About the Authors:
Jordy K. van Zandwijk
Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing
* E-mail: [email protected]
Affiliations Magnetic Detection & Imaging, TechMed Centre, University of Twente, Enschede, Netherlands, Department of Vascular Surgery, Medisch Spectrum Twente, Enschede, Netherlands
ORCID logo https://orcid.org/0000-0002-6393-0139
Frank F. J. Simonis
Roles Conceptualization, Investigation, Methodology, Resources, Supervision, Validation, Writing – review & editing
Affiliation: Magnetic Detection & Imaging, TechMed Centre, University of Twente, Enschede, Netherlands
ORCID logo https://orcid.org/0000-0002-0734-1778
Friso G. Heslinga
Roles Conceptualization, Data curation, Writing – review & editing
Affiliation: Magnetic Detection & Imaging, TechMed Centre, University of Twente, Enschede, Netherlands
Elfi I. S. Hofmeijer
Roles Data curation, Writing – review & editing
Affiliation: Magnetic Detection & Imaging, TechMed Centre, University of Twente, Enschede, Netherlands
ORCID logo https://orcid.org/0000-0002-4492-311X
Robert H. Geelkerken
Roles Supervision, Writing – review & editing
Affiliations Department of Vascular Surgery, Medisch Spectrum Twente, Enschede, Netherlands, Multimodality Medical Imaging Group, TechMed Centre, University of Twente, Enschede, Netherlands
Bennie ten Haken
Roles Conceptualization, Methodology, Supervision, Writing – review & editing
Affiliation: Magnetic Detection & Imaging, TechMed Centre, University of Twente, Enschede, Netherlands
Introduction
Contrast agents (CAs) have been used in MRI for decades with a great use for angiographic purposes [1]. Cardiovascular diseases as peripheral arterial disease, aortic aneurysms, and cardiomyopathy benefit from enhanced imaging possibilities due to CA administration with excellent signal-to-noise (SNR) ratios. Gadolinium based contrast agents (GBCAs) are used for generating positive contrast in millions of MRI examinations because of their unique magnetic properties [2]. An important characteristic of CAs in MRI is that their effect depends on the used field strength [3]. At lower magnetic field strengths the molecular tumbling rate is reduced which increases effect of a T1 CA [4].
Low-field MRI (0.25-1T) has advantages of lower costs, smaller footprint, and better subject accessibility with respect to higher field strengths [5]. MR on these lower field strengths could also benefit from developments in higher-field MRI over the past decades. The benefits of a high-performance low-field MR system (0.55T) in MRI-guided catheterizations, high susceptibility regions, and efficient image acquisition strategies have been demonstrated [6]. Because of this renewed interest, it has become relevant which CA is best suited for low-field MRI.
The use of CAs in low-field MRI up to now has mainly focused on examinations with gadolinium, which generates positive contrast [3]. This was mainly done in intraoperative MR systems for neurosurgery, which often employ low field strengths [7, 8]. Further applications that may require CA administration at lower field strengths can be found in endovascular interventions [9, 10], enhanced vascular imaging [11], or even in gravity dependent investigations [12]. The lowered field will probably not affect the GBCA behavior [13]. However, not much work has been published on optimizing contrast agents for low-field MRI applications. Besides, when the characteristics of a CA that influence relaxation times stay equal at lower field strengths, the contrast-to-noise ratio will drop because of an inherent lower T1 on lower field strengths. This raises the question whether at low field strengths CAs other than GBCAs might be more suitable, such as ultra-small super paramagnetic iron-oxides (USPIOs).
USPIOs possess different physiological and relaxation characteristics compared to GBCAs. When shortening the T1, the signal becomes larger (positive contrast) because of faster longitudinal relaxation. Likewise, shortening the T2* results in less signal (negative contrast) since the net magnetization decreases faster with shorter T2*. Where GBCAs are mostly used for generating positive contrast on T1-weighted imaging, USPIOs are frequently used in T2* weighted imaging as negative CA because the magnetic susceptibility of their iron core greatly shortens T2* [14]. However, when administered at low concentrations, the T1 shortening effects of USPIOs dominate the T2* shortening effects, leading to a positive contrast. Next to that, the relaxivity of USPIOs (i.e. the amount of change in relaxation rate per concentration) increases greatly and possibly triples at field strengths lower than 1T [15], meaning their effect is larger at lower concentrations. At these lower field strengths their r2/r1 ratio is also more favorable to achieve positive contrast [15, 16]. This makes USPIOs an interesting option for a low-field MR system. Besides their improved signal enhancement properties, USPIOs are often given a dextran coating resulting in surface properties that ensure vasculature retention times with half-times up to 21 hours [17] versus 1.5 hours for GBCAs [18]. Next to that, GBCAs are associated with nephrogenic systemic fibrosis and gadolinium accumulation in organs that could be harmful for patients in longer procedures [19, 20]. Ferumoxytol, an USPIO previously used for magnetic resonance angiography has therefore been considered a useful alternative to GBCAs [21–23].
This work combines simulations and experimental testing of an USPIO and a GBCA on low magnetic field strength (0.25T). There have been investigations [13, 24–26] in relaxivities of both contrast agent types on field strengths like 1.5T and 3.0T. An additional signal intensity (SI) gain because of higher relaxivities for the clinically interesting USPIO ferumoxytol [17] on a field strength lower than 1.5T is to be expected. At 0.55T only slightly higher relaxivity was found for ferumoxytol [6], but for ultra-low field MRI (<10mT) USPIOs show more pronounced advantages in terms of enhanced signal and shorter scan times [16, 27]. Forthcoming, at 0.25T the added value of USPIO enhanced imaging should be investigated.
The aim of this study was to investigate whether a gadolinium-based or an iron-oxide based CA is more suited to provide signal enhancement in low-field MRI. For this comparison, SIs were simulated for three different MR sequences that are generally employed in an angiographic or vascular interventional setting. Initial scan parameters were chosen to provide optimal positive contrast in T1-weighted images and to avoid saturation. Subsequently, the signal enhancement generated by both CAs was measured in phantom samples on a low-field MRI-scanner and compared to the simulations.
Materials and methods
The GBCA gadoterate acid (Gd-DOTA, Dotarem®, Guerbet, France) and USPIO ferumoxytol (Feraheme, AMAG Pharmaceuticals, USA) were used as contrast agents in this research. A 0.25T MRI scanner (G-scan Brio, Esaote, Italy) equipped with a coil used for wrist examinations was used in all experiments. For MR angiography, values of 2.67 mM for gadoterate (0.2 mmol/kg) [28] and 0.96 mM of iron (4 mg/kg) [29] are conventional on field strengths equal or higher than 1.5T. Therefore, a phantom both for ferumoxytol and gadoterate was built containing concentrations of 0.15, 0.30, 1.2, 7.0, and 14 mM in bovine blood in a circular setup of 15 ml vials with a fish oil marker to indicate the orientation of the samples. A sample of bovine blood without CA served as reference. First, the relaxivities of both CA’s on 0.25T were experimentally estimated. Secondly, their SIs were simulated for common MR sequences. Lastly, phantom experiments were performed on the 0.25T MRI scanner to validate the simulated signal intensities.
Relaxivity
Since the r1 and r2 values of both ferumoxytol and gadoterate were unknown on 0.25T, this was measured on a 0.25T MRI scanner using the concentrations range of CAs and the NOVIFAST method [30] for r1 and DESPOT2 [31] for r2. Scanning parameters for these methods are given in Table 1. First, with NOVIFAST we used a spoiled gradient echo sequence (SGE) with six varying flip angles between 10° and 90° to obtain T1 maps of the samples. Subsequently, the r1 values were calculated using Eq 1.
[1]
[Figure omitted. See PDF.]
Table 1. Parameters of the NOVIFAST and DESPOT2 methods that were applied to obtain consecutively T1 and T2 maps.
https://doi.org/10.1371/journal.pone.0256252.t001
For calculation of the r2 values with the DESPOT2 method the T1 values resulting from NOVIFAST were required as input. We obtained balanced steady-state free precession (bSSFP) scans with varying flip angles ranging from 20° to 90°. The resulting T2 maps were used for determination of the r2 values using Eq 1.
The r1 and r2 values of ferumoxytol on 1.5T are known from literature (r1 = 15 mM-1 s-1 and r2 = 89 mM-1 s-1) [32]. Relaxivity for gadoterate on 1.5T has been described by Rohrer [13] as r1 = 2.9 mM-1 s-1 and r2 = 3.2 mM-1 s-1.
Sequences & simulations
SIs were simulated for three relevant sequences to study the SI as a function of the CA concentration. Simulations were performed at low field (0.25T) and at common field strength (1.5T) for comparison.
First, a regular spin echo (SE) that is often used for anatomical reference was simulated. Eq 2 states the SI in a SE in relation to the proton density (PD), repetition time (TR), echo time (TE), and T1 and T2 values of the tissue [33]. To obtain maximum T1-weighted contrast for optimal contrast agent visibility, TR and TE were set to the lowest possible values of the MRI scanner (Table 2). Furthermore, a flip angle of 90° was chosen.
[2]
[Figure omitted. See PDF.]
Table 2. Parameters of the MR sequences that were used to analyze the samples with different concentrations of ferumoxytol and gadoterate.
Acq. res. = acquired resolution.
https://doi.org/10.1371/journal.pone.0256252.t002
Second, an SGE sequence that can be used for angiographic purposes was simulated. Its signal is defined by Eq 3 [34].
[3]
The measured signal in this sequence is also dependent on the flip angle (θ). To avoid saturation of the sample with large flip angle, we used a flip angle of 40° in the simulations and experiments.
Third, a bSSFP was simulated because of its favorable SNR characteristics. Less contrast difference due to CAs is expected for this sequence since its contrast is known to be T2/T1-weighted. Eq 4 gives the SI expression for bSSFP [35].
[4]
Signal of all sequences was normalized with respect to the situation where no CA was added (C = 0) to accentuate the effect of the CA. The main outcome is the ratio of increase in SI with respect to SI(0). Table 2 shows the used scan parameters for all sequences. All simulations were validated by scanning both ferumoxytol and gadoterate phantoms.
Analysis
The data were analyzed with Matlab (Mathworks Inc., Natick, USA). Signal intensities were measured from the average of automatically selected regions of interest with a radius of 6 pixels around the center of each sample to avoid signal affected by Gibbs ringing. Subsequently, the values were normalized with respect to the intensity of the reference sample. The coefficient of variation of the signal intensity in the samples was obtained by dividing the SD by the mean SI in each sample.
Results
Fig 1 shows the MRI scans of ferumoxytol and gadoterate samples for the three sequences. The coefficient of variation in the samples was 1.3% (SE), 0.9% (SGE), and 1.1% (bSSFP). Noticeable is the susceptibility artefact around the samples of 7 and 14 mM ferumoxytol, which partly distorted the reference sample (see Fig 1, top row). Moreover, the magnetic field inhomogeneities due to the highly paramagnetic samples cause banding artefacts to appear in the bSSFP scans of the ferumoxytol samples. There was no signal in the samples with high concentration (7 mM and 14 mM) of ferumoxytol.
[Figure omitted. See PDF.]
Fig 1.
Setup of the CA phantom (left) for both ferumoxytol (upper row; Fe) and gadoterate (lower row; Gd), with 0 = reference (no CA), 1 = 0.15 mM, 2 = 0.3 mM, 3 = 1.2 mM, 4 = 7 mM, 5 = 14 mM. ‘M’ indicates the marker for orientation. MRI scans of the SE (middle left), SGE (middle right), and bSSFP (right) sequence show the amount of signal compared to the reference sample which is in the center of the setup. Deformation of the reference sample can be seen in the SE and SGE scans of ferumoxytol. Due to scaling only images made with the same sequence can be compared with respect to signal intensity.
https://doi.org/10.1371/journal.pone.0256252.g001
The T1 and T2 maps together with the fit relaxivity values are shown in Fig 2. For ferumoxytol we found an r1 of 40.3 mM-1 s-1 and an r2 of 259.5 mM-1 s-1 at 0.25T (both with R-squared = 0.99). For gadoterate, we found an r1 of 3.58 mM-1 s-1 and an r2 of 21.6 mM-1 s-1 at 0.25T (both with R-squared = 0.96). These values were used as input for simulations of SI curves for SE, SGE, and bSSFP.
[Figure omitted. See PDF.]
Fig 2.
T1 maps (left column) that were calculated with NOVIFAST and used as input for the DESPOT2 method to generate the T2 maps (middle column). T1 and T2 values in the samples were used to calculate relaxation rates as function of the concentration (right column). For ferumoxytol (top row) and gadoterate (bottom row) relaxivity values were fit based on the relaxation rates.
https://doi.org/10.1371/journal.pone.0256252.g002
Fig 3 shows the simulated SI curves for different concentrations of ferumoxytol and gadoterate on 0.25T and 1.5T. The curves of SE and SGE have similar shapes for ferumoxytol with a sharp peak of signal enhancement at low concentrations. For gadoterate there is a plateau at higher concentrations indicating a broader range of concentrations that gives similar signal enhancement on both field strengths. The bSSFP simulation shows for both CAs a decrease in signal compared to the reference, meaning that no signal enhancement could be achieved. A concentration of around 0.16 mM ferumoxytol yielded the maximum signal increase of 3.3 times the reference at 0.25T, whereas for gadoterate this was 3.5 times at a concentration of 1.9 mM on SE and SGE on 0.25T. Maximum increases are thus comparable for both CAs, but occurred earlier for ferumoxytol. The amount of potential maximal signal increase was similar on 0.25T and 1.5T for ferumoxytol, but higher for gadoterate on 1.5T with a 10-fold (SE) and 6.8-fold (SGE) increase.
[Figure omitted. See PDF.]
Fig 3.
Simulated SIs calculated from 0–15 mM for ferumoxytol (Fe) and gadoterate (Gd) for an SE, SGE, and bSSFP sequence at 0.25T (left column) and 1.5T (right column). SIs are shown as normalized values as ratio to no CA (which corresponds with a value of 1). These simulations contributed to identifying relevant concentrations (0–0.15–0.30–1.20–7.0–14.0 mM) that were prepared to perform the experiments. Note the difference in y-axis between 0.25T and 1.5T and the lack of signal enhancement in the bSSFP sequence on 0.25T.
https://doi.org/10.1371/journal.pone.0256252.g003
Fig 4 shows the measured SIs for both CAs on all three sequences on 0.25T. Visual comparison with the simulated curves shows similar results with a peak in SI at low concentrations of ferumoxytol and a wider peak at higher concentrations for gadoterate on SE and SGE. The measured bSSFP shows for ferumoxytol a maximum increase of 1.6 times the reference in the 0.30 mM sample, whereas all gadoterate samples show increase with a maximum of 2.3 times for the 7 mM sample. This is in contrast with simulations which predicted no enhancement at all.
[Figure omitted. See PDF.]
Fig 4. Measured signals (SIs) for the prepared concentrations (0–0.15–0.30–1.20–7.0–14.0 mM) for an SE, SGE, and bSSFP sequence at 0.25T.
SIs are shown as normalized values as ratio to no CA (which corresponds with a value of 1). These results were compared with the normalized simulated SIs from Fig 3 (left column).
https://doi.org/10.1371/journal.pone.0256252.g004
Discussion
The goal of this research was to measure whether a GBCA (gadoterate) or an USPIO (ferumoxytol) would be more suitable for providing positive contrast at low magnetic field strengths. This research confirmed that a low concentration of ferumoxytol (<0.40 mM) leads to similar signal enhancement as much higher concentrations of gadoterate (around 5 mM) on these field strengths.
Relaxivities at 0.25T were measured based on T1 and T2 maps that were calculated with the NOVIFAST [30] and DESPOT2 [31] methods respectively. Both methods seemed to be accurate based on the R-squared values when fitting through the different samples. The T1 maps seemed to produce r1 results in line with expected relaxivities after extrapolation of 0.5T and 1.5T data [6, 13]. However, the calculated r2 was much higher than expected. Based on known relaxivities on 0.5T and 1.5T, these r2 of 259.5 mM-1 s-1 (ferumoxytol) and 21.6 mM-1 s-1 (gadoterate) were expected to be in the range of 80–120 mM-1 s-1 and 3–5 mM-1 s-1 respectively [6, 13].We hypothesize that this is due to signal loss in bSSFP; when this sequence is incorrectly balanced due to background gradients and a relatively long TR, the signal can become T2* (instead of T2) dependent [36]. This results in a severe underestimation of T2 leading to an overestimated r2. Since the simulations were based on these overestimated r2 values, simulated signal in the bSSFP sequence is lower than the measured signal. This was seen and confirmed in the experiments, where a slight increase in SI for both CAs was observed. When simulations were performed with the r2 values expected from literature, simulated signals were more like the experiments.
The relaxation effects that were investigated in this study are harnessed differently for an SE, SGE and bSSFP sequence. An important factor when selecting and applying a CA to specific environments that should be taken into account is the weighting of the sequence [4, 37]. For angiographic purposes, often T1-weighted sequences are used like the SE and SGE. Contrast of those sequences can be enhanced by using a CA with high r1 to strongly influence signal intensity in the blood. However, for bSSFP sequences that are often used in abdominal and cardiac imaging, the contrast depends on the T2/T1 ratio. Administration of CAs is then only beneficial when the T2/T1 ratio does not approach unity [37].
The chosen imaging parameters were determined by the lowest possible combination of TR/TE on the used MRI system. Although these parameters could be further minimized on other MR systems, the experiments demonstrated higher SI at low concentrations of ferumoxytol (<1.2 mM) than at higher concentrations (> 7.0 mM) of gadoterate.
Although ferumoxytol seems to be advantageous at low field strengths when used in low concentrations, it can have certain drawbacks. For example, in dynamic contrast enhanced MRI the application uptake curve is fit to characterize certain tissue properties, and it is often required to run more than one CA dose [38]. To this aim, a faster CA washout is required. Since USPIOs show considerably slower washout times, it may be less convenient for such applications. Furthermore, the narrow concentration range at which the SI peaks has the potential drawback that the concentration should not be too high after administration since this will weaken SI. However, careful administration of USPIO doses is in line with its safety regulations, meaning that high doses are already not allowed. Besides, ferumoxytol blood-pool residence time in comparison with GBCAs is much longer, which negates the need for administration of subsequent doses [29].
The observed narrow peak in Fig 3 in the simulations for ferumoxytol in SE and SGE implicates that only low concentrations cause signal increase, whereas higher concentrations would lead to signal decrease. Besides, the measurements may be missing the absolute maximum of the peak because sampling points are scarce. The optimal concentration range for gadoterate is achieved at higher concentrations under a wider range due to moderate r2 relaxivity. Literature supports our findings by stating that ferumoxytol is an interesting choice as intravascular contrast agent at lower field strengths [6].
Clinical examples that could benefit from lowered CA administration are MR interventions, vascular imaging, and situations where GBCAs are not desired or even impossible to use because of kidney disease. Besides, the difference in CA excretion mechanism between ferumoxytol and GBCAs can also be exploited [39]. Whereas the half-life of ferumoxytol is more than 15 hours, gadolinium CAs have half-lives of around 70–120 minutes [40]. This longer blood circulation time of ferumoxytol will be an advantage in a situation where a constant SI increase is required over a longer time, like in an endovascular intervention with a length of hours [41]. A high peak in SI for low concentrations is then even more useful, since it means that less CA has to be administered over time, decreasing toxicity. Furthermore, lower doses of ferumoxytol at lower field strengths have clinical benefits in terms of reduced adverse reactions, and less hepatic uptake giving less confounding signal changes in other MRI scans of the patient.
The additional benefit of the long blood circulation time of ferumoxytol is that it enables the user to exploit longer imaging times. This allows for more averaging during acquisition resulting in higher SNR, which is also desirable when scanning at low field strengths. Further research could address SNR optimization in low-field MRI to facilitate clinical application of USPIO enhanced imaging.
Conclusions
In conclusion, solely based on its relaxivity characteristics ferumoxytol is more beneficial in generating positive contrast at low magnetic field strengths than gadoterate where lower concentrations yield almost equal signal enhancement. MR sequence optimalisation with respect to specific USPIO behavior in vivo addressing both excretion mechanism and retention time should be the subject of subsequent research.
Supporting information
S1 Dataset.
https://doi.org/10.1371/journal.pone.0256252.s001
(ZIP)
Citation: van Zandwijk JK, Simonis FFJ, Heslinga FG, Hofmeijer EIS, Geelkerken RH, ten Haken B (2021) Comparing the signal enhancement of a gadolinium based and an iron-oxide based contrast agent in low-field MRI. PLoS ONE 16(8): e0256252. https://doi.org/10.1371/journal.pone.0256252
1. Scott LJ. Gadobutrol: A review of its use for contrast-enhanced magnetic resonance imaging in adults and children. Clin Drug Investig. 2013;33: 303–314. pmid:23435930
2. Beomonte B, Carlo Z, Quattrocchi C, Errante Y, Grasso RF. Gadolinium ‑ based contrast agents: did we miss something in the last 25 years? Radiol Med. 2015; 1–4. pmid:26706453
3. Desai NK, Runge VM. Contrast Use at Low Field—A Review. Top Magn Reson Imaging. 2003;14: 360–364. pmid:14625464
4. Wahsner J, Gale EM, Rodríguez-Rodríguez A, Caravan P. Chemistry of MRI contrast agents: Current challenges and new frontiers. Chem Rev. 2019;119: 957–1057. pmid:30350585
5. Marques JP, Simonis FFJ, Webb AG. Low-Field MRI: An MR Physics Perspective. 2019; 1–15. pmid:30637943
6. Campbell-Washburn AE, Ramasawmy R, Restivo MC. Opportunities in Interventional and Diagnostic Imaging by Using High-Performance Low-Field-Strength MRI. Radiology. 2019; 2–11. pmid:31769746
7. Wirtz CR, Aras N. Low-field interventional MRI in neurosurgery: finding the right dose of contrast medium. Interv Neuroradiol. 2001;43: 254–258. pmid:11305762
8. Yrjana SK, Tuominen J, Koivukangas J. Intraoperative Magnetic Resonance Imaging in Neurosurgery. Acta radiol. 2007;48: 540–549. pmid:17520431
9. Blanco RT, Ojala R, Kariniemi J, Perälä J, Niinimäki J, Tervonen O. Interventional and intraoperative MRI at low field scanner—A review. Eur J Radiol. 2005;56: 130–142. pmid:15908156
10. Campbell-washburn AE, Rogers T, Khan JM, Basar B, Ramasawmy R, Grant LP, et al. Clinical MRI-guided right heart catheterization with standard metallic devices using a high performance 0.55T system. Proc ISMRM 2019. 2019.
11. Simonetti OP, Ahmad R. Low-Field Cardiac Magnetic Resonance Imaging: A Compelling Case for Cardiac Magnetic Resonance’s Future. Circ Cardiovasc Imaging. 2017;10: 1–7. pmid:28611117
12. Qureshi AI, Saleem MA. Upright Catheter-Based Cerebral Angiography. J Vasc Interv Neurol. 2017;9: 14–19. pmid:29445433
13. Rohrer M, Bauer H, Mintorovitch J, Requardt M, Weinmann H-J. Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths. Invest Radiol. 2005;40: 715–724. pmid:16230904
14. Wang Y-XJ. Superparamagnetic iron oxide based MRI contrast agents: Current status of clinical application. Quant Imaging Med Surg. 2011;1: 35–40. pmid:23256052
15. Gossuin Y, Gillis P, Hocq A, Vuong QL, Roch A. Magnetic resonance relaxation properties of superparamagnetic particles. Wiley Interdiscip Rev nanobiotechnology. 2009;1: 299–310. pmid:20049798
16. Yin X, Russek SE, Zabow G, Sun F, Mohap J, Keenan KE, et al. Large T1 contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI. Nat Sci reports. 2018;8: 1–10. pmid:30089881
17. Toth GB, Varallyay CG, Horvath A, Bashir MR, Choyke PL, Daldrup-Link HE, et al. Current and potential imaging applications of ferumoxytol for magnetic resonance imaging. Kidney Int. 2016; 1–20. pmid:28434822
18. Aime S, Caravan P. Biodistribution of gadolinium-based contrast agents, including gadolinium deposition. J Magn Reson Imaging. 2009;30: 1259–1267. pmid:19938038
19. Weller A, Barber JL, Olsen ØE. Gadolinium and nephrogenic systemic fibrosis: an update. 2014; 1927–1937. pmid:24146299
20. Vergauwen E, Vanbinst A, Brussaard C, Janssens P, Clerck D De, Lint M Van, et al. Central nervous system gadolinium accumulation in patients undergoing periodical contrast MRI screening for hereditary tumor syndromes. 2018; 1–9. pmid:29312473
21. Bashir MR, Bhatti L, Marin D, Nelson RC. Emerging applications for ferumoxytol as a contrast agent in MRI. J Magn Reson Imaging. 2015;41: 884–898. pmid:24974785
22. Lehrman ED, Plotnik AN, Hope T, Saloner D. Ferumoxytol-enhanced MRI in the peripheral vasculature. Clin Radiol. 2019;74: 37–50. pmid:29731126
23. Nguyen KL, Yoshida T, Kathuria-Prakash N, Zaki IH, Varallyay CG, Semple SI, et al. Multicenter safety and practice for off-label diagnostic use of ferumoxytol in MRI. Radiology. 2019;293: 554–564. pmid:31638489
24. Knobloch G, Colgan T, Wiens CN, Wang X, Schubert T, Hernando D, et al. Relaxivity of Ferumoxytol at 1.5 T and 3.0 T. Invest Radiol. 2017;00: 1. pmid:29215401
25. Hagberg GE, Scheffler K. Effect of r 1 and r 2 relaxivity of gadolinium-based contrast agents on the T 1 -weighted MR signal at increasing magnetic field strengths. Contrast Media Mol Imaging. 2013;8: 456–465. pmid:24375901
26. Gossuin Y, Disch S, Vuong QL, Gillis P, Hermann RP, Park J-H, et al. NMR relaxation and magnetic properties of superparamagnetic nanoworms. Contrast Media Mol Imaging. 2010;5: 318. pmid:21190269
27. Waddington DEJ, Boele T, Maschmeyer R, Kuncic Z, Rosen MS. High-sensitivity in vivo contrast for ultra-low field magnetic resonance imaging using superparamagnetic iron oxide nanoparticles. Sci Adv. 2020;6: 1–10. pmid:32733998
28. Nacif MS, Arai AE, Lima JAC, Bluemke DA. Gadolinium-enhanced cardiovascular magnetic resonance: administered dose in relationship to united states food and drug administration (FDA) guidelines. J Cardiovasc Magn Reson. 2012;14: 18. pmid:22376193
29. Vasanawala SS, Nguyen KL, Hope MD, Bridges MD, Hope TA, Reeder SB, et al. Safety and technique of ferumoxytol administration for MRI. Magn Reson Med. 2016;75: 2107–2111. pmid:26890830
30. Ramos-Llorden G, Vegas-Sanchez-Ferrero G, Bjork M, Vanhevel F, Parizel PM, San Jose Estepar R, et al. NOVIFAST: A Fast Algorithm for Accurate and Precise VFA MRI T 1 Mapping. IEEE Trans Med Imaging. 2018;37: 2414–2427. pmid:29993537
31. Deoni SCL, Rutt BK, Peters TM. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magn Reson Med. 2003;49: 515–526. pmid:12594755
32. Weinstein JS, Varallyay CG, Dosa E, Gahramanov S, Hamilton B, Rooney WD, et al. Superparamagnetic iron oxide nanoparticles: diagnostic magnetic resonance imaging and potential therapeutic applications in neurooncology and central nervous system inflammatory pathologies, a review. J Cereb Blood Flow Metab. 2010;30: 15–35. pmid:19756021
33. Perman WH, Hilal SK, Simon HE, Maudsley AA. Contrast manipulation in NMR imaging. Magn Reson Imaging. 1984;2: 23–32. pmid:6530914
34. Ernst R and, Anderson W. Application of Fourier transform spectroscopy to magnetic resonance. Rev Sci Instrum. 1966;37: 93. Available: http://dx.doi.org/10.1063/1.1719961
35. Oppelt A, Graumann R, Barfuss H, Fischer H, Hartl W, Schajor W. FISP—a new fast MRI sequence. Electromedica. 1986;54: 15–18. Available: http://mri-q.com/uploads/3/2/7/4/3274160/oppelt_fisp_1986.pdf
36. Scheffler K, Hennig J. Is TrueFISP a gradient-echo or a spin-echo sequence? Magn Reson Med. 2003;49: 395–397. pmid:12541263
37. Graves MJ. Pulse sequences for contrast-enhanced magnetic resonance imaging. Radiography. 2007;13: 20–30.
38. Yan Y, Sun X, Shen B. Contrast agents in dynamic contrast-enhanced magnetic resonance imaging. Oncotarget. 2017;8: 43491–43505. Available: www.impactjournals.com/oncotarget/ pmid:28415647
39. Finn JP, Nguyen K-L, Hu P. Ferumoxytol vs. Gadolinium agents for contrast-enhanced MRI: Thoughts on evolving indications, risks, and benefits. J Magn Reson Imaging. 2017; 1–5. pmid:28160356
40. Ersoy H, Rybicki FJ. Biochemical safety profiles of gadolinium-based extracellular contrast agents and nephrogenic systemic fibrosis. J Magn Reson Imaging. 2007;26: 1190–1197. pmid:17969161
41. Krombach GA. MRI Guidance of Vascular Applications. Interventional Magnetic Resonance Imaging. 2012. pp. 175–188.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2021 van Zandwijk et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Recently, there has been a renewed interest in low-field MRI. Contrast agents (CA) in MRI have magnetic behavior dependent on magnetic field strength. Therefore, the optimal contrast agent for low-field MRI might be different from what is used at higher fields. Ultra-small superparamagnetic iron-oxides (USPIOs), commonly used as negative CA, might also be used for generating positive contrast in low-field MRI. The purpose of this study was to determine whether an USPIO or a gadolinium based contrast agent is more appropriate at low field strengths. Relaxivity values of ferumoxytol (USPIO) and gadoterate (gadolinium based) were used in this research to simulate normalized signal intensity (SI) curves within a concentration range of 0–15 mM. Simulations were experimentally validated on a 0.25T MRI scanner. Simulations and experiments were performed using spin echo (SE), spoiled gradient echo (SGE), and balanced steady-state free precession (bSSFP) sequences. Maximum achievable SIs were assessed for both CAs in a range of concentrations on all sequences. Simulations at 0.25T showed a peak in SIs at low concentrations ferumoxytol versus a wide top at higher concentrations for gadoterate in SE and SGE. Experiments agreed well with the simulations in SE and SGE, but less in the bSSFP sequence due to overestimated relaxivities in simulations. At low magnetic field strengths, ferumoxytol generates similar signal enhancement at lower concentrations than gadoterate.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer