About the Authors:
Takaaki Tanaka
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Investigation, Resources, Software, Validation, Visualization, Writing – original draft, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Masashi Hirooka
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Methodology, Project administration, Resources, Software, Visualization, Writing – original draft, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Yohei Koizumi
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Investigation, Resources, Software, Validation, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Takao Watanabe
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Resources, Software, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Osamu Yoshida
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Resources, Software, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Yoshio Tokumoto
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Resources, Software, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Yoshiko Nakamura
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Resources, Software, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Koutarou Sunago
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Resources, Software, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Atsushi Yukimoto
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Data curation, Formal analysis, Resources, Software, Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Masanori Abe
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Writing – review & editing
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
Yoichi Hiasa
Contributed equally to this work with: Takaaki Tanaka, Masashi Hirooka, Yohei Koizumi, Takao Watanabe, Osamu Yoshida, Yoshio Tokumoto, Yoshiko Nakamura, Koutarou Sunago, Atsushi Yukimoto, Masanori Abe, Yoichi Hiasa
Roles Conceptualization, Funding acquisition, Project administration, Writing – original draft, Writing – review & editing
* E-mail: [email protected]
Affiliation: Department of Gastroenterology and Metabology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan
ORCID logo https://orcid.org/0000-0003-4117-339X
Introduction
Liver cirrhosis is the final stage in the evolution of chronic liver disease, and the outcome of affected patients is regulated by the degree of portal hypertension (PH). PH is a clinical syndrome with hallmarks of portosystemic collaterals, splenomegaly, ascites, and encephalopathy [1]. Esophagogastric varices (EGVs) resulting from PH are a serious complication of cirrhosis. The estimated prevalence of EGVs in patients with cirrhosis has been reported to be 50% [2], and the mortality rate of variceal bleeding ranges from 20% to 35% [3]. To detect EGVs, screening endoscopy is recommended for all patients with cirrhosis [4,5]. However, a general program of routine endoscopic screening of these patients may be expensive and unnecessary. Therefore, non-invasive measures for diagnosing EGVs in cirrhotic patients before performing invasive screening endoscopy are required to avoid these challenges.
Over the past few decades, measurement of the hepatic venous pressure gradient (HVPG) has been established as a gold standard for the diagnosis and staging of PH [6–11]. It was previously demonstrated that an HVPG value higher than 10 mmHg predicts the presence of EGVs, whereas a value higher than 12 mmHg is predictive of variceal bleeding [12]. In contrast, HVPG measurement in routine clinical practice is limited by the invasiveness of the procedure and the technical expertise required for its performance. Consequently, different biomarkers and imaging modalities have been evaluated for their usefulness as non-invasive surrogate markers of PH [13–18].
Recently, Colecchia et al. showed that spleen stiffness (SS) is more accurate than other non-invasive parameters in identifying patients with EGVs and those with different degrees of PH [19,20]. Moreover, SS measurement (SSM) is useful as a non-invasive method to evaluate PH. We have previously proposed the splenic elasticity for PH (SEP) score (HVPG = splenic elasticity x 1.63–2.88) and reported a correlation between HVPG and SS [21]. However, the SEP score was measured using the real-time tissue elastography (RTE) method. Transient elastography (TE) is an easy and common method to evaluate organ stiffness. However, TE does not have B-mode imaging; hence, performing SSM is difficult and results in a high rate of inaccurate measurements. Thus, a new method of TE is needed for SSM. We developed a new device and printed on a three-dimensional (3D) printer, with TE and ultrasound-fusion method, to resolve this limitation. We aimed to assess the diagnostic performance of this method for SSM and determine its usefulness for estimating the risk of bleeding caused by EGVs.
Materials and methods
Study design
The study was approved by the Accredited Clinical Research Review Committee National University Corporation Ehime University Clinical Research Review Committee (IRB No. jRCTs062190005). The nature of the study was explained to the patients, and written informed consent was obtained from each patient according to the principles of the Declaration of Helsinki. This was a single-centre, translational, cross-sectional study of patients admitted to the Department of Gastroenterology and Metabology, Ehime University Hospital from June 2019 to September 2019. Patients who underwent abdominal ultrasonography and whose liver stiffness and SS were measured using TE were recruited. All patients underwent computed tomography (CT) scans from the upper abdomen to the pelvic cavity. The selection criteria were: 1) 20 to 90 years of age and 2) Eastern Cooperative Oncology Group (ECOG) Performance Status of 0 to 2. Notably, the patients’ sex was not considered in the course of selection. Abdominal ultrasonography was performed in the laboratory of the inpatient ward; on the same day, patients underwent B-mode examination and measurements using TE on an empty stomach more than 2 hours after eating. All participants were provided with a written explanation of the purpose, method, and details regarding handling personal information of the research.
The primary endpoint was set by comparing the test success rates. We compared the "success rate of the conventional method without attachments and that of the new method with attachments" using the paired t-test. When the change was 0.06, the standard deviation of the conventional and new methods was 0.4, respectively, and the correlation coefficients were set to 0.75, α 0.05, and 1-β 0.8. Although there were 177 cases, the total number of cases was 200, considering those who dropped out. The primary endpoint was the measurement success rate, which was defined based on the following criteria: 1) estimation of SS until 10 measurement values are obtained; 2) estimation of SS 20 times in an event 10 measurement values are not obtained. The measurement rate was calculated as: The number of times the measured value was obtained / the number of measurements. The secondary endpoints were examination time, measured interquartile range (IQR)/ median, and diagnostic ability for EGV. The analysis of the primary and secondary endpoints was based on the Full Analysis Set (FAS). Similarly, we conducted an analysis targeting the Per Protocol Set (PPS) and confirmed the stability of the analysis results. Values of p<0.05 were considered statistically significant.
SSM was performed in 292 patients with chronic liver disease between June 2019 and September 2019. Exclusion criteria were as follows: 1) pregnancy in females, 2) previous splenectomy, and 3) previous open abdominal surgery. Ninety-two patients were excluded from this study because TE image could not be obtained by the M probe due to severe obesity. The remaining 200 enrolled patients were classified into two groups (training and validation sets). Consecutive patients enrolled from June 2019 to August 2019 constituted the training set to evaluate the diagnostic accuracy of TE for EGVs and determine the cut-off level for screening for high-risk EGVs. In the training set, valid SSM could not be obtained in 28 (21.5%) of the 130 patients who were initially evaluated due to unreliable results (success rate <60%; IQR/median >30%), leaving a total of 102 patients. Subsequently, to validate the diagnostic accuracy of TE for SSM, 70 patients were enrolled between August 2019 and September 2019 and included in the validation set. Twenty patients (28.6%) were excluded due to unreliable results, leaving a total of 50 patients. Clinical and laboratory data were collected just before TE.
TE with virtual B-mode imaging using the ultrasound-fusion method
Spleen and liver stiffness were measured using vibration-controlled transient elastography (VCTE) with an M probe from FibroScan502 (EchoSens, Paris, France), by two operators (T.T. and Y.K., with 3 and 10 years of experience, respectively). To confirm the reproducibility, only 20 of 200 patients were measured by both T.T and Y.K. The remaining 180 patients were measured by T.T. only. The liver stiffness measurement was obtained as in previous reports [22]. The criteria for reliability were 10 validated measurements and a median (IQR) success rate of at least 60% or <0.3 of the obtained liver and spleen stiffness values. For assessing the SSM, the patient was placed in the supine position with the left arm in maximum abduction, and the transducer was placed in the left intercostal spaces. To confirm the B-mode images, we followed the ultrasound-fusion method [23,24], which was performed with TE with the newly developed device (Fig 1). Although two different types of devices were used to adjust the measurements, for the attached M and convex probes, each distance was matched between the tip of the probe and the position sensors (a = a’; b = b’; and c = c’). First, the new device with the electromagnetic position sensors was attached to the convex probe (4.0-MHz curvilinear C1-6, GE Healthcare) (S1 Movie). By tilting the convex probe with both the new devices and the sensors, 3D ultrasound volume data were acquired (S2 Movie). An ultrasound machine (LOGIQ E9, GE Healthcare, Chalfont St. Giles, UK) and a low magnetic field generator were used. Both the low magnetic field generator and position sensors were connected to a position-sensing unit embedded in the ultrasound machine. The 3D ultrasound volume data contained spatial information in the generated magnetic field. Next, the sensors were removed from the new device with the convex probe and attached to the other new device with an M probe (Fig 1B) (S1 Movie). Fusion imaging revealed the multiplanar reconstruction (MPR)-ultrasound images, which were constructed from the 3D ultrasound volume data by the direction of the M probe (Fig 2) (S2 Movie). SSM was performed while the ultrasound-fusion image was displayed. Two dimensional (2D)-shear wave elastography (SWE) was performed with the same ultrasound machine.
[Figure omitted. See PDF.]
Fig 1. New devices to attach to the ultrasound probes.
A) To attach the electromagnetic position sensors (black arrowhead) with the convex probe, we developed a device using the three-dimensional (3D) printer. B) To attach the sensors to the M probe, another new device (black arrowhead) was similarly developed. Each distance was matched between the tip of the probe and the position sensors (a = a’, b = b’, c = c’).
https://doi.org/10.1371/journal.pone.0246315.g001
[Figure omitted. See PDF.]
Fig 2. B-mode image of the spleen constructed by ultrasound-fusion image.
The ultrasound-fusion image shows the synchronized view from the real-time B-mode imaging (A) and multiplanar reconstruction (MPR)-ultrasound images (B: Past B-mode imaging, so-called virtual B-mode imaging) simultaneously, side-by-side, in the display of the ultrasound device.
https://doi.org/10.1371/journal.pone.0246315.g002
Esophagogastroduodenoscopy and hepatic venous pressure gradient
Two independent, experienced endoscopists performed the examinations with good agreement to evaluate the size of EGVs (κ-value, 0.95). If discrepancy occurred, final reports, based on a consensus, were adopted. Endoscopic findings were recorded according to the following criteria: F1: Linear dilation, meandering varices; varices were flattened by insufflation. F2: Varices-like prayer beads; varices were not flattened by insufflation. F3: Nodular or mass-like thick varices; varices were not flattened by insufflation. Similarly, whether it was accompanied by a red colour (RC) sign was recorded. Varices with a high risk of rupture were identified as F1 with positive RC sign, or F2, or higher.
For measuring HVPG, the right hepatic vein was catheterized through the right femoral vein, and pressure in both the wedged and free positions was measured using a 5-Fr balloon-tipped catheter. The HVPG was calculated by subtracting the free hepatic venous pressure from the wedged venous pressure.
Statistical analysis
All statistical analyses were performed using SAS version 9.4 (SAS Institute, Cary, NC). Data are presented as medians and IQR. The Kruskal-Wallis test was used with nonparametric data. To predict EGVs, receiver operating characteristic (ROC) curve analysis was performed. The area under the ROC curve (AUC) was calculated by the trapezoidal rule. Optimal cut-off values for predicting EGVs were selected to maximize the sensitivity, specificity, and accuracy. Spearman’s rank correlation coefficient was used to evaluate the correlation between SSM and HVPG. The prediction of the EGVs was calculated, and the performance of SSM for the evaluation of the training set compared with that of validation set was set as the gold standard using a κ agreement test.
Results
The reliability of SSM and patient characteristics
After excluding the patients based on the XL probe measurements, the median (IQR) success rates were compared with and without using the new device (n = 200, Fig 3). The median value of SSM was not significantly different between the 2 researchers (Y.K and T.T.) (p = 0.899). The median SSM success rate was 83.3% (55.6–100%) with the new device and 57.6% (0–83.3%; p <0.001) without the new device (Fig 4A). Especially in patients without splenomegaly (<100 mL), the median TE success rates were 76.9% and 35.0% with and without the new device, respectively (p <0.001). The median value (IQR) of SSM was 0.19 (0.10–0.30) and 0.21 (0.06–0.31; p = 0.785) with and without the new device, respectively (Fig 4B). Moreover, the characteristics of the two independent study groups (training and validation sets) after adequate SSM in the enrolled patients are shown in Table 1. Age, sex, etiology, platelet counts, prothrombin time (PT), total bilirubin, fasting plasma glucose, and haemoglobin A1c were not significantly different between the training and validation set.
[Figure omitted. See PDF.]
Fig 3. Clinical profiles of patients for the spleen stiffness measurement with and without the new device.
IQR = interquartile range.
https://doi.org/10.1371/journal.pone.0246315.g003
[Figure omitted. See PDF.]
Fig 4. Median success rate and IQR/median of SSM with and without the new device.
A) Median success rate of the SSM with the new device was higher than that of the SSM without one. B) No significant difference is observed between SSM with and without the new device. IQR = interquartile range, SSM = spleen stiffness measurement.
https://doi.org/10.1371/journal.pone.0246315.g004
[Figure omitted. See PDF.]
Table 1. Clinical characteristics and laboratory data of patients.
https://doi.org/10.1371/journal.pone.0246315.t001
Diagnostic accuracy for predicting EGVs and complication of PH in the training set
Diagnostic accuracy in predicting high-risk varices in the training set are shown in Fig 5A and 5B and Table 2. SSM with the new device had higher AUC (AUC, 0.880; 95% confidence interval (CI): 0.714–0.917; P <0.0001) than with other methods. The diagnostic accuracy of EGVs analyzed using the corresponding SSM with the new device from the ROC curves corresponded to the cut-off values of SSM at 46.4. The diagnostic accuracy of the cut-off values for SSM with the new device in the validation set had sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), positive likelihood ratio (LR+), and negative likelihood ratio (LR-) of 92.3%; 72.4%; 53.3%; 74.5%; 3.34; and 0.11, respectively. Both SSM with and without the device had low LR-. Regarding other factors, the cut-off values of platelet and liver stiffness measurement were 119,000 /mm3 and 20.7 kPa, respectively. The diagnostic accuracy of complications of PH (presence of ascites and/or encephalopathy) was similar to that of EVs (Fig 5C and 5D and Table 3). SSM with the new device had higher AUC (AUC, 0.887; P <0.0001) than with other methods.
[Figure omitted. See PDF.]
Fig 5. Diagnostic accuracy for predicting EGVs and complications of portal hypertension in the training set.
A) ROC curves for the diagnostic accuracy for predicting EGVs by imaging. B) ROC curves for the diagnostic accuracy for predicting EGVs by biochemical markers. C) ROC curves for the diagnostic accuracy for predicting complications of PH by imaging. D) ROC curves for the diagnostic accuracy for predicting complications of portal hypertension by biochemical markers. SSM = spleen stiffness measurement, LSM = liver stiffness measurement, SWE = shear wave elastography, M2BPGi = Mac-2 binding protein, ALBI = albumin-bilirubin score, ROC = receiver operating characteristic, EGVs = esophagogastric varices, PH = portal hypertension.
https://doi.org/10.1371/journal.pone.0246315.g005
[Figure omitted. See PDF.]
Table 2. Diagnostic accuracy for predicting high-risk varices in the training set.
https://doi.org/10.1371/journal.pone.0246315.t002
[Figure omitted. See PDF.]
