1. Introduction
The effectiveness of fixed orthodontic treatment requires an adequate bonding between brackets and enamel surfaces [1]. Orthodontic brackets are fixed appliances that are bonded to the tooth and should remain in place until the end of treatment [2], to achieve this, the bond strength between bracket base and enamel surfaces should be strong enough to resist orthodontic forces and masticatory loads [3]. In this sense, many factors can lead to bracket–enamel bond failure, including the type of enamel conditioner, composition of adhesive, bracket base design, bracket material, as well as clinician skills [4].
Debonding of orthodontics from teeth is a common occurrence during orthodontic treatment, varying between 1.8% [5] and 20.1% [6]. Debonding of brackets during treatment is an unpleasant occurrence for the clinician and the patient resulting in increased treatment costs and duration [7]. During orthodontic treatment, the clinician may decide to debond one bracket intentionally and to rebond it on the tooth in a better position [8]. Therefore, clinicians have often to deal with what is the best option for treating with unintentional/intentional debonded brackets, and regardless of the cause of debonding, the orthodontist must decide whether to rebond the same bracket or to bond a new one [9].
One solution is to recycle or re-condition these brackets to reuse them for the same patient during the same visit. The re-condition process consists of removing bonding agent remnants from the bracket base, thus allowing the brackets to be rebonded [10]. Once a bracket is rebonded for its use again, it should exhibit sufficient bond strength. Thus, the main challenge in rebonding brackets is restoring the bracket base to a retentive pattern without damaging the bracket itself [11].
Adhesive remnants of the dislodged brackets had been conventionally removed in-office by using green stones [12], direct flame [13], tungsten-carbide bur [14], sandblasting [15], silica coating [16], or laser application [17]. Even though these methods can be easily performed out in the dental office with minimal cost, there is no consensus as to which is the best method to remove adhesive remnants from the bracket base. Accordingly, this systematic review and meta-analysis aims to evaluate the bond strength of rebonded brackets after different residual adhesive removal methods. The hypothesis to be tested is that different residual adhesive removal methods would provide similar bond strength of recycled/reused brackets when compared to new orthodontic brackets.
2. Materials and Methods
This systematic review and meta-analysis was reported by following the guidelines of the PRISMA statement [18]. The following PICOS framework was used: population, debonded orthodontic brackets; intervention, methods for residual adhesive removal; control, new orthodontic brackets; outcomes, bond strength; and study design, in vitro studies. The research question was: is there an optimal method to remove the residual adhesive of debonded orthodontic brackets?
2.1. Literature Search
The literature search was performed by two independent reviewers until 15 December 2020. The following five electronic databases were screened: PubMed (MedLine), ISI Web of Science, Cochrane Library, SciELO, and Scopus. The search strategy used is listed in Table 1. The reviewers also hand-searched the reference lists of included articles for the identification of additional manuscripts. After the initial screening, all studies were imported into Mendeley Desktop 1.17.11 software to remove duplicates.
2.2. Study Selection
Two reviewers independently assessed the titles and abstracts of all the manuscripts. Manuscripts for full-text review were selected according to the following eligibility criteria: (1) evaluated the bond strength of new orthodontic metallic brackets; (2) evaluated the bond strength of debonded orthodontic metallic brackets after using a method to remove the adhesive of the orthodontic metallic bracket base; (3) evaluated the bond strength of debonded orthodontic metallic bracket on new intact enamel; (4) included mean and standard deviation data in MPa; (5) published in the English language. Case reports, case series, pilot studies, and reviews were excluded. Full copies of all the potentially relevant studies were analyzed. Those that appeared to meet the inclusion criteria or had insufficient data in the title and abstract to make a clear decision were selected for full analysis. The full-text papers were independently assessed by two authors. Any disagreement regarding the eligibility of the included studies was resolved through discussion and consensus by a third reviewer.
2.3. Data Extraction
Data of interest from the manuscripts included was extracted using Microsoft Office Excel 2019 sheets (Microsoft Corporation, Redmond, WA, USA). These data included the year of publication, country, type of bracket, type of tooth, orthodontic adhesive used, the method for adhesive removal, the mean and standard deviation of the bond strength, and storage conditions.
2.4. Quality Assessment
The methodological quality of each study was assessed by two reviewers, according to the parameters of a previous systematic review of in vitro studies [19]. The risk of bias in each article was evaluated according to the description of the following parameters: specimen randomization, single-operator protocol implementation, blinding of the operator, the presence of a control group, standardization of the sample preparation, adhesive remnant index evaluation (ARI), use of all materials according to the manufacturer’s instructions, and description of the sample size calculation. If the authors reported the parameter, the study received a “YES” for that specific parameter. In case of missing information, the parameter received a “NO.” The risk of bias was classified according to the sum of “YES” answers received: 1 to 3 indicated a high bias, 4 to 6 medium, and 7 to 8 indicated a low risk of bias.
2.5. Statistical Analysis
Meta-analyses were carried out by using a software program (Review Manager Software version 5.4, The Cochrane Collaboration, Copenhagen, Denmark). The analyses were carried out using a random-effect model, and pooled-effect estimates were obtained by comparing the mean difference between bond strength values obtained using new orthodontic brackets or after removing the adhesive resin. Bond strength comparisons were made considering the method used for removing the residual adhesive on the bracket base. A p-value < 0.05 was considered statistically significant. Statistical heterogeneity of the treatment effect among studies was assessed using the Cochran Q test and the inconsistency I2 test.
3. Results
A total of 3748 publications were collected from all databases (Figure 1).
After duplicates were removed, the literature review yielded 3337 manuscripts for initial examination. From these studies, 3300 studies were excluded after reviewing their titles and summaries. In total, 37 studies were examined by full-text reading. Of these studies, 23 were not included in the qualitative analysis: 2 studies evaluated the bond strength to other substrates different than enamel [20,21], 1 study combined several methods for adhesive removal in the same group [22], and 20 studies performed the rebonding process in the same tooth where the initial bonding process was performed [9,10,12,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39], of the remaining 14 studies, 2 were excluded from the quantitative analysis because the mean and standard deviation was not available [40,41], totalizing 12 studies for the quantitative analysis.
Four different adhesive removal methods were identified in this review. These included air abrasion [13,14,15,17,42,43,44,45,46,47], laser [17,42], mechanical grinding [14,43,44], and direct flame [12,41,43]. The characteristics of these studies are summarized in Table 2.
A meta-analysis was performed with 12 in vitro studies. Separate analyses for each adhesive removal method were performed (Figure 2). As the control for each study, the SBS value of new orthodontic brackets was considered. Direct flame methods for removing the residual adhesive were evaluated (Figure 2A). The meta-analysis demonstrated that these methods achieved significantly lower bond strength values of rebonded brackets when compared with the new bonded brackets (p < 0.001). With regards to the use of mechanical grinding methods to remove the residual adhesive from the base of orthodontic brackets, significantly lower SBS values were also observed (Figure 2B; p = 0.007). SBS of rebonded orthodontic brackets after adhesive removal with sandblasting was analyzed (Figure 2C). The meta-analysis performed demonstrated that bond strength values after adhesive removal through sandblasting were significantly lower than the bond strength of new orthodontic brackets (p = 0.006). With regards to the use of laser, two different types of laser were identified (Figure 2D). When a CO2 laser was used for adhesive removal, the SBS of rebonded brackets was lower than the bond strength of new orthodontic brackets (p < 0.001). On the other hand, the use of an Er:YAG laser for adhesive removal showed similar SBS values when compared with those of new orthodontic brackets (p = 0.71).
According to the parameters considered in the risk of bias assessment, the majority of studies were classified with a medium risk of bias (Table 3). Several of the studies failed to report single-operator, operator-blinded, and sample size calculation parameters.
4. Discussion
This systematic review and meta-analysis aimed to evaluate the bond strength of debonded brackets after different residual adhesive removal methods. Direct flame, mechanical grinding, sandblasting, and laser were the methods found in the literature used for this purpose. Except for the Er:YAG laser, none of the methods evaluated managed to restore SBS values of new orthodontic brackets values, thus our hypothesis was partially rejected.
One of the methods proposed to remove the adhesive remnant after bracket debonding is direct flaming of the bracket base. Under the use of this method, removal of the bonding agent is the most critical part of the recycling process and requires long exposure to heat [44]. The results of the present meta-analysis helped to demonstrate that this method was unable to recover the original values achieved by new orthodontic brackets. Several explanations may be suggested to explain this behavior. First, direct flaming increases the temperature of the bracket base to a temperature in the range of 600–800 °C, which can lead to the disintegration of the metal alloy, and consequently weakens its structure, making it more vulnerable to damage [30]. Also, as most of the metallic orthodontic brackets are made of austenitic stainless steel, application of heat leads to the formation of chrome-carbide compounds, which can render them more susceptible to tarnish and corrosion, and this, in turn, could be responsible for its failure in the mouth [48]. Finally, it has been found that the heat treatment could lead to a decrease in the diameter of the support mesh, which is caused by the presence of large amounts of adhesive residues on the base [30].
When observing the data about mechanical grinding, four studies reported this method. For this purpose, a green stone [43,48], or a carbide bur [14,41] at slow speed were used to grind the bracket surface. The meta-analysis revealed that, when this method was used for the removing of adhesive residual, significantly lower values in the bond strength of rebonded brackets were achieved. During the adhesive removal from the bracket, the preservation of the integrity of the bracket mesh is crucial to ensure an adequate bond strength to the enamel. By grinding the bracket base using a green stone or a carbide bur, there is a high risk of damage or grinding off the mesh base itself, resulting in a decrease in bond strength. [12] Also, grinding the bracket mesh has been proved to leave a considerable amount of the adhesive, obliterating the mesh and decreasing the contact area, thus eliminating virtually any mechanical retention [10,12,14,44].
Sandblasting has been described as a viable procedure for rebonding accidentally lost brackets. This method was the most used in the studies included in the present systematic review. The findings obtained by the meta-analysis suggest that the bond strength observed by debonded, cleaned brackets with sandblasting is significantly lower when compared with new brackets. Previous research has demonstrated that sandblasting of the bracket base could provoke distortion of the mesh [33]. In this sense, the air abrasion procedure causes macro and microscopic alterations in the structure of the bonding surface, consequently affecting the bond strength outcomes [12]. Also, it has been described that after sandblasting, some abrasive particles adhere to the blasted surface, and it is possible that bond strength between any luting material and the abrasive particle remnants might exceed the bond strength of the abrasive particles and the bonding surface, causing premature debonding [49]. On the other hand, the sandblasting process is not able to remove all the resin attached to the bonding mesh [17], directly affecting the bond strength outcomes.
