Abstract
Background
The Siemens high-resolution research tomograph (HRRT - a dedicated brain PET scanner) is to this day one of the highest resolution PET scanners; thus, it can serve as useful benchmark when evaluating performance of newer scanners. Here, we report results from a cross-validation study between the HRRT and the whole-body GE SIGNA PET/MR focusing on brain imaging.
Phantom data were acquired to determine recovery coefficients (RCs), % background variability (%BG), and image voxel noise (%). Cross-validation studies were performed with six healthy volunteers using [11C]DTBZ, [11C]raclopride, and [18F]FDG. Line profiles, regional time-activity curves, regional non-displaceable binding potentials (BPND) for [11C]DTBZ and [11C]raclopride scans, and radioactivity ratios for [18F]FDG scans were calculated and compared between the HRRT and the SIGNA PET/MR.
Results
Phantom data showed that the PET/MR images reconstructed with an ordered subset expectation maximization (OSEM) algorithm with time-of-flight (TOF) and TOF + point spread function (PSF) + filter revealed similar RCs for the hot spheres compared to those obtained on the HRRT reconstructed with an ordinary Poisson-OSEM algorithm with PSF and PSF + filter. The PET/MR TOF + PSF reconstruction revealed the highest RCs for all hot spheres. Image voxel noise of the PET/MR system was significantly lower. Line profiles revealed excellent spatial agreement between the two systems. BPND values revealed variability of less than 10% for the [11C]DTBZ scans and 19% for [11C]raclopride (based on one subject only). Mean [18F]FDG ratios to pons showed less than 12% differences.
Conclusions
These results demonstrated comparable performances of the two systems in terms of RCs with lower voxel-level noise (%) present in the PET/MR system. Comparison of in vivo human data confirmed the comparability of the two systems. The whole-body GE SIGNA PET/MR system is well suited for high-resolution brain imaging as no significant performance degradation was found compared to that of the reference standard HRRT.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details

1 University of British Columbia, Department of Physics and Astronomy, Vancouver, Canada (GRID:grid.17091.3e) (ISNI:0000 0001 2288 9830); Eberhard-Karls University Tuebingen, Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, Tuebingen, Germany (GRID:grid.10392.39) (ISNI:0000 0001 2190 1447); University of Tuebingen, Cluster of Excellence iFIT (EXC 2180) “Image Guided and Functionally Instructed Tumor Therapies”, Tuebingen, Germany (GRID:grid.10392.39) (ISNI:0000 0001 2190 1447)
2 University of British Columbia, Department of Physics and Astronomy, Vancouver, Canada (GRID:grid.17091.3e) (ISNI:0000 0001 2288 9830); University of British Columbia, Pacific Parkinson’s Research Centre, Vancouver, Canada (GRID:grid.17091.3e) (ISNI:0000 0001 2288 9830)
3 University of British Columbia, Pacific Parkinson’s Research Centre, Vancouver, Canada (GRID:grid.17091.3e) (ISNI:0000 0001 2288 9830)
4 University of British Columbia & Vancouver Coastal Health, Djavad Mowafaghian Centre for Brain Health, Pacific Parkinson’s Research Centre, Vancouver, Canada (GRID:grid.17091.3e) (ISNI:0000 0001 2288 9830)
5 GE Healthcare Canada, Global MR Applications & Workflow, Vancouver, Canada (GRID:grid.17091.3e)
6 University of British Columbia, UBC MRI Research Centre, Vancouver, Canada (GRID:grid.17091.3e) (ISNI:0000 0001 2288 9830)
7 University of British Columbia, Department of Physics and Astronomy, Vancouver, Canada (GRID:grid.17091.3e) (ISNI:0000 0001 2288 9830)