for CPC characterizations are not chosen with respect to chemical interactions with vapor molecules of the working fluid by default. Here, we present experiments on the influence of the seed particle material on the detection efficiencies and the 50 % cutoff diameters of commonly used CPCs for the detection of sub-10 nm particles. A remarkable set consisting of six different and commercially available particle detectors, including the newly developed TSI V-WCPC 3789 and a tuned TSI 3776, was tested. The corresponding working fluids of the instruments are
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details






1 University of Vienna, Aerosol Physics and Environmental Physics, Boltzmanngasse 5, 1090 Vienna, Austria
2 University of Vienna, Aerosol Physics and Environmental Physics, Boltzmanngasse 5, 1090 Vienna, Austria; University of Helsinki, Institute for Atmospheric and Earth System Research/Physics, 00014 Helsinki, Finland
3 TSI GmbH, Neukoellner Straße 4, 52068 Aachen, Germany