When observing the data about lasers, two different methods were analyzed separately [17,43]. This technology selectively ablates composite by high pulse repetition rates [49]. When analyzing the CO2 laser, it was found that it is not a suitable method for recycling brackets because considerable amounts of adhesive remnants were left on the base of CO2 laser-irradiated brackets [49]. As explained before, the remaining adhesive on the bracket base lessens the contact area between the meshwork and adhesive and leads to a decrease in bond strength values. On the other hand, the analysis of the results from the Er:YAG laser method demonstrated that this method is efficient for removing the residual adhesive, being that the values obtained were similar to those of new orthodontic brackets. This result could be explained due to the complete removal of the residual adhesive from the bracket bases without altering the micro and macrostructure of the mesh, resembling the appearance and bonding performance of new brackets [17]. Nevertheless, it should be advised that the use of the Er:YAG laser could melt the meshwork of the bracket base due to heat, and some precautions should be taken when using this method [17].
Regarding the limitations of this systematic review, it is important to highlight that all analyses performed showed high heterogeneity values, which could be attributed to the lack of standardization of the methods for evaluation of the SBS; actually, none of the included studies indicated the following of the international standards for bond strength tests to dental tissues. Future research with more standardized methods is desired to reduce the heterogeneity between the studies focused on this topic and also to establish the optimal protocol for the adhesive removal for rebonding of debonded orthodontic brackets. Also, it is important to encourage researchers for designing and conducting clinical trials evaluating this outcome.
On the other hand, it should be pointed out that despite the meta-analysis showing statistical differences between the SBS of debonded and new brackets, such differences are not clinically relevant, this is because the mean bond strength values of the methods evaluated succeeded to achieve at least 6 MPa, which is the minimum bond strength required for successful orthodontic treatment [17,23]. This could also lead to the perspective that rebonding of debonded orthodontic brackets in the same patient is a reliable treatment option, as long as the adhesive residual within the orthodontic base is completely removed using the above-mentioned procedures.
5. Conclusions
Within the limitations of this systematic review, it could be concluded that the Er:YAG laser could be considered as an optimal method for promoting the bond of debonded orthodontic brackets, this conclusion is based on the fact that the bond strength of rebonded orthodontic brackets was the same as that of the new brackets. Nevertheless, the data suggest that the use of direct flame, mechanical grinding, or sandblasting are suitable options for the removal of residual adhesive from the orthodontic bracket base, obtaining clinically acceptable bond strength values.
Conceptualization, G.G. and L.N.; methodology, G.G., L.N. and C.E.C.-S.; software, L.H., R.B. and M.L.-S.; validation, R.B., G.G., P.P. and E.A.; formal analysis, C.E.C.-S. and R.B.; investigation, L.H. and P.P.; resources, M.Z., N.J., L.H. and R.B.; data curation, C.E.C.-S.; writing—original draft preparation G.G., L.N., R.B., M.Z., E.A. and L.H.; writing—review and editing, M.L.-S., L.H. and C.E.C.-S.; visualization, E.A., R.B., M.L.-S., N.J., L.H. and M.Z.; supervision, M.L.-S., L.H. and C.E.C.-S.; project administration, L.H. and C.E.C.-S. All authors have read and agreed to the published version of the manuscript.
This research received no external funding.
Not applicable.
Not applicable.
The data that support the findings of this study are available from the corresponding author upon reasonable request.
The authors declare no conflict of interest.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Figure 2. Results of the meta-analysis of bond strength of debonded orthodontics brackets after residual adhesive removal using; (A) Direct flame; (B) Mechanical grinding; (C) Sandblasting; and (D) Laser.
Keywords used in the search strategy.
Search Strategy | |
---|---|
# 1 | Orthodontic bracket OR bracket OR braces OR stainless steel bracket OR recycled bracket. |
# 2 | Rebonded OR rebonding OR reconditioning OR recycling OR recycling methods OR recycled brackets OR rebonded brackets OR electropolishing OR sandblasting OR ultrasonic scaling OR heating OR Er:YAG laser OR CO2 laser |
# 3 | #1 and #2 |
Demographic data of included studies.
Study | Bracket Used | Tooth Used | Orthodontic Adhesive Used | Storing Conditions | Residual Adhesive Removal Method Used | Secondary Outcome |
---|---|---|---|---|---|---|
Achio, 2015 | Stainless-steel premolar bracket (UnitekTM Gemini Bracket, 3M Unitek, Monorovia, CA, USA) | Human premolar | Transbond™ Plus Self Etching Primer (3M Unitek)/TransbondTM XT Light Cure Composite (3M Unitek) | Thermocycling (500 cycles between 5 °C and 55 °C) | Sandblasting (Al2O3; 50 µm, 90 psi, 10 mm, 10–15 s) | Adhesive remnant index |
Bahnasi, 2013 | Stainless steel upper premolar bracket (UnitekTM Gemini Bracket (3M Unitek) | Human premolar | Light Cure Orthodontic Adhesive Primer (3M Unitek)/TransbondTM XT Light Cure Composite (3M Unitek) | Thermocycling (500 cycles between 5 °C and 55 °C) | Sandblasting (Al2O3; 50 µm, 90 psi, 10 mm, 20–30 s) | Adhesive remnant index |
Bahnasi, 2013 (b) | Stainless steel upper premolar bracket (UnitekTM Gemini Bracket, 3M Unitek) | Human premolar | Light Cure Orthodontic Adhesive Primer (3M Unitek)/TransbondTM XT Light Cure Composite (3M Unitek) | Thermocycling (500 cycles between 5 °C and 55 °C) | Sandblasting (Al2O3; 50 µm, 90 psi, 10 mm, 20–30 s). |
Qualitative analysis of the distortion of the base with SEM |
Egan, 1996 | Stainless steel upper premolar brackets (GAC International Inc., New York, NY, USA) | Human premolar | Rely a Bond (Reliance Orthodontic Products Inc., Itasca, IL, USA) and Phase II paste-paste (Reliance Orthodontic Products Inc.) | Distilled water at 37 °C for 1 week | Mechanical grinding with a green stone | Failure mode |
Harini, 2011 | Stainless steel premolar brackets * | Human premolar | All Bon-2 (Bisco Inc., Schaumburg, IL, USA. | Distilled water for 24 h | Direct flame with a soldered torch for 5 s. | Adhesive remnant index |
Heravi, 2006 | Standard Edgewise metal brackets (Dentarum Corp., Ispringen, Germany) | Human upper premolar | No-mix composite (Dentarum Corp., Germany) | Distilled water at 37 °C for 48 h | Mechanical grinding with a tungsten carbide bur with high-speed hand piece | Adhesive remnant index |
Ishida, 2011 | Metal premolar bracket (UnitekTM Victory series, 3M Unitek) | Human premolar | Transbond™ Plus Self Etching |
Artificial saliva at 37 °C for 24 h | Er,Cr:YSGG laser (Power output of 3.75 W, wavelength of 2.78 µm, a pulse duration of 140 µs, a frequency of 20 Hz, and air and water |
Adhesive remnant index |
Kachoei, 2016 | Maxillary central incisors (Ortho-Organizer, Carlsbad, CA, USA) | Bovine upper central incisors | Unite Bonding System (3M Unitek, USA) | Distilled water at 37 °C |
Sandblasting (Al2O3; 50 um, 5 mm). |
Adhesive remnant index |
Kamissety, 2015 | Stainless steel lower premolar brackets (Gemini, 3M Unitek) | Lower human premolar | Transbond XT adhesive (3M Unitek). | Artificial saliva for 24 h at 37 °C | Mechanical grinding with a green stone with low-speed hand piece. |
UV/Vis transmittance analysis |
Maaitah, 2013 | Premolar brackets (Omni 0.022′’ Roth, GAC International Inc, New York, NY, USA) | Human premolar teeth | TransbondTM XT Adhesive (3M Unitek) | Thermocycling (500 cycles between 5 °C and 55 °C) | Mechanical grinding with slow speed round tungsten carbide bur. |
Adhesive remnant index |
Montero, 2015 | Upper central incisor brackets (UnitekTM Victory series, 3M Unitek) | Bovine upper central incisors | Transbond Plus Self Etching |
Distilled |
Sandblasting (Al2O3; 25 µm, 50 µm, or 110 µm at 5 mm) | SEM observation |
Shahamfar, 2014 | Premolar bracket (Equilibrium, Dentaaurum Inc., Ispringen, Germany) | Human premolar teeth | Light BondTM (Reliance Orthodontic products, IL, USA) | Distilled |
Mechanical grinding with slow speed multi blade tungsten carbide bur. | Adhesive remnant index |
Sonis, 1996 | Lower premolar brackets (GAC International, Inc., Central Islip, Long Island, NY, USA) | Lower human premolar | Rely-a-bond (Reliance, Inc., Itasca, IL, USA) | Thermocycling (1000 cycles between 10 °C and 50 °C) | Sandblasting (90 µm; 90 PSI, 15 to 30 s) | Scanning electron micrograph of base surface |
Wheeler, 1983 | Stainless steel premolar brackets | Human premolar | Dyna Bond II Series B (Unitek Corporation, Monrovia, CA, USA) | Non-specified | Heating in an oven for 50 min at 454 °C |
Qualitative synthesis (risk of bias assessment).
