1. Introduction
Keratoconus is a progressive corneal ectasia resulting from localized corneal thinning and steepening. Patients with this condition usually demonstrate irregular astigmatism, an increased aberration, and a deteriorated visual acuity at an age ranging from 9 to 28 years [1]. The prevalence of keratoconus is approximately 0.14%; however, its prognosis is poor if an effective treatment is not available before its deterioration [2]. At present, there are two main treatment methods for keratoconus. One is to slow the progression of the disease, such as corneal collagen cross-linking (CXL), and the other one is to improve vision, such as rigid gas-permeable (RGP) contact lens and intracorneal ring segment implantation (ICRS). Corneal graft is the traditional treatment for advanced keratoconus, which carries a risk of postsurgical infection, stromal opacification, and host-corneal rejection [3]. Additionally, approximately 53% of patients worldwide are unable to receive a corneal transplantation due to lack of corneal grafts [4].
Femtosecond laser has been developed for the correction of refractive errors with minimal invasiveness. It can create both a donor lenticule with expected specifications and a stromal pocket at an expected depth within the host cornea [5]. It has been suggested in some clinical studies that the lenticule from small incision lenticule extraction (SMILE) has a great potential for application in corneal surgery [6, 7]. Femtosecond laser-assisted lenticule implantation has already been used to correct hyperopia in humans [8, 9] and also provides a new surgical approach for the treatment of corneal ulcer [10, 11].
In a previous preliminary study, we have demonstrated that SFII has a faster postoperative recovery than penetrating keratoplasty (PKP) in patients with progressive keratoconus [12, 13]. This study aimed to assess the morphological changes and biomechanical stability of the host cornea in patients with progressive keratoconus over a period of 5 years after SFII.
2. Materials and Methods
2.1. Study Design
This prospective study was approved by the Ethics Committee of Hainan Eye Hospital at Zhongshan Ophthalmic Center (Sun Yat-sen University, China) and followed the ethical guidelines of the Declaration of Helsinki. A written informed consent was obtained from all patients after the nature of the study, and potential risks were explained (ethics acceptance number: 2016-007).
This prospective study included patients who were diagnosed as stage II to III progressive keratoconus based on the Amsler–Krumeich classification system in one of the eyes [14]. All of the eyes were free of acute inflammation and previous surgery in the anterior segment of the eyes. Inclusion criteria were those who had met the following conditions in the past year: (1) the maximum curvature value of corneal topographic map increased by 0.75 diopters (D) or more; (2) the astigmatism of optometry increased by 0.75D or more, or the spherical equivalent increased by 0.5D or more; (3) the corneal thickness was less than 480 μm; (4) keratoconus apex within the central 2 mm (central keratoconus); (5) physical intolerance to any type of contact lenses [15]. A total 11 eyes of 11 patients (male: 7, female: 4) with the age ranging from 18 to 28 years were recruited to the Hainan Eye Hospital at the ZhongShan Ophthalmic Center between March 2016 and December 2017.
2.2. Clinical Examinations
Ophthalmic examinations were performed preoperatively and at different time points postsurgically (1 week and 1, 3, 6, 12, 24, 36, and 60 months). The clinical measurements taken included uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA) and manifest refraction with a synthetical optometer (Nikon, Tokyo, Japan), assessment of ocular anterior segment by a slit-lamp microscope (Topcon, Tokyo, Japan), and corneal photography with a slit-lamp digital image processing system (Topcon, Tokyo, Japan). Corneal topography and anterior chamber evaluation were recorded using the Pentacam HR Scheimpflug camera (OCULUS Optikgeräte GmbH, Wetzler, Germany). Anterior segment optical coherence tomography (Topcon, Tokyo, Japan) and confocal microscopy (HRT III; Heidelberg Engineering GmbH, Heidelberg, Germany) were also performed. The biomechanically corrected intraocular pressure and biomechanical properties of the patients were measured using a Corvis ST system (OCULUS, Wetzlar, Germany).
2.3. Surgical Procedures
All donor tissues were collected from the eye bank of the Hainan Eye Hospital in accordance with local guidelines (Hainan Entry-Exit Inspection and Quarantine Bureau of the People’s Republic of China). Topical anaesthesia consisted of proparacaine hydrochloride eye drops (Alcaine; Alcon Laboratories, Fort Worth, TX). A myopic correction of −0.75 diopters (28 μm thickness) using a femtosecond laser system under SMILE treatment was performed on the recipient cornea, and the surgical incision was 2.5 mm long. The surgeon unfolded the lenticule with microforceps and created a “stromal pocket” at a depth of 160 μm from the cornea epithelium, followed by implantation of the lenticule into the “stromal pocket.” The surgical procedures were detailed in our previous studies [13, 16]. Following the surgery, all patients received tobramycin and dexamethasone (Alcon Laboratories, US) eye drops 4 times per day for 2 weeks and recombinant bovine basic fibroblast growth factor (Fusion Protein, Bausch and Lomb, Rochester, NY) eye gel 3 times per day for 6 months.
2.4. Statistical Analysis
For statistical purposes, UDVA and CDVA data were converted by minimum resolution (logMAR) according to the standard logarithmic visual acuity chart. Statistical analysis was conducted using SPSS 23.0 statistical software (Stanford University, USA). All data that met normal distribution by the Shapiro–Wilk test were expressed as mean ± SD. ANOVA (analysis of variance) with multiple comparisons between groups was compared using LSD pairwise comparison with
3. Results
3.1. Visual Acuity and Intraocular Pressure
All 11 patients returned on schedule for their postoperative examinations. The mean values of the manifest sphere went from −9.25 ± 0.47 D at the preoperative level to −8.17 ± 0.52 D at 60 months postsurgically (F = 2.744,
The UDVA was found not to have improved (Figure 1) at different time points postsurgically (1 week and 1, 3, 6, 12, 24, 36, 60 months). The CDVA showed improved significantly from 1.00 ± 0.19 to 0.48 ± 0.13 logMAR at 3 months postsurgically (
3.2. Slit-Lamp Microscopy
All operated eyes did not show corneal ulceration and corneal neovascularisation and exudates in the anterior chamber (Figure 2) throughout the entire follow-up period. At 60 months postsurgically, a scar line was observed at the periphery of the lenticule, while the central region remained transparent.
[figures omitted; refer to PDF]
3.3. Pentacam HR Scheimpflug Camera Scanning
There were no significant changes in anterior surface curvature K1, K2, posterior surface curvature K1, K2, anterior chamber (AC) depth, AC volume, and posterior surface elevation at different time points postsurgically (1 week and 1, 3, 6, 12, 24, 36, 60 months), as compared to the preoperative results (Table 1). The corneal densitometry (CD) increased within 1 week postsurgically, recovered to the preoperative baseline after 6 months, and remained constant over the following 5 years (F = 0.009,
Table 1
Postoperative changes in corneal morphology.
Parameters | Time | ||||||||||
Preop | 1wk Postoperatively | 1 mo Postoperatively | 3 mo Postoperatively | 6 mo Postoperatively | 12 mo Postoperatively | 24 mo Postoperatively | 36 mo Postoperatively | 60 mo Postoperatively | F | P60mo | |
Anterior, K1, D | 51.60 ± 1.79 | 52.37 ± 2.86 | 52.27 ± 2.63 | 51.16 ± 2.31 | 51.19 ± 2.33 | 51.14 ± 2.25 | 50.73 ± 2.13 | 50.37 ± 2.12 | 50.11 ± 2.06 | 0.715 | 0.677 |
Anterior, K2, D | 58.54 ± 2.47 | 59.56 ± 2.76 | 59.76 ± 2.78 | 59.04 ± 2.81 | 58.20 ± 2.62 | 58.40 ± 2.71 | 57.66 ± 2.65 | 58.01 ± 2.71 | 57.24 ± 2.71 | 1.372 | 0.233 |
Posterior, K1, D | −7.74 ± 0.43 | −8.16 ± 0.66 | −8.31 ± 0.61 | −7.94 ± 0.39 | −8.04 ± 0.53 | −8.04 ± 0.55 | −7.96 ± 0.52 | −8.04 ± 0.49 | −7.98 ± 0.50 | 1.377 | 0.231 |
Posterior, K2, D | −8.96 ± 0.48 | −9.39 ± 0.61 | −9.43 ± 0.55 | −9.19 ± 0.48 | −9.16 ± 0.52 | −9.18 ± 0.60 | −9.10 ± 0.58 | −9.21 ± 0.52 | −9.11 ± 0.55 | 1.470 | 0.193 |
Elevation, (Back), μm | 94.14 ± 13.67 | 117.00 ± 19.98 | 100.57 ± 14.29 | 106.28 ± 14.89 | 101.57 ± 15.28 | 99.29 ± 14.95 | 101.29 ± 13.09 | 99.28 ± 12.60 | 96.01 ± 12.96 | 1.776 | 0.105 |
AC volume, mm3 | 190.75 ± 11.03 | 184.50 ± 12.19 | 183.25 ± 12.24 | 185.00 ± 10.43 | 182.63 ± 11.47 | 182.88 ± 10.66 | 184.63 ± 12.45 | 186.63 ± 13.52 | 185.38 ± 13.13 | 0.752 | 0.646 |
AC depth, mm | 3.58 ± 0.07 | 3.33 ± 0.09 | 3.45 ± 0.10 | 3.50 ± 0.10 | 3.53 ± 0.10 | 3.51 ± 0.10 | 3.54 ± 0.13 | 3.59 ± 0.14 | 3.57 ± 0.13 | 2.79 | 0.10 |
Total CD, GSU | 15.98 ± 0.57 | 19.58 ± 0.89 | 18.98 ± 0.73 | 17.68 ± 0.68 | 17.10 ± 0.67 | 16.21 ± 0.45 | 15.40 ± 0.22 | 15.81 ± 0.39 | 15.87 ± 0.53 | 0.09 | 0.777 |
The anterior K1, DK2; posterior K1, DK2; and elevation back remained stable postoperatively. The
3.4. Corneal Biomechanical Observation
The SP-A1 of the cornea and CCT increased within 1 week postsurgically (FSPA1 = 16.666,
Table 2
Postoperative changes in corneal biomechanics.
