Abstract
Objective
This study evaluated the biomechanical changes in the adjacent vertebrae under a physiological load (500 N) when the clinically relevant amount of bone cement was injected into fractured cadaver vertebral bodies.
Methods
The embalmed cadaver thoracolumbar specimens in which each vertebral body (T12–L2) had a BMD of < 0.75 g/cm2 were used for the experiment. For establishing a fracture model, the upper one third of the L1 vertebra was performed wedge osteotomy and the superior endplate was kept complete. Stiffness of specimens was measured in different states. Strain of the adjacent vertebral body and intervertebral disc were measured in pre-fracture, post-fracture, and after augmentation by non-contact optical strain measurement system.
Results
The average amount of bone cement was 4.4 ml (3.8–5.0 ml). The stiffness of after augmentation was significantly higher than the stiffness of post-fracture (p < 0.05), but still lower than pre-fracture stiffness (p < 0.05). After augmentation, the adjacent upper vertebral strain showed no significant difference (p > 0.05) with pre-fracture, while the strain of adjacent lower vertebral body was significantly higher than that before fracture (p < 0.05). In flexion, T12/L1 intervertebral disc strain was significantly greater after augmentation than after the fracture (p < 0.05), but there was no significant difference from that before the fracture (p > 0.05); L1/2 vertebral strain after augmentation was significantly less than that after the fracture (p < 0.05), but there was no significant difference from that before the fracture (p > 0.05).
Conclusions
PVP may therefore have partially reversed the abnormal strain state of adjacent vertebral bodies which was caused by fracture.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer