© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background and Objectives: Fracture healing is currently assessed through qualitative evaluation of radiographic images, which is highly subjective in nature. Radiographs can only provide snapshots in time, which are limited due to logistics and radiation exposure. We recently proposed assessing the bone healing status through continuous monitoring of the implant load, utilizing an implanted sensor system, the Fracture Monitor. The device telemetrically transmits statistically derived implant parameters via the patient’s mobile phone to assist physicians in diagnostics and treatment decision-making. This preclinical study aims to systematically investigate the device safety and performance in an animal setting. Materials and Methods: Mid-shaft tibial osteotomies of different sizes (0.6–30 mm) were created in eleven Swiss mountain sheep. The bones were stabilized with either a conventional Titanium or stainless-steel locking plate equipped with a Fracture Monitor. Data were continuously collected over the device’s lifetime. Conventional radiographs and clinical CT scans were taken longitudinally over the study period. The radiographs were systematically scored and CTs were evaluated for normalized bone volume in the defect. The animals were euthanized after 9 months. The sensor output was correlated with the radiologic parameters. Tissue samples from the device location were histologically examined. Results: The sensors functioned autonomously for 6.5–8.4 months until energy depletion. No macroscopic or microscopic adverse effects from device implantation were observed. The relative implant loads at 4 and 8 weeks post-operation correlated significantly with the radiographic scores and with the normalized bone volume metric. Conclusions: Continuous implant load monitoring appears as a relevant approach to support and objectify fracture healing assessments and carries a strong potential to enable patient-tailored rehabilitation in the future.

Details

Title
Continuous Implant Load Monitoring to Assess Bone Healing Status—Evidence from Animal Testing
Author
Windolf, Markus  VIAFID ORCID Logo  ; Varjas, Viktor; Gehweiler, Dominic; Schwyn, Ronald; Arens, Daniel  VIAFID ORCID Logo  ; Constant, Caroline; Zeiter, Stephan  VIAFID ORCID Logo  ; Richards, Robert Geoff  VIAFID ORCID Logo  ; Ernst, Manuela
First page
858
Publication year
2022
Publication date
2022
Publisher
MDPI AG
ISSN
1010660X
e-ISSN
16489144
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2694026046
Copyright
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.