1 Introduction
Aerosol–cloud interactions are a source of uncertainty in predicting future radiative forcing . One important aerosol–cloud interaction is the ice nucleation of supercooled liquid water droplets caused by ice-nucleating particles (INPs). Heterogeneous freezing can occur at temperatures as warm as 1 C for certain bacterial
Figure 1
(a) Photograph of FINC. (b) Computer-aided design (CAD) software model of the aluminum and stainless-steel movable structure placed inside the LAUDA bath. The circled numbers correspond to the following parts: (1) a chip-on-board LED array, (2) a thin polytetrafluoroethylene sheet, (3) a camera, (4) three clear polypropylene Piko™ PCR trays, (5) a level sensor, (6) a peristaltic pump, (7) an Arduino board and a stepper motor driver, (8) an ethanol reservoir, and (9) a Plexiglas plate. (c) Flowchart of the bath leveler setup. (d) Image of two of the three trays taken by the FINC camera showing the difference in light intensity between the liquid (circled in red) and frozen (circled in blue) wells used for freezing temperature detection.
[Figure omitted. See PDF]
This prediction is challenging due to spatial and temporal heterogeneity in the source, number, and composition of INPs and INMs. In order to reduce uncertainties, advanced methods are needed to quantify and characterize INPs and INMs from ambient and laboratory samples. A variety of laboratory instruments has been developed to measure INPs in the immersion freezing mode. Methods include continuous-flow diffusion chambers
Each bench-top immersion freezing method has its advantages and disadvantages, which vary depending on the samples of interest. Herein, we compiled a summary of multi-drop bench-top immersion freezing instruments used for atmospheric ice nucleation measurements published between 2000 and 2020 (Table 1). Included in this summary is a brief description of the operation of each instrument, the water background with the reported protocol, the average drop size, and the average number of droplets per experiment. Generally, advantageous qualities include large operating temperature ranges, low background freezing temperatures, and a high number of drops per experiment. As these types of instruments are not yet commercial, we also built our own drop Freezing Ice Nuclei Counter (FINC) using a cooling bath and an optical detection method. In comparison to the existing methods, FINC fits well within the range of operating parameters with drop sizes of 5–60 L, 288 drops per experiment, and background freezing at 25 C (Table 1). Noteworthy features of FINC compared to existing methods are its automation of the ethanol level, its use of 288 wells to increase statistics, and its improved code for well detection and for harmonizing the output data.
With an increasing number of research groups developing DFTs, there is an ongoing search for suitable standards for freezing temperature intercomparisons. A typical standard used to compare immersion freezing instruments is the mineral dust NX-illite, a known ice-active mineral and a cheap and readily available material . However, NX-illite measurements can differ by orders of magnitude across different instruments . This discrepancy may be due to NX-illite's insolubility in water, creating a suspension rather than a homogeneous solution. In practice, NX-illite suspensions settle quickly, potentially leading to a range of freezing temperatures. Cellulose has also been used as an intercomparison standard , as it is the most abundant biopolymer in the environment and can contribute to ice nucleation in clouds below about 21 C . However, cellulose is also a suspension in water. Snomax® has additionally been used as a bacterial ice-nucleating standard and consists of freeze-dried, irradiated cells from P. syringae . Unfortunately, found that solutions of Snomax can have irreproducible ice-nucleating activity over time, making it a rather poor standard.
An alternative to mineral dust, cellulose, and Snomax is the use of a water-soluble organic material as a standard in immersion freezing experiments. Here, we show that commercial lignin, a complex organic polymer from the cell wall structure of vascular plants , can serve as a reproducible standard for ice nucleation across different immersion freezing techniques. Indeed, lignin is a water-soluble macromolecule with ice-nucleating activity, thereby qualifying it as an INM . Furthermore, lignin and its oxidation products are present in the atmosphere and are emitted, for example, during agricultural harvesting and biomass burning in Houston, Texas, with typical plume concentrations of 149 ng m . Lignin is also produced as a by-product of an industrial process in which wood is converted to wood pulp and subsequently used for paper products . Recent research has shown lignin to be ice-active, albeit with colder freezing temperatures than leaf litter and agricultural dust . Several other studies have shown that plant materials, which may have included lignin, can be ice-active in immersion freezing
Herein, we present (1) the development, characterization, and validation of our home-built FINC for the quantification of INP and INM concentrations in the immersion freezing mode, (2) an intercomparison of DFTs to validate FINC, and (3) the use of soluble lignin as an intercomparison standard. As part of our intercomparison study with lignin, we show results with two other drop freezing instruments: ETH Zurich's DRoplet Ice Nuclei Counter Zurich
List of droplet freezing techniques (DFTs) ordered by year of first publication. Information includes the instrument name, a brief description of the instrument, the freezing temperature of a water background freezing experiment according to the reported protocol (median where available), the droplet size used, the number of droplets per experiment (drops/expt.), and the main references associated with the instrument, including the reference in which the instrument was first published and, if applicable, a following reference with updates; “n.f.” indicates information not found.
Instrument name | Brief description | Water background | Drop size | Drops/expt. | References |
---|---|---|---|---|---|
using reported | |||||
protocol (C) | |||||
Flow cell microscopy technique for aerosol phase transitions | Vapors condensed onto the bottom of a sample cell on aluminum cooling block; freezing monitored via microscope | 37 | 7–33 m diam. | , | |
Droplet freezing technique (DFT) | Particles deposited on a glass slide in a sample cell on a cold stage, with droplets grown by water vapor; freezing monitored via microscope | 37 | 120 m diam. | 100 | , |
Microfluidic apparatus | Flow-focusing nozzle continuously produces droplets in a stream of fluorocarbon across a seven-temperature-zone cold plate; freezing monitored via microscope | 37 | 80 m diam. | 10 000 | |
FRankfurt Ice Deposition freezinG Experiment – Tel Aviv University (FRIDGE-TAU) | Pipetted drops onto Vaseline-coated Peltier cold stage in low-pressure diffusion chamber; freezing monitored via CCD camera | 35 | 2 L | 120 | , |
Picoliter and Nanoliter Nucleation by Immersed Particle Instrument (pico-NIPI, nano-NIPI) | Nebulized droplets encased in silicon oil on hydrophobic glass slides on aluminum cold stage; freezing monitored via microscope | 37 | 0.25–1.7 pL; 0.1–6 nL | 135 (pL); 51 (nL) | , , |
Vienna Optical Droplet Crystallization Analyzer (VODCA) | Water–oil emulsion pipetted onto a glass slide on a Peltier cold stage, all contained in an airtight cell; freezing monitored via microscope | 36 | 10–200 m diam. | n.f. | |
Drop freezing apparatus for filters | Filter cutouts placed inside small tubes with water, cooled in a water bath; freezing monitored by manual inspection | 12 | 0.1 mL | 108 | |
Microliter Nucleation by Immersed Particle Instrument (microL-NIPI) | Drops pipetted onto a hydrophobic glass slide in humidity-controlled enclosure on a cold stage; freezing monitored via camera | 26 | 1 L | 40 | , |
North Carolina State University cold stage (NC State-CS) | Emulsion of water in squalene placed on a glass slide resting in an aluminum dish on a thermoelectric element; freezing monitored via camera | 34 to 36 | 400 pL–150 nL | 300–1500 | , |
Microfluidic device for homogeneous ice nucleation | Microfluidically produced water-in-oil emulsion on cryo-microscopy cold stage; freezing monitored via microscope (alternatively frozen with differential scanning calorimetry) | 36 to 37 | 53–96 m diam. | 1000 | |
LED-based Ice Nucleation Detection Apparatus (LINDA) | Sample in tubes held in polycarbonate tray atop an LED array submersed in a water–glycerin cooling bath; freezing monitored via camera | 15 | 40–400 L | 52 | |
Colorado State University Ice Spectrometer (CSU-IS) | Sample aliquots pipetted into two 96-well PCR trays cooled on custom cold blocks with N2 flow; freezing monitored via camera | 25 | 50 L | 192 | , , |
Bielefeld Ice Nucleation ARraY (BINARY) | Droplets pipetted onto a glass slide with separated compartments atop a Peltier cold stage, all enclosed in a, N2-purged chamber; freezing monitored via camera | n.f. | 0.5–5 L | 36 | |
Continued.
Instrument name | Brief description | Water background | Drop size | Drops/expt. | References |
---|---|---|---|---|---|
using reported | |||||
protocol (C) | |||||
Water-Activity-Controlled Immersion Freezing Experiment (WACIFE) | Droplets pipetted onto a glass plate in a humidity-controlled aerosol conditioning cell, then sealed from ambient air and cooled on a cold stage; freezing monitored via microscope | 37 | 60–129 m diam. | 30–50 | |
National Institute of Polar Research Cryogenic Refrigerator Applied to Freezing Test (NIPR-CRAFT) | Drops pipetted onto a Vaseline-coated aluminum plate cooled on a cryogenic refrigerator stage; freezing monitored via camera | 33 | 5 L | 49 | |
Karlsruhe Institute of Technology Cold Stage (KIT-CS) | Droplets printed on a silicon substrate by piezo-driven drop-on-demand generator, and drops then covered in silicone oil and placed on cold stage; freezing monitored via CCD camera | 36 | 215 70 pL | 1500 | |
Microplate partially submersed in cooling liquid | Droplets contained in 96-well microplates partially submersed in a cooling water–alcohol bath; freezing monitored via infrared camera | 17.3 | 150 L | 96–768 | |
Carnegie Mellon University Cold Stage (CMU-CS) | Droplets of water in oil on a substrate in aluminum chamber cooled with a thermoelectric element; freezing monitored via microscope | 27 to 28 | 1 or 0.1 L | 30–40 | , |
Microfluidic device and cold stage | Microfluidically generated drops in oil on a glass slide on a cryostage; freezing monitored via microscope camera | 37 | 35 m diam. | 200 | |
Automated Ice Spectrometer (AIS) | Drops in two 96-well PCR trays fitted into aluminum blocks fixed in a liquid cooling bath, all enclosed in an acrylic box; freezing monitored via camera | 25 to 27 | 50 L | 192 | |
National Oceanic and Atmospheric Administration Drop Freezing Cold Plate (NOAA-DFCP) | Drops pipetted onto Vaseline-coated copper disk placed on a thermoelectric cold plate and covered in a plastic dome; freezing monitored optically | 30 | 2.5 L | 100 | |
Peking University Ice Nucleation Array (PKU-INA) | Drops pipetted into compartments on a glass slide atop a cold stage in a N2-purged box; freezing monitored via CCD camera | 26 | 1 L | 90 | |
WeIzmann Supercooled Droplets Observation on Microarray (WISDOM) | Microfluidically produced droplet array on a PDMS surface placed on a cryostage purged with N2; freezing monitored via microscope camera | 36 | 40 or 100 m diam. | 550 (40 m) or 120 (100 m) | |
Twin-plate Ice Nucleation Assay (TINA) | Droplets contained in four multiwell plates ( and ) placed on two custom aluminum cooling blocks; freezing monitored via infrared camera | 25(96-well), 28 (384-well) | 3 L | 960 | |
Freezing on a chip | Drops loaded on a silicon plate with etched cavities and set on thermoelectric cooler in an N-flushed cell; freezing monitored via camera | 37.5 | 4–300 pL | 25 | |
InfraRed Nucleation by Immersed Particles Instrument (IR-NIPI) | Drops pipetted into 96-well plate on a cold stage enclosed in a chamber; freezing monitored via infrared camera | 22 | 50 L | 96 | |
Continued.
Instrument name | Brief description | Water background | Drop size | Drops/expt. | References |
---|---|---|---|---|---|
using reported | |||||
protocol (C) | |||||
Ice Nucleation Droplet Array (INDA) | Samples placed in wells of a 96-well PCR tray cooled in a cooling bath; freezing monitored via CCD camera | 25 | 50 L | 96 | |
Leipzig Ice Nucleation Array (LINA) | Droplets pipetted into compartments on a glass slide and cooled on a Peltier element; freezing monitored via CCD camera | 30 | 1 L | 90 | |
Microfluidic droplet freezing assay | Microfluidically produced droplets in oil collected in microwells on glass slides placed on a Peltier cold stage in an airtight chamber; freezing monitored via microscope camera | 35 | 83–99 m diam. | 250–500 | |
Drop freeze assay experiment directly on exposed filters | Droplets pipetted onto filters placed on a glass slide and cold stage in an N2-purged chamber; freezing monitored via camera | 30 | 1 L | ||
West Texas Cryogenic Refrigerator Applied to Freezing Test (WT-CRAFT) | Drops pipetted onto Vaseline-coated aluminum plate and cooled on a cryogenic refrigerator stage; freezing monitored via camera | 26 | 3 L | 49 | |
DRoplet Ice Nuclei Counter Zurich (DRINCZ) | Droplets pipetted into 96-well PCR tray submersed in ethanol cooling bath; freezing monitored via camera | 22.5 | 50 L | 96 | |
Cold stage to detect the most active INP in single crystals | Ice crystals placed with ultrapure water on a copper cold plate, melted, and refrozen; freezing monitored via manual inspection | 25 | 3 L | 4 | |
“Store and create” microfluidic device | Microfluidically generated droplets in oil in microwells of a PDMS chip placed on a cold plate sealed with an acrylic lid; freezing monitored via microscope camera | 34 | 6 nL | 720 | |
Pyroelectric thermal sensor for ice nucleation | Drops pipetted onto Vaseline-coated pyroelectric polymer atop a cooling block; freezing monitored via pyroelectric thermal sensor | 27 | 1 L | 30 | |
University of Toronto Drop Freezing Technique (UT-DFT) | Drops pipetted into multiwell PCR trays cooled in an ethylene glycol water bath; freezing monitored via camera | 23 | 50 L | 48 | |
Ice Nucleation SpEctrometer of the Karlsruhe Institute of Technology (INSEKT) | Drops pipetted into two 96-well PCR trays cooled in custom cooling blocks; freezing monitored via camera | 20 | 50 L | 192 | |
Lab-On-a-Chip Nucleation by Immersed Particle Instrument (LOC-NIPI) | Water-in-oil droplets mirofluidically generated in continuous flow and passed over a series of Peltier cold plates in an N-purged container; freezing monitored via microscope camera | 36 | 80–100 m diam. | 1000 | |
Freezing Ice Nuclei Counter (FINC) | Drops pipetted into three 96-well Piko PCR trays submersed in ethanol cooling bath; freezing monitored via camera | 25 | 5–60 L | 288 | this work |
2.1 Building components
2.1.1 Hardware
The hardware design of FINC is based on DRINCZ , its predecessors , and earlier descriptions of water droplets placed on oil-covered aluminum sheets over a cold plate, as described, for example, in and . FINC has a temperature-controlled ethanol cooling bath (LAUDA Proline RP 845, Lauda-Königshofen) (Fig. a) in which a commercially available chip-on-board LED array (50W COB Panel Light, Cooleeon Lighting Tech) is submerged 20 cm deep (Fig. b-1). A thin polytetrafluoroethylene sheet (Fig. b-2) acts as a diffuser mounted at a distance of 2 cm above the light source. A camera (IMX179 CMOS 8MP, ELP Free Driver) (Fig. b-3) mounted above the bath images clear Piko™ PCR trays made of polypropylene (SPL0960, Thermo Fisher Scientific) (Fig. b-4) resting on the frame at the bath's surface. A level sensor (LLE102000, Honeywell) (Fig. b-5) measures the height of the ethanol in the bath, and a peristaltic pump (KAS-S10-SE, Kamoer) (Fig. b-6) controlled by a micro-controller (Leonardo, Arduino®) (Fig. b-7) and powered by a stepper motor driver (TB6612, Adafruit) (Fig. b-7) moves ethanol between the LAUDA chiller bath and the ethanol reservoir (Fig. b-8) (Fig. c). A transparent 6 mm thick polymethyl methacrylate (Plexiglas) plate (Fig. b-9) covers the Piko PCR trays to avoid contamination of the wells, minimize evaporative loss of ethanol, and limit condensation of water vapor into the ethanol (see Sect. ). The components are mounted to a removable aluminum frame with stainless-steel rods, which is recommended to avoid corrosion over time (Fig. a, b). Consumer-grade hardware was used as building components when possible to reduce the overall instrument cost while ensuring measurement accuracy and reproducibility. The bulk of the building costs are constrained to the ethanol cooling bath. The per-measurement cost is dominated by laboratory consumables, such as the Piko PCR trays.
2.1.2 PCR trays
The use of Piko PCR trays in FINC is an updated feature of existing DFTs (Table 1). Specifically, standard PCR trays contain 96 wells of 200 L volume with dimensions of 127.76 mm by 85.34 mm, whereas the Piko PCR trays used in FINC contain 96 wells of 65 L volume with dimensions of 80.55 mm by 26.75 mm. The smaller dimensions allow for the use of up to four 96-well Piko PCR trays instead of one standard PCR tray, resulting in improved freezing temperature statistics per experiment. The trays are heated in an oven at 120 C for at least 1 h before use; this procedure improves reproducibility of background water experiments (Fig. S3). In FINC, we use three Piko PCR trays to optimize the ethanol circulation and temperature spread across the 288 sample wells (see Sect. ).
2.1.3 Bath leveler
An automated bath leveler system was built to account for the temperature dependence of the density of ethanol in FINC's cooling bath. For example, ethanol's density at 0 and at 25 C is 806.7 and 826.3 g L, respectively. This increase in density at lower temperatures consequently translates to a decrease in volume in the cooling bath and corresponds to a 4.9 mm decrease in height, equivalent to approximately 300 mL of ethanol, between 0 and 25 C. To achieve reproducible measurements, the ethanol level must submerge the well throughout the experiment to avoid the formation of vertical temperature gradients within the well . In FINC, a constant ethanol level is maintained by adding and removing ethanol to the cooling bath via a peristaltic pump (Fig. c), thereby automating this process. The binary level sensor outputs either a submerged or emerged status signal to the micro-controller, which then turns the pump in the corresponding direction: either moving ethanol from the reservoir to the cooling bath during the experiment or removing ethanol from the bath back to the reservoir during warm-up at the end of the experiment (Fig. c). The leveler does not need to be adjusted depending on the well volume used, as capillary action between the wells of the Piko PCR tray ensures that all wells are submerged in ethanol as long as the ethanol reaches the bottom of the wells. In all, freezing measurements in FINC occur without manual intervention between measurements except for removing and placing new Piko PCR trays.
2.2 Cooling rate
The cooling bath temperature is controlled by a MATLAB® script ramping down at 1 C min while the script records an image every 0.2 C. We prefer to record images as a function of temperature rather than as a function of time to ensure that all measurements have identical increments and can be easily averaged without interpolation. The cooling rate has been previously reported to have a negligible effect on immersion freezing temperatures , although much faster cooling can lead to temperature assignment uncertainties . Thus, we chose a cooling rate matching atmospheric updraft velocities and within the capacity of the LAUDA bath's cooling mechanism.
2.3 Freezing detection
The freezing of the solutions inside the wells of the Piko PCR trays is detected by a change in light intensity passing through the wells. Light passes through liquid water but is scattered by ice, resulting in dark pixels in the image (Fig. d). Once the images are recorded over the course of one experiment, a circular Hough transform algorithm, described in , is implemented to locate the wells on the images (Fig. S1). In the case of failure of the automatic well detection, we also developed a filtered algorithm output to identify the well positions, as well as a manual well alignment grid, by selecting two wells in opposing corners. After determining the well positions, the average pixel intensity is calculated for each well per image. This data analysis generates an intensity profile as a function of temperature. Then, the greatest change in intensity is attributed to the freezing temperature and can be visualized as a temporal map (Fig. ). For example, multiple changes in light intensity for one well over the course of an experiment or neighboring wells freezing at the same temperature can flag an error (for example, an object in front of the camera), requiring manual deletion of some images or a rerun of the measurement. A color map of the freezing temperature is also generated to visually inspect any well location bias (Fig. S2) . These data verification steps increase confidence in the measurement.
The data output of FINC is a vector containing the freezing temperature of each well. The vector is sorted by well column from top to bottom and left to right (see Fig. S1). In addition, a frozen fraction graph can be plotted by sorting the freezing temperature vector and plotting it versus a linearly spaced vector with values ascending from 0 to 1 in 288 steps (example in Fig. S3). Based on the recommendation by , the data are not trimmed, and all 288 freezing temperature data points are plotted. These data manipulations retain all information available from the experiment: freezing temperature and well location in one vector.
