Abstract

Background

Temporary epicardial pacing wires, implemented in patients during heart transplantation, are routinely removed before discharge. However, in some cases, these wires may remain in situ and are often considered as a contraindication for cardiovascular magnetic resonance (CMR) imaging in the future. Therefore, we aimed to provide data about safety and image quality of CMR in these patients.

Methods

This is a report on a subpopulation out of 88 patients after heart transplantation that were included in a prospective cohort study and underwent multiple CMR in their post-transplant course. During CMR, patients were monitored by electrocardiogram and all examinations were observed by a physician to document potential adverse events. Additionally, image quality was assessed by an imaging specialist.

Results

Nineteen of 88 patients included had temporary pacing wires in situ. These patients underwent a total of 51 CMR studies. No major adverse event and only one single, mild sensory event could be documented. All CMR studies showed preserved diagnostic image quality. Temporary pacing wires were visible in 100% of HASTE and cine sequences. In less than 50% of the examinations, temporary pacing wires were also visible in T1 and T2 mapping, short tau inversion recovery (STIR), and late gadolinium enhancement (LGE) sequences, without any impairment of image quality.

Conclusions

With a low event rate of only one mild adverse event during 51 CMR examinations (2%), CMR appears to be safe in patients with retained temporary epicardial pacing wires after heart transplantation. Moreover, image quality was not impaired by the presence of pacing wires.

Details

Title
Safety and image quality of cardiovascular magnetic resonance imaging in patients with retained epicardial pacing wires after heart transplantation
Author
Gatterer, Constantin  VIAFID ORCID Logo  ; Marie-Elisabeth Stelzmüller  VIAFID ORCID Logo  ; Kammerlander, Andreas  VIAFID ORCID Logo  ; Zuckermann, Andreas  VIAFID ORCID Logo  ; Krššák, Martin  VIAFID ORCID Logo  ; Loewe, Christian  VIAFID ORCID Logo  ; Beitzke, Dietrich  VIAFID ORCID Logo 
Pages
1-8
Section
Research
Publication year
2021
Publication date
2021
Publisher
BioMed Central
ISSN
10976647
e-ISSN
1532429X
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2502625608
Copyright
© 2021. This work is licensed under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.