© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

In radiography, the exposure index (EI), as per the International Electrotechnical Commission standard, depends on the incident beam quality and exposure dose to the digital radiography system. Today automatic exposure control (AEC) systems are commonly employed to obtain the optimal image quality. An AEC system can maintain a constant incident exposure dose on the image receptor regardless of the patient thickness. In this study, we investigated the relationship between body thickness, entrance surface dose (ESD), EI, and the exposure indicator (S value) with the aim of using EI as the dose optimization tool in digital chest radiography (posterior–anterior and lateral projection). The exposure condition from the Korean national survey for determining diagnostic reference levels and two digital radiography systems (photostimulable phosphor plate and indirect flat panel detector) were used. As a result, ESD increased as the phantom became thicker with constant exposure indicator, which indicates similar settings to an AEC system, but the EI indicated comparatively constant values without following the tendency of ESD. Therefore, body thickness should be considered under the AEC system for introducing EI as the dose optimization tool in digital chest radiography.

Details

Title
Proper Management of the Clinical Exposure Index Based on Body Thickness Using Dose Optimization Tools in Digital Chest Radiography: A Phantom Study
Author
Yoon, Yongsu 1   VIAFID ORCID Logo  ; Park, Hyemin 2   VIAFID ORCID Logo  ; Kim, Jungmin 2 ; Kim, Jungsu 3   VIAFID ORCID Logo  ; Roh, Younghoon 2 ; Tanaka, Nobukazu 4 ; Morishita, Junji 4 

 Department of Radiological Science, Dongseo University, 47 Jurye-ro, Sasang-gu, Busan 47011, Korea; [email protected] 
 Department of Health and Safety Convergence Sciences, Korea University, 145 Anam-ro, Seongbuk-gu, Seoul 02841, Korea; [email protected] 
 Department of Radiological Technology, Daegu Health College, 15 Youngsong-Ro, Buk-gu, Daegu 41453, Korea; [email protected] 
 Department of Health Sciences, Faculty of Health Sciences, Kyushu University, 774 Motooka, Nishi-ku, Fukuoka 819-0395, Japan; [email protected] (N.T.); [email protected] (J.M.) 
First page
5203
Publication year
2021
Publication date
2021
Publisher
MDPI AG
ISSN
1661-7827
e-ISSN
1660-4601
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2532502408
Copyright
© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.