PURPOSE
We aimed to evaluate and compare the shapeability and stability of five microcatheters commonly used in interventional radiology after steam shaping and manual shaping.
METHODS
Steam shaping was performed using three mandrels of different angles: L(S) shape (90°), U(S) shape (180°), and O(S) shape (360°). Three manual shapes—L(M), U(M), and O(M)—were made to have a similar angle to their steam-shaped counterparts. The stability of the microcatheters was evaluated by passing them through a 5 F catheter and inserting microguidewires. The tip angles of the microcatheters and the angle change rates were compared between groups.
RESULTS
The mean angle of the microcatheters after steam shaping was 42.4°–54.1° for L(S) shape, 80.2°–96.7° for U(S) shape, and 130.7°–150.8° for O(S) shape. Five microcatheters showed significantly different mean angle reductions after passing through the 5 F catheter (17.4%–30.3%) and inserting microguidewires (24.1%–61.2%). Different microguidewires also caused significantly different mean angle reductions (34.6%–50.8%). The reduced angle caused by the guidewire was almost completely recovered after withdrawing it (93.2%–101.6%). Although manual-shaped microcatheters showed a 4.2%–6.3% greater angle reduction than steam-shaped microcatheters after passing through the 5 F catheter, the final tip angle was not significantly different between the two groups and was within 10%.
CONCLUSION
The tip angle of the microcatheters after steam shaping using mandrels may differ depending on the shape of the mandrel and the type of microcatheter used, and the stability varies depending on the type of microcatheter. The manual shaping of microcatheters can be a good alternative to steam shaping.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer