Abstract

The Radixact® linear accelerator contains the motion Synchrony system, which tracks and compensates for intrafraction patient motion. For respiratory motion, the system models the motion of the target and synchronizes the delivery of radiation with this motion using the jaws and multi‐leaf collimators (MLCs). It was the purpose of this work to determine the ability of the Synchrony system to track and compensate for different phantom motions using a delivery quality assurance (DQA) workflow. Thirteen helical plans were created on static datasets from liver, lung, and pancreas subjects. Dose distributions were measured using a Delta4® Phantom+ mounted on a Hexamotion® stage for the following three case scenarios for each plan: (a) no phantom motion and no Synchrony (M0S0), (b) phantom motion and no Synchrony (M1S0), and (c) phantom motion with Synchrony (M1S1). The LEDs were placed on the Phantom+ for the 13 patient cases and were placed on a separate one‐dimensional surrogate stage for additional studies to investigate the effect of separate target and surrogate motion. The root‐mean‐square (RMS) error between the Synchrony‐modeled positions and the programmed phantom positions was <1.5 mm for all Synchrony deliveries with the LEDs on the Phantom+. The tracking errors increased slightly when the LEDs were placed on the surrogate stage but were similar to tracking errors observed for other motion tracking systems such as CyberKnife Synchrony. One‐dimensional profiles indicate the effects of motion interplay and dose blurring present in several of the M1S0 plans that are not present in the M1S1 plans. All 13 of the M1S1 measured doses had gamma pass rates (3%/2 mm/10%T) compared to the planned dose > 90%. Only two of the M1S0 measured doses had gamma pass rates > 90%. Motion Synchrony offers a potential alternative to the current, ITV‐based motion management strategy for helical tomotherapy deliveries.

Details

Title
Evaluation of radixact motion synchrony for 3D respiratory motion: Modeling accuracy and dosimetric fidelity
Author
Ferris, William S 1   VIAFID ORCID Logo  ; Kissick, Michael W 2 ; Bayouth, John E 3   VIAFID ORCID Logo  ; Culberson, Wesley S 1 ; Smilowitz, Jennifer B 4   VIAFID ORCID Logo 

 Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin‐Madison, Madison, WI, USA 
 Accuray Incorporated, Madison, WI, USA 
 Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin‐Madison, Madison, WI, USA 
 Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin‐Madison, Madison, WI, USA; Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin‐Madison, Madison, WI, USA 
Pages
96-106
Section
RADIATION ONCOLOGY PHYSICS
Publication year
2020
Publication date
Sep 2020
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2443588169
Copyright
© 2020. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.