Study | Specimen Randomization | Single Operator | Operator Blinded | Control Group | Standardized Specimens | ARI | Manufacturer’s Instructions | Sample Size Calculation | Risk of Bias |
---|---|---|---|---|---|---|---|---|---|
Achio, 2015 | Yes | No | No | Yes | Yes | Yes | Yes | No | Medium |
Bahnasi, 2013 | No | No | No | Yes | Yes | Yes | Yes | No | Medium |
Bahnasi (b), 2013 | Yes | No | No | Yes | Yes | No | Yes | No | Medium |
Egan, 1996 | No | No | No | Yes | Yes | Yes | Yes | No | Medium |
Harini, 2011 | No | No | No | Yes | Yes | Yes | Yes | No | Medium |
Heravi, 2006 | No | No | No | Yes | Yes | Yes | Yes | No | Medium |
Ishida, 2011 | Yes | Yes | No | Yes | Yes | Yes | Yes | No | Medium |
Kachoei, 2016 | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Low |
Kamissety, 2015 | No | No | No | Yes | Yes | No | Yes | No | High |
Maaitah, 2013 | Yes | Yes | No | Yes | Yes | Yes | Yes | No | Medium |
Montero, 2015 | Yes | No | No | Yes | Yes | No | Yes | Yes | Medium |
Shahamfar, 2014 | No | No | No | Yes | Yes | Yes | Yes | No | Medium |
Sonis, 1996 | Yes | No | No | Yes | Yes | No | Yes | No | Medium |
Wheeler, 1983 | Yes | Yes | No | Yes | Yes | No | Yes | No | Medium |
References
1. Ahmed, T.; Rahman, N.A.; Alam, M.K. Assessment of in Vivo Bond Strength Studies of the Orthodontic Bracket-Adhesive System: A Systematic Review. Eur. J. Dent.; 2018; 12, pp. 602-609. [DOI: https://dx.doi.org/10.4103/ejd.ejd_22_18] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30369810]
2. Tsichlaki, A.; Chin, S.Y.; Pandis, N.; Fleming, P.S. How Long Does Treatment with Fixed Orthodontic Appliances Last? A Systematic Review. Am. J. Orthod. Dentofac. Orthop.; 2016; 149, pp. 308-318. [DOI: https://dx.doi.org/10.1016/j.ajodo.2015.09.020]
3. Knox, J.; Hubsch, P.; Jones, M.L.; Middleton, J. The Influence of Bracket Base Design on the Strength of the Bracket–Cement Interface. J. Orthod.; 2000; 27, pp. 249-254. [DOI: https://dx.doi.org/10.1179/ortho.27.3.249] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/11099557]
4. Bakhadher, W.; Halawany, H.; Talic, N.; Abraham, N.; Jacob, V. Factors Affecting the Shear Bond Strength of Orthodontic Brackets—A Review of In Vitro Studies. Acta Med.; 2015; 58, pp. 43-48. [DOI: https://dx.doi.org/10.14712/18059694.2015.92]
5. Roelofs, T.; Merkens, N.; Roelofs, J.; Bronkhorst, E.; Breuning, H. A Retrospective Survey of the Causes of Bracket- and Tube-Bonding Failures. Angle Orthod.; 2017; 87, pp. 111-117. [DOI: https://dx.doi.org/10.2319/021616-136.1] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/27304230]
6. Barbosa, I.V.; de Miranda Ladewig, V.; Almeida-Pedrin, R.R.; Cardoso, M.A.; Santiago Junior, J.F.; de Castro Ferreira Conti, A.C. The Association between Patient’s Compliance and Age with the Bonding Failure of Orthodontic Brackets: A Cross-Sectional Study. Prog. Orthod.; 2018; 19, 11. [DOI: https://dx.doi.org/10.1186/s40510-018-0209-1]
7. Sukhia, H.R.; Sukhia, R.H.; Mahar, A. Bracket De-Bonding and Breakage Prevalence in Orthodontic Patients. Pak. Oral. Dent. J.; 2011; 31, pp. 1-5.
8. Koo, B.C.; Chung, C.-H.; Vanarsdall, R.L. Comparison of the Accuracy of Bracket Placement between Direct and Indirect Bonding Techniques. Am. J. Orthod. Dentofac. Orthop.; 1999; 116, pp. 346-351. [DOI: https://dx.doi.org/10.1016/S0889-5406(99)70248-9]
9. Mui, B.; Rossouw, P.E.; Kulkarni, G.V. Optimization of a Procedure for Rebonding Dislodged Orthodontic Brackets. Angle Orthod.; 1999; 69, pp. 276-281. [DOI: https://dx.doi.org/10.1043/0003-3219(1999)0692.3.CO,2]
10. Yassaei, S.; Aghili, H.; KhanPayeh, E.; Goldani Moghadam, M. Comparison of Shear Bond Strength of Rebonded Brackets with Four Methods of Adhesive Removal. Lasers Med. Sci.; 2014; 29, pp. 1563-1568. [DOI: https://dx.doi.org/10.1007/s10103-013-1310-9]
11. Sohrabi, A.; Jafari, S.; Kimyai, S.; Rikhtehgaran, S. Er,Cr:YSGG Laser as a Novel Method for Rebonding Failed Ceramic Brackets. Photomed. Laser Surg.; 2016; 34, pp. 483-486. [DOI: https://dx.doi.org/10.1089/pho.2014.3868]
12. Halwai, H.K.; Kamble, R.H.; Hazarey, P.V.; Gautam, V. Evaluation and Comparision of the Shear Bond Strength of Rebonded Orthodontic Brackets with Air Abrasion, Flaming, and Grinding Techniques: An in Vitro Study. Orthodontics; 2012; 13, pp. e1-e9. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22567639]
13. Bahnasi, F.I.; Abd-Rahman, A.N.; Abu-Hassan, M.I. Effects of Recycling and Bonding Agent Application on Bond Strength of Stainless Steel Orthodontic Brackets. J. Clin. Exp. Dent.; 2013; 5, pp. e197-e202. [DOI: https://dx.doi.org/10.4317/jced.51113] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/24455081]
14. Al Maaitah, E.F.; Alomari, S.; Abu Alhaija, E.S.; Safi, A.A.M. The Effect of Different Bracket Base Cleaning Method on Shear Bond Strength of Rebonded Brackets. J. Contemp. Dent. Pract.; 2013; 14, pp. 866-870. [DOI: https://dx.doi.org/10.5005/jp-journals-10024-1417] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/24685789]
15. Sanchez Achio, T. A Comparative Study of Shear Debonding Strengh between New Brackets, Air-Abrasion and Recycled Brackets: An In Vitro Analysis. Odovtos-Int. J. Dent. Sci.; 2015; 17, pp. 59-69.
16. Guarita, M.K.; Moresca, A.H.K.; Losso, E.M.; Moro, A.; Moresca, R.C.; Correr, G.M. Effect of Different Surface Treatments for Ceramic Bracket Base on Bond Strength of Rebonded Brackets. Braz. Dent. J.; 2015; 26, pp. 61-65. [DOI: https://dx.doi.org/10.1590/0103-6440201300234]
17. Ishida, K.; Endo, T.; Shinkai, K.; Katoh, Y. Shear Bond Strength of Rebonded Brackets after Removal of Adhesives with Er,Cr:YSGG Laser. Odontology; 2011; 99, pp. 129-134. [DOI: https://dx.doi.org/10.1007/s10266-011-0012-7]
18. Elshafay, A.; Omran, E.S.; Abdelkhalek, M.; El-Badry, M.O.; Eisa, H.G.; Fala, S.Y.; Dang, T.; Ghanem, M.A.T.; Elbadawy, M.; Elhady, M.T. et al. Reporting Quality in Systematic Reviews of in Vitro Studies: A Systematic Review. Curr. Med. Res. Opin.; 2019; 35, pp. 1631-1641. [DOI: https://dx.doi.org/10.1080/03007995.2019.1607270]
19. Bourgi, R.; Hardan, L.; Rivera-Gonzaga, A.; Cuevas-Suárez, C.E. Effect of Warm-Air Stream for Solvent Evaporation on Bond Strength of Adhesive Systems: A Systematic Review and Meta-Analysis of in Vitro Studies. Int. J. Adhes. Adhes.; 2021; 105, 102794. [DOI: https://dx.doi.org/10.1016/j.ijadhadh.2020.102794]
20. Regan, D.; van Noort, R.; O’Keeffe, C. The Effects of Recycling on the Tensile Bond Strength of New and Clinically Used Stainless Steel Orthodontic Brackets: An In Vitro Study. Br. J. Orthod.; 1990; 17, pp. 137-145. [DOI: https://dx.doi.org/10.1179/bjo.17.2.137]
21. Basudan, A.M.; Al-Emran, S.E. The Effects of In-Office Reconditioning on the Morphology of Slots and Bases of Stainless Steel Brackets and on the Shear/Peel Bond Strength. J. Orthod.; 2001; 28, pp. 231-236. [DOI: https://dx.doi.org/10.1093/ortho/28.3.231]
22. Wendl, B.; Muchitsch, P.; Pichelmayer, M.; Droschl, H.; Kern, W. Comparative Bond Strength of New and Reconditioned Brackets and Assessment of Residual Adhesive by Light and Electron Microscopy. Eur. J. Orthod.; 2011; 33, pp. 288-292. [DOI: https://dx.doi.org/10.1093/ejo/cjq081] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/21135035]
23. Abe, R.; Endo, T.; Shimooka, S. Effects of Tooth Bleaching on Shear Bond Strength of Brackets Rebonded with a Self-Etching Adhesive System. Odontology; 2011; 99, pp. 83-87. [DOI: https://dx.doi.org/10.1007/s10266-010-0138-z] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/21271331]
24. Atsu, S.; Catalbas, B.; Gelgor, I.E.; Atsü, S.; Çatalbaş, B.; Gelgör, I.E. Effects of Silica Coating and Silane Surface Conditioning on the Bond Strength of Rebonded Metal and Ceramic Brackets. J. Appl. Oral. Sci.; 2011; 19, pp. 233-239. [DOI: https://dx.doi.org/10.1590/S1678-77572011000300010] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/21625739]
25. Bansal, N.; Valiathan, A.; Bansal, K. The Effects of Various In-Office Reconditioning Methods on Shear Bond Strength, Morphology of Slots and Bases of Stainless Brackets: An in Vitro Study. J. Indian Orthod. Soc.; 2011; 45, pp. 175-182. [DOI: https://dx.doi.org/10.5005/jp-journals-10021-1031]
26. Bishara, S.E.; VonWald, L.; Laffoon, J.F.; Warren, J.J. The Effect of Repeated Bonding on the Shear Bond Strength of a Composite Resin Orthodontic Adhesive. Angle Orthod.; 2000; 70, pp. 435-441. [DOI: https://dx.doi.org/10.1043/0003-3219(2000)0702.0.CO,2] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/11138647]
27. Chacko, P.K.; Kodoth, J.; John, J.; Kumar, K. Recycling Stainless Steel Orthodontic Brackets with Er:YAG Laser—An Environmental Scanning Electron Microscope and Shear Bond Strength Study. J. Orthod. Sci.; 2013; 2, pp. 87-94. [DOI: https://dx.doi.org/10.4103/2278-0203.119680]
28. Chung, C.H.; Fadem, B.W.; Levitt, H.L.; Mante, F.K. Effects of Two Adhesion Boosters on the Shear Bond Strength of New and Rebonded Orthodontic Brackets. Am. J. Orthod. Dentofac. Orthop.; 2000; 118, pp. 295-299. [DOI: https://dx.doi.org/10.1067/mod.2000.104810] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/10982930]
29. Grabouski, J.K.; Staley, R.N.; Jakobsen, J.R. The Effect of Microetching on the Bond Strength of Metal Brackets When Bonded to Previously Bonded Teeth: An in Vitro Study. Am. J. Orthod. Dentofac. Orthop.; 1998; 114, pp. 452-460. [DOI: https://dx.doi.org/10.1016/S0889-5406(98)70192-1]
30. Gupta, N.; Kumar, D.; Palla, A. Evaluation of the Effect of Three Innovative Recyling Methods on the Shear Bond Strength of Stainless Steel Brackets-an in Vitro Study. J. Clin. Exp. Dent.; 2017; 9, pp. e550-e555. [DOI: https://dx.doi.org/10.4317/jced.53586]
31. Kumar, M.; Maheshwari, A.; Lall, R.; Navit, P.; Singh, R.; Navit, S. Comparative Evaluation of Shear Bond Strength of Recycled Brackets Using Different Methods: An In Vitro Study. Int. J. Oral Health Dent.; 2014; 6, pp. 5-11.
32. Lew, K.K.; Chew, C.L.; Lee, K.W. A Comparison of Shear Bond Strengths between New and Recycled Ceramic Brackets. Eur. J. Orthod.; 1991; 13, pp. 306-310. [DOI: https://dx.doi.org/10.1093/ejo/13.4.306]
33. Lunardi, N.; Gameiro, G.H.; de Araújo Magnani, M.B.; Nouer, D.F.; de Siqueira, V.C.; Consani, S.; Pereira-Neto, J.S. The Effect of Repeated Bracket Recycling on the Shear Bond Strength of Different Orthodontic Adhesives. Braz. J. Oral Sci.; 2008; 7, pp. 1648-1652.
34. Montasser, M.A.; Drummond, J.L.; Roth, J.R.; Al-Turki, L.; Evans, C.A. Rebonding of Orthodontic Brackets. Angle Orthod.; 2008; 78, pp. 537-544. [DOI: https://dx.doi.org/10.2319/022707-102.1]
35. Jimenez, E.E.O.; Hilgenberg, S.P.; Rastelli, M.C.; Pilatti, G.L.; Orellana, B.; Coelho, U. Rebonding of Unused Brackets with Different Orthodontic Adhesives. Dent. Press J. Orthod.; 2012; 17, pp. 69-76. [DOI: https://dx.doi.org/10.1590/S2176-94512012000200014]
36. Quick, A.N.; Harris, A.M.P.; Joseph, V.P. Office Reconditioning of Stainless Steel Orthodontic Attachments. Eur. J. Orthod.; 2005; 27, pp. 231-236. [DOI: https://dx.doi.org/10.1093/ejo/cjh100] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/15947221]
37. Regan, D.; LeMasney, B.; van Noort, R. The Tensile Bond Strength of New and Rebonded Stainless Steel Orthodontic Brackets. Eur. J. Orthod.; 1993; 15, pp. 125-135. [DOI: https://dx.doi.org/10.1093/ejo/15.2.125]
38. Tavares, S.W.; Consani, S.; Nouer, D.F.; de Araújo Magnani, M.B.B.; Nouer, P.R.A.; Martins, L.M. Shear Bond Strength of New and Recycled Brackets to Enamel. Braz. Dent. J.; 2006; 17, pp. 44-48. [DOI: https://dx.doi.org/10.1590/S0103-64402006000100010] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/16721464]
39. Harini, T.; Sreedhar, R. Effect of an Adhesion Booster on the Bond Strength of New and Recycled Brackets. Ann. Dent.; 2011; III, pp. 20-22. [DOI: https://dx.doi.org/10.5368/aedj.2011.3.3.1.4]
40. Wheeler, J.J.; Ackerman, R.J. Bond Strength of Thermally Recycled Metal Brackets. Am. J. Orthod. Dentofac. Orthop.; 1983; 83, pp. 181-186. [DOI: https://dx.doi.org/10.1016/0002-9416(83)90081-7]
41. Bahnasi, F.I.; Rahman, A.N.A.A.; Abu-Hassan, M.I. The Impact of Recycling and Repeated Recycling on Shear Bond Strength of Stainless Steel Orthodontic Brackets. Orthod. Waves; 2013; 72, pp. 16-22. [DOI: https://dx.doi.org/10.1016/j.odw.2012.10.001]
42. Kachoei, M.; Mohammadi, A.; Esmaili Moghaddam, M.; Rikhtegaran, S.; Pourghaznein, M.; Shirazi, S. Comparison of Multiple Rebond Shear Strengths of Debonded Brackets after Preparation with Sandblasting and CO2 Laser. J. Dent. Res. Dent. Clin. Dent. Prospects; 2016; 10, pp. 148-154. [DOI: https://dx.doi.org/10.15171/joddd.2016.024]
43. Kamisetty, S.K.; Verma, J.K.; Arun, S.; Chandrasekhar, S.; Kumar, A. SBS vs Inhouse Recycling Methods-An Invitro Evaluation. J. Clin. Diagn. Res.; 2015; 9, ZC04-8. [DOI: https://dx.doi.org/10.7860/JCDR/2015/13865.6432]
44. Haro Montero, M.M.; Vicente, A.; Alfonso-Hernández, N.; Jiménez-López, M.; Bravo-González, L.A. Comparison of Shear Bond Strength of Brackets Recycled Using Micro Sandblasting and Industrial Methods. Angle Orthod.; 2015; 85, pp. 461-467. [DOI: https://dx.doi.org/10.2319/032414-221.1] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25230006]
45. Heravi, F.; Naseh, R.A. Comparative Study between Bond Strength of Rebonded and Recycled Orthodontic Brackets. Dent. Res. J.; 2006; 2, pp. 1-6.
46. Sonis, A.L. Air Abrasion of Failed Bonded Metal Brackets: A Study of Shear Bond Strength and Surface Characteristics as Determined by Scanning Electron Microscopy. Am. J. Orthod. Dentofac. Orthop.; 1996; 110, pp. 96-98. [DOI: https://dx.doi.org/10.1016/S0889-5406(96)70094-X]
47. Huang, T.-H.; Yen, C.-C.; Kao, C.-T. Comparison of Ion Release from New and Recycled Orthodontic Brackets. Am. J. Orthod. Dentofac. Orthop.; 2001; 120, pp. 68-75. [DOI: https://dx.doi.org/10.1067/mod.2001.113794]
48. Egan, F.R.; Alexander, S.A.; Cartwright, G.E. Bond Strength of Rebonded Orthodontic Brackets. Am. J. Orthod. Dentofac. Orthop.; 1996; 109, pp. 64-70. [DOI: https://dx.doi.org/10.1016/S0889-5406(96)70164-6]
49. Yassaei, S.; Aghili, H.; Firouzabadi, A.H.; Meshkani, H. Effect of Er: YAG Laser and Sandblasting in Recycling of Ceramic Brackets. J. Lasers Med. Sci.; 2017; 8, pp. 17-21. [DOI: https://dx.doi.org/10.15171/jlms.2017.04]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Debonding of orthodontic brackets is a common occurrence during orthodontic treatment. Therefore, the best option for treating debonded brackets should be indicated. This study aimed to evaluate the bond strength of rebonded brackets after different residual adhesive removal methods. This systematic review and meta-analysis was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement. PubMed, Web of Science, The Cochrane Library, SciELO, Scopus, LILACS, IBECS, and BVS databases were screened up to December 2020. Bond strength comparisons were made considering the method used for removing the residual adhesive on the bracket base. A total of 12 studies were included for the meta-analysis. Four different adhesive removal methods were identified: sandblasting, laser, mechanical grinding, and direct flame. When compared with new orthodontic metallic brackets, bond strength of debonded brackets after air abrasion (p = 0.006), mechanical grinding (p = 0.007), and direct flame (p < 0.001) was significantly lower. The use of an erbium-doped yttrium aluminum garnet (Er:YAG) laser showed similar shear bond strength (SBS) values when compared with those of new orthodontic brackets (p = 0.71). The Er:YAG laser could be considered an optimal method for promoting the bond of debonded orthodontic brackets. Direct flame, mechanical grinding, or sandblasting are also suitable, obtaining clinically acceptable bond strength values.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details







1 Department of Dental Materials, School of Dentistry, Universidad de la República. Av. General Las Heras 1925, Montevideo 11300, Uruguay;
2 Department of Restorative Dentistry, School of Dentistry, Saint-Joseph University, Beirut 1107 2180, Lebanon;
3 Graduate Program in Dentistry, School of Dentistry, Federal University of Pelotas, Rua Gonçalves Chaves, 457, Pelotas 96015560, Brazil;
4 Department of Orthodontics, School of Dental Medicine, Saint Joseph University, Beirut 1107 2180, Lebanon;
5 “NZOZ SPS Dentist” Dental Clinic and Postgraduate Course Centre—pl. Inwalidow 7/5, 30-033 Cracow, Poland;
6 Department of Pediatric Dentistry, Medical University of Lodz, Pomorska 251, 92-213 Lodz, Poland;
7 Dental Materials Laboratory, Academic Area of Dentistry, Autonomous University of Hidalgo State, Circuito Ex Hacienda La Concepción S/N, San Agustín Tlaxiaca 42160, Mexico
8 Department of General Dentistry, Medical University of Lodz, 251 Pomorska St., 92-213 Lodz, Poland