This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
1. Introduction
Molecular absorption spectrometry incorporating chemometrics is often used to determine components in a mixture with overlapping spectra simultaneously. The general principle of this combined method is that the full-spectral data are loaded into software programmed with algorithms and statistical methods, from which each component in the mixture is quantified. Due to its simplicity, the method is considered environmentally friendly and economical. Besides, full-spectral data for calculations combined with algorithms help to increase the accuracy of the determination. Standard chemometrics methods include classical least squares (CLS), partial least squares (PLS), Kalman filter, principal component regression (PCR), artificial neural networks (ANNs), and derivative spectrophotometry. These methods have been studied and applied to various matrices [1–7].
In our previous study [4], we succeeded in employing the Kalman filter algorithm with full-spectral measurement to build a program (written on the Microsoft Excel 2016 and Visual Basic for Applications (VBA) platform) for determining the concentration of substances in a mixture. Inheriting these results, we continue modifying and applying this combined method (from now on, Kalman-Excel method) to determine Hydrochlorothiazide (HCT) and Losartan Potassium (LSP) in Splozarsin Plus simultaneously. Hydrochlorothiazide (C7H8ClN3O4S2, M = 297.7 g/mol) is a diuretic medicine (or water pills) used to treat hydronephrosis, renal failure, or nephropathy. In addition, HCT is also used effectively to prevent and treat hypertension, liver disorders, myocardial infarction, heart failure, or stroke in patients with cardiovascular problems. Losartan Potassium (C22H22ClKN6O, M = 461.0 g/mol) belongs to a class of medicine called angiotensin receptor blockers that dilate blood vessels, facilitating blood to flow more easily. Losartan Potassium is used to treat high blood pressure and heart failure and help to protect the kidney damage caused by diabetes [8–10].
For the determination of HCT and LSP in medicine, various techniques have been developed elsewhere. Maggio et al. [8] established a procedure based on the multivariate analysis of spectral data in the 220–274 nm region, with the partial least square algorithm. The results agreed well with those determined with HPLC. Rathee et al. [9] developed Vierordt’s and Q absorbance ratio methods. They were successfully applied to the assay of LSP and HCT in tablet formulations. Rao et al. [10] built three techniques for this purpose: (1) to solve the equations simultaneously; (2) to determine the absorbance ratio at 272 nm (the maximum absorption of HCT) and 266.5 nm (the isosbestic wavelength); (3) to use the first-order derivative method. Thube et al. [11] also proposed three spectrophotometric methods for the simultaneous determination of LSP and HCT in a binary mixture: (1) to measure the absorbance at 235 nm and 271 nm, corresponding to the absorbance maxima of LSP and HCT, respectively (the concentration of each drug was obtained by using the absorption values calculated for both drugs at these wavelengths); (2) to use the dual-wavelength principle in which the absorbance difference between two points on the mixture spectra was directly proportional to the concentration of the component of interest and independent of the interfering components; and (3) to describe and calculate the area under the curve between 265 and 282 nm for LSP and between 229 and 242 nm for HCT. The proposed methods were applied to marketed formulation. High-performance liquid chromatography was also applied to quantify these two compounds in pharmaceutical products [12–15]. Although HPLC delivers simultaneous determination, high cost prevents its application.
As the absorption spectra of HCT and LSP are overlapped, the Kalman filter algorithm using full spectra (Kalman-Excel method) [4] is employed in this study to quantify these two substances in pharmaceutical products simultaneously. The proposed method opens up the possibility of a fast and cheap technique for pharmaceutical analysis and testing.
2. Materials and Methods
2.1. Instrument and Chemicals
2.1.1. Instruments
A UV-Vis spectrophotometer Cary 60 (Agilent, USA) for spectral scanning (190–990 nm) is connected to a computer with Cary WinUV software for spectral data storage (in Excel spreadsheets). All the measurements were carried out at the Laboratory of Applied Chemistry, Department of Chemistry, University of Sciences, Vietnam
2.1.2. Chemicals
Hydrochlorothiazide standard (C7H8ClN3O4S2), 99.8% purity, was purchased from the National Institute of Drug Quality Control, standard Vietnamese pharmacopoeia (Code: C0219308.02). Losartan Potassium standard (C22H22ClKN6O), 99.9% purity, was obtained from the Institute of Drug Quality Control, Ho Chi Minh City, standard Vietnamese pharmacopoeia (Lot No. 166090520). All other chemicals used were of analytical grade.
The stock solution of HCT (50.0 μg/mL) was prepared by dissolving 12.5 mg of HCT standard in 150 mL of distilled water and ultrasonicated for 15 min. The solution was then made up to 250 mL with distilled water. Twenty-five millilitres of the stock solution was diluted with 75 mL of distilled water to make a working solution of HCT (12.5 μg/mL). The stock solution of LSP (50.0 μg/mL) (also acted as a working solution) was prepared by dissolving 12.5 mg LSP standard in distilled water and making up to 250 mL.
2.2. Sample Preparation
In this study, Splozarsin Plus tablets (Co. Ltd. SHINPOONG DAEWOO, Bien Hoa industrial zone, Dong Nai, Vietnam, batch code 9003, manufactured date 5 June 2019, expired date 5 June 2021) were used for the method validation. The reference amount of LSP and HCT on the label is 50.0 mg and 12.5 mg, respectively.
Twenty tablets with an average weight of 253.4 mg were ground to a fine powder and mixed well. A particular amount of the powder (m) was dissolved in 150 mL distilled water and ultrasonicated for 30 min before filtering to make up a 250 mL solution. Ten millilitres of the solution was diluted to 100 mL with distilled water for measurements.
The amount of each target compound in a tablet is expressed via mg/tablet and calculated according to the following equation:
2.3. Analytical Procedure
To determine HCT and LSP simultaneously, we utilized the Kalman filter algorithm with full-spectral measurements. The Visual Basic for Applications, a programming language written on the basis of Microsoft Excel 2016, was used for data treatment [4].
Briefly, the standard solutions of individual compounds and their actual sample solutions were subjected to record molecular absorption spectra (scanned from 220 to 300 nm wavelength at 1.0 nm intervals). The achieved data were saved as .csv files and extracted to the computer as Excel sheets. The developed Kalman-Excel program was applied to determine the concentrations and respective errors.
2.4. Method Validation
2.4.1. Relative Error (RE%)
The relative error (RE%) between concentrations is calculated according to the following equation [16]:
2.4.2. Limit of Detection (LOD) and Limit of Quantification (LOQ)
Measurements are made on a minimum of seven aliquots (n ≥ 7) of a prepared sample solution that has a concentration near the expected limit of detection [16]. The developed Kalman-Excel program was applied to determine the concentrations of HCT and LSP in the seven aliquots when using their spectrum data and spectrum data of the standard solutions of HCT and LSP. The LOD and LOQ were then calculated by the following formula:
where SD is the standard deviation of seven replicates.
2.4.3. Repeatability (RSD%)
The repeatability of the analytical method is assessed via relative standard deviation [16–18]:
The RSD value is acceptable if lower than 1/2RSDHorwitz [19], in which RSDHorwitz is calculated according to the following equation:
2.4.4. Trueness
The trueness of the analytical method was verified via the method’s recovery and the mean values of the developed method and a standard method.
(i) Recovery. The mean recovery is determined by assessing spiked samples at three different levels (three replicates per level) of each target compound, namely, 0.500, 1.000, and 1.500 μg/mL for HCT and 2.000, 4.000, and 6.000 μg/mL for LSP. The recovery rate is calculated by the following equation:
In this study, Splozarsin Plus tablets were used as the original sample. Three-fifths of the average tablet weight (3/5 × M) were taken for each sample. The standard solutions of HCT (5.000 μg/mL) and LSP (20.000 μg/mL) were also prepared at the same time for spectral measurements. The developed Kalman-Excel program was applied to quantify the concentration of HCT and LSP in the samples simultaneously. The acceptable recovery rate is from 85% to 110% [18].
(ii) Comparing Two Sample Means Performed By Kalman-Excel Method and Standard HPLC Method. According to Harvey [20], the trueness of an analytical method could be evaluated by analyzing n1 and n2 replicates with the developed method and the standard method. The mean values achieved from the two analytical methods are compared with the t-test. The Splozarsin Plus tablets were sent to the Drug, Cosmetic and Food Quality control Center of Thua Thien Hue Province for analysis with the HPLC method. Hydrochlorothiazide and Losartan Potassium were quantified following the Vietnam Pharmacopoeia V guideline [20]. The Splozarsin Plus solutions were prepared (described in Section 2.2) and analyzed with the developed Kalman-Excel method (described in Section 2.3) with three replicates. The t-test was used for evaluation: when the variances for the two analyses are not significantly different if the experimental t value is lower than the t value at risk of 0.05 and (n1 + n2−2) degrees of freedom, the mean values achieved from the two methods are not significantly different.
3. Results and Discussion
3.1. Relative Errors and Repeatability
Seven mixtures of HCT and LSP standards with different concentration ratios were prepared from the standard solutions of HCT (12.500 μg/mL) and LSP (50.000 μg/mL). The mixtures and the standard solutions were subjected to analysis with the Kalman-Excel method, as described in Section 2.3. The absorption spectra are shown in Figure 1.
[figure omitted; refer to PDF]
At different HCT and LSP concentration ratios, the RE values between reference concentrations and their measured concentrations vary from −1.55 to 1.93% (Table 1). In addition, the RSD values are far lower than the respective 1/2RSDH values. This finding indicates the good performance of the Kalman-Excel method for the simultaneous determination of HCT and LSP in standard mixtures.
Table 1
Relative errors and repeatability of the Kalman-Excel method.
| Sample name | Conc. ratios (HCT/LSP) | n | HCT | LSP | ||||
| CHCT (μg/mL) | RE (%) | RSD (1/2RSDH) (%) | CLSP (μg/mL) | RE (%) | RSD (1/2RSDH) (%) | |||
| M1 | 1 : 24 | 1 | 1.002 | 0.20 | 0.30 | 24.206 | 0.86 | 0.07 |
| 2 | 0.996 | −0.40 | (8.00) | 24.238 | 0.99 | (4.96) | ||
| 3 | 0.999 | −0.10 | 24.229 | 0.95 | ||||
| Average | 0.990 | −0.10 | 24.224 | 0.93 | ||||
| M2 | 2 : 20 | 1 | 1.981 | −0.95 | 0.175 | 20.115 | 0.57 | 0.05 |
| 2 | 1.981 | −0.95 | (7.21) | 20.112 | 0.56 | (5.10) | ||
| 3 | 1.975 | −1.25 | 20.131 | 0.66 | ||||
| Average | 1.979 | −1.05 | 20.119 | 0.6 | ||||
| M3 | 3 : 16 | 1 | 2.982 | −0.60 | 0.166 | 15.961 | −0.24 | 0.06 |
| 2 | 2.973 | −0.90 | (6.78) | 15.971 | −0.18 | (5.27) | ||
| 3 | 2.974 | −0.87 | 15.979 | −0.13 | ||||
| Average | 2.976 | −0.79 | 15.970 | −0.18 | ||||
| M4 | 4 : 12 | 1 | 3.975 | −0.62 | 0.14 | 12.067 | 0.56 | 0.02 |
| 2 | 3.966 | −0.85 | (6.49) | 12.071 | 0.59 | (5.50) | ||
| 3 | 3.965 | −0.87 | 12.072 | 0.60 | ||||
| Average | 3.969 | −0.78 | 12.070 | 0.58 | ||||
| M5 | 5 : 08 | 1 | 5.008 | 0.16 | 0.20 | 8.019 | 0.24 | 0.10 |
| 2 | 4.991 | −0.18 | (6.28) | 8.035 | 0.44 | (5.85) | ||
| 3 | 4.990 | −0.20 | 8.031 | 0.39 | ||||
| Average | 4.996 | −0.07 | 8.028 | 0.36 | ||||
| M6 | 6 : 04 | 1 | 5.921 | −1.32 | 0.21 | 4.064 | 1.60 | 0.17 |
| 2 | 5.897 | −1.72 | (6.11) | 4.074 | 1.85 | (6.49) | ||
| 3 | 5.904 | −1.60 | 4.077 | 1.93 | ||||
| Average | 5.907 | −1.55 | 4.072 | 1.79 | ||||
| M7 | 5 : 20 | 1 | 4.990 | −0.20 | 0.15 | 20.203 | 1.02 | 0.06 |
| 2 | 4.975 | −0.50 | (6.28) | 20.224 | 1.12 | (5.10) | ||
| 3 | 4.980 | −0.40 | 20.220 | 1.10 | ||||
| Average | 4.982 | −0.37 | 20.216 | 1.08 | ||||
RSDH: RSD value calculated from the Horwitz function; M1: mixture of HCT 1.000 μg/mL + LSP 24.000 μg/mL; M2: mixture of HCT 2.000 μg/mL + LSP 20.000 μg/mL; M3: mixture of HCT 3.000 μg/mL + LSP 16.000 μg/mL; M4: mixture of HCT 4.000 μg/mL + LSP 12.000 μg/mL; M5: mixture of HCT 5.000 μg/mL + LSP 8.000 μg/mL; M6: mixture of HCT 6.000 μg/mL + LSP 4.000 μg/mL; M7: mixture of HCT 5.000 μg/mL + LSP 20.000 μg/mL.
3.2. Method Validation on Pharmaceutical Tablets
3.2.1. Quantification of HCT and LSP in Splozarsin Plus Tablet and Method Repeatability
A Splozarsin Plus powder weighting 253.4 mg (after grinding and mixing) was prepared as described in Section 2.2 and analyzed as described in Section 2.3. The amount of HCT and LSP in the Splozarsin Plus tablet was determined following equation (1). The absorption spectra of the sample and standard solutions are illustrated in Figure 2.
[figure omitted; refer to PDF]
The results in Table 2 show that the Kalman-Excel method to determine HCT and LSP in a pharmaceutical product simultaneously yields good repeatability (RSDs vary from 0.27 to 0.3% for two studied compounds). The average amounts of HCT and LSP quantified in Splozarsin Plus tablet are 12.42 ± 0.09 mg and 50.02 ± 0.34 mg, respectively. These values agree well with the reference content on the label (12.5 mg and 50 mg, respectively). They also comply with the quality standards of the Vietnamese Ministry of Health, which allows the content of the ingredients in this type of tablets as follows: HCT: 12.50 mg ± 7.5% (11.56–13.44 mg), LSP: 50.00 mg ± 5% (47.5–52.5 mg) [21].
Table 2
Measured amounts of HCT and LSP in Splozarsin Plus tablet and the respective RSD values.
| Sample name | HCT | LSP | ||
| CHCT (μg/mL) | Content (mg/tablet) | CLSP (μg/mL) | Content (mg/tablet) | |
| SP1 | 4.978 | 12.45 | 20.033 | 50.08 |
| SP2 | 4.975 | 12.44 | 20.042 | 50.11 |
| SP3 | 4.953 | 12.38 | 19.942 | 49.86 |
| Average | 4.969 | 12.42 | 20.01 | 50.02 |
| RSD% | 0.29 | 0.30 | 0.27 | 0.27 |
3.2.2. Trueness, LOD, and LOQ of Method
(1) Recovery of the Method. The absorption spectra of standard solutions and spiked samples are demonstrated in Figure 3. The results of recovery are shown in Table 3.
[figure omitted; refer to PDF]Table 3
Recovery of HCT and LSP in Splozarsin Plus.
| Sample | HCT | LSP | ||||||
| Cspiked (μg/mL) | Cmeasured (μg/mL) | Rev (%) | RSD (%) | Cspiked (μg/mL) | Cmeasured (μg/mL) | Rev (%) | RSD (%) | |
| S01 | 0 | 3.003 | — | 0.167 | 0 | 11.997 | — | 0.029 |
| S02 | 3.009 | — | 0 | 11.991 | — | |||
| S03 | 3.013 | — | 0 | 11.997 | — | |||
| S11 | 0.500 | 3.512 | 101.80 | 0.066 | 2.000 | 13.962 | 98.25 | 0.014 |
| S12 | 3.508 | 99.80 | 2.000 | 13.964 | 98.65 | |||
| S13 | 3.508 | 99.00 | 2.000 | 13.966 | 99.45 | |||
| S21 | 1.000 | 3.996 | 99.30 | 0.100 | 4.000 | 15.887 | 97.25 | 0.017 |
| S22 | 4.004 | 99.50 | 4.000 | 15.882 | 97.28 | |||
| S23 | 4.000 | 98.70 | 4.000 | 15.886 | 97.73 | |||
| S31 | 1.500 | 4.515 | 100.80 | 0.056 | 6.000 | 17.973 | 99.60 | 0.010 |
| S32 | 4.520 | 100.73 | 6.000 | 17.973 | 99.73 | |||
| S33 | 4.517 | 100.27 | 6.000 | 17.970 | 99.88 | |||
The recovery of the Kalman-Excel method is approximately 100% for the studied compounds with good repeatability (RSD% < 0.167, n = 3). This result is comparable with that of the previous studies [8–14], which employed different spectrophotometry methods for simultaneous determination of HCT and LSP in tablets (Table 4).
Table 4
The validation results of other methods in comparison with this Kalman-Excel method for the determination of HCT and LSP in tablets.
| Sources | Method | Solvent | Range of concentration (μg/mL) | LOD (μg/mL) | LOQ (μg/mL) | Rev (%) | RSD (%) |
| [8] | PLS, spectrophotometry | H2O | 1.06–5.70 | 0.06 | 0.14 | 95.6 | 0.6 |
| [9] | 1/Vierordt’s method | 0.01N HCl | 0.563–27.27 | 0.233 | 0.186 | 98.13–99.69 | <0.246 |
| [10] | 1/simultaneous equation | Methanol | 1/1–20 | 0.5 | 1.97 | ≈99 | 0.57 |
| [11] | 1/simultaneous equation | 5–30 | 0.30 | 0.90 | 98.7–100.3 | 0.49–0.57 | |
| [12] | Microemulsion | Mobile | 2.5–12.5 | 0.03 | 0.1 | 98.9 | <1.4 |
| [13] | RP-HPLC | Acetonitrile and formic acid | 2–10 | 0.5 for HCT and LSP | 1.5 for HCT and LSP | 96 | <0.61 |
| [14] | RP-HPLC | Acetonitrile and formic acid | 1.0–300.3 | 0.304 | 0.921 | 99.55-100.67 | 0.24–0.61 |
| This study | Spectrophotometry with Kalman filter algorithm | H2O | 1.0–6.0 | 0.037 | 0.12 | 98.7–101.8 | <0.167 for HCT and LSP |
In addition to that, the LOD and LOQ values achieved in this method were 0.037 and 0.12 μg/mL for HCT and 0.228 and 0.76 μg/mL for LSP (Table 4). Accordingly, it was an advance over the results from the studies of Rao et al. [10] who used three different spectrophotometry methods including simultaneous equation, ratio of absorbance, and first-order derivative method to quantify HCT and LSP in tablets, or the studies of Mohammed et al. [13] and Srinivasu et al. [14] who applied RP-HPLC (Table 4).
Besides the good performance of the developed method, another highlight was its simplicity in sample preparation, in which H2O was the only solvent used for sample dilution and preparation, while most of the other methods consumed various organic solvents (Table 4).
(2) Comparing Two Sample Means Performed by Kalman-Excel Method and HPLC Method. To continue evaluating the method trueness, we compare the concentration of HCT and LSP in the Splozarsin Plus tablets quantified with the Kalman-Excel method with that of the HPLC method. The t-test is used for data comparison.
The data in Table 5 demonstrate that texp is lower than the t (0.05; 4), implying that the analytical results of the two methods are statistically consistent with the significance level of 0.05. In other words, the Kalman-Excel method exhibits good trueness. In addition, the Fexp values are lower than the F (0.05, 2, 2) when α = 0.05, revealing that the developed analytical method and the standard method (HPLC) produce the same exact repeatability.
Table 5
Performance of Kalman-Excel method versus HPLC method to quantify HCT and LSP in Splozarsin Plus tablet simultaneously (n1 = n2 = 3).
| Amount of HCT in a tablet (mg/tablet) | Amount of LSP in a tablet (mg/tablet) | ||
| Kalman-Excel | HPLC | Kalman-Excel | HPLC |
| 12.4450 | 12.45 | 50.0825 | 49.93 |
| 12.4375 | 12.38 | 50.105 | 50.30 |
| 12.3825 | 12.42 | 49.855 | 50.45 |
| Mean = 12.422 ± 0.034 | Mean = 12.417 ± 0.035 | Mean = 50.014 ± 0.138 | Mean = 50.227 ± 0.268 |
| Fexp = 1.059 | Fexp = 3.745 | ||
| F (0.05; 2; 2) = 39.00 | F (0.05; 2; 2) = 39.00 | ||
| SDpool = 0.0346 | SDpool = 0.2130 | ||
| texp = 0.177 | texp = 1.222 | ||
| t (0.05, 4) = 2.776 | t (0.05, 4) = 2.776 | ||
Fexp: experimental variance of the method; F (0.05, 2, 2): F distribution at alpha 0.05 and the respective degrees of freedom of numerator and denominator; Spool: pooled variance; texp: experimental t value, t (0.05, 4): t value at risk of 0.05 and 4 degrees of freedom.