Table 3. Diagnostic accuracy for predicting the complication of portal hypertension in the training set.
https://doi.org/10.1371/journal.pone.0246315.t003
A previous study reported that platelet count (150,000/mm3), liver stiffness measurement (LSM) (20 kpa), and SSM (46 kpa) were useful factors for the screening of EGVs [25]. When the platelet count changed from 150,000/mm3 to 120,000/mm3, the kappa coefficient improved (Table 4). Although both SSM with and without the new devices were linearly correlated with HVPG, the r-value was higher for SSM with the new device than without the new device (Fig 6).
[Figure omitted. See PDF.]
Fig 6. SSM with and without the new device linearly correlated with HVPG.
A) SSM with the new device, B) SSM without the new device. SSM = spleen stiffness measurement, HPVG = hepatic venous pressure gradient.
https://doi.org/10.1371/journal.pone.0246315.g006
[Figure omitted. See PDF.]
Table 4. Screening for high-risk esophago-gastric varices.
https://doi.org/10.1371/journal.pone.0246315.t004
Discussion
The SSM success rate was remarkably increased when the B-mode image for the spleen was confirmed using the newly developed device. Splenic size and configuration varied with typical parameters commonly quoted for the normal splenic size of 12 × 7 × 4 cm and weight of 150 g (range 100–200 g).
The spleen is an organ in the abdominal cavity with a smooth serosal surface, which is attached to the retroperitoneum by a falciform-fat ligament with blood vessel supply. The splenic surface is described in conjunction with anatomical positions, including the phrenic nerve innervating the diaphragm and some internal organs’ surface. The internal organ surface is classified into the upheaval of a front part or the stomach and the rear part or the kidney. Thus, visualization of the spleen by ultrasound is difficult, while that of the liver is easy. The rates of SSM failure were similar in two previous studies (14% vs. 11%) [19,20]. This was because of the high body mass index (BMI) in some patients; however, this was more frequently because of the success rate of measurement, which was <60%. SSM failed less frequently in patients with larger measurable spleen volume. In fact, it was easier to localize, under ultrasound guidance, the middle of the region of interest (ROI) in a spleen with a longitudinal diameter of at least 12 cm and a mean anteroposterior diameter of 4 cm [26].
Recently, RTE and acoustic radiation force impulse (ARFI) imaging are suggested as new alternative methods to evaluate SS. We can integrate the imaging modality with real-time B mode imaging. A promising result regarding the precision of the RTE and ARFI technology for non-invasive assessments of organization-induced fibrosis has been reported [21,27]. Because measurement by TE is based on M- and A-mode imaging, it is influenced by the experience of the operator. RTE and the ARFI imaging can be performed under the clear observation of the real measurement position by B-mode imaging; therefore, this can be applied in obese patients. The important clinical advantage of ARFI imaging is that it has a higher success rate of elastic measurement than TE [28,29]. Thus, we developed a new method of TE with B-mode imaging. In an environment with a B-mode image signal by an ultrasound machine other than TE, tissue stiffness measurement cannot be performed. For this reason, TE cannot be simultaneously performed with real-time B-mode imaging. The ultrasound-fusion image shows the synchronized view of the simultaneous real-time B-mode imaging and MPR-ultrasound images (past B-mode imaging, so-called virtual B-mode imaging) side-by-side in the display of the ultrasound device (Fig 3B). In a low magnetic field, the MPR-ultrasound plane, defined by electromagnetic position sensors, was similarly reconstructed as the plane in real-time B-mode imaging. Thus, if the 3D ultrasound volume containing spatial information in the generated magnetic field was acquired by scanning in a manual sweeping manner and electromagnetic position sensors are attached to the M or XL probes, ultrasound-fusion imaging could be used in TE. To attach the electromagnetic position sensors, new devices were developed using a 3D printer. Even when using the new device, whether the push pulse is adequately generated to the spleen is unknown. However, the operator, assisted by a time-motion image, can locate the spleen portion at least 6 cm deep and free of large vascular structures. Thus, this establishes the case for acquired SSM, as the spleen is the only organ in the left hypochondrium whose stiffness can be measured by TE. Thus, the spleen can be measured appropriately from the acquired SSM numerical values. SSM was mainly measured by T.T. (with only 3 years of TE experience). There was a high reproducibility between the non-expert and the expert. To detect the high-risk EGV group, the cut-off value of SSM was reported as 46 kPa in a previous report [24]. In this study, a cut-off level similar to that in the previous report was adopted.
