Abstract

Background

A fluid warmer can prevent hypothermia during the perioperative period. This study evaluated the heating capabilities of Hotline and Barkey S-line under different flow rates and initial fluid temperatures, as well as after the extension line installation.

Methods

We measured the temperature of a 0.9% sodium chloride solution at the fluid warmer outlet (TProx) and the extension line end (TDistal) with three different initial fluid temperatures (room, warm, and cold) and two flow rates (250 ml/hr and 100 mL/hr).

Results

At a 250 ml/hr flow rate, the TProx and TDistal values were observed to be higher in Hotline than in S-line when using room-temperature or cold fluid. Administering of the warm fluid at the same flow rate significantly increased the TProx and TDistal values in S-line more than the cold and room-temperature fluids. At flow rates of 100 ml/hr, TDistal values were significantly lower than TProx values in both devices regardless of the initial fluid temperature.

Conclusions

Hotline outperformed S-line for warming fluids at a high flow rate with cold or room-temperature fluids. Administering warm fluid in S-line prevented a decrease in the fluid temperature at a high flow rate. However, at a low flow rate, the fluid temperature significantly decreased in both devices after passing through an extension line.

Details

Title
Influence of flow rate, fluid temperature, and extension line on Hotline and S-line heating capability: an in vitro study
Author
Kim, Hosu; Kim, Tae Kyong; Yoo, Sukha; Jin-Tae, Kim  VIAFID ORCID Logo 
Pages
1-6
Section
Research article
Publication year
2021
Publication date
2021
Publisher
BioMed Central
e-ISSN
14712253
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2478825296
Copyright