1. Introduction
Q-switched solid-state lasers manifest the capability of generating nanosecond-level pulses and high peak power light sources. Since the concept of Q-switching was proposed by Gordon Gould in 1958 and the first Q-switched device of ruby laser was demonstrated by R.W. Hellwarth and F.J. McClung in 1961 [1,2], there has been a wide range of applications to the fields of lidar, rangefinder, medicine, material processing, holography, and nonlinear optics [3,4,5,6,7]. The technique is achieved by tuning the intracavity losses of the resonator and can be further classified into active Q-switched (AQS) lasers and passive Q-switched (PQS) lasers. The operation of AQS lasers requires active control devices such as spinning mirrors, and acousto-optic attenuators to modulate the cavity losses. For PQS lasers, the saturable absorbers like Cr4+:YAG for 1-μm lasers and V3+:YAG for 1.3-μm lasers can be used to modulate the cavity losses; the crystals of these material are much lighter in comparison with the active control devices. Thus, PQS lasers possess the advantages of compactness, reliability, and light weight in the cavity design. Owing to the concern of reliability and cost reduction, PQS had been evaluated to be more applicable for the space mission in GLAS and MLA of NASA [8,9].
Since the coupled rate equations of passive Q-switching were proposed [10], several models and approaches of optimization have been developed. Many works have shown the dependence of performance on several factors in the cavity, such as the selection of gain medium, initial transmission of the absorber, reflectivity of output coupler, and intracavity mode area [11,12,13,14,15]. The transverse intensities of the pump and laser might affect the performance as well. In the classic PQS model, the transverse mode of lasing was assumed to be uniform. However, the mode of a resonator is more like a Hermite-Gaussian distribution in real cases instead; thus, the influences of the spatial effect must be considered into the coupled rate equations.
In this article, we start from considering the lasing photon density to be a constant TEM00 Gaussian distribution along the optical direction and the pump density of a laser diode to be a top-hat distribution which diffracts along the optical direction, respectively, then incorporate the spatial dependence into the coupled rate equations for four-level PQS lasers. A transcendental equation of energy utilization is derived and approximated by an analytical function for calculating the output energy in the condition of no saturation in the gain crystal. To validate the developed spatially-dependent model, an end-pumped Nd:YVO4/Cr4+:YAG laser in a concave-plano cavity with output coupling reflectivity ranging from 10% to 92% was demonstrated, and two different pumping spot sizes of laser diode were adopted.
2. Modeling of Spatially Dependent PQS
The spatially-dependent lasing photon density and the population inversion density can be expressed as and where and are the spatially-normalized lasing and pumping photon densities, respectively, Φ and N are the number of the photon in the cavity and the number of the inversion population in the gain medium. For a single fundamental transverse TEM00 mode, the lasing photon density can be expressed as:
(1)
(2)
where is the length of the cavity, is the mode size in the gain crystal, is the beam waist of lasing, is the distance from the beam waist to the middle of the length of the gain crystal, and is the wavelength of lasing. We assume a constant beam size without diffraction here if the variation of the beam size in the gain medium is small enough and neglectable. For a fiber-coupled laser diode pumping, the population inversion density in the gain medium can be described as a top-hat distribution [16,17]:(3)
(4)
where δ is the absorption coefficient at the wavelength of the pumping source, is the length of the gain crystal, is the pump size in the gain crystal, H is the Heaviside step function, is the beam waist of the pump, is the wavelength of the pump, is the beam quality factor, is the refractive index of the gain crystal, and is the location of the focal plane of the pump. We introduce the spatial dependence stated above to the coupled rate equations of the four-level PQS laser. The coupled rate equations can be expressed as:(5)
(6)
(7)
where is the stimulated emission cross section of the gain medium, and are the ground-state and excited-state absorption cross section of the absorber, and are the number of the population in the ground state and in the excited state of the absorber, and are the lengths of the gain medium and the absorber respectively, is the volume of the absorber, is the output coupler reflectivity, is the dissipative loss in a roundtrip for a photon, is the roundtrip time for a photon in the cavity, is the inversion reduction factor, and and are the cross areas of lasing in the gain medium and in the absorber, respectively. According to the Lambert-Beer absorption law, the initial transmission of the absorber can be expressed as where . By integrating coupled rate equations with respect to the cavity volume, we have(8)
(9)
(10)
where the effective mode volume and the overlap efficiency S(11)
(12)
(13)
From Equation (8), with the condition that and , the initial population in the gain medium for generating a pulse is given by
(14)
Dividing the Equation (9) by the Equation (10) and integrating the result can give as a function of N, that is where . Dividing the rate Equation (8) by the rate Equation (9) and using the Equation (14) give
(15)
where(16)
and . The second derivative of Φ with respect to N can be expressed as(17)
The condition at is referred to as the second threshold for generating a stable pulse train in PQS lasers and can be given by where is the ratio of the initial inversion to the critical inversion. Integrating the equation of yields
(18)
The final inversion population can be determined by the condition :
(19)
Defining as the energy-utilization factor, the Equation (19) can be expressed as a transcendental equation:
(20)
In the condition of no gain saturation, the larger value of the means the more efficient population depleted in the saturable absorber, since cannot be directly solved as a function of and . A powerful expression derived from numerical fitting [18] for the energy utilization efficiency is given by
(21)
with . The transcendental equation quantitatively determines the characteristics of PQS. The output energy can be expressed as(22)
where is the lasing photon energy and is the effective mode area that is equal to . The transmittance of output coupler can be well approximated to be for high reflectivity. However, for those reflectivity lower than , it is more general and appropriate to use term for a wide range of reflectivity [18]. So far, we have completed the derivation of the output energy equation, and the term and S in the Equation (22) show the significance of considering the spatial effect. Without considering the effect, the energy factor term would be according to the previous works, where A is the mode size area in the gain medium.3. Experimental Setup and Theoretical Fitting
To verify the model, we set up a concave-plano cavity for the PQS operation, and the experimental scheme is shown in Figure 1. The pumping source was a fiber-coupled laser diode with a core diameter of 0.2 mm, which was able to offer a maximum power of 40 W at 808 nm. The pumping energy from the laser diode was configured as pulse-pumping with the frequency 10 kHz and duty cycle 30% for reducing the thermal lensing effect. The input side of the resonator was a concave mirror with a radius of curvature 30 mm, and its entrance facet was coated anti-reflection films at 808 nm with reflectivity less than 0.1%; the second facet was coated high-reflection films at 1064 nm with reflectivity 99.8% and transmittance 96% at 808 nm. Since our model was based on the condition of no gain saturation, we chose a low-doped concentration 0.3 at. % Nd:YVO4 as the gain medium to avoid saturation cases. Furthermore, the gain medium was an a-cut rectangular crystal with the dimensions of mm3, and both ends were coated anti-reflection films at 1064 nm with reflectivity less than 0.1%. The saturable absorbers were two Cr4+:YAG with the initial transmission and 50%, and both were rectangular crystals with the dimensions of mm3. The coating for the absorbers on both ends were the same as those of the gain medium. The gain medium and the saturable absorbers were wrapped with indium foils and mounted in copper holders that were connected to a water-cooling system keeping a temperature of 22 °C. The indium foils were used as high-efficiency heat sinks and filling material for tiny gaps between the crystal and the holder. Several flat mirrors with different reflectivity in a range from 10% to 92% were employed as the output couplers in this experiment. The distance in Figure 1 is the average distance from the output coupler to the gain crystal, and is the distance from the output coupler to the absorber.
The cavity length was set to be 29 mm to satisfy the criterion of the second threshold condition. By , the mode sizes of the two ends and the middle point in the gain crystal can be evaluated to be 0.180 mm2, 0.226 mm2, and 0.202 mm2 by substituting the mm, 22 mm, and 25mm, respectively. In modeling, we assumed the variation of the mode size in the crystal is small, then we ignored the effect of diffraction inside the gain crystal. Since the mode size in the middle is very close to the average mode size of the two ends, the approximation still holds in this case, so here we take mm2. By the experimental conditions that mm, the mode sizes in the saturable absorber can be obtained to be 0.046 mm. Adding the spontaneous emission coefficient of the gain crystal cm2 and the absorption coefficient of the ground state of the absorber cm2, the value of can be evaluated to be .
To investigate the influence of the spatial overlap effect on the output energy, two sets of focusing lens were adopted for different scales of beam expansion. Consequently, the average pump beam radii of 0.1 mm and 0.3 mm were reimaged into the gain crystal with 90% coupling efficiency, respectively. Given the following experimental parameters mm, mm, mm, nm, mm−1, , mm, the integration in Equation (13) can be evaluated numerically, then substitute the result into the Equations (11) and (12) to obtain the effective mode area and overlap efficiency mm2 and for the pump size mm, mm2, and for mm. Adding the inversion reduction factor , the output energy factor can be evaluated to be 52 μJ and 65 μJ. We will discuss the experimental results in the following section.
4. Results and Discussions
Figure 2 shows the plot of the average output power versus the input power in the pulse pumping scheme for and 50% with the output coupler . The results show a characteristic staircase structure where the plateaus indicate the locking effect at various simple frequency ratio. This phenomenon is consistent with the previous results [19].
Figure 3 shows a typical stable pulse train and the pulse width of a single pulse with time span division of 8 ms and 40 ns in the case of to the frequency-locked states of 1:1, i.e., frep = 10 KHz. The overall pulse-to-pulse amplitude fluctuation was found to be within 5%. The pulse duration was measured approximately to be 2 ns, as shown in Figure 3b. We also observed that there was no additional pulse or satellite pulse in every single Q-switched shot.
The output energy versus the output coupler reflectivity of the theoretical fitting and the experimental results with for two pumping foci are shown in Figure 4a. It can be clearly seen that the values of the curve of the case mm is higher than that of the case mm by a factor of 1.25 on average, which is exactly corresponding to the ratio of the energy factor 65 μJ to the energy factor 52 μJ. The results for other cases of is also shown in Figure 4b. The agreement of the theoretical calculations and the experimental data confirms the validity of the developed spatial model.
By the condition from Equation (12), the inversion population when the maximum photon number occurs is almost equal to in our cases with . Thus, the spatially dependent peak power of the PQS model can be evaluated by
(23)
and the space-related terms S and are included in the maximum photon number . The theoretical fitting and experimental results for the peak powers are shown in the Figure 5 for the cases mm and mm with . The peak powers for both pumping foci possess the similar pattern to that of the output energies owing to the spatial overlap effect too. From that, again, we see the good agreement between the theoretical fitting and experimental results.The above result and discussion are based on the ratios of the output frequency to the input frequency to be 1:1, which means that the input energy during one pulse-pumping cycle is enough to generate a corresponding lasing pulse. From the experimental results, we find that the model is still valid at various simple frequency ratios such as 1:2 and 2:3, which is the frequency locking effect as the first and second plateau shown in Figure 2. The time-domain pulse train is shown in Figure 6 for the case that mm, , with pulse-pumping frequency 10 kHz, and the output energy factor for the pump size is still around 52 μJ.
5. Conclusions
We have considered the spatial dependency of the laser and the pump into the coupled rate equations for an end-pumped four-level PQS laser in the condition of no gain saturation, then derived an expression for calculating the output energy. The experimental setup for verifying the model has been demonstrated. The good agreements of the theoretical fittings and experimental results on the output energy and the peak power have validated the accuracy of the model.
Conceptualization, Y.-C.T. and M.-X.H.; validation, Y.-C.T. and M.-X.H.; formal analysis, M.-X.H. and H.-C.L.; resources, Y.-C.T.; writing—original draft preparation, Y.-C.T.; writing—review and editing, H.-C.L. and Y.-T.Y.; supervision, H.-C.L. and Y.-T.Y. All authors have read and agreed to the published version of the manuscript.
This work was supported by the Ministry of Science and Technology of Taiwan (Contract No. 109-2119-M-009-015-MY3).
All of the data reported in the paper are presented in the main text. Any other data will be provided on request.
The authors declare no conflict of interest.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Figure 2. The plot of the output power versus the input power of two pumping foci with the initial transmission (a) 35% and (b) 50% of the saturable absorbers.
Figure 3. (a) A stable pulse train with fluctuation deviation less than 5%. (b) A single pulse of 2 nanoseconds generated by a PQS Nd:YVO4/Cr4+:YAG laser.
Figure 4. Theoretical fitting and experimental results of the output energy of the two pumping foci [Forumla omitted. See PDF.] mm and [Forumla omitted. See PDF.] mm with (a) [Forumla omitted. See PDF.] and (b) [Forumla omitted. See PDF.].
Figure 5. Theoretical fitting and experimental results of the peak power of the two pumping foci [Forumla omitted. See PDF.] mm and [Forumla omitted. See PDF.] mm with (a) [Forumla omitted. See PDF.] and (b) [Forumla omitted. See PDF.].
Figure 6. The pulse trains at frequency ratios (a) fout/fin = 1/2 and (b) fout/fin = 2/3 for the case [Forumla omitted. See PDF.] mm, [Forumla omitted. See PDF.], [Forumla omitted. See PDF.] with pulse-pumping frequency 10 kHz.
References
1. Taylor, N. LASER: The Inventor, the Nobel Laureate, and the Thirty-Year Patent War; Simon & Schuster: New York, NY, USA, 2002.
2. McClung, F.J.; Hellwarth, R.W. Giant optical pulsations from ruby. Appl. Opt.; 1962; 1, pp. 103-105. [DOI: https://dx.doi.org/10.1364/AO.1.S1.000103]
3. Zayhowski, J.J. Passively Q-switched Nd:YAG microchip lasers and applications. J. Alloys Compd.; 2000; 303, pp. 393-400. [DOI: https://dx.doi.org/10.1016/S0925-8388(00)00647-2]
4. De Young, R.J.; Barnes, N.P. Profiling atmospheric water vapor using a fiber laser lidar system. Appl. Opt.; 2010; 49, pp. 562-567. [DOI: https://dx.doi.org/10.1364/AO.49.000562] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/20119001]
5. Skorczakowski, M.; Swiderski, J.; Pichola, W.; Nyga, P.; Zajac, A.; Maciejewska, M.; Galecki, L.; Kasprzak, J.; Gross, S.; Heinrich, A. et al. Mid-infrared Q-switched Er:YAG laser for medical applications. Laser Phys. Lett.; 2010; 7, 498. [DOI: https://dx.doi.org/10.1002/lapl.201010019]
6. Meier, M.; Romano, V.; Feurer, T. Material processing with pulsed radially and azimuthally polarized laser radiation. Appl. Phys. A; 2007; 86, pp. 329-334. [DOI: https://dx.doi.org/10.1007/s00339-006-3784-9]
7. Harigel, G.; Baltay, C.; Bregman, M.; Hibbs, M.; Schaffer, A.; Bjelkhagen, H.; Hawkins, J.; Williams, W.; Nailor, P.; Michaels, R. et al. Pulse stretching in a Q-switched ruby laser for bubble chamber holography. Appl. Opt.; 1986; 25, pp. 4102-4110. [DOI: https://dx.doi.org/10.1364/AO.25.004102] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/18235751]
8. Krebs, D.J.; Novo-Gradac, A.M.; Li, S.X.; Lindauer, S.J.; Afzal, R.S.; Anthony, W.Y. Compact, passively Q-switched Nd: YAG laser for the MESSENGER mission to Mercury. Appl. Opt.; 2005; 44, pp. 1715-1718. [DOI: https://dx.doi.org/10.1364/AO.44.001715] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/15813276]
9. Afzal, R.S.; Anthony, W.Y.; Dallas, J.L.; Melak, A.; Lukemire, A.T.; Ramos-Izqueirdo, L.; Mamakos, W. The geoscience laser altimeter system (GLAS) laser transmitter. IEEE J. Sel. Topics Quantum Electron.; 2007; 13, pp. 511-536. [DOI: https://dx.doi.org/10.1109/JSTQE.2007.896051]
10. Szabo, A.; Stein, R.A. Theory of laser giant pulsing by a saturable absorber. J. Appl. Phys.; 1965; 36, pp. 1562-1566. [DOI: https://dx.doi.org/10.1063/1.1703087]
11. Degnan, J.J. Optimization of passively Q-switched lasers. IEEE J. Quantum Electron.; 1995; 31, pp. 1890-1901. [DOI: https://dx.doi.org/10.1109/3.469267]
12. Xiao, G.; Bass, M. A generalized model for passively Q-switched lasers including excited state absorption in the saturable absorber. IEEE J. Quantum Electron.; 1997; 33, pp. 41-44. [DOI: https://dx.doi.org/10.1109/3.554875]
13. Chen, Y.P.; Lan, Y.P.; Chang, H.L. Analytical model for design criteria of passively Q-switched lasers. IEEE J. Quantum Electron.; 2001; 37, pp. 462-468. [DOI: https://dx.doi.org/10.1109/3.910458]
14. Pavel, N.; Saikawa, J.; Kurimura, S.; Taira, T. High average power diode end-pumped composite Nd: YAG laser passively Q-switched by Cr4+: YAG saturable absorber. Jpn. J. Appl. Phys.; 2001; 40, 1253. [DOI: https://dx.doi.org/10.1143/JJAP.40.1253]
15. Tuan, P.H.; Chang, C.C.; Chang, F.L.; Lee, C.Y.; Sung, C.L.; Cho, C.Y.; Chen, Y.F.; Su, K.W. Modelling end-pumped passively Q-switched Nd-doped crystal lasers: Manifestation by a Nd: YVO 4/Cr 4+: YAG system with a concave-convex resonator. Opt. Express; 2017; 25, pp. 1710-1722. [DOI: https://dx.doi.org/10.1364/OE.25.001710] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/29519025]
16. Chen, Y.F. cw dual-wavelength operation of a diode-end-pumped Nd: YVO4 laser. Appl. Phys. B.; 2000; 70, pp. 475-478. [DOI: https://dx.doi.org/10.1007/s003400050847]
17. Huang, Y.J.; Tzeng, Y.S.; Tang, C.Y.; Chen, Y.F. Efficient dual-wavelength synchronously mode-locked picosecond laser operating on the 4 F 3/2→ 4 I 11/2 transition with compactly combined dual gain media. IEEE J. Sel. Top. Quantum Electron.; 2014; 21, pp. 56-62. [DOI: https://dx.doi.org/10.1109/JSTQE.2014.2324758]
18. Chen, Y.F.; Hsieh, M.X.; Tu, Y.C.; Lee, C.C.; Yu, Y.T.; Tsou, C.H.; Liang, H.C. Pedagogically fast model to evaluate and optimize passively Q-switched Nd-doped solid-state lasers. Opt. Lett.; 2021; 46, pp. 1588-1591. [DOI: https://dx.doi.org/10.1364/OL.422382] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33793494]
19. Chen, Y.F.; Chien, P.Y.; Lee, C.C.; Huang, K.F.; Liang, H.C. Timing jitter reduction of passively Q-switched solid-state lasers by coupling resonance between pumping and firing rates. Opt. Lett.; 2020; 45, pp. 2902-2905. [DOI: https://dx.doi.org/10.1364/OL.394613] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32412497]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
In this work, the spatial effects of the pumping and lasing were considered into the coupled rate equations of passively Q-switched lasers for the condition of no saturation in the gain medium. A transcendental equation of energy utilization was derived, and the solution to the equation was approximated by an analytic model developed in previous work. The reflection factor of output coupler with low reflectivity was modified for more accurate output energy evaluations. Experimentally, an end-pumped Nd:YVO4/Cr4+:YAG laser in a concave-plano cavity with output coupling reflectivity ranging from 10% to 92% was demonstrated, and two different pumping spot sizes of laser diode were adopted for validating the developed spatial model. The experimental results showed good agreements with the theoretical calculations.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details

1 Department of Electrophysics, National Yang Ming Chiao Tung University, 1001 Ta-Hsueh Rd., Hsinchu 30010, Taiwan;
2 Department of Optoelectronics and Materials Technology, National Taiwan Ocean University, Keelung 20224, Taiwan;