SSM with the new device is a better predictive marker of patients in the high-risk EGV, ascites, or hepatic encephalopathy group. However, SSM without the new device was a better marker when the SSM value could be acquired. Furthermore, the result of 2D-SWE was unexpected; however, the reason could not be clarified. The coloration heterogeneity within the ROI was evident compared to the measurement in the liver; therefore, the 2D-SWE may not have been measured properly. In this study, the cut-off value of the platelet count was 120,000/mm3; thus, the reason for the difference in etiology.
This study had some limitations. First, data collection was performed in a single center, primarily to confirm the efficacy of the newly developed device. To validate the clinical efficacy, multicenter studies need to be performed. Secondly, unlike the liver, histological diagnosis cannot be easily performed in the spleen. Thirdly, this was a cross-sectional study; thus, a prospective study should be performed.
In conclusion, SSM can be precisely performed using this new device with TE and the ultrasound-fusion method. Moreover, the risk of bleeding caused by EGV was estimated by the method.
Supporting information
S1 Movie. Setting a self-made device and replacement.
A self-made device is set to the convex probe, and a position sensor is set to the device. At this time, a mark is made on the right side to prevent a mix up between the left and right sides of the position sensor. (A) Removal of the position sensor from the convex probe device. (B) Setting the self-made device to the M probe during TE. A position sensor is set to the device. At this time, the marking is made on the right side.
https://doi.org/10.1371/journal.pone.0246315.s001
(ZIP)
S2 Movie. Ultrasound-fusion image creation and spleen stiffness measurement.
A). A convex probe is used to visualize the spleen on the left side. At this time, the splenic hilar portion is drawn with the maximum diameter. An ultrasound-fusion image is created by sliding the convex probe from head to tail between the same ribs. (B) The spleen stiffness is measured on the same ribs by TE, a self-made device, and position sensor. The Logic E9 monitor is split into two images, with the ultrasound-fusion image on the right. By vibrating the fusion image, the possibility of measuring the stiffness by TE is determined. TE = transient elastography.
https://doi.org/10.1371/journal.pone.0246315.s002
(ZIP)
S1 File.
https://doi.org/10.1371/journal.pone.0246315.s003
(XLSX)
S2 File.
https://doi.org/10.1371/journal.pone.0246315.s004
(XLSX)
Citation: Tanaka T, Hirooka M, Koizumi Y, Watanabe T, Yoshida O, Tokumoto Y, et al. (2021) Development of a method for measuring spleen stiffness by transient elastography using a new device and ultrasound-fusion method. PLoS ONE 16(2): e0246315. https://doi.org/10.1371/journal.pone.0246315
1. Koh C, Heller T. Approach to the diagnosis of portal hypertension. Clin Liver Dis. 2012;1:133–135.
2. Garcia-Tsao G. Current management of the complications of cirrhosis and portal hypertension: variceal haemorrhage, ascites, and spontaneous bacterial peritonitis. Gastroenterology. 2001;120:726–748. pmid:11179247
3. Jensen DM. Endoscopic screening for varices in cirrhosis: findings, implications, and outcomes. Gastroenterology. 2002;122:1620–1630. pmid:12016427
4. Grace ND. Practice guidelines: diagnosis and treatment of gastrointestinal bleeding secondary to portal hypertension. Am J Gastroenterol. 1997;92:1081–1091.
5. de Franchis R; Faculty Baveno V. Revising consensus in portal hypertension: report of the Baveno V consensus workshop on methodology of diagnosis and therapy in portal hypertension. J Hepatol. 2010;53:762–768. pmid:20638742
6. Bosch J, Abraldes JG, Berzigotti A, Garcia-Pagan JC. The clinical use of HVPG measurements in chronic liver disease. Nat Rev Gastroenterol Hepatol. 2009;6:573–582. pmid:19724251
7. Groszmann RJ, Garcia-Tsao G, Bosch J, Grace ND, Burroughs AK, Planas R, et al. Portal Hypertension Collaborative Group. Beta-blockers to prevent gastroesophageal varices in patients with cirrhosis. N Engl J Med. 2005;353:2254–2261. pmid:16306522
8. Groszmann RJ, Wongcharatrawee S. The hepatic venous pressure gradient: anything worth doing should be done right. Hepatology. 2004;39:280–282. pmid:14767976
9. Perello A, Escorsell A, Bru C, Gilabert R, Moitinho E, Garcia-Pagan JC, et al. Wedged hepatic venous pressure adequately reflects portal pressure in hepatitis C virus-related cirrhosis. Hepatology. 1999;30:1393–1397. pmid:10573517
10. Ripoll C, Groszmann R, Garcia-Tsao G, Grace N, Burroughs A, Planas R, et al.; Portal Hypertension Collaborative Group. Hepatic venous pressure gradient predicts clinical decompensation in patients with compensated cirrhosis. Gastroenterology. 2007;133:481–488. pmid:17681169
11. Ripoll C, Groszmann RJ, Garcia-Tsao G, Bosch J, Grace N, Burroughs A, et al. Portal Hypertension Collaborative Group. Hepatic venous pressure gradient predicts development of hepatocellular carcinoma independently of severity of cirrhosis. J Hepatol. 2009;50:923–928. pmid:19303163
12. Bosch J, Garcia-Pagan JC, Berzigotti A, Abraldes JG. Measurement of portal pressure and its role in the management of chronic liver disease. Semin Liver Dis. 2006;26:348–362. pmid:17051449
13. Buck M, Garcia-Tsao G, Groszmann RJ, Stalling C, Grace ND, Burroughs AK, et al. Novel inflammatory biomarkers of portal pressure in compensated cirrhosis patients. Hepatology. 2014;59:1052–1059. pmid:24115225
14. Castera L, Pinzani M, Bosch J. Noninvasive evaluation of portal hypertension using transient elastography. J Hepatol. 2012;56:696–703. pmid:21767510
15. Hametner S, Ferlitsch A, Ferlitsch M, Etschmaier A, Sch€ofl R, Ziachehabi A, et al. The VITRO score (Von Willebrand factor antigen/thrombocyte ratio) as a new marker for clinically significant portal hypertension in comparison to other non-invasive parameters of fibrosis including ELF test. PLoS One. 2016;11:e0149230. pmid:26895398
16. Sgouros SN, Vasiliadis KV, Pereira SP. Systematic review: endoscopic and imaging-based techniques in the assessment of portal haemodynamics and the risk of variceal bleeding. Aliment Pharmacol Ther. 2009;30:965–976. pmid:19735231
17. Sirli R, Sporea I, Popescu A, Danila M. Ultrasound-based elastography for the diagnosis of portal hypertension in cirrhotics. World J Gastroenterol. 2015;21:11542–11551. pmid:26556985
18. Thabut D, Imbert-Bismut F, Cazals-Hatem D, Messous D, Muntenau M, Valla DC, et al. Relationship between the Fibrotest and portal hypertension in patients with liver disease. Aliment Pharmacol Ther. 2007;26:359–368. pmid:17635370
19. Stefanescu H, Grigorescu M, Lupsor M, Procopet B, Maniu A, Badea R. Spleen stiff ness measurement using Fibroscan for the noninvasive assessment of esophageal varices in liver cirrhosis patients. J Gastroenterol Hepatol. 2011;26:164–170. pmid:21175810
20. Colecchia A, Montrone L, Scaioli E, Bacchi–Reggiani ML, Colli A, Casazza G, et al. Measurement of spleen stiffness to evaluate portal hypertension and the presence of esophageal varices in patients with HCV-related cirrhosis. Gastroenterology. 2012;143:646–654. pmid:22643348
21. Hirooka M, Ochi H, Koizumi Y, Kisaka Y, Abe M, Ikeda Y, et al. Splenic elasticity measured with real-time tissue elastography is a marker of portal hypertension. Radiology. 2011;261:960–968. pmid:21926379
22. Kudo M, Shiina T, Moriyasu F, Iijima H, Tateishi R, Yada N, et al. JSUM ultrasound elastography practice guidelines: liver. J Med Ultrason (2001). 2013;40:325–357. pmid:27277450
23. Hiraoka A, Hirooka M, Koizumi Y, Hidaka S, Uehara T, Ichikawa S, et al. Modified technique for determining therapeutic response to radiofrequency ablation therapy for hepatocellular carcinoma using US-volume system. Oncol Rep. 2010;23:493–497. pmid:20043112
24. Minami Y, Minami T, Hagiwara S, Ida H, Ueshima K, Nishida N, et al. Ultrasound-ultrasound image overlay fusion improves real-time control of radiofrequency ablation margin in the treatment of hepatocellular carcinoma. Eur Radiol. 2018;28:1986–1993. pmid:29196855
25. Colecchia A, Ravaioli F, Marasco G, Colli A, Dajti E, Di Biase AR, et al. A combined model based on spleen stiffness measurement and Baveno VI criteria to rule out high-risk varices in advanced chronic liver disease. J Hepatol. 2018;69:308–317. pmid:29729368
26. Calvaruso V, Bronte F, Conte E, Simone F, Craxì A, Di Marco V. Modified spleen stiffness measurement by transient elastography is associated with presence of large oesophageal varices in patients with compensated hepatitis C virus cirrhosis. J Viral Hepat. 2013;20:867–874. pmid:24304456
27. Takuma Y, Morimoto Y, Takabatake H, Toshikuni N, Tomokuni J, Sahara A, et al. Measurement of spleen stiffness by acoustic radiation force impulse imaging identifies cirrhotic patients with esophageal varices. Gastroenterology. 2013;144:92–101. pmid:23022955
28. Rifai K, Cornberg J, Mederacke I, Bahr MJ, Wedemeyer H, Malinski P, et al. Clinical feasibility of liver elastography by acoustic radiation force impulse imaging (ARFI). Dig Liver Dis. 2011;43:491–497. pmid:21439919
29. Rizzo L, Calvaruso V, Cacopardo B, Alessi N, Attanasio M, Petta S, et al. Comparison of transient elastography and acoustic radiation force impulse for non-invasive staging of liver fibrosis in patients with chronic hepatitis C. Am J Gastroenterol. 2011;106:2112–2120. pmid:21971536
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2021 Tanaka et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
[...]a general program of routine endoscopic screening of these patients may be expensive and unnecessary. [...]non-invasive measures for diagnosing EGVs in cirrhotic patients before performing invasive screening endoscopy are required to avoid these challenges. [...]different biomarkers and imaging modalities have been evaluated for their usefulness as non-invasive surrogate markers of PH [13–18]. Recently, Colecchia et al. showed that spleen stiffness (SS) is more accurate than other non-invasive parameters in identifying patients with EGVs and those with different degrees of PH [19,20]. [...]SS measurement (SSM) is useful as a non-invasive method to evaluate PH. [...]TE does not have B-mode imaging; hence, performing SSM is difficult and results in a high rate of inaccurate measurements. [...]a new method of TE is needed for SSM.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer