Full text

Turn on search term navigation

© 2022. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Purpose

The Leksell Gamma Knife Icon unit (IU) was introduced recently as an upgrade to the Perfexion unit (PU) at our Gamma Knife practice. In the current study, we sought mainly to characterize dosimetry and targeting accuracy of the IU treatment deliveries using both invasive frame and frameless treatment workflows.

Methods

Relative output factors were measured by delivering single‐shot 4, 8 and 16 mm radiation profiles in the manufacturer's acrylonitrile butadiene styrene spherical phantom in coronal and sagittal planes using EBT3 film. Resultant dosimetry was compared with the manufacturer's dose calculation and derived output factors were compared with the manufacturer's published value. Geometric consistency of stereotactic coordinates based on cone‐beam computed tomography (CBCT) versus the traditional conventional CT‐based method was characterized using a rigid phantom containing nine fiducial indicators over four separate trials. End‐to‐end (E2E) testing using EBT3 film was designed to evaluate both dosimetric and geometric accuracy for hypothetical framed and frameless workflows.

Results

Relative output factors as measured by the manufacturer were independently confirmed using EBT3 film measurements to within 2%. The mean 3D radial discrepancy in stereotactic space between CBCT and CT‐based definition over the sampled locations in our rigid geometry phantom was demonstrated to be between 0.40 mm and 0.56 mm over the set of trials, larger than prior reported values. E2E performed in 2D demonstrates sub‐mm (and typically < 0.5 mm) accuracy for framed and frameless workflows; geometric accuracy of framed treatments using CBCT‐defined stereotactic coordinates was shown to be slightly improved in comparison with those defined using conventional CT. Furthermore, in phantom, frameless workflows exhibited better accuracy than framed workflows for fractionated treatments, despite large magnitudes of introduced interfraction setup error. Accuracy of dosimetric delivery was confirmed in terms of qualitative comparisons of dose profiles and in terms of 2D gamma pass rates based on 1%/1 mm criteria.

Conclusion

The IU was commissioned for clinical use of frameless and framed treatment protocols. The present study outlines an extensive E2E methodology for confirmation of dosimetric and geometric treatment accuracy.

Details

Title
Characterization and commissioning of a Leksell Gamma Knife ICON system for framed and frameless stereotactic radiosurgery
Author
Yue‐Houng Hu 1 ; Hickling, Susannah V 1 ; Qian, Jing 1 ; Blackwell, C Robert 1 ; McLemore, Luke B 1 ; Tryggestad, Erik J 1 

 Department of Radiation Oncology, Division of Medical Physics, Mayo Clinic, Rochester, Minnesota 
Section
RADIATION ONCOLOGY PHYSICS
Publication year
2022
Publication date
Mar 2022
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2637295789
Copyright
© 2022. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.