Figure 2
Map of the change in light intensity for each of the 288 wells ( axis) as a function of the image number ( axis), recorded every 0.2 C and corresponding to temperature decreasing and experimental time progressing. (The first 90 images corresponding to an interval of 18 C are excluded here for simplicity). The color scale represents the first derivative of the light intensity, with bright colors indicating a sudden drop in light intensity associated with freezing of the well. The algorithm assigns the freezing temperature of the well to the greatest change in light intensity and is marked by a white “x”. The dashed horizontal lines indicate the tray boundaries between the three Piko PCR trays. The freezing experiment depicts the results of a background water measurement.
[Figure omitted. See PDF]
2.4 Sample preparation and data analysisSample solutions were prepared within a laminar flow hood (Labculture Class II BioSafety Cabinet, ESCO) to prevent contamination from lab air. All glassware was pre-rinsed three times with deionized water and acetone, and it was subsequently dried in an oven at 120 C for at least 1 h. Sample solutions were then prepared using molecular-biology-free reagent water (W4502, Sigma Aldrich, hereafter referred to as background water). The solutions were then pipetted from sterile plastic reservoirs (10141-922, VWR) with an electronic multi-pipettor (4671040BT, Thermo Fisher Scientific, USA) into pre-cleaned Piko PCR trays (see Sect. ). We recommend pipetting the solution volume in one dispensing volume to minimize the formation of bubbles (see Sect. ). In addition, recently suggested working with clean-room vinyl or polyethylene gloves and aluminum-foil-coated surfaces.
When transporting the Piko PCR trays from the laminar flow hood to FINC, we placed cover films (Z369667, Sigma Aldrich) over the trays to avoid contamination from ambient air. We removed the cover film before placing the trays inside the ethanol bath in order to avoid problems with ethanol penetrating between the cover film and the Piko PCR tray by capillary effects. Furthermore, the Plexiglass hinged cover over the trays prevents contamination from the air depositing into the wells.
Note that our sample preparation procedure does not include dilution series; we made the solutions with the required concentration from weighed solid. We have previously proposed that lignin may be aggregating in solution, leading to concentration-dependant ice nucleation behavior of INMs . Based on this hypothesis, we also do not conduct any data merging procedure for different concentrations of INMs or drop volumes (for example in Figs. S6 and S7). Furthermore, we do not subtract background water values and prefer to show raw data in box plot formats (as in Fig. 7; see also ).
3 FINC's uncertainties
3.1 Temperature uncertainty
A temperature calibration is necessary to correct for the difference between the recorded temperature of the LAUDA bath and the temperature within each well. This calibration was done using a multichannel thermocouple data logger (HH-4208SD, Thermosense, UK), whereby each K-type thermocouple was placed inside a well filled with ethanol, with nine thermocouples evenly spaced across the three trays. With the probes inside the wells, the bath temperature was ramped down to 30 C at a rate of 1 C min. The temperature of each thermocouple was recorded every 10 s () and was plotted against the bath temperature recorded by the LAUDA system (), as shown for one calibration experiment in Fig. S4. We then obtained the following calibration equation (Eq. ), wherein the slope and intercept are averaged values across three independent calibration experiments.
1
Multiple temperature calibrations conducted several months apart and by multiple users led to identical slopes, confirming reproducible temperature gradients and constant ethanol circulation inside the bath. The mean of the standard deviations of the nine evenly spaced thermocouples across three Piko PCR trays was 0.5 C for temperatures down to 25 C (Table S1). We therefore report the temperature uncertainty for each well to be 0.5 C.
3.2 Temperature spread across wellsFurthermore, we tested different numbers of Piko PCR trays and different bath pump speeds to reduce the temperature bias across the three trays within an experiment (Sect. S3, Table S1). Tests with four trays led to the identification of a vortex in the upper left corner of the ethanol bath, yielding higher temperature biases across the trays (Table S1). Overall, we determined that using three trays placed to the right to avoid the vortex, as well as pump speed setting 8, resulted in the smallest temperature biases of 0.46 C between 10 and 15 C and of 0.55 C between 20 and 25 C (Table S1, last row). Since the average of these values is precisely 0.5 C, we consider this value to be equal to the temperature uncertainty described in Sect. .
3.3 FINC's limit of detection
It is necessary to accurately characterize the background of the instrument to determine the lowest trustworthy freezing temperature of a sample. FINC's limit of detection (LOD) for the freezing control experiments (Sect. ) was calculated as the mean of 10 replicates of background water experiments using 5 L droplets (Fig. S5). We calculated the mean temperature and 1 standard deviation (a spread of 1) for each of the 288 values, resulting in an LOD of 25.4 0.2 C. We note that a value of 3 can also be used and would lead to a similar background of 25.4 0.4 C (Fig. S5). Furthermore, we show the LOD as a box plot of the mean frozen fraction (Fig. S5). The LOD depends on the experiment type, but as long as the appropriate background characterizations are measured, e.g., artificial salt water, background water, and water through a laboratory setup, the instrument can be used to measure freezing temperatures of 288 wells at a time (see Sect. for a further discussion). Finally, we add that no background corrections are made in our data analysis.
4 Freezing control experiments
To test the capabilities of FINC and to characterize its sources of uncertainties, we conducted several freezing control experiments. We considered (1) the nonhomogeneous freezing of the background water (Sects. and ), (2) the roles of tray material and of droplet shape (Sect. ), (3) eight different sources of contamination (Sect. ), (4) the choice of well volume (Sect. ), and (5) the freezing-point depression of a dissolved organic matter solution with different salt concentrations (Sect. ).
4.1 Nonhomogeneous freezing in FINC
The use of Piko PCR trays allows for a range of sample volumes of 5–60 L to be measured in FINC. Theoretically, freezing rates of water droplets are dependent on the volume of the droplet; smaller droplets freeze at lower temperatures . Classical nucleation theory approximates interfacial tension between ice and water, the activation energy of the phase transfer, and the size of clusters and embryos . In the atmosphere, 50 % of a droplet population of 5 L volume is predicted to freeze spontaneously ( 1 s) and thus homogeneously at 31.81 C, whereas 60 L volume is predicted to freeze at 31.41 C (equations from ). However, these temperatures were never reached during FINC experiments with background water. Over the range of possible sample volumes in FINC (5–60 L), the mean value of background water was 24.5 0.8 C. DFTs tabulated in Table 1, which use drops in the microliter range, also show this nonhomogeneous freezing behavior. As argued in the following sections, the difference in temperatures between atmospheric homogeneous freezing and background water freezing in FINC is likely due to a combination of tray material (Sect. ), nonspherical drop shapes within the wells (Sect. ), and different sources of contamination (Sect. ).
4.2 Volume dependence on nonhomogeneous freezing
We further attempted to quantify these uncertainties by comparing the values of background water over a range of background water volumes. The mean background water values in FINC were 25.4 0.1 C, 24.7 0.1 C, 23.4 0.6 C, 24.5 C, 24.4 0.2 C, 25.2 C, and 25.3 0.1 C for 5, 10, 20, 30 40, 50, and 60 L, respectively, across one to five replicates (Table S2; Fig. S6). Furthermore, if we collect all the freezing temperatures from different volumes described in Sect. and convert the data into INPs per volume, we observe a large spread in the freezing behavior of the water (Fig. S7). Through both these data analyses, we observe that experiments with 20 L of background water freezes warmer and with a larger spread than all other volumes. In Sects. and , we investigate and discuss the potential sources of contamination involved in creating a volume-dependent nonhomogeneous freezing temperature, and we attempt to reason the unpredictable behavior of the 20 L freezing experiments.
4.3 Potential uncontrollable factors affecting nonhomogeneous freezing
4.3.1 Effect of tray material
As described in detail in and , the material interacting with the supercooled water droplets impacts their freezing. Indeed, the Piko PCR trays used in FINC are made of polypropylene and are therefore a hydrophobic surface. Interestingly, hydrophobic surfaces have previously been observed to freeze at a warmer temperature than hydrophilic surfaces . It is likely that the material of the tray is contributing to warmer temperatures than expected for homogeneous freezing, but it is difficult to quantify the extent or percentage of this contribution to the overall nonhomogeneous freezing behavior. Imperfections on the surface of each well could also induce nonhomogeneous freezing behavior , and the use of a larger number of wells could help provide reliable statistics.
4.3.2 Effect of drop shape
Due to the narrow width of the Piko PCR tray wells, the water in FINC's wells is subjected to capillary forces. Indeed, a concave meniscus is evident when examining the solution in the Piko PCR trays, thereby exerting negative pressure on the solution. It has been previously shown that the homogeneous nucleation rate of water can be significantly increased when water is subjected to negative pressure . However, the negative pressure associated with the radius of the meniscus in the Piko wells is on the order of 1–3 (see Table S3 and Fig. S10) and is likely negligible in terms of participating in nonhomogeneous freezing in FINC . Nevertheless, classical nucleation theory assumes spherical supercooled water droplets, whereas DFTs using a cold stage or PCR trays rather investigate the freezing behavior of half-spheres or cone-shaped supercooled droplets.
4.4 Potential controllable sources of contamination
4.4.1 Effect of background water contamination
Background water is a contentious issue in the field of atmospheric ice nucleation. An excellent overview of the challenges of “cleaning up our water” is described in . Upon their recommendation, we experimented with different types of purified water and found that the molecular-biology-free reagent water (i.e., background water) gave the most reproducible measurements, consistent with . Filtration of the background water through a 0.02 m filter led to no difference in freezing behavior for the background water, except for our lab's Milli-Q water, which is known to contain higher levels of organic carbon (Fig. S8). We therefore chose to work with commercial molecular-biology-free reagent water from Sigma Aldrich without further purification.
4.4.2 Effect of condensing water vapor
We considered the possibility of water vapor condensing into the wells from the air between the Plexiglass and the trays during a measurement (Fig. S9). We calculated a maximum amount of condensable water vapor or, in other words, a worst-case scenario in which if the volume of water vapor corresponding to 90 % RH at 0 C between the Plexiglass and the trays were to condense into liquid water inside a well, we would arrive at a value of 2 nL (Sect. S7). This volume is small and unlikely to be able to condense into one well, freeze on a colder wall, and trigger nucleation within the well. We therefore conclude that this condensation process has a negligible effect on the well volume and thus on freezing temperatures.
4.4.3 Effect of the surface area of the tray
We considered whether a difference in the surface area to volume ratio could explain the different values observed for different volumes depicted in Fig. S6 (see Sect. S8 for the calculations of surface areas of the wells). In particular, we hypothesized that the warmer freezing of the 20 L could be due to the cone-like shape of the well (see the shape diagram in Fig. S10). However, the surface area to volume ratios of the different volumes shown in Fig. S11 could not explain the variability or higher freezing temperatures observed, particularly with 20 L. Note that the water vapor condensation is inconsequential to changing the surface area to volume ratio (Sect. ). Our running hypothesis for the peculiar higher freezing of the 20 L experiments is the presence of microscopic bubbles generated at the intersection of the two truncated cones when using this volume (Sect. ).
4.4.4 Effect of air bubbles in the wells
Air bubbles in the solutions within a well can be generated during the pipetting process, particularly if the solution mixture has surfactant-like properties (see the example image in Fig. S12). Bubbles are defined here as visible pockets of air in the well. This problem is likely the result of a combination of trapped air in the narrow wells during dispensing. Wells containing bubbles in FINC (Fig. S12) froze at warmer temperatures (Fig. S13). When the bubbles collapsed within the well, they may have created a spike in negative pressure, inducing freezing at warmer temperatures than in the absence of bubbles . Therefore, bubbles should be avoided by careful introduction of the solution into the well. Since bubbles associated with pipetting can be seen in the images, we can confirm the absence of bubbles in all images used for the data plotted in Fig. S6. As no visible bubbles were present, then this potential effect is unlikely to explain the observed nonhomogeneous freezing behavior for different well volumes. Nevertheless, microscopic bubbles invisible to the human eye and to the camera could be affecting the nonhomogeneous freezing, including for the 20 L volumes, but to an unknown degree.
4.4.5 Effect of contamination from the tray
A series of experiments was conducted to test the leaching of potential ice-active material from the Piko PCR trays into the background water. To test this hypothesis, we placed 60 L of background water into one tray and then pipetted 50 L of the water out of each well and into another tray. If leaching was indeed a problem, we would have observed higher freezing temperatures in the tray containing the transferred solution. However, we found no effect of freezing from leaching over duplicate sets of experiments (Fig. S14). We conclude that material leaching from the Piko PCR trays is not a problem.
4.4.6 Effect of lab air contamination
As discussed by and , the concept of an open droplet system can be prone to further contamination from the surrounding air. We have built upon the authors' recommendations by using a laminar flow hood, placing a cover film on the top of the Piko PCR tray between the flow hood and FINC, and having a Plexiglass cover on FINC to minimize deposition of airborne contaminants. Furthermore, we calculated the surface area of the solution exposed to air but observed no trend related to the nonhomogeneous freezing behavior for different well volumes (see Table S3).
Figure 3
The freezing-point depression effect of solutions of 20 mg C of Jericho Ditch dissolved organic matter with increasing concentrations of sodium chloride (NaCl). The axis represents the concentration (mol L) of NaCl, and the axis represents the values, where 50 % of the wells were frozen. Triplicate experiments were done for concentrations of 0, 1, 2, and 3 M of NaCl, marked with red crosses. Note that the values for the experiments at 0 mol L are identical and thus overlap. Experimental data were compared to Blagden's law (grey dashed line), which describes the freezing-point depression phenomenon. All temperature measurements have an uncertainty of 0.5 C.
[Figure omitted. See PDF]
4.5 Volume of solution per well discussion and recommendationBased on Sects. , , , and , it remains difficult to quantify the contribution of each possible source of contamination to nonhomogeneous freezing. This difficulty appears to be a combination of uncontrollable (Sect. ) and controllable factors (Sect. ), and this difficulty has been previously addressed in . Nevertheless, the pre-treatment of the trays (see Fig. S3), the use of a laminar flow hood, and the use of molecular-biology-free reagent water allowed for reproducible background water measurements. In all, the recommended well volume while working with FINC depends on the research question. One should use a larger volume to study active, but less abundant, INPs and INMs, whereas one should use a smaller volume to study less active, but more abundant, INPs and INMs. If the samples have high concentrations of salts leading to freezing-point depression, for instance, then a larger volume might be necessary to remain in the operating range of the LAUDA bath.
Figure 4
(a) The active surface-site density of NX-illite suspensions measured on FINC (yellow, orange, red dots), DRINCZ (grey downward triangles), and LINDA (grey dots) during the intercomparison. The wet suspension parameterization by is plotted for reference. Concentrations used were 0.1, 0.05, and 0.01 g L of NX-illite. The DRINCZ illite measurements as reported in are also shown for comparison (grey upwards triangles). was calculated using a BET surface area of 124.4 m g . Error bars in the and direction are shown on the value for one sample each from DRINCZ, LINDA, and FINC (see Supplement Sect. S11 for details). Note that the -direction uncertainties do not extend beyond the marker. (b) INP concentrations (L) of an ambient aerosol sample solution from Zurich, Switzerland, measured on FINC (red dots), DRINCZ (grey triangles), and LINDA (grey dots). Ambient aerosols were collected with a Coriolis impinger for 20 min at 300 L min. The sample was split for measurements on each of the immersion freezing instruments. One measurement run was done with FINC; two runs each were done with DRINCZ and LINDA. The grey shaded area denotes the range of INP concentrations from precipitation and cloud water samples compiled by . Error bars in the and direction are shown on the value for one sample each from DRINCZ, LINDA, and FINC (see Supplement Sect. S11 for details). Note that the -direction uncertainties do not extend beyond the marker.
[Figure omitted. See PDF]
4.6 Freezing-point depressionInorganic salts have a freezing-point depression effect on water's freezing temperature
2 where is the freezing-point depression, is the cryoscopic constant of the solvent ( 1.86 K kg ), and is the molality of the solute . To further characterize FINC, we tested the ice nucleation activity of solutions of dissolved organic matter (DOM) and sodium chloride (NaCl; 31 434, Sigma Aldrich). The DOM solutions were at concentrations of 20 mg C (milligrams of carbon per liter) and were samples obtained from Jericho Ditch, part of the Great Dismal Swamp in Suffolk, Virginia, USA
To validate FINC against other similar and peer-reviewed DFTs, we conducted an intercomparison study with ETH's DRINCZ and the University of Basel's LINDA (Sect. ). Here, we report the comparison measurements using NX-illite (Sect. ) and an ambient aerosol sample (Sect. ).
5.1 Experimental details of DRINCZ and of LINDA
DRINCZ was operated using a freshly shaken sample poured into a sterile reservoir. Then, 96 droplets of 50 L were transferred into a 96-well PCR polypropylene tray (732-2386, VWR, USA) using an eight-channel multi-pipette. The tray was sealed with a transparent foil and immediately analyzed with DRINCZ as described in . LINDA was operated using a total of 5.2 mL per experiment, and each sample was shaken by hand and immediately pipetted under a laminar flow hood (AURA Mini, EuroClone) into 52 microtubes (0.5 mL Eppendorf Safe-Lock Tubes) using a repeater pipette (Stepper TM 411, Socorex) and bioproof syringes (Ecostep TM, sterilized, single wrapped, bioproof; range 50–500 L). Each droplet contained 100 L of the sample solution and was measured with LINDA according to .
All samples for the intercomparison were prepared at ETH Zurich on 10 July 2019. On the following day, measurements with all three instruments were conducted. As LINDA was located in Basel, one batch of aliquots was transported in a cooler by train in the morning from the preparation location (in Zurich) to the measurement location (in Basel). For background measurements, purchased, bottled molecular-biology-free reagent water (Sigma-Aldrich, W4502-1L) was used (background measurements reported in Fig. S15), and no background corrections were made.
5.2 NX-illite intercomparison
NX-illite has been repeatedly used as a standard to compare ice nucleation instruments
3 where is the frozen fraction at each freezing temperature, SA is the BET-determined surface area of the NX-illite particles (124.4 m g; ), is the mass concentration of NX-illite, and is the volume in each well ( L for FINC, 50 L for DRINCZ, and 100 L for LINDA). The uncertainties in include a 1 % error in SA , 1 % error in , 1 % error in , 0.5 % error in , 8 % error in (well volume errors are based on error in pipettes), and an error of in the number of frozen wells. Uncertainty calculations are detailed in Sect. S11 and plotted in Fig. a as vertical error bars on the values for one sample from each instrument.
We additionally compare these measurements with NX-illite solutions measured on DRINCZ in 2018 and with the parameterization from (Fig. a).
The FINC measurement spread at 15 C is from 68–209 and grows to 2600–9200 at 20 C. The parameterization fits well within the FINC measurements above 17 C and deviates up to a factor of 4.8 at 21 C. Between FINC, DRINCZ, and LINDA, the spread was a factor of 6 at 15 C and a factor of 7.5 at 20 C. The DRINCZ measurements from are up to 1 order of magnitude higher than those reported here (Fig. a). Indeed, the spread of is a common outcome of NX-illite suspensions
Next, we measured the INP concentration of an ambient aerosol sample during the intercomparison study with DRINCZ and LINDA. Ambient aerosols were collected with a Coriolis air sampler (Bertin Technologies, France), an instrument designed for outdoor monitoring of bio-aerosols such as pollen and spores . The sample was collected for min at a flow rate of L min on 10 July 2019 on the terrace of the Institute for Atmospheric and Climate Science at ETH Zurich in Switzerland. The initial sample volume mL was further diluted with a dilution factor of DF 3.7 to obtain enough volume to split into sterile Falcon tubes and brought to FINC, DRINCZ, and LINDA for measurement the following day. The INP concentration per liter of air was calculated by Eq. ().
4 The uncertainties associated with the INP concentrations are plotted in Fig. b as vertical error bars on the values for one sample from each instrument, and the calculations are described in Section S11. The uncertainties include 1 % error in , 0.5 % error in , and 8 % error in (well volume errors are based on error in pipettes), as well as an error of in the number of frozen wells, an error in the flow rate of the Coriolis impinger of L min, and an error of mL in the final Coriolis sample volume.
The INP concentration in the aerosol sample as a function of temperature agreed well between the instruments, with the advantage of FINC being able to achieve lower temperature measurements than DRINCZ and LINDA due to the smaller well volume (Fig. b, and see Sect. ). The measurements also fit within the range of INP concentrations per liter of air from cloud water and precipitation samples compiled by . Furthermore, since LINDA was part of the intercomparison study of ice nucleation instruments reported by , which used cellulose as a standard, FINC and DRINCZ are by extension expected to have representative freezing temperature ranges. These results further validate the capabilities of FINC to be used for measuring ice nuclei concentrations.
