© 2022 Masuda et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

Intracranial stents are used to treat aneurysms by diverting the blood flow from entering into the aneurysmal dome. Although delayed rupture is rare, clinical outcomes are extremely poor in such cases. Hemodynamics after stent deployment may be related to delayed rupture and a better understanding of the basic characteristics of pressure changes resulting from stent deployment is needed; therefore, this study investigated the relationships between hemodynamics in aneurysms of different sizes treated using stents of different wire mesh densities.

Methods

Using computational fluid dynamics analysis, parameters related to velocity, volume flow rate, pressure, and residual volume inside the aneurysm were evaluated in digital models of 5 basic aneurysms of differing sizes (Small, Medium, Medium-Large, Large, and Giant) and using 6 different types of stent (varying number of wires, stent pitch and wire mesh density) for each aneurysm.

Results

Regardless of the aneurysm size, the velocity inside the aneurysm and the volume flow rate into the aneurysm were observed to continuously decrease up to 89.2% and 78.1%, respectively, with increasing stent mesh density. In terms of pressure, for giant aneurysms, the pressure on the aneurysmal surface elevated to 10.3%, then decreased to 5.1% with increasing stent mesh density. However, in smaller aneurysms, this pressure continuously decreased with increasing stent mesh density. The flow-diverting effect of the stents was limited when a stent with low mesh density (under 20%) was used with a giant aneurysm.

Conclusions

The present results indicate that the selection of appropriate stents according to aneurysm size may contribute to reduced risks of hemodynamic alternations related to stent deployment, which could reduce the incidence of delayed rupture.

Details

Title
Effects of different stent wire mesh densities on hemodynamics in aneurysms of different sizes
Author
Masuda, Shunsuke; Contributed equally to this work with: Shunsuke Masuda; Soichiro Fujimura Soichiro Fujimura; Soichiro Fujimura Hiroyuki Takao  VIAFID ORCID Logo  ; Takeshita, Kohei; Suzuki, Takashi  VIAFID ORCID Logo  ; Uchiyama, Yuya  VIAFID ORCID Logo  ; Karagiozov, Kostadin; Ishibashi, Toshihiro; Fukudome, Koji; Yamamoto, Makoto  VIAFID ORCID Logo  ; Murayama, Yuichi
First page
e0269675
Section
Research Article
Publication year
2022
Publication date
Jun 2022
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2687693651
Copyright
© 2022 Masuda et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.