Abstract

Background

For oblique metacarpal shaft fracture, if anatomical reduction is achieved through conservative cast immobilization rather than stable fixation, bone malrotation can easily occur, resulting in severe loss in hand prehensile function. However, whether bone plate fixation or only lag screw fixation is more preferable remains unclear. Few studies have evaluated whether screw fixation can provide biomechanical fixation strength similar to bone plate fixation.

Objective

We assessed the difference in fixation strength between fixtation with two lag screws and bone plate for oblique metacarpal shaft fractures.

Materials and methods

We created oblique metacarpal shaft fractures on 21 artificial bones and fixated them using (1) double lag screw (2LS group), (2) regular plate (RP group), or (3) locked plate (LP group). To obtain the force–displacement data, a cantilever bending test was conducted for each specimen through a material testing machine. One-way analysis of variance and a Tukey test were conducted to compare the maximum fracture force and stiffness of the three fixation methods.

Results

The maximum fracture force of the 2LS group (mean + SD: 153.6 ± 26.5 N) was significantly lower than that of the RP (211.6 ± 18.5 N) and LP (227.5 ± 10.0 N) groups (p < 0.001). However, no significant differences were discovered between the RP and LP groups. The coefficient of variation for the maximum fracture force of the 2LS group (17.3%) was more than twice as high as that of the RP (8.7%) and LP (4.4%) groups. In addition, the stiffness of the three fixation methods was similar.

Conclusion

Compared with bone plate fixation, double lag screw fixation yielded slightly lower maximum bearable fracture force but similar stiffness. Therefore, this technique could be used for treating oblique metacarpal shaft fractures. However, using double lag screw fixation alone is technically demanding and requires considerable surgical experiences to produce consistent results.

Details

Title
Comparison of the fixation ability between lag screw and bone plate for oblique metacarpal shaft fracture
Author
Yung-Cheng, Chiu; Tsung-Yu, Ho; Cheng-En Hsu; Yen-Nien Ting; Tsai, Ming-Tzu; Jui-Ting Hsu  VIAFID ORCID Logo 
Pages
1-6
Section
Research article
Publication year
2022
Publication date
2022
Publisher
BioMed Central
e-ISSN
1749-799X
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2630530710
Copyright
© 2022. This work is licensed under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.