6 Lignin as an ice nucleation standardWe describe here the use of commercial and water-soluble lignin as a standard for intercomparing droplet freezing techniques (DFTs). Its applicability is demonstrated through the intercomparison measurements of lignin with three different DFTs (Sect. ), the comparison of different batches of a commercial lignin product (Sect. ), the stability of lignin solutions over time (Sect. ), and the recalcitrance of lignin towards chemical and atmospheric processing (Sect. ).
Figure 5
Example of the chemical structure of a lignin monomer known as a lignol. The commercial lignin material used in this work has a thiol group in the benzylic position instead of an alcohol group (highlighted in bold).
[Figure omitted. See PDF]
6.1 Chemical composition of ligninLignin is a high-molecular-weight natural polymer accounting for approximately 30 % of the organic carbon present in the biosphere . It is composed of three different types of hydroxycinnamyl alcohol monomers referred to as monolignols, specifically para-coumaryl alcohol, coniferyl alcohol, and sinapyl alcohol (Fig. ). The relative quantity of monolignols within the lignin polymer varies depending on taxa, cell type, and cell layers, as well as on the development stage of the tree, the climate, and the habitat , leading to a large number of possible permutations within the natural lignin polymer. The polymerization of monolignols occurs through a stepwise chemically controlled process linking ether bonds (-O-4, -O-4) and carbon–carbon bonds . Lignin indeed constitutes a mixture of monolignols but specifically represents repeating units of known chemical functional groups such as oxygen-substituted arenes, conjugated double bonds, and alcohols. Lignin's van Krevelen diagrams, determined using high-resolution mass spectrometry, typically have ratios between 0.6 and 1.6 and ratios between 0.1 and 0.7 . These ratios are distinct from other classes of organic matter, namely carbohydrates, lipids, peptides, and tannins . Interestingly, lignin has recently been modeled to aggregate with itself . Furthermore, the commercial lignin used in this work is a by-product of the pulp and paper industry and therefore has a reproducible chemical composition, arguably important for a standard, and a molecular weight average of 10 000 g mol (Fig. 5). Section confirms this reproducibility specifically for ice nucleation. We further acknowledge that this commercial lignin contains a thiol group, which differs from atmospherically relevant monolignols yet is used here as a soluble and cheap standard for the intercomparison of DFTs (Fig. 5).
Figure 6
Ice-active mass site density () of lignin measured by FINC, DRINCZ, and LINDA. Lignin was measured during the intercomparison on FINC (30 L well volume; three replicates; dark red dots), DRINCZ (50 L well volume; four replicates; orange triangles), and LINDA (100 L well volume; one replicate; red triangles). Additionally, different product batches of lignin were measured by FINC (5 L well volume; six experiments, blue dots). Note that the markers of batches 1a and 1b are partially hidden behind the orange intercomparison markers. All lignin solutions were of concentration 20 mg C . The black line is the empirical fit of the intercomparison data only (). The parameterization of the sea surface microlayer organic matter from is also shown (dashed lined). Error bars in the and direction are shown on the value for one sample each from DRINCZ, LINDA, and FINC batch experiments, as well as FINC intercomparison experiments (see Supplement Sect. S11 for details). Note that the -direction uncertainties do not extend beyond the marker.
[Figure omitted. See PDF]
6.2 Instrument intercomparison with lignin and IN parameterizationThe measurements on FINC, DRINCZ, and LINDA were conducted on the same day (11 July 2019) with identical lignin solutions. Specifically, one solution of 20 mg C of lignin (of batch 1; see Sect. ) diluted in background water was divided into sterile Falcon tubes and brought to each instrument.
We calculated the ice-active mass site density, the values, for each measurement according to Eq. () and analogous to Eq. (), also following :
5 where TOC is the total organic carbon of the lignin solution, and is the volume inside each well ( is 30 L for FINC during the intercomparison study, 5 L for FINC during batch experiments, 50 L for DRINCZ, and 100 L for LINDA). TOC values were determined using the mass of lignin weighed and the vendor's description of carbon content (50.13 %), and these were further validated using a Shimadzu TOC analyzer with the prior pre-treatment addition of nitric acid and sulfuric acid. Uncertainties in include a 1 % error in TOC, 1 % error in , 0.5 % error in , 8 % error in (well volume errors are based on error in pipettes), and finally an error of in the number of frozen wells. The uncertainty calculations are presented in Sect. S11 and displayed in Fig. as error bars on the of one sample from each instrument.
The freezing temperatures of 20 mg C L compared well between FINC, DRINCZ, and LINDA, with overlapping traces falling within a factor of 3 (Fig. ). Specifically, at 14.8 C, lignin's values from the intercomparison range from 77 to 223 , and at 19.5 C they range from 1760 to 4560 . This spread is an improvement relative to NX-illite, which has values that span an order of magnitude (Fig. 4a). Moreover, the data from the intercomparison were fitted to obtain a parameterization for lignin's ice-nucleating ability in immersion freezing, shown as the solid black line in Fig. . In addition, and values were 17.5 0.9 and 12.4 1.3 C, respectively. Unlike , who observed reproducible initial freezing temperatures with Snomax, we observe lignin's freezing temperatures to be reproducible starting at values (with 5 % of wells frozen). Note that the FINC batch measurements, also shown in Fig. , were not included in the empirical fit to avoid over-weighting the FINC data. The fit is represented in Eq. (), where is temperature in degrees Celsius (C). 6 Note that this parameterization is specifically valid for lignin solutions of 20 mg C . In fact, lignin is a macromolecule suggested to adopt different solution aggregation properties depending on its concentration , and thus our parameterization is exclusively for concentrations of 20 mg C , equivalent to 40 mg of lignin per liter. For reference to other INMs, the parameterization by of sea surface microlayer organic matter is included (Fig. ). While the slope is similar for both parameterizations (Fig. ), the Wilson parameterization is interestingly 2 orders of magnitude higher. Lignin is not the most active INM in the sea surface microlayer, in sea spray, or in bioaerosols , so we were therefore surprised to notice the similarity in slope. However, organic matter data from several other studies are inconsistent with this slope, highlighting the difficulties in predicting the ice-nucleating ability of organic matter .
6.3 Lignin batch comparisonA desirable quality of a suitable ice nucleation standard is its ability to be reproducible across commercial product batches. The lignin used here is alkali low-sulfonate kraft lignin (CAS 8068-05-1, Sigma Aldrich, product code 471003). We tested the ice-nucleating ability of three different batches of this product: batch numbers 04414PE, MKCL2371, and MKCK3344 (hereafter referred to as batches 1, 2, and 3, respectively). Specifications on their production dates and their carbon and sulfur contents are provided in Table S4. The values, determined according to Eq. (), of batches 2 and 3 overlapped with each other (Fig. ). The values of batch 1 overlapped with the other batches below 19 C but above 19 C had fewer ice-active mass site densities (Fig. ). Still, all batches were within 1 order of magnitude of each other and of the empirical fit from the intercomparison data (Fig. ). These results indicate that lignin's ice-nucleating activity is reproducible across different production batches.
6.4 Lignin solution stability
In addition to reproducibility across product batches, another quality of a standard is its stability in solution over time towards ice nucleation. Therefore, we performed multiple FINC experiments with the same 20 mg C solution of lignin (using batch 2) over 4 months and found no change in ice-nucleating activity (Fig. ), unlike for NX-illite (see Sect. and Fig. ). We additionally performed several freezing experiments on a more concentrated solution of 200 mg C of lignin (using batch 2) and also found no significant change in ice-nucleating activity over 8 weeks (Fig. S16). Furthermore, we tested the effect of the storage temperature of the lignin solution in the fridge and on the lab bench and found no difference (Fig. S17). This solution stability makes lignin a complementary ice nucleation standard over NX-illite and Snomax.
Figure 7
The solution stability of a 20 mg C lignin solution (batch 2) over 125 d, stored in the fridge in the dark. Each box plot represents the frozen temperatures of 288 wells; the middle line is the median (), the outer edges of the box are the 25th and 75th percentiles, the whiskers extend to the 10th and 90th percentiles, and the red dot is the mean. The water box plot represents the background water solution. Duplicates were taken on day 1 (overlapping box plots). All temperature measurements have an uncertainty of 0.5 C.
[Figure omitted. See PDF]
6.5 Lignin's macromolecular size and reactivityLignin's recalcitrance towards atmospheric processing has recently been demonstrated by . Indeed, photochemical exposure, ozonation, heating, sonication, and hydrogen peroxide treatments had negligible effects on lignin's ice nucleation ability. Also note that the lignin recommended herein is soluble; accordingly, it passed through a 0.2 m filter without loss of ice activity but lost some activity after filtering to 0.02 m . Lignin could therefore be termed a nano-INP or an INM and operationally defined as soluble . The ability of lignin to be chemically recalcitrant further demonstrates its value as a standard for DFTs.
7 Conclusions
We describe herein our home-built drop Freezing Ice Nuclei Counter (FINC) to measure the ice-nucleating ability of INPs and INMs in the immersion freezing mode. FINC is a complementary instrument and part of a growing list of droplet freezing techniques (Table 1). Its capabilities include (1) simultaneously measuring three trays of 96 wells each, (2) supercooling to 0.2 with 5 L droplets with a temperature uncertainty of 0.5 C, and (3) automated controls for running an experiment. We further explored eight different possible explanations for the nonhomogeneous freezing behavior observed in FINC and other DFTs. As part of the development and validation of FINC, we intercompared measurements with NX-illite suspensions and an ambient aerosol sample collected in Zurich with other DFTs, specifically with DRINCZ and LINDA. Additionally, we present evidence for the use of soluble lignin as a reproducible, easy to use, and commercial intercomparison standard for future ice nucleation studies in immersion freezing. Indeed, we demonstrated that solutions of 20 mg C measured on FINC, DRINCZ, and LINDA yielded a spread of values within a factor of 3 between 8 and 25 C. We subsequently fitted a parameterization () through our empirical data for further use in intercomparison and validation studies. Finally, we showed that lignin solutions are stable over several months, and their ice-nucleating activity is well reproduced across different batches of the same product, making lignin a competitive solution standard.
Code and data availability
Data sets for all data presented in figures in the text are deposited in the ETH Research Collection data repository at
The supplement related to this article is available online at:
Author contributions
KPB and NBD designed the instrument. KPB wrote the code to operate the instrument. AJM validated the instrument. AJM, KPB, CM, and JW acquired data during the intercomparisons; NBD, AJM, and KPB wrote the paper with contributions from all authors.
Competing interests
The authors declare that they have no conflict of interest.
Acknowledgements
We acknowledge the technical help of Marco Vecellio and Michael Rösch for advice, ordering hardware, and machining structural instrument components. We also acknowledge the assistance of Max Aragón Cerecedes during the intercomparison experiment day. We thank Sophie Bogler for help with the tray washing experiments and Silvan Müller for help with the TOC analyzer. We are grateful to Paul DeMott and Tom Hill for reviewing our compiled list of droplet freezing techniques.
Financial support
This research was supported by the Swiss National Science Foundation (grant no. PZ00P2_179703). Furthermore, Claudia Mignani and Jörg Wieder acknowledge funding from the Swiss National Science Foundation (grant nos. 200021_169620 and 200021_17582, respectively), and Robert O. David acknowledges funding from the European Research Council (grant StG 758005).
Review statement
This paper was edited by Mingjin Tang and reviewed by three anonymous referees.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2021. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Aerosol–cloud interactions, including the ice nucleation of supercooled liquid water droplets caused by ice-nucleating particles (INPs) and macromolecules (INMs), are a source of uncertainty in predicting future climate. Because INPs and INMs have spatial and temporal heterogeneity in source, number, and composition, predicting their concentration and distribution is a challenge requiring apt analytical instrumentation. Here, we present the development of our drop Freezing Ice Nuclei Counter (FINC) for the estimation of INP and INM concentrations in the immersion freezing mode. FINC's design builds upon previous droplet freezing techniques (DFTs) and uses an ethanol bath to cool sample aliquots while detecting freezing using a camera. Specifically, FINC uses 288 sample wells of 5–60
As part of the validation of FINC, an intercomparison campaign was conducted using an NX-illite suspension and an ambient aerosol sample from two other drop freezing instruments: ETH's DRoplet Ice Nuclei Counter Zurich (DRINCZ) and the University of Basel's LED-based Ice Nucleation Detection Apparatus (LINDA). We also tabulated an exhaustive list of peer-reviewed DFTs, to which we added our characterized and validated FINC.
In addition, we propose herein the use of a water-soluble biopolymer, lignin, as a suitable ice-nucleating standard. An ideal INM standard should be inexpensive, accessible, reproducible, unaffected by sample preparation, and consistent across techniques. First, we compared lignin's freezing temperature across different drop freezing instruments, including on DRINCZ and LINDA, and then determined an empirical fit parameter for future drop freezing validations. Subsequently, we showed that commercial lignin has consistent ice-nucleating activity across product batches and demonstrated that the ice-nucleating ability of aqueous lignin solutions is stable over time. With these findings, we present lignin as a good immersion freezing standard for future DFT intercomparisons in the research field of atmospheric ice nucleation.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details




1 Institute for Biogeochemistry and Pollutant Dynamics, ETH Zurich, Zurich, 8092 Switzerland
2 Institute for Atmospheric and Climate Science, ETH Zurich, Zurich, 8092 Switzerland
3 Department of Environmental Sciences, University of Basel, Basel, 4056 Switzerland
4 Department of Geosciences, University of Oslo, Oslo, 0315 Norway
5 Institute for Biogeochemistry and Pollutant Dynamics, ETH Zurich, Zurich, 8092 Switzerland; Institute for Atmospheric and Climate Science, ETH Zurich, Zurich, 8092 Switzerland; now at: Department of Chemistry, University of British Columbia, Vancouver, V6T 1Z1, Canada