4. Conclusions
We successfully apply full-spectral measurements with the modified Kalman filter algorithm method for the simultaneous determination of Hydrochlorothiazide and Losartan Potassium in pharmaceutical formulation tablets. The method minimizes the time needed for sample preparation as well as the amounts of solvent used and the waste generated. The analytical procedure exhibits good repeatability, good trueness, and low relative errors for the studied compounds. There is no statistically significant difference in analytical results between the Kalman-Excel method and the standard HPLC procedure used as reference. The simplicity and good sample throughput of this method make it suitable for routine quality control testing of the pharmaceutical products containing simultaneously Hydrochlorothiazide and Losartan Potassium.
Acknowledgments
This paper was funded by Vietnam Ministry of Education and Training with code no. B2020-DHH-16.
[1] H. R. Pouretedal, B. Shafiee, M. H. Keshavarz, "Simultaneous determination of trace amounts of thorium and zirconium using spectrophotometric partial least-squares calibration method," Bulgarian Chemical Communications, vol. 41 no. 3, pp. 230-235, 2009.
[2] S. Grangeiro, M. M. Albuquerque, D. Santana, M. H. Pimentel, R. H. da Silva, S. Simone, "Simultaneous spectrophotometric determination of lamivudine and zidovudine in fixed dose combinations using multivariate calibration," Quimica Nova, vol. 34 no. 5, pp. 859-863, DOI: 10.1590/s0100-40422011000500022, 2011.
[3] R. Bhaskar, R. Bhaskar, M. K. Sagar, V. Saini, K. M. Bhat, "UV-Spectrophotometric-assisted chemometric methods for the simultaneous determination of metformin hydrochloride and gliclazide in pharmaceutical formulations," Pharmaceutica Analytica Acta, vol. 3 no. 4,DOI: 10.4172/2153-2435.1000158, 2012.
[4] N. T. Quynh Trang, N. Van Hop, N. Dang Giang Chau, T. Binh Tran, "Simultaneous determination of amlodipine, hydrochlorothiazide, and valsartan in pharmaceutical products by a combination of full spectrum measurement and kalman filter algorithm," Advances in Materials Science and Engineering, vol. 2019,DOI: 10.1155/2019/5719651, 2019.
[5] H. W. Darwish, A. H. Bakheit, A. S. Abdelhameed, R. A. Mothana, "Application of classical least squares, principal component regression and partial least squares methods for simultaneous spectrophotometric determination of rutin and ascorbic acid in their combined dosage form," Life Science Journal, vol. 10 no. 4, 2013.
[6] V. R. Galande, K. G. Baheti, S. Indraksha, M. H. Dehghan, "Estimation of amlodipine besylat, valsartan and hydrochlorothiazide in bulk mixture and tablet by UV spectrophotometry," Indian Journal of Pharmaceutical Sciences, vol. 74 no. 1, pp. 18-23, DOI: 10.4103/0250-474X.102538, 2012.
[7] M. H. Abdel-Hay, S. M. Galal, M. A. A. Ragab, "Application of derivative and derivative-ratio spectrophotometry in the determination of a binary mixture of naproxen and diflunisal and a ternary mixture of orphenadrine, caffeine and aspirin," Taiwan Pharmaceutical Journal, vol. 59 no. 4, pp. 157-170, 2007.
[8] R. M. Maggio, P. M. Castellano, T. S. Kaufman, "A multivariate approach for the simultaneous determination of losartan potassium and hydrochlorothiazide in a combined pharmaceutical tablet formulation," Analytical and Bioanalytical Chemistry, vol. 391 no. 8, pp. 2949-2955, DOI: 10.1007/s00216-008-2180-z, 2008.
[9] P. Rathee, S. Rathee, D. Rathee, H. Chaudhary, "Stability indicating UV-spectrophotometric methods for simultaneous determination of losartan potassium and hydrochlorothiazide in pharmaceuticals," Eurasian Journal of Analytical Chemistry, vol. 4 no. 1, pp. 98-109, 2009.
[10] K. Rao, N. Keshar, M. Panda, "Spectrophotometric methods for the simultaneous estimation of losartan potassium and hydrochlorothiazide in tablet dosage forms," Chronicles of Young Scientists, vol. 2 no. 3,DOI: 10.4103/2229-5186.90893, 2011.
[11] C. Thube, J. Dhagude, P. Y. Pawar, "Simultaneous estimation and validation of losartan potassium and hydrochlorothiazide in bulk and tablet dosage form by using different spectrophotometric method," Der Pharma Chemica, vol. 6 no. 2, pp. 24-30, 2014.
[12] L. Li, C. Lai, X. Xuan, C. Gao, N. Li, "Simultaneous determination of hydrochlorothiazide and losartan potassium in osmotic pump tablets by microemulsion liquid chromatography," Journal of Chromatographic Science, vol. 54 no. 8, pp. 1415-1420, DOI: 10.1093/chromsci/bmw101, 2016.
[13] M. Nidhal, S. Mohammed, H. R. Abdo, H. M. Hassan, "Method development and validation of simultaneous determination of hydrochlorothiazide and losartan in tablet dosage form by RP-HPLC," International Journal of Pharmaceutical Sciences and Research, vol. 10 no. 1, pp. 227-231, DOI: 10.13040/IJPSR.0975-8232.10(1).1000-05, 2019.
[14] T. Srinivasu, M. Mathrusri Annapurna, "A new stability indicating RP-HPLC method for the analysis of losartan and hydrochlorothiazide tablets," Journal of Chemical and Pharmaceutical Sciences, vol. 10 no. 2, pp. 798-802, 2017.
[15] N. Erk, "Analysis of binary mixtures of losartan potassium and hydrochlorothiazide by using high performance liquid chromatography, ratio derivative spectrophotometric and compensation technique," Journal of Pharmaceutical and Biomedical Analysis, vol. 24, pp. 603-611, DOI: 10.1016/s0731-7085(00)00434-9, 2001.
[16] P. M. Berthouex, L. C. Brown, Statistics for Environmental Engineers, pp. 123-127, 2002.
[17] I. Taverniers, M. De Loose, E. Van Bockstaele, "Trends in quality in the analytical laboratory. II. analytical method validation and quality assurance," TrAC Trends in Analytical Chemistry, vol. 23 no. 8, pp. 535-552, DOI: 10.1016/j.trac.2004.04.001, 2004.
[18] AOAC International, AOAC ® Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals, 2012.
[19] M. Thompson, P. J. Lowthian, "A horwitz-like function describes precision in a proficiency test," The Analyst, vol. 120 no. 2, pp. 271-272, DOI: 10.1039/an9952000271, 1995.
[20] D. Harvey, Morden Analytical Chemistry, pp. 80-93, 2000.
[21] Ministry of Health, Vietnamese Pharmacopoeia V, vol. 483, pp. 588-589, 2017.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2021 Tran Thuc Binh et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. https://creativecommons.org/licenses/by/4.0/
Abstract
In this paper, hydrochlorothiazide (HCT) and losartan potassium (LSP) in tablets are simultaneously determined with UV-Vis spectrophotometry and chemometrics without separation. The spectra of standard and sample solutions were recorded in the wavelength from 220 to 300 nm at 1.0 nm intervals. The concentrations of HCT and LSP in the sample solutions were computed with the Kalman filter algorithm written on the Microsoft Excel 2016 and Visual Basic for Applications (VBA) platform. The method validation was determined via the accuracy and repeatability of measurements when analyzing HCT and LSP in the Splozarsin Plus tablet and comparing the mean values of their contents in the sample with those analyzed with HPLC. The proposed method is simple with a low cost compared with the standard HPLC method.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details
; Le Thi Phuong Tram 1 ; Nguyen Van Hop 1 ; Nguyen Dang Giang Chau 1 ; Luu, Nguyen Duy 2 ; Nguyen Thi Quynh Trang 3 1 University of Sciences, Hue University, Hue 530000, Vietnam
2 Da Nang University of Medical Technology and Pharmacy, Da Nang 550000, Vietnam
3 Faculty of Environmental Sciences, Saigon University, Ho Chi Minh City 700000, Vietnam