© 2020 Ng et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Introduction

The COVID-19 pandemic has unveiled widespread shortages of personal protective equipment including N95 respirators. Several centers are developing reusable stop-gap respirators as alternatives to disposable N95 respirators during public health emergencies, using techniques such as 3D-printing, silicone moulding and plastic extrusion. Effective sealing of the mask, combined with respiratory filters should achieve 95% or greater filtration of particles less than 1um. Quantitative fit-testing (QNFT) data from these stop-gap devices has not been published to date. Our team developed one such device, the “SSM”, and evaluated it using QNFT.

Methods

Device prototypes were iteratively evaluated for comfort, breathability and communication, by team members wearing them for 15-30min. The fit and seal were assessed by positive and negative pressure user seal checks. The final design was then formally tested by QNFT, according to CSA standard Z94.4–18 in 40 volunteer healthcare providers. An overall fit-factor >100 is the passing threshold. Volunteers were also tested by QNFT on disposable N95 masks which had passed qualitative fit testing (QLFT) by institutional Occupational Health and Safety Department.

Results

The SSM scored 3.5/5 and 4/5 for comfort and breathability. The median overall harmonic mean fit-factors of disposable N95 and SSM were 137.9 and 6316.7 respectively. SSM scored significantly higher than disposable respirators in fit-test runs and overall fit-factors (p <0.0001). Overall passing rates in disposable and SSM respirators on QNFT were 65% and 100%. During dynamic runs, passing rates in disposable and SSM respirators were 68.1% and 99.4%; harmonic means were 73.7 and 1643.

Conclusions

We present the design and validation of a reusable N95 stop-gap filtering facepiece respirator that can match existent commercial respirators. This sets a precedence for adoption of novel stop-gap N95 respirators in emergency situations.

Details

Title
Subject validation of reusable N95 stop-gap filtering facepiece respirators in COVID-19 pandemic
Author
Ng, William C K; Arnaud Romeo Mbadjeu Hondjeu; Syrett, Andrew; Caragata, Rebecca; Rozenberg, Dmitry; Xiao, Zixuan; Anwari, Vahid; Trac, Jessica; Azad Mashari
First page
e0242304
Section
Research Article
Publication year
2020
Publication date
Nov 2020
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2460374041
Copyright
© 2020 Ng et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.