Parameters | Time | ||||||||||
Preop | 1 wk Postoperatively | 1 mo Postoperatively | 3 mo Postoperatively | 6 mo Postoperatively | 12 mo Postoperatively | 24 mo Postoperatively | 36 mo Postoperatively | 60 mo Postoperatively | F | P60mo | |
CCT, μm | 434.75 ± 14.18 | 616.75 ± 34.26 | 622.88 ± 35.84 | 613.25 ± 34.83 | 607.50 ± 29.07 | 617.12 ± 28.03 | 620.13 ± 31.84 | 601.25 ± 28.56 | 610.88 ± 27.40 | 24.311 | 0.000 |
SP-A1, mmHg/mm | 42.73 ± 8.93 | 80.08 ± 8.94 | 89.09 ± 10.51 | 84.26 ± 9.28 | 83.95 ± 9.53 | 81.56 ± 7.65 | 76.75 ± 7.01 | 79.38 ± 7.26 | 79.83 ± 6.55 | 16.666 | 0.000 |
DA Ratio, mm | 6.13 ± 0.39 | 5.15 ± 0.24 | 5.40 ± 0.29 | 5.48 ± 029 | 5.58 ± 0.32 | 7.30 ± 2.01 | 6.77 ± 1.14 | 6.65 ± 0.72 | 6.35 ± 0.60 | 1.159 | 0.347 |
SSI | 0.63 ± 0.03 | 0.55 ± 0.03 | 0.53 ± 0.05 | 0.47 ± 0.04 | 0.46 ± 0.05 | 0.52 ± 0.05 | 0.53 ± 0.05 | 0.56 ± 0.06 | 0.60 ± 0.03 | 2.580 | 0.103 |
bIOP, mmHg | 11.37 ± 2.32 | 16.85 ± 3.14 | 12.75 ± 2.18 | 10.87 ± 2.31 | 11.19 ± 2.11 | 11.93 ± 1.97 | 12.03 ± 1.97 | 12.65 ± 1.89 | 12.53 ± 1.84 | 0.480 | 0.502 |
CCT, central corneal thickness; SP-A1, stiffness parameter at applanation 1; DA ratio, deformation amplitude ratio; SSI, stress-strain index, bIOP, biomechanically corrected intraocular pressure; bold entries indicate
3.5. AS-OCT Observation
The lenticule was transparent and well biocompatible within the host cornea at a depth of 160 μm throughout the follow-up in all 11 cases. Keratitis was not observed in any of the host cornea after surgery. The corneal thickness of the host increased compared with the preoperative level, and no obvious tissue reaction, such as infiltrates, ulcer, and neovascularisation, from the host cornea was observed (Figure 5).
[figure omitted; refer to PDF]3.6. Confocal Microscopy
At 1 week postsurgically, corneal edema and stromal cell activation were observed in the host cornea (Figure 6(a)) with bright reflected spots and spiky keratocytes along the lenticule-stromal interface (Figure 6(f)). These reactions gradually subsided over time (3 months postsurgically: Figures 6(b) and 6(g); 6 months postsurgically: figures 6(c) and 6(h); 24 months postsurgically: Figures 6(d) and 6(i)). At 60 months postsurgically, edema subsided and stromal cell remained stable (Figure 6(e)), but high reflectivity still appeared in stromal-implant interfaces (Figure 6(j)). Lenticule vascularization, stromal infiltration, and endothelial cell damage were not observed in the cornea throughout the postoperative period.
[figures omitted; refer to PDF]
4. Discussion
A successful treatment for keratoconus should be able to stabilize the cornea biologically and mechanically in order to inhibit the progression of corneal ectasias. The human corneal stroma is mainly composed of collagen lamellae, which accounts for approximately 90% of the overall thickness of the cornea [17]. The corneal strength derives from the meshwork of the collagen lamellae, which play a crucial role in the maintenance of normal corneal functions [18, 19]. Our previous animal study has confirmed that the surgically designed “stromal pocket” at a depth of 160 μm from the cornea epithelium facilitates the recovery of the surgical incision of 2.5 mm without damaging the biomechanical stability of the cornea [20, 21].
In this study, all operated eyes achieved a stable condition during the 5-year follow-up period and showed an improvement in CDVA (11 eyes) without progression of ectasia or host corneal rejection. The UDVA at 60 months postsurgically did not deteriorate as compared with the preoperative level, while SE and manifest refraction remained stable from the preoperative level (
In this study, bIOP increased rapidly within 1 week postsurgically and returned to preoperative levels at 1 month postsurgically. This short-term bIOP elevation could be related to topical use of corticosteroids, damage to the angle of the anterior chamber, suturing method, and peripheral anterior synechiae in the presence of inflammation [25]. The surgical incision in the present study was 2.5 mm long with no sutures applied. After ruling out surgical factors such as leakage from the surgical incision, angular injury, and suturing method as causes of ocular hypertension, the most likely cause for the elevated bIOP in this study was most likely due to the use of corticosteroids. This hypothesis was further supported by a recovery of the bIOP after the corticosteroid was ceased. In addition, we suspect that another reason may be the reduction of corneal edema and the decline of intraocular pressure caused by the decrease of corneal thickness. Smedowsk et al. [26] reported that there is a medium-strength correlation between bIOP and CCT.
Progression of keratoconus is usually accompanied with an increase in corneal densitometry [27, 28]. Jiménez-García et al. [29] reported that changes in corneal densitometry were consistent with the progression of keratoconus and are one of the indicators for monitoring the deterioration of keratoconus disease. In the present study, we found that the CD value of total corneal thickness increased within 1 week postsurgically, recovered to the preoperative level after 6 months, and remained constant thereafter. This short-term change was likely due to corneal edema after surgery. As the edema subsided, the CD value of total corneal thickness recovered to the preoperative level and remained constant from 6 months to 60 months postsurgically. Although the scarring was observed at the edge of the lenticule through the slit-lamp microscope at 60 months postsurgically, the stable CD value of the total corneal thickness indicates that the transparency of the central area of the lenticules was not significantly affected. During the follow-up period, corneal densitometry remained constant, which indicate the transparency of the cornea remains stable after SFII.
The monitoring of corneal biomechanics is one of the most important indicators in the analysis of changes in the course of keratoconus [30]. It has been shown that corneal thinning in patients with keratoconus reduces corneal stiffness, and this leads to a reduction in the structural integrity of the cornea, which in turn promotes further corneal thinning [31]. Linear regression analysis indicates that SP-A1 is highly accurate in the diagnosis of keratoconus and monitoring of the disease progress [32, 33]. SP-A1 can be used to quantify the resistance of the cornea to deformation, which is the ratio of the displacement between the pressure load on the cornea to the apex of the undeformed cornea and the deflection of the initial flattening [34]. The value of SP-A1 in normal eyes is between 89.32 and 148.95 mmHg/mm [35, 36], which is higher than in keratoconus eyes, ranging from 46.6 to 77.16 mmHg/mm [37, 38]. In the present study, the SP-A1 value was 42.73 ± 8.93 mmHg/mm before surgery. It reached a maximum value of 89.09 ± 10.51 mmHg/mm at 1 month postsurgically and 79.83 ± 6.55 mmHg/mm at 60 months postsurgically. These results indicate that the lenticule implanted increased the corneal strength.
Apart from SP-A1, the DA ratio is highly sensitive and specific in the diagnosis of keratoconus [39]. As corneal stiffness decreases, the less the resistance to deformation, the greater the DA ratio value [40]. We observed a slight decrease in DA ratio from 1 week to 60 months postsurgically, but there was no significant difference compared to the preoperative levels. Hassan et al. [41] found an increase in SP-A1 when looking at corneal biomechanical changes 2 years after corneal collagen cross-linking, but the DA ratio did not change significantly from the preoperative period.
In addition, to eliminate the influence of other factors on corneal biomechanical examination, we observed a new detection index: the SSI. Zhang et al. [42] reported that changes in SSI can be used as an assessment of the effectiveness of treatment in keratoconus. Unlike other stiffness parameters, the change of SSI is not affected by bIOP or CCT and can be used to evaluate the mechanical properties of corneal tissue accurately [43]. Zhao et al. [44] found that the SSI of patients with keratoconus was 0.64 ± 0.12, which was comparable to our preoperative SSI (0.63 ± 0.03). With the implantation of the lenticule, we also found a stable trend of SSI before and at 60 months postsurgically. This suggested that the corneal resistance to deformation had been improved.
To further understand the real-time changes in the recipient cornea, postoperative patients were observed regularly by AS-OCT and confocal microscopy. AS-OCT showed that the corneal thickness of the recipient increased 1 week postsurgically (Figure 5). Confocal microscopy documented that lenticule-stromal interfaces were characterized by high reflectivity with bright dots (Figure 6(f)), and edema, stromal cell activation was observed at the corneal stroma in the 1 week postsurgically (Figure 6(a)). Edema, stromal cell activation and reflectivity gradually decreased over time (3 months postsurgically: Figures 6(b) and 6(g); 6 months postsurgically: Figures 6(c) and 6(h); 24 months postsurgically: Figures 6(d) and 6(i)). At 60 months postsurgically, edema subsided and stromal cell remained stable (Figure 6(e)), but high reflectivity still appeared in stromal-implant interfaces (Figure 6(j)). No evidence of leukocyte infiltration at lenticule-stromal interface was observed during the follow-up. This observation is similar to those of studies by other scholars [45].
We evaluated the morphological changes and biomechanical stability of the host cornea after SFII and demonstrated the potential of this surgical technique for anterior corneal profile restoration. The major limitations of our study include the small sample size, and no effective grouping or comparison of patients with corneal thicknesses >400 μm and ≤400 μm was conducted to observe the different effects of lenticule implantation on patients with different thicknesses.
Our findings suggest that SFII is a safe surgical approach, and stromal lenticule can be well tolerated by the recipient cornea, with a steady state of corneal morphology and biomechanics. It could provide an effective therapeutic approach for patients with progressive keratoconus.
Acknowledgments
Funding support was provided by the National Natural Science Foundation of China (81870681), Hainan Province Clinical Medical Center, and Science and Technology Planning Project of Hainan Province (ZDYF2020151).
[1] Y. S. Rabinowitz, "Keratoconus," Survey of Ophthalmology, vol. 42 no. 4, pp. 297-319, DOI: 10.1016/s0039-6257(97)00119-7, 1998.
[2] H. Hashemi, S. Heydarian, E. Hooshmand, M. Saatchi, A. Yekta, M. Aghamirsalim, M. Valadkhan, M. Mortazavi, A. Hashemi, M. Khabazkhoob, "The prevalence and risk factors for keratoconus: a systematic review and meta-analysis," Cornea, vol. 39 no. 2, pp. 263-270, DOI: 10.1097/ico.0000000000002150, 2020.
[3] F. Arnalich-Montiel, J. L. A. D. Barrio, J. L. Alio, "Corneal surgery in keratoconus: which type, which technique, which outcomes?," Eye and Vision, vol. 3 no. 1,DOI: 10.1186/s40662-016-0033-y, 2016.
[4] P. Gain, R. Jullienne, Z. He, M. Aldossary, S. Acquart, F. Cognasse, G. Thuret, "Global survey of corneal transplantation and eye banking," JAMA Ophthalmology, vol. 134 no. 2, pp. 167-173, DOI: 10.1001/jamaophthalmol.2015.4776, 2016.
[5] D. Z. Reinstein, T. J. Archer, M. Gobbe, "Small incision lenticule extraction (SMILE) history, fundamentals of a new refractive surgery technique and clinical outcomes," Eye and Vision, vol. 1 no. 1,DOI: 10.1186/s40662-014-0003-1, 2014.
[6] M. Moshirfar, T. J. Shah, M. Masud, S. H. Linn, Y. Ronquillo, P. C. Hoopes, "Surgical options for retreatment after small-incision lenticule extraction: advantages and disadvantages," Journal of Cataract & Refractive Surgery, vol. 44 no. 11, pp. 1384-1389, DOI: 10.1016/j.jcrs.2018.07.047, 2018.
[7] I. B. Pedersen, A. Ivarsen, J. Hjortdal, "Three-year results of small incision lenticule extraction for high myopia: refractive outcomes and aberrations," Journal of Refractive Surgery, vol. 31 no. 11, pp. 719-724, DOI: 10.3928/1081597x-20150923-11, 2015.
[8] S. Jacob, D. A. Kumar, A. Agarwal, A. Agarwal, R. Aravind, A. I. Saijimol, "Preliminary evidence of successful near vision enhancement with a new technique: presbyopic allogenic refractive lenticule (PEARL) corneal inlay using a SMILE lenticule," Journal of Refractive Surgery, vol. 33 no. 4, pp. 224-229, DOI: 10.3928/1081597x-20170111-03, 2017.
[9] L. Sun, P. Yao, M. Li, Y. Shen, J. Zhao, X. Zhou, "The safety and predictability of implanting autologous lenticule obtained by SMILE for hyperopia," Journal of Refractive Surgery, vol. 31 no. 6, pp. 374-379, DOI: 10.3928/1081597x-20150521-03, 2015.
[10] Y. Jiang, Y. Li, X.-W. Liu, J. Xu, "A novel tectonic keratoplasty with femtosecond laser intrastromal lenticule for corneal ulcer and perforation," Chinese Medical Journal, vol. 129 no. 15, pp. 1817-1821, DOI: 10.4103/0366-6999.186639, 2016.
[11] V. Bhandari, S. Ganesh, S. Brar, R. Pandey, "Application of the SMILE-derived glued lenticule patch graft in microperforations and partial-thickness corneal defects," Cornea, vol. 35 no. 3, pp. 408-412, DOI: 10.1097/ico.0000000000000741, 2016.
[12] C. Zhang, H. Ding, M. He, L. Liu, L. Liu, G. Li, B. Niu, X. Zhong, "Comparison of early changes in ocular surface and inflammatory mediators between femtosecond lenticule extraction and small-incision lenticule extraction," PLoS One, vol. 11 no. 3,DOI: 10.1371/journal.pone.0149503, 2016.
[13] H. Jin, M. He, H. Liu, X. Zhong, J. Wu, L. Liu, H. Ding, C. Zhang, X. Zhong, "Small-incision femtosecond laser-assisted intracorneal concave lenticule implantation in patients with keratoconus," Cornea, vol. 38 no. 4, pp. 446-453, DOI: 10.1097/ico.0000000000001877, 2019.
[14] J. H. Krumeich, J. Daniel, A. Knülle, "Live-epikeratophakia for keratoconus," Journal of Cataract & Refractive Surgery, vol. 24 no. 4, pp. 456-463, DOI: 10.1016/s0886-3350(98)80284-8, 1998.
[15] R. Ishii, K. Kamiya, A. Igarashi, K. Shimizu, Y. Utsumi, T. Kumanomido, "Correlation of corneal elevation with severity of keratoconus by means of anterior and posterior topographic analysis," Cornea, vol. 31 no. 3, pp. 253-258, DOI: 10.1097/ico.0b013e31823d1ee0, 2012.
[16] H. Jin, L. Liu, H. Ding, M. He, C. Zhang, X. Zhong, "Comparison of femtosecond laser-assisted corneal intrastromal xenotransplantation and the allotransplantation in rhesus monkeys," BMC Ophthalmology, vol. 17 no. 1,DOI: 10.1186/s12886-017-0595-z, 2017.
[17] Y. Qazi, G. Wong, B. Monson, J. Stringham, B. K. Ambati, "Corneal transparency: genesis, maintenance and dysfunction," Brain Research Bulletin, vol. 81 no. 3, pp. 198-210, DOI: 10.1016/j.brainresbull.2009.05.019, 2010.
[18] H. R. Vellara, D. V. Patel, "Biomechanical properties of the keratoconic cornea: a review," Clinical and Experimental Optometry, vol. 98 no. 1, pp. 31-38, DOI: 10.1111/cxo.12211, 2015.
[19] Y. Li, Z. Xu, Q. Liu, Y. Wang, K. Lin, J. Xia, S. Chen, L. Hu, "Relationship between corneal biomechanical parameters and corneal sublayer thickness measured by Corvis ST and UHR-OCT in keratoconus and normal eyes," Eye and Vision, vol. 8 no. 1,DOI: 10.1186/s40662-020-00225-z, 2021.
[20] M. He, H. Jin, H. He, H. Ding, W. Wang, L. Liu, C. Zhang, X. Zhong, "Femtosecond laser-assisted small incision endokeratophakia using a xenogeneic lenticule in rhesus monkeys," Cornea, vol. 37 no. 3, pp. 354-361, DOI: 10.1097/ico.0000000000001470, 2018.
[21] M. He, W. Wang, H. Ding, X. Zhong, "Comparison of two cap thickness in small incision lenticule extraction: 100 μ m versus 160 μ m," PLoS One, vol. 11 no. 9,DOI: 10.1371/journal.pone.0163259, 2016.
[22] L. Mastropasqua, M. Nubile, N. Salgari, R. Mastropasqua, "Femtosecond laser-assisted stromal lenticule addition keratoplasty for the treatment of advanced keratoconus: a preliminary study," Journal of Refractive Surgery, vol. 34 no. 1, pp. 36-44, DOI: 10.3928/1081597x-20171004-04, 2018.
[23] A. Vega-Estrada, J. L. Alio, "Keratoconus corneal posterior surface characterization according to the degree of visual limitation," Cornea, vol. 38 no. 6, pp. 730-736, DOI: 10.1097/ico.0000000000001909, 2019.
[24] S. A. Greenstein, D. Chung, L. Rosato, J. D. Gelles, P. S. Hersh, "Corneal higher-order aberrations after crosslinking and intrastromal corneal ring segments for keratoconus," Journal of Cataract & Refractive Surgery, vol. 46 no. 7, pp. 979-985, DOI: 10.1097/j.jcrs.0000000000000209, 2020.
[25] Y. B. A. Eta, O. Tomkins-Netzer, M. Mimouni, S. H. Azzam, R. S. Mashour, "Predicting factors of ocular hypertension following keratoplasty: indications versus the procedure," European Journal of Ophthalmology, vol. 31 no. 4, pp. 1749-1753, DOI: 10.1177/1120672120948757, 2021.
[26] A. Smedowski, B. Weglarz, D. Tarnawska, K. Kaarniranta, E. Wylegala, "Comparison of three intraocular pressure measurement methods including biomechanical properties of the cornea," Investigative Opthalmology & Visual Science, vol. 55 no. 2, pp. 666-673, DOI: 10.1167/iovs.13-13172, 2014.
[27] B. Lopes, I. Ramos, R. Ambrósio, "Corneal densitometry in keratoconus," Cornea, vol. 33 no. 12, pp. 1282-1286, DOI: 10.1097/ico.0000000000000266, 2014.
[28] M. A. Anayol, M. A. Sekeroglu, B. B. Ceran, M. Dogan, S. Gunaydin, P. Yilmazbas, "Quantitative assessment of corneal clarity in keratoconus: a case control study of corneal densitometry," European Journal of Ophthalmology, vol. 26 no. 1, pp. 18-23, DOI: 10.5301/ejo.5000644, 2016.
[29] M. Jiménez-García, S. Ní Dhubhghaill, A. Consejo, S. Hershko, C. Koppen, J. J. Rozema, "Scheimpflug densitometry in keratoconus: a new method of visualizing the cone," Cornea, vol. 40 no. 2, pp. 194-202, DOI: 10.1097/ico.0000000000002458, 2021.
[30] S. Kling, F. Hafezi, "Corneal biomechanics-a review," Ophthalmic and Physiological Optics, vol. 37 no. 3, pp. 240-252, DOI: 10.1111/opo.12345, 2017.
[31] C. J. Roberts, W. J. Dupps, "Biomechanics of corneal ectasia and biomechanical treatments," Journal of Cataract & Refractive Surgery, vol. 40 no. 6, pp. 991-998, DOI: 10.1016/j.jcrs.2014.04.013, 2014.
[32] R. Vinciguerra, R. Ambrósio, A. Elsheikh, C. J. Roberts, B. Lopes, E. Morenghi, C. Azzolini, P. Vinciguerra, "Detection of keratoconus with a new biomechanical index," Journal of Refractive Surgery, vol. 32 no. 12, pp. 803-810, DOI: 10.3928/1081597x-20160629-01, 2016.
[33] R. Ambrósio, B. T. Lopes, F. Faria-Correia, M. Q. Salomão, J. Bühren, C. J. Roberts, A. Elsheikh, R. Vinciguerra, P. Vinciguerra, "Integration of scheimpflug-based corneal tomography and biomechanical assessments for enhancing ectasia detection," Journal of Refractive Surgery, vol. 33 no. 7, pp. 434-443, DOI: 10.3928/1081597x-20170426-02, 2017.
[34] C. J. Roberts, A. M. Mahmoud, J. P. Bons, A. Hossain, A. Elsheikh, R. Vinciguerra, P. Vinciguerra, R. Ambrósio, "Introduction of two novel stiffness parameters and interpretation of air puff-induced biomechanical deformation parameters with a dynamic Scheimpflug analyzer," Journal of Refractive Surgery, vol. 33 no. 4, pp. 266-273, DOI: 10.3928/1081597x-20161221-03, 2017.
[35] M. Wang, Y. Zhang, W. Wu, J. A. Young, K. M. Hatch, R. P. Ii, T. Elze, Y. Wang, "Predicting refractive outcome of small incision lenticule extraction for myopia using corneal properties," Translational Vision Science & Technology, vol. 7 no. 5,DOI: 10.1167/tvst.7.5.11, 2018.
[36] J. Fernández, M. Rodríguez-Vallejo, J. Martínez, A. Tauste, P. Salvestrini, D. P. Piñero, "New parameters for evaluating corneal biomechanics and intraocular pressure after small-incision lenticule extraction by Scheimpflug-based dynamic tonometry," Journal of Cataract & Refractive Surgery, vol. 43 no. 6, pp. 803-811, DOI: 10.1016/j.jcrs.2017.03.035, 2017.
[37] P. Kataria, P. Padmanabhan, A. Gopalakrishnan, V. Padmanaban, S. Mahadik, R. Ambrósio, "Accuracy of scheimpflug-derived corneal biomechanical and tomographic indices for detecting subclinical and mild keratectasia in a South Asian population," Journal of Cataract & Refractive Surgery, vol. 45 no. 3, pp. 328-336, DOI: 10.1016/j.jcrs.2018.10.030, 2019.
[38] R. Vinciguerra, V. Romano, E. M. Arbabi, M. Brunner, C. E. Willoughby, M. Batterbury, S. B. Kaye, "In vivo early corneal biomechanical changes after corneal cross-linking in patients with progressive keratoconus," Journal of Refractive Surgery, vol. 33 no. 12, pp. 840-846, DOI: 10.3928/1081597x-20170922-02, 2017.
[39] M.-R. Sedaghat, H. Momeni-Moghaddam, R. Ambrósio, H.-R. Heidari, N. Maddah, Z. Danesh, F. Sabzi, "Diagnostic ability of corneal shape and biomechanical parameters for detecting frank keratoconus," Cornea, vol. 37 no. 8, pp. 1025-1034, DOI: 10.1097/ico.0000000000001639, 2018.
[40] M. A. Henriquez, M. Hadid, L. Izquierdo, "A systematic review of subclinical keratoconus and forme fruste keratoconus," Journal of Refractive Surgery, vol. 36 no. 4, pp. 270-279, DOI: 10.3928/1081597x-20200212-03, 2020.
[41] H. Hashemi, R. Ambrósio, R. Vinciguerra, P. Vinciguerra, C. J. Roberts, R. Ghaffari, S. Asgari, "Two-year changes in corneal stiffness parameters after accelerated corneal cross-linking," Journal of Biomechanics, vol. 93, pp. 209-212, DOI: 10.1016/j.jbiomech.2019.06.011, 2019.
[42] H. Zhang, A. Eliasy, B. Lopes, A. Abass, R. Vinciguerra, P. Vinciguerra, R. Ambrósio, C. J. Roberts, A. Elsheikh, "Stress-strain index map: a new way to represent corneal material stiffness," Frontiers in Bioengineering and Biotechnology, vol. 9,DOI: 10.3389/fbioe.2021.640434, 2021.
[43] A. Eliasy, K.-J. Chen, R. Vinciguerra, B. T. Lopes, A. Abass, P. Vinciguerra, R. Ambrósio, C. J. Roberts, A. Elsheikh, "Determination of corneal biomechanical behavior in-vivo for healthy eyes using CorVis ST tonometry: stress-strain index," Frontiers in Bioengineering and Biotechnology, vol. 7,DOI: 10.3389/fbioe.2019.00105, 2019.
[44] W. Zhao, Y. Shen, W. Jian, J. Shang, V. Jhanji, A. Aruma, X. Zhou, "Comparison of corneal biomechanical properties between post-LASIK ectasia and primary keratoconus," Journal of Ophthalmology, vol. 2020,DOI: 10.1155/2020/5291485, 2020.
[45] L. Mastropasqua, N. Salgari, E. D’Ugo, M. Lanzini, J. L. Alió Del Barrio, J. L. Alió, "“In vivo confocal microscopy of stromal lenticule addition keratoplasty for advanced keratoconus," Journal of Refractive Surgery, vol. 36 no. 8, pp. 544-550, DOI: 10.3928/1081597x-20200527-01, 2020.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2022 Qi Wei et al. This work is licensed under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Purpose. The purpose of this study was to evaluate the long-term prognosis of small-incision femtosecond laser-assisted intracorneal concave lenticule implantation (SFII) in correction of human keratoconus. Methods. This was a prospective study for 11 patients who received SFII after being diagnosed as progressive keratoconus based on the Amsler–Krumeich classification system. Clinical assessment was performed for all the patients prior to and postsurgically at different time points for 5 years. These included uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), biomechanically corrected intraocular pressure (bIOP), corneal topography, anterior segment optical coherence tomography (AS-OCT), confocal microscopy, and biomechanical assessment with Corvis ST. Results. Comparison of preoperative and 60-month postoperative UDVA and CDVA (
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details










1 State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Guangdong Provincial Clinical Research Center for Ocular Diseases, Guangzhou 510060, China
2 Hainan Eye Hospital and Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, Haikou 570300, China
3 Affiliated Hospital of Guilin Medical University, Guilin Medical University, Guilin 541000, China
4 State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Guangdong Provincial Clinical Research Center for Ocular Diseases, Guangzhou 510060, China; Hainan Eye Hospital and Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, Haikou 570300, China