1 Introduction
Long-term ground-based observations of atmospheric composition are essential for monitoring the evolution of the Earth–atmosphere system. Among current atmospheric measurement techniques, FTIR (Fourier transform infrared) spectrometry has an outstanding importance for climate research, since most atmospheric molecules interact with solar electromagnetic radiation in the infrared spectral region. By analysing the measured solar absorption spectra, this technique can provide atmospheric concentrations of many different gases simultaneously and with high precision (e.g. ).
Within NDACC (Network for the Detection of Atmospheric Composition Change,
One of the factors that is most dependent on each NDACC FTIR site is the treatment of the spectrometer response through the instrumental line shape (ILS) function (e.g. ). A precise knowledge of the ILS is essential to properly characterize the instrument performance, since the ILS affects the absorption line shape on which the retrieved information is based on. This is critical for estimating the vertical distribution of an absorber because such retrievals largely rely on the shape of the absorption lines (pressure broadening effect). Therefore, uncertainties in the ILS can also affect the quality of the total column (TC) amounts. Furthermore, the temporal stability of the ILS is an important requisite for precise gas retrievals .
Nowadays, different approaches are used to deal with the ILS characterization.
-
In the first approach, ILS is assumed to be ideal. However, in cases where there are misalignments in the FTIR spectrometer, a considerable systematic error may be introduced on the gas retrievals .
-
In the second approach, ILS is retrieved simultaneously with the gas vertical distribution from the measured solar absorption spectra. This is a superior strategy when compared to the first approach (e.g. ); however, part of the actual gas variability may be wrongly mapped into changes of the ILS, since the ILS and the absorber profiles have similar effect on the absorption line shapes (i.e. changing the shape and width of the line). In addition, overlapping lines (i.e. due to interfering species) may introduce asymmetry in the absorption lines that may be indistinguishable from an ILS phase deviation .
-
In the third approach, ILS is monitored through independent, regular, and calibrated low-pressure gas cell measurements (; Fig. 5 of ), whereby both the independence of FTIR gas retrievals and the instrumental characterization are ensured.
The stratospheric gases are more sensitive to the ILS treatment and its temporal behaviour than the tropospheric ones, since the full width at half maximum of their sharp absorption lines (absorptions taking place at low pressure) and of ILS have similar magnitudes . Given its key role in atmospheric chemistry, the ILS effects are of special relevance for the analysis of ozone (O) (e.g. ). Recently, have documented that a typical ILS degradation of 10 % may produce changes in the O TCs by 2 %, but the resulting disturbances of the vertical profile are considerably larger (between 20 %) and dependent on altitude. These values become especially important when compared to the rather small signals of O recovery obtained from long-term observations or projected from chemistry climate models. As summarized in the latest WMO/UNEP (World Meteorological Organization/United Nations Environment Programme) report , no significant trend has been detected in global (60 S–60 N) O TCs over the 1997–2016 period, and outside the polar regions only upper stratospheric O has been found to increase significantly by 0.1 % yr–0.3 % yr since 2000. Consistent estimates are predicted by climate models (e.g. ), pointing to a vertical stratification of the O recovery in response to the combined effects of an acceleration of Brewer–Dobson circulation, the stratospheric cooling induced by increasing greenhouse gases concentrations, and the levelling off of anthropogenic O depleting substances.
Within the FTIR community considerable efforts have been made toward the development, optimization, and validation of retrieval strategies for atmospheric O monitoring (e.g. ). These works were mainly focused on investigating the optimal selection of the O spectral absorption lines or the inversion settings used, such as the a priori information, the different constraints, or the inclusion of an additional atmospheric temperature profile retrieval. Nonetheless, the influence of instrumental characterization has not been addressed in detail yet. This is the focus of this paper, where the impact of the different approaches for characterizing the ILS (listed above as approaches 1–3) are examined using FTIR O products as an example. The study has been performed at the subtropical Izaña Observatory (IZO), where since 1999 ground-based FTIR observations have been carried out coincidentally with other independent O measurement techniques. In addition, the ILS function of the IZO FTIR spectrometers has been routinely monitored by means of independent gas cell measurements since 1999. These two facts make IZO a unique place for developing and documenting the reliability of new O retrieval strategies from ground-based FTIR spectrometry. In this context, this paper is structured as follows: Sect. describes the Izaña Observatory and its O programme, while Sect. presents the IZO FTIR observations, describing the monitoring of the ILS function, the O and ILS retrieval strategies used in this work, and their theoretical characterization in terms of vertical sensitivity and expected uncertainties. Section addresses the comparison of the ILS time series determined from gas cell measurements and those simultaneously retrieved with the O concentrations from the measured solar absorption spectra and also describes how the differences between the ILS approaches are transferred to the FTIR O products. Section assesses the impact of the different ILS treatments on the FTIR O products by comparing them to independent datasets (Brewer TC observations and electrochemical concentration cell, ECC, sondes). Finally, Sect. summarizes the main results and conclusions drawn from this work.
2 Izaña Observatory and its ozone programme
The Izaña Observatory (IZO) is a subtropical high-mountain station managed by the Izaña Atmospheric Research Center (IARC,
2.1 FTIR programme
Within the IZO's research activities, the FTIR programme was established in 1999 in the framework of a collaboration between the AEMet-IARC and the KIT (Karlsruhe Institute of Technology), with the main goals of the long-term monitoring of atmospheric gas composition and the validation of space-based observations and climate models. Since then two Bruker high-resolution FTIR systems have been operated at IZO contributing to NDACC: an IFS 120M from 1999 to 2005 and an IFS 120/5HR from 2005 onward.
The FTIR O measurements within NDACC are retrieved from the measured solar absorption spectra in the 990–1015 cm spectral region by using a potassium bromide (KBr) beam splitter and a cooled mercury cadmium telluride (MCT) detector. To resolve the narrow O absorption lines, the solar spectra have been acquired at the high spectral resolution of 0.0036 cm (250 cm of maximum optical path difference, OPD, OPD) until April 2000 and 0.005 cm (OPD cm) from that point on. The IFS 120M's field-of-view (FOV) angle has varied between 0.17 and 0.29 depending on the measurement period, while for the IFS 120/5HR it has been always limited to 0.2, both of which are considerably lower than the solar diameter of 0.5. For this study, the 20-year O measurements taken from 1999 to 2018 have been used. Refer to for a detailed description of the FTIR spectrometry activities at IZO.
2.2 Brewer and ECC sonde programmes
At IZO O TC observations have been also continuously taken by Brewer spectrometers since 1991. In 2001 the IZO Brewer activities were accepted by NDACC, and 2 years later the RBCC-E (Regional Brewer Calibration Center Europe,
Although the full uncertainty budget is still in development, provisional results indicate that the IZO RBCC-E reference instruments can provide O TCs with a total uncertainty (standard uncertainty, ) between 1.2 % and 1.5 % . For the current study, the O TCs of the permanent instrument Brewer#157 have been used, which were computed following the data processing of the EUBREWNET network .
The latest extension of the IZO O programme is the uptake of regular O sonde observations, which were initiated in November 1992 and included within NDACC in 2001. O soundings were performed once a week from the Santa Cruz Station (30 km north-east of IZO, 36 m a.s.l.) from 1992 until 2011, when they were moved to a launch site at the Botanic Observatory (13 km north of IZO, 114 m a.s.l.). The O sounding is based on ECC that senses O as it reacts with a dilute solution of potassium iodide (KI) to produce an electrical current proportional to the atmospheric O concentration . The O sonde model SPC-5A was used until September 1997, after which it was updated to the SPC-6A model with a cathode sensing solution type SST1.0 (1.0 % KI and full pH buffer). The sounding provides O (mPa) profiles, from the ground to the burst level (generally between 30 and 35 km), with a resolution of 0.01 mPa and accuracy of 5 %–15 % in the troposphere and 5 % in the stratosphere .
Note that for the purposes of this paper both Brewer and ECC sonde databases fully cover the entire FTIR 1999–2018 period.
Although they are not used in this work, the IZO O programme also includes DOAS (differential optical absorption spectroscopy) observations, performed within NDACC since 1999, and ground-level O records, taken in the framework of the WMO/GAW programme since 1987. More details about these measurement techniques are given in and .
3 FTIR ILS and ozone observations
3.1 Monitoring of the ILS function
The ILS function is the Fourier transform of the weighting applied to the interferogram . In the case of ideal instruments, the ILS is affected only by modulation loss that is due to the self-apodization of the interferometer (accepting a finite FOV) and is symmetric . From the physical viewpoint, interferometric misalignments, deviations from the desired circular interferometric FOV, OPD-dependent vignetting effects, and mismatch between the wave fronts of the reference laser wave front and the infrared beam are the main drivers of ILS imperfections. For real instruments, the ILS is equivalent to a complex modulation efficiency (ME) in the interferogram. The phase-corrected interferogram generated by a spectral line is of the following form :
1 where is the interferogram (with as the mirror displacement), is the wavenumber, MEA is the modulation efficiency amplitude, and PE is the modulation efficiency phase error. Note that the MEA and PE correspond to the real and imaginary part of the complex ME, respectively.
Figure 1
Example of the (a) normalized MEA and (b) PE (rad) retrievals as a function of the OPD for the IFS 120M (NO cell measurement taken on 3 July 2003) and for the IFS 120/5HR (NO cell measurement performed on 20 March 2013). (c, d) Averaging kernel rows of the MEA and PE retrievals for the cell example of the IFS 120M. Panels (e, f) are the same as (c) and (d) but for the IFS 120/5HR. The degrees of freedom for signal (DOFS) are also shown. (g) Time series of the normalized MEA and PE values at four OPD (38, 85, 133 and 180 ) between 1999 and 2018. Data points represent individual NO cell measurements, and solid lines depict the smoothed MEA and PE curves. The solid grey arrows indicate punctual interventions made regarding the IZO FTIR instruments: the changes in the field stop instruments between 1999 and 2004, there is a switch from the IFS 120M to the IFS 120/5HR in January 2005, optical re-alignments are performed in June 2008 and February 2013, and internal laser replacements are performed in August 2016 and June 2017. (g) Adapted from .
[Figure omitted. See PDF]
At IZO much effort has been made toward the instrumental characterization of the two FTIR spectrometers. Since 1999 the ILS function has been routinely monitored about every 2 months using low-pressure NO cell measurements and the software LINEFIT (v14.5). The LINEFIT package allows the deviation of the measured ILS from the ideal one to be determined through estimating the MEA and PE parameters . Because the measurement process of an FTIR spectrometer is performed in the interferogram domain, this parameterization refers to the interferogram, LINEFIT uses 20 equidistant grid points up to OPD and assumes a smooth variation of MEA and PE along OPD. MEA (normalized to unity at zero path difference, ZPD) is a measure of ILS width, and a decline of MEA indicates a broader ILS, while a rise indicates a narrower ILS with stronger sidelobes. A curving PE indicates ILS asymmetry (while a linear rise is equivalent to a spectral shift of the ILS but does not indicate a distortion of its shape).
The ILS parameters used in the FTIR O retrievals at IZO are estimated from the NO cell measurements using two broad micro-windows, combining saturated and unsaturated NO absorbing lines between 1235.0–1279.5 and 1291.8–1301.9 cm, and following the approach suggested by . In Hase's approach, the MEA at zero OPD (ZPD) is kept fixed to unity, while the PE is not constrained at ZPD. Here Hase's method has been modified by considering a clamped PE retrieval at ZPD (i.e. the PE is kept fixed to zero at ZPD). This modification has proved to allow for a superior interpretation of the measured O absorption lines, and thus it offers a better reconstruction of ILS than the retrievals with free PE at ZPD. The ILS time series evaluated assuming a non-clamped PE retrieval (i.e. free PE) at ZPD and the comparison of the theoretical performance of the O retrievals between the clamped and non-clamped PE retrievals are both included and discussed in Appendix .
Figure a–f displays two exemplary contrasting ILS retrievals, showing, on the one hand, the noticeable ILS degradation of the IFS 120M spectrometer in 2003 and, on the other hand, the proper alignment of the IFS 120/5HR after 2008. For both instruments the overall response of the ILS estimates on the actual instrumental performance is rather consistent. As observed in the rows of the averaging kernel matrix obtained in the ILS inversion procedure, the sensitivity of the MEA retrieval slightly increases towards maximum OPDs, which ensures enough sensitivity to resolve the narrower O absorption lines, while the PE sensitivity is larger towards ZPD to properly characterize the instrumental asymmetry issues. Nonetheless, the comparison between the two instruments also confirms the improved sensitivity of the IFS 120/5HR as compared to the IFS 120M due to a poorer signal-to-noise ratio in the cell spectra of the latter. As a result, the number of independent pieces of ILS information that can be estimated (given by the trace of matrix, the so-called “degrees of freedom for signal”, DOFS) for the IFS 120/5HR are greater than those obtained for the IFS 120M instrument by almost a factor of 2. For the cell examples of Fig. , the MEA DOFS is 0.73 and 1.61 for the IFS 120M and the IFS 120/HR, respectively, while the PE DOFS is 0.50 and 1.23 for the IFS 120M and the IFS 120/HR respectively.
The continuous monitoring of the ILS function is critical to control the instrumental alignment and temporal stability and to detect instrumental issues. The ILS time series at IZO (Fig. g) shows a few jumps due to punctual interventions made regarding the FTIR instruments between 1999 and 2018: (1) different changes to the field stop instruments between 1999 and 2004, (2) the switch from the IFS 120M to the IFS 120/5HR instrument in January 2005, (3) two optical re-alignments of the IFS 120/5HR system in June 2008 and February 2013, and (4) two replacements of the spectrometer's internal laser in July 2016 and July 2017. Figure also illustrates that, besides suffering from a greater level of spectral noise in the cell and atmospheric measurements, the ILS of the IFS 120M spectrometer is less stable over time than the ILS of the IFS 120/5HR. After the first re-alignment in 2008, the IFS 120/5HR's MEA shows a deviation of an ideal instrument of only % throughout the whole OPD range that remains until time of writing, while deviations of up to 5 % are observed for the IFS 120M. Regarding the PE time series, subtle asymmetries ( rad) are observed from 2003 onward in the IFS 120M and between 2010 and 2016 in the IFS 120/5HR. The former was due to the IFS 120M instrumental degradation, while the latter was likely introduced in the replacement of the interferometer's scanner motor at the end of 2009 and was properly corrected with the internal laser switch in July 2016. As discussed by , finite divergence and misalignment of the internal reference laser might slightly distort the sampling positions and thus impact PE.
From the analysis of the ILS evolution, three periods with different features affecting the IZO FTIR measurements clearly emerge: (1) 1999–2004, when despite the NO cell measurements being routinely carried out, the ILS estimation is imprecise due to the instability of the IFS 120M spectrometer; (2) 2005–May 2008, when the ILS function is well assessed despite the IFS 120/5HR instrument exhibiting a gradual temporal drift; and (3) June 2008–2018, when the IFS 120/5HR instrument is well characterized and optically well aligned (ILS is nearly nominal). Thereby, these three periods will be independently analysed in the present work in order to examine the impact of the instrument's status on the FTIR O products.
3.2 Ozone and ILS retrieval strategiesIn order to determine how the ILS characterization affects the FTIR O products, the retrieval strategy that theoretically and experimentally showed the best performance in has been modified to assess different ILS treatments (the so-called 5MWs set-up). In this strategy, the O volume mixing ratio (VMR) profiles are estimated from the measured O absorption lines in five single micro-windows between 991 and 1014 cm by means of the inversion code PROFFIT
The strategies considered to deal with the ILS function are listed in Table and are based on different approaches traditionally used by the FTIR community
2 where and are the linear and square parameters, respectively, OPD is equal to 180 cm, and the MEA is sampled at 20 equidistant positions in the interval , …, OPD. Set-up 5D assumes that the MEA linearly varies with the OPD, whereby only the linear parameter is retrieved simultaneously with the O concentrations, while set-up 5F considers a second-order polynomial dependency, meaning that both the linear and square parameters are estimated. Set-ups 5E and 5G optimize set-ups 5D and 5F, respectively, by including the PE fit, which is retrieved similarly to set-up 5C (i.e. a constant value throughout the whole OPD range). Set-ups 5C, 5D, 5E, 5F, and 5G assume an ideal ILS as a priori information. Finally, in order to account for uncertainties in the assessment of the independent cell measurements, set-up 5H uses the cell-derived MEA values, but a simultaneous fit of a PE offset is superimposed on the retrieved path-dependent cell-derived PE.
Table 1Description of the FTIR O retrieval strategies with different ILS characterizations and with and without a simultaneous atmospheric temperature profile retrieval ( the set-ups fitting the temperature profile are given in brackets). The 5A, 5AT, 5B, and 5BT configurations consider the cell-derived and the ideal ILS, respectively. The remaining set-ups fit some of the ILS parameters in the O retrieval procedure. Note that the level of refinement of the ILS retrieval increases from set-up 5C to set-up 5H.
Retrieval | ILS | Linear | Square | Phase Error | Temperature | A priori |
---|---|---|---|---|---|---|
strategy | approach | parameter () | parameter () | (PE) | fit | ILS |
5A(T) | NO cell retrieval | No | No | No | No (yes) | Ideal |
5B(T) | Ideal | No | No | No | No (yes) | – |
5C(T) | O retrieval | No | No | Yes | No (yes) | Ideal |
5D(T) | O retrieval | Yes | No | No | No (yes) | Ideal |
5E(T) | O retrieval | Yes | No | Yes | No (yes) | Ideal |
5F(T) | O retrieval | Yes | Yes | No | No (yes) | Ideal |
5G(T) | O retrieval | Yes | Yes | Yes | No (yes) | Ideal |
5H(T) | NO cell O retrieval | No | No | Yes | No (yes) | NO cell |
As reported in previous work , the quality of the FTIR O products can be significantly improved by including a simultaneous atmospheric temperature profile retrieval. However, this enhancement can only be achieved provided the FTIR spectrometer is well characterized and stable over time. For unstable instruments, the temperature fit has been found to exhibit a strong negative impact on the O retrievals by increasing the cross-interference between the instrumental performance and the temperature retrieval . In order to examine the combined effect of a simultaneous ILS and temperature retrieval for different instrument's status, an optimal estimation of the atmospheric temperature profile has also been considered in this study. For this purpose, four isolated CO absorption lines between 962.80 and 969.60 cm have been added to the O micro-windows, and the temperature a priori information and inversion settings have been defined according to
The FTIR spectra are only recorded when the line of sight (LOS) between the instrument and the Sun is cloud-free. However, to avoid possible contamination of thin clouds or unstable measured spectra, once the O retrievals have been computed they are filtered according to the number of iterations at which the convergence is reached and the fitting residuals between the simulated and measured spectra. Following this, all O datasets are temporally paired to ensure a fair comparison. The coincident and quality-filtered FTIR O retrievals amount to 4924 in the 1999–2018 period ( % of the measured dataset).
3.3 Theoretical quality assessmentThis section presents the theoretical characterization of the different ILS retrieval strategies based on the evaluation of their performance and expected uncertainties. Table provides an overview of the vertical sensitivity of the different retrievals on real O variations (i.e. total O DOFS) and of the interpretation of the measured spectra (i.e. fitting residuals).
Table 2
Summary of statistics of the DOFS and fitting residuals for the set-ups 5A and 5AT, 5B and 5BT, 5C and 5CT, 5D and 5DT, 5E and 5ET, 5F and 5FT, 5G and 5GT, and 5H and 5HT for the periods 1999–2004, 2005–May 2008, and June 2008–2018 and for the entire time series (1999–2018). The fitting residuals are computed as the signal-to-noise ratio for a common spectral region contained in all set-ups (1001.47–1003.04 cm). Shown are the median () and standard deviation () for each period. The number of quality-filtered measurements is 466, 683, and 3775 for the three periods, respectively, and 4924 for the whole dataset. The strategies showing the best performance (largest DOFS and smallest residuals) are highlighted in bold for each period.
DOFS | Residuals () | |||||||
---|---|---|---|---|---|---|---|---|
1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | 1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | |
Set-up | , | , | , | , | , | , | , | , |
5A | 4.29, 0.29 | 4.56, 0.15 | 4.52, 0.13 | 4.51, 0.18 | 3.51, 1.96 | 2.57, 0.86 | 2.70, 0.56 | 2.72, 0.93 |
5B | 4.25, 0.28 | 4.55, 0.14 | 4.49, 0.12 | 4.49, 0.17 | 3.68, 1.94 | 2.62, 0.84 | 2.78, 0.55 | 2.79, 0.93 |
5C | 4.33, 0.30 | 4.57, 0.14 | 4.50, 0.12 | 4.50, 0.17 | 3.49, 1.91 | 2.61, 0.83 | 2.76, 0.54 | 2.77, 0.89 |
5D | 4.29, 0.29 | 4.57, 0.15 | 4.49, 0.12 | 4.49, 0.17 | 3.44, 1.97 | 2.55, 0.85 | 2.76, 0.54 | 2.77, 0.92 |
5E | 4.35, 0.31 | 4.58, 0.15 | 4.49, 0.12 | 4.50, 0.17 | 3.27, 1.93 | 2.54, 0.84 | 2.75, 0.54 | 2.75, 0.88 |
5F | 4.20, 0.31 | 4.49, 0.16 | 4.41, 0.13 | 4.41, 0.18 | 3.43, 1.97 | 2.55, 0.85 | 2.75, 0.54 | 2.76, 0.92 |
5G | 4.27, 0.32 | 4.50, 0.16 | 4.42, 0.13 | 4.42, 0.18 | 3.27, 1.93 | 2.54, 0.84 | 2.73, 0.53 | 2.73, 0.88 |
5H | 4.35, 0.30 | 4.58, 0.15 | 4.52, 0.13 | 4.52, 0.17 | 3.37, 1.93 | 2.54, 0.85 | 2.70, 0.55 | 2.71, 0.90 |
5AT | 4.09, 0.34 | 4.42, 0.20 | 4.35, 0.15 | 4.35, 0.21 | 3.44, 1.96 | 2.55, 0.86 | 2.68, 0.56 | 2.70, 0.93 |
5BT | 4.08, 0.34 | 4.43, 0.20 | 4.33, 0.15 | 4.33, 0.21 | 3.56, 1.95 | 2.59, 0.84 | 2.75, 0.54 | 2.77, 0.92 |
5CT | 4.13, 0.35 | 4.45, 0.20 | 4.34, 0.15 | 4.34, 0.21 | 3.38, 1.92 | 2.57, 0.82 | 2.74, 0.53 | 2.74, 0.88 |
5DT | 4.06, 0.36 | 4.42, 0.20 | 4.31, 0.15 | 4.31, 0.22 | 3.42, 1.97 | 2.53, 0.85 | 2.75, 0.54 | 2.76, 0.92 |
5ET | 4.12, 0.37 | 4.43, 0.20 | 4.32, 0.15 | 4.32, 0.21 | 3.25, 1.93 | 2.52, 0.84 | 2.73, 0.53 | 2.73, 0.88 |
5FT | 3.95, 0.36 | 4.32, 0.21 | 4.22, 0.16 | 4.22, 0.22 | 3.42, 1.97 | 2.53, 0.85 | 2.74, 0.53 | 2.75, 0.92 |
5GT | 4.01, 0.38 | 4.34, 0.21 | 4.23, 0.16 | 4.23, 0.22 | 3.25, 1.93 | 2.52, 0.83 | 2.72, 0.53 | 2.72, 0.88 |
5HT | 4.14, 0.36 | 4.44, 0.20 | 4.36, 0.15 | 4.36, 0.21 | 3.28, 1.93 | 2.51, 0.85 | 2.67, 0.55 | 2.68, 0.90 |
Figure 2
(a) Estimated total statistical (ST) and (b) systematic (SY) errors (%) for O TCs retrieved from the 5A and 5AT, 5B and 5BT, and 5G and 5GT set-ups as a function of O slant column (SC) (DU) for measurements taken on 31 August 2007 from solar zenith angles (SZAs) between 84 ( 07:00 UT) and 21 ( 13:30 UT). (c) Example of estimated total statistical (ST) and (d) systematic (SY) error profiles (%) for the same set-ups for the spectrum taken on 31 August 2007 at a SZA of 50 and an O SC of 390 DU. The total errors are computed as the square root of the quadratic sum of all ST and SY error sources considered.
[Figure omitted. See PDF]
For all ILS set-ups an averaged total DOFS of 4 has been consistently obtained, whereby four atmospheric O altitude regions can be distinguished by both FTIR instruments (i.e. troposphere, upper troposphere–lower stratosphere (UTLS), middle stratosphere, and upper stratosphere). Although the differences between the periods and retrieval strategies lie within the overall variance, the set-ups considering the cell-derived ILS (5A and 5AT) seem to present the best performance when the FTIR system is very stable over time (i.e. 2008–2018 period). However, the opposite behaviour is observed for the more unstable periods, which is likely due to uncertainties in the cell-derived PE estimates, as documented by the enhancement of sensitivity and the reduction of fitting residuals when the PE is simultaneously fitted in the O retrieval procedure (5H and 5HT). For the remaining set-ups, the total vertical sensitivity slightly decreases as the ILS set-ups become more sophisticated (from 5C and 5CT to 5G and 5GT) because the information contained in the measured spectra is then split into the ILS and O retrievals (the retrieved state vector space is not perfectly orthogonal). However, in return the measured spectra are better reproduced. Similar patterns are observed when the temperature profile fit is also included in the retrieval strategy (lower total DOFS and lower fitting residuals are obtained as compared to those set-ups that do not retrieve the temperature).
The different ILS strategies have been also evaluated by performing an uncertainty analysis, which follows the formalism given in and the NDACC IRWG recommendations and has been analytically performed by the PROFFIT package. The error budget includes the impact of the spectral measurement noise and errors due to uncertainties in the input parameters, accounting for instrumental and model features, which are split into statistical (ST) and systematic (ST) contributions. More specifically, the ILS function estimates are assumed to have an uncertainty of 1 % and 0.01 rad for the MEA and PE, respectively. The baseline parameters, accounting for channelling effects and the intensity offsets, are expected to show an error of 0.1 %, while the solar pointing issues are limited to 0.001 rad. Regarding atmospheric temperatures, an uncertainty of 2 K up to 50 km a.s.l. and 1 K for higher altitudes has been considered. Possible uncertainties in the determination of solar lines have been also included (an error of 1 % and 10 for intensity and spectral position, respectively). Finally, the O spectroscopy data (intensity and pressure broadening parameters) are assumed to be uncertain by 3 %. All of these error sources and values can be considered typical for FTIR measurements (e.g. ). Refer to
The propagation of these uncertainty sources for a typical measurement day of the IFS 120/5HR instrument, considering the cell-derived (5A and 5AT), ideal (5B and 5BT), and the most refined ILS characterization (5G and 5GT), is displayed in Fig. . The error estimates reveal consistent results for all ILS set-ups. On the one hand, both the statistical and systematic uncertainties do depend on the O spectroscopic signatures due to the increase in the pointing errors and measurement noise at larger O slant column (SC) amounts. On the other hand, the inclusion of a simultaneous temperature retrieval significantly improves the theoretical performance for all FTIR O products, as mentioned above. The total statistical errors of O TCs are reduced to one-third when applying the temperature fit (from 1.5 %–3.0 % to 0.5 %–1.0 % for O SCs between 250 and 3000 DU), while the total systematic contributions drop by 0.1 %–0.4 % at low- and high-O SCs, respectively. However, the uncertainty analysis also illustrates the cross-interference between the simultaneous ILS and temperature retrievals, which becomes more evident in the altitude-resolved error patterns (Fig. c and d). For set-ups without the temperature fit, the ILS retrieval (5G) increases the statistical errors of the O TCs by significantly worsening the error profiles beyond the middle stratosphere. However, when the temperature and ILS are jointly retrieved with the O concentrations (5GT), more precise O TCs are expected due to an error reduction in the middle stratosphere layer around 30 km (visible for O SCs smaller than 1500 DU in Fig. a). Although only the error estimation for the set-ups 5G and 5GT is depicted in Fig. for simplicity, the remaining strategies fitting the ILS parameters are found to be consistent with these results overall.
It is worth highlighting that the uncertainty analysis carried out in this work assumes the same uncertainty values for the MEA and PE parameters for all strategies. However, the ILS errors for set-ups 5B and 5BT (ideal MEA and PE) or 5C and 5CT (ideal MEA) are expected to be significantly larger than for the 5A and 5AT retrievals (cell-derived ILS), especially for the MEA parameter as displayed in Fig. . As shown by , an uncertainty of 5 % for the MEA parameter may double the expected total statistical errors if the temperature fit is taken into account in the O retrieval procedure.
4 Comparison of the cell-derived and retrieved ILS and ozone observationsFigure synthesizes the comparison between the cell-derived ILS time series for the two IZO FTIR instruments, evaluated from NO cell measurements (Fig. ), and those retrieved simultaneously with the O concentrations from the most refined ILS set-ups (5G and 5GT). The MEA time series as a function of the OPD and the MEA and the PE time series at an OPD of 133 cm are shown as examples of the ILS behaviour at large OPDs. Note that the ILS evaluation becomes determinant towards maximum OPDs due to the high spectral resolution required to resolve the narrow O absorption lines.
Figure 3
Time series of the MEA as a function of the OPD from 1999 to 2018 evaluated from (a) cell measurements, (b, c) O retrievals (5G and 5GT). Panels (d), (e), and (f) are the same as (a), (b), and (c) but for the MEA and PE at an OPD of 133 cm. The PE values retrieved from the 5H set-up are also included in (d). Note that for set-ups 5G, 5GT, 5H, and 5HT the retrieved PE is constant throughout the whole OPD range.
[Figure omitted. See PDF]
The ILS comparison proves that the general shape of the cell-derived MEA is reproduced well by the ILS retrieval strategies: the estimated MEA values capture the increasing drift for the IFS 120M instrument and for the IFS 120/5HR spectrometer between 2005 and May 2008 and its stable evolution since then. However, the fitted MEA shows an artificial annual cycle, which is more noticeable when the temperature is not simultaneously retrieved and the instrument is properly aligned and stable (after the first re-alignment in 2008; a zoom in of this period is shown in Fig. e). In addition, fitting the ILS parameters produces, together with a noticeable daily variability, unrealistic MEA retrievals (see, for example, the values at the beginning of 2000 and at the end of 2002 in Fig. e and f). Thereby, the ILS fit seems to misinterpret extreme or anomalous O events as instrumental deficiencies. These results point out the existence of a significant cross-interference between the ILS and the O concentrations when both are simultaneously retrieved, meaning that part of the O variability on a daily and annual basis are damped when fitting the ILS parameters.
Similarly to the MEA retrievals, the estimated PE values are also able to capture the overall evolution of the IFS 120/5HR instrument and the punctual interventions made regarding the spectrometer (e.g. the scanner's motor change at the end of 2009 and the internal laser switches in 2016 and 2017). As expected, the IFS 120M instrument exhibits more variable retrieved PE values from the atmospheric spectra than those evaluated from the NO cell measurements. Figure d also includes the retrieved PE values from the set-up 5H, which superimposes an PE offset on the cell-derived PE values. The excellent agreement between the PE retrievals from atmospheric spectra using an ideal and the cell-derived ILS as a priori information becomes evident (Fig. e and f). However, as shown in Appendix , larger PE corrections are retrieved from the atmospheric spectra when the cell-derived ILS is evaluated assuming a non-clamped PE retrieval (especially for the IFS 120M instrument). This fact further corroborates that a clamped PE retrieval at ZPD when evaluating the NO cell measurements is a superior choice to characterize the instrumental performance of the NDACC FTIR spectrometers, as outlined in Sect. .
Figure 4
Averaged annual cycle of the MEA (at 133 cm) and PE (rad) from 2009 to 2018 for the set-ups 5A, 5AT, 5G, and 5GT. For 5A and 5AT the PE is given at 133 cm, while for the other strategies it corresponds to a unique value throughout the whole OPD range. The annual range (annual maximum minus annual minimum) is shown in the legend for each set-up. The error bars in the seasonal cycles are the standard errors of the mean, , where is the standard deviation and is the number of monthly measurements.
[Figure omitted. See PDF]
Figure 5
Summary of the O profile comparison for the set-ups 5B, 5BT, 5G, and 5GT with respect to 5A and 5AT for the periods 1999–2004, 2005–May 2008 and June 2008–2018, and for the entire time series (1999–2018). Panels (a), (c), and (e) display the vertical profiles of the median () RD (5X and 5XT to 5A and 5AT, in %) for the three periods, respectively. Panels (b), (d), and (f) are the same as (a), (c), and (e) but for the standard deviation of RD distributions (, in %).
[Figure omitted. See PDF]
Table 3Summary of statistics for the O TC comparison for the set-ups 5B, 5BT, 5C, 5CT, 5D, 5DT, 5E, 5ET, 5F, 5FT, 5G, 5GT, 5H, and 5HT with respect to 5A and 5AT: median (, in %) and standard deviation (, in %) of the relative differences (RD, 5X and 5XT to 5A and 5AT), and Pearson correlation coefficient () for the periods 1999–2004, 2005–May 2008, and June 2008–2018 and for the entire time series (1999–2018). The number of quality-filtered measurements is 466, 683, and 3775 for the three periods, respectively, and 4924 for the whole dataset.
1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | |
---|---|---|---|---|
Set-up | (%), (%), | (%), (%), | (%), (%), | (%), (%), |
5B | 0.13, 0.12, 1.000 | 0.27, 0.07, 1.000 | 0.11, 0.04, 1.000 | 0.12, 0.08, 1.000 |
5C | 0.10, 0.13, 1.000 | 0.27, 0.07, 1.000 | 0.10, 0.04, 1.000 | 0.12, 0.08, 1.000 |
5D | 0.05, 0.47, 0.998 | 0.01, 0.57, 0.994 | 0.01, 0.32, 0.999 | 0.01, 0.38, 0.998 |
5E | 0.08, 0.47, 0.998 | 0.02, 0.59, 0.993 | 0.01, 0.32, 0.999 | 0.00, 0.38, 0.998 |
5F | 0.00, 0.53, 0.998 | 0.08, 0.70, 0.991 | 0.16, 0.64, 0.993 | 0.15, 0.64, 0.994 |
5G | 0.01, 0.53, 0.998 | 0.08, 0.69, 0.991 | 0.16, 0.64, 0.993 | 0.15, 0.64, 0.993 |
5H | 0.06, 0.05, 1.000 | 0.01, 0.01, 1.000 | 0.00, 0.01, 1.000 | 0.00, 0.03, 1.000 |
5BT | 0.65, 0.59, 0.997 | 1.62, 0.39, 0.998 | 0.36, 0.17, 1.000 | 0.36, 0.17, 0.997 |
5CT | 0.63, 0.62, 0.997 | 1.63, 0.40, 0.998 | 0.36, 0.17, 1.000 | 0.42, 0.49, 0.997 |
5DT | 0.30, 1.88, 0.976 | 0.29, 1.82, 0.940 | 0.11, 0.85, 0.991 | 0.14, 1.16, 0.982 |
5ET | 0.24, 1.87, 0.976 | 0.30, 1.74, 0.942 | 0.11, 0.89, 0.990 | 0.14, 1.17, 0.982 |
5FT | 0.30, 1.87, 0.976 | 0.33, 1.65, 0.950 | 0.25, 1.09, 0.983 | 0.26, 1.28, 0.977 |
5GT | 0.23, 1.85, 0.976 | 0.34, 1.76, 0.943 | 0.25, 1.10, 0.983 | 0.26, 1.30, 0.977 |
5HT | 0.18, 0.25, 1.000 | 0.00, 0.02, 1.000 | 0.02, 0.03, 1.000 | 0.01, 0.09, 1.000 |
In order to examine the presence of seasonality in the retrieved ILS time series in more detail, Fig. displays the averaged annual cycle of the MEA and PE parameters for the more stable IFS 120/5HR period (2009–2018). Similarly to Fig. , the annual cycles at an OPD of 133 cm for the set-ups 5A, 5AT, 5G, and 5GT are shown. As expected, the cell-derived MEA and PE values derived from the NO cells exhibit no seasonality. By contrast, a noticeable annual cycle for the retrieved MEA is observed for the 5G and 5GT set-ups, although it is partially corrected by the temperature fit. The MEA seasonal range is (0.9 % with respect to the mean retrieved MEA) and (0.5 % with respect to the mean retrieved MEA) for 5G and 5GT, respectively, which are about 10 and 5 times that of the reference, (0.1 % with respect to the mean cell-derived MEA). Regarding the PE parameter, although a subtle seasonal dependence is detected in the retrieved PE values, no significant differences are observed between the ILS strategies, and no influence of temperature retrieval is found.
Consistent results were obtained for the less sophisticated ILS set-ups: the cell-derived ME and PE values are well reproduced overall for both FTIR instruments, especially for the IFS 120/5HR periods (data not shown). However, the artificial ILS seasonality for these configurations is less noticeable than for the most refined ILS set-ups (5G and 5GT), particularly when the temperature retrieval is not considered. The MEA seasonal range is limited to 0.4 % (5D and 5E) and 0.5 % (5DT and 5ET) with respect to the mean retrieved MEA for the 5E and 5ET set-ups, respectively.
In cases where gas cell measurements are not available and the independence of ILS and target gas retrievals is pursued, an alternative approach might be to retrieve the ILS information from atmospheric trace gas retrievals with well-known vertical distribution . To assess the viability of this strategy, the ILS parameters have been also evaluated from the measured absorbing lines of very stable tropospheric and stratospheric gases, such as carbon dioxide () and hydrogen fluoride (HF), respectively. Nonetheless, as discussed in Appendix , both strategies produce unrealistic ILS estimates and a strong artificial annual cycle in the retrieved MEA and PE values; therefore, they have been discarded in the subsequent O assessment. Consequently, in the absence of independent ILS measurements, the most promising approach to characterize the instrumental performance might be to apply a retrieval of the ILS parameters simultaneously to the atmospheric temperature and O profiles (provided the FTIR instrument is stable over time).
The differences between the ILS treatments are transferred to the O TCs and profiles as summarized in Table and Fig. , respectively. The set-ups not retrieving MEA information provide the largest bias with respect to the cell-derived ILS O TCs (i.e. up to 0.3 % for 5B and 5C in the 2005–May 2008 period), whereas the most significant variability is observed for the most refined ILS set-ups (i.e. up to 0.7 % for 5F and 5G in the 2005–May 2008 period). As expected, the simultaneous temperature retrieval strongly affects the differences between the ILS treatments due to the cross-interference between the ILS and the O and temperature profiles (especially beyond the lower stratosphere, as illustrated in Fig. ).
5 Comparison to reference observations5.1 FTIR and Brewer ozone total columns
The quality assessment of the different ILS set-ups has been addressed by a straightforward comparison to coincident Brewer observations, acquired within a temporal window of min around the FTIR measurements. This matching criterion leads to 1958 coincidences in the 1999–2018 period. As observed in the 20-year time series of the relative differences (RD, FTIR–Brewer) for the set-ups 5A, 5AT, 5B, 5BT, 5G, and 5GT (Fig. ), the differences between the ILS set-ups become more significant when the temperature is simultaneously estimated, corroborating the cross-interference between both retrievals. The most remarkable discrepancies occur for the IFS 120M period and between 2005 and mid-2008 for the IFS 120/5HR instrument, when the behaviour of both systems was far from ideal. In fact, when assuming the ILS to be ideal (5BT set-up) the FTIR O TCs seem to absorb the actual ILS temporal degradation, as documented by the consistent increasing drift observed in the RD and cell-derived MEA time series (2003–2008 period in Figs. and b). By contrast, when the IFS 120/5HR behaves almost as an ideal instrument (2008–2018 period), no significant discrepancies were found between using the cell-derived or ideal ILS (with and without temperature retrieval). This fact further confirms the reliability of the ILS estimates from independent gas cell measurements.
Figure 6
Summary of Brewer–FTIR comparison from 1999 to 2018. (a) RD time series (O TC–O TC, in %) for the set-ups 5A, 5B, and 5G. Panel (b) is the same as (a) but for the set-ups 5AT, 5BT, and 5GT. (c) Averaged annual cycles of the RD for the 2009–2018 period for the set-ups 5A, 5B, and 5E. Panel (d) is the same as (c) but for the set-ups 5AT, 5BT, and 5GT. The averaged annual cycles for the ILS strategies 5G and 5GT have also been included in (c) and (d), respectively. The solid lines in (a) and (b) correspond to the monthly median values. The legend in (c) and (d) shows the seasonal amplitude of each cycle.
[Figure omitted. See PDF]
Figure also illustrates that although the scatter of RD is significantly improved by the temperature retrieval, these strategies present more extreme RD values, which are larger for those set-ups without fitting of the ILS. Note, for example, the correlation between the unrealistic retrieved MEA values at the beginning of 2000 and at the end of 2002 (Fig. e and f) and the extreme RD values only obtained for those set-ups not retrieving the ILS parameters (especially in Fig. b). This behaviour is likely due to the ILS retrievals absorbing the most imprecise FTIR O retrievals, as mentioned in Sect. . However, the extreme RD values may also indicate measurement days with an unusual temperature vertical stratification, which might be wrongly captured by the Brewer and FTIR products assuming a fixed temperature (and pressure) profile .
An overview of the FTIR–Brewer comparison for all ILS set-ups, distinguishing between the instruments and periods, is given in Table . Consistently for both instruments, when the temperature fit is not considered in the retrieval strategy, the set-ups using the cell-derived MEA values (5A and 5H) seem to offer the best agreement with respect to Brewer data (less dispersion, greater accuracy, and correlation). Nonetheless, the inclusion of the simultaneous temperature estimation enhances the performance of those strategies fitting the MEA parameter (5DT, 5ET, 5FT, and 5GT), confirming the results obtained in the theoretical quality assessment (Sect. ). Note that the PE retrieval also has a minor impact on the O products for both FTIR instruments. Considering the IFS 120/5HR periods as reference, the combined strategies provide the most precise O TCs with a scatter in the RD of 0.5 %–0.6 % (5DT, 5ET, 5FT, and 5GT), while it increases up to 0.6 %–0.7 % when using the cell-derived MEA values (5AT and 5HT) and up to 0.7 %–0.8 % when the MEA is assumed to be ideal (5BT and 5CT). However, this improvement is not good enough to completely cancel out the negative cross-interference between the temperature retrieval and instrumental performance for an unstable FTIR spectrometer such as the IFS 120M. In that case, the most precise O TCs are obtained considering the cell-derived ILS data and without performing a simultaneous temperature fit in the O retrieval procedure.
Table 4Summary of statistics for the Brewer–FTIR comparison for the set-ups 5A, 5AT, 5B, 5BT, 5C, 5CT, 5D, 5DT, 5E, 5ET, 5F, 5FT, 5G, 5GT, 5H, and 5HT: median (, in %) and standard deviation (, in %) of the RD (FTIR–Brewer) and Pearson correlation coefficient () for the periods 1999–2004, 2005–May 2008, and June 2008–2018 and for the entire time series (1999–2018). The number of coincident FTIR–Brewer measurements is 89, 169, and 1722 for the three periods, respectively, and 1980 for the whole dataset. Note that the set-ups 5G and 5GT stand for the FTIR O data estimated using the retrieved ILS parameters from the set-ups 5G and 5GT, respectively, but averaged on a annual basis (see details in the text). The strategies showing the best performance (smallest and , and largest ) are highlighted in bold for each period.
1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | |
---|---|---|---|---|
Set-up | (%), (%), | (%), (%), | (%), (%), | (%), (%), |
5A | 4.33, 1.14, 0.972 | 4.47, 0.81, 0.975 | 3.39, 0.82, 0.982 | 3.49, 0.91, 0.977 |
5B | 4.50, 1.14, 0.972 | 4.75, 0.82, 0.974 | 3.51, 0.83, 0.982 | 3.62, 0.93, 0.976 |
5C | 4.45, 1.14, 0.971 | 4.75, 0.82, 0.974 | 3.51, 0.83, 0.982 | 3.61, 0.93, 0.976 |
5D | 4.22, 1.17, 0.968 | 4.45, 0.89, 0.970 | 3.43, 0.86, 0.980 | 3.51, 0.94, 0.976 |
5E | 4.15, 1.17, 0.968 | 4.45, 0.89, 0.969 | 3.42, 0.86, 0.980 | 3.50, 0.94, 0.976 |
5F | 4.17, 1.15, 0.969 | 4.37, 0.88, 0.971 | 3.23, 0.83, 0.981 | 3.34, 0.94, 0.975 |
5G | 4.18, 1.16, 0.969 | 4.38, 0.88, 0.971 | 3.24, 0.83, 0.981 | 3.34, 0.94, 0.975 |
5G | 4.29, 1.19, 0.967 | 4.42, 0.84, 0.973 | 3.22, 0.81, 0.982 | 3.33, 0.92, 0.976 |
5H | 4.25, 1.14, 0.972 | 4.46, 0.81, 0.975 | 3.38, 0.82, 0.982 | 3.48, 0.91, 0.977 |
5AT | 4.91, 1.60, 0.953 | 4.13, 0.61, 0.985 | 3.43, 0.67, 0.988 | 3.52, 0.80, 0.982 |
5BT | 5.63, 1.72, 0.943 | 5.88, 0.79, 0.974 | 3.81, 0.70, 0.987 | 3.91, 1.00, 0.972 |
5CT | 5.65, 1.73, 0.941 | 5.89, 0.80, 0.974 | 3.81, 0.70, 0.987 | 3.91, 1.00, 0.972 |
5DT | 5.20, 1.47, 0.947 | 4.34, 0.52, 0.989 | 3.59, 0.64, 0.988 | 3.68, 0.79, 0.982 |
5ET | 5.18, 1.47, 0.948 | 4.35, 0.53, 0.989 | 3.59, 0.64, 0.988 | 3.68, 0.79, 0.982 |
5FT | 5.17, 1.48, 0.947 | 4.37, 0.52, 0.989 | 3.73, 0.63, 0.989 | 3.82, 0.76, 0.984 |
5GT | 5.13, 1.47, 0.947 | 4.38, 0.52, 0.989 | 3.73, 0.63, 0.989 | 3.82, 0.76, 0.984 |
5GT | 4.90, 1.82, 0.926 | 4.41, 0.55, 0.988 | 3.71, 0.66, 0.988 | 3.78, 0.81, 0.982 |
5HT | 4.87, 1.61, 0.953 | 4.14, 0.61, 0.985 | 3.43, 0.67, 0.988 | 3.52, 0.80, 0.982 |
Regarding systematic differences, the FTIR–Brewer bias ranges from 3 % to 6 %, which is compatible with errors of 3 % in the O infrared spectroscopy as documented by the theoretical uncertainty analysis. Such differences are consistent with previous studies (e.g. ; ) and with ultraviolet, infrared, and microwave intercomparison experiments carried out in laboratory conditions (e.g. ). Nonetheless, the simultaneous ILS and temperature retrievals and the instrument status also contribute to the observed offset. As shown in Table , the combined (ILS and temperature) approaches increase the differences with respect to Brewer data as compared to the usage of the cell-derived ILS values (consistent with that documented in Sect. , Table ). Note that the reduction of the FTIR–Brewer bias between the two IFS 120/5HR periods is likely introduced by punctual interventions on the FTIR spectrometer .
The O concentrations are expected to be impacted by the different ILS approaches at a seasonal scale. To examine this effect, Fig. also includes the averaged annual cycle of the RD for the more stable IFS 120/5HR period (2009–2018). Independently of the temperature treatment, the O strategies using the cell-derived MEA values show the lowest RD seasonal amplitudes. They vary from 0.62 % (5A) to 0.69 % (5AT) and from 0.57 % (5H) to 0.62 % (5HT), while for the most refined ILS set-up the annual peak-to-peak amplitude ranges from 0.74 % (5G) to 0.69 % (5GT). The latter is even larger than those observed when assuming an ideal MEA value and not considering the simultaneous temperature retrieval (0.66 % for the set-ups 5B and 5C). These results suggest that the artificial annual cycle of the retrieved ILS parameters comes along with a misinterpretation of the retrieved seasonal cycle of O, as already pointed out by the comparison of the cell-derived and the retrieved ILS time series (Sect. ): the simultaneous ILS fitting leads to an underestimation of the actual O seasonality.
In order to reduce this seasonal artefact an additional approach has been assessed, which is based on the retrieved MEA and PE parameters from the most sophisticated ILS configuration (5G and 5GT) (so-called 5G and 5GT set-ups). Firstly, the estimated ILS parameters are averaged on an annual basis in the three periods analysed independently. Following this, the annually averaged ILS values are used to compute the O retrievals in a second step. As shown in Table , this approach indeed improves the agreement with respect to Brewer data for the IFS 120/5HR instrument, when the simultaneous temperature fit is not applied and the seasonality of the retrieved ILS values is therefore more evident. At a seasonal scale, the peak-to-peak amplitude of RD is also reduced when the ILS parameters are averaged annually, from 0.74 % (5G) to 0.63 % (5G) (Fig. ). However, this strategy does not offer similar results if the temperature fit is also considered in the retrieval procedure. It slightly increases the dispersion in the straightforward comparison to the Brewer database (Table ) and the seasonal amplitude of RD (from 0.69 % to 0.70 %). Finally, for the IFS 120M spectrometer this approach has not been found to improve the comparability to Brewer data due to the high variability of the retrieved ILS parameters and their remarkable temporal degradation.
5.2 FTIR and ECC ozone vertical profilesThe ILS characterization is especially critical in the retrieval of the vertical distribution, as illustrated in Fig. . It depicts the median and standard deviation of the relative difference between the FTIR profiles obtained with different ILS set-ups and coincident ECC sondes for the three periods considered. As ECC sondes usually burst between 30 and 35 km, only those sondes with information up to 29 km have been paired to the FTIR profiles within a temporal window of 3 h around the sonde launch ( 12:00 UT). These matching criteria provide 277 coincidences in the 1999–2018 period. In addition, to account for the limited FTIR vertical sensitivity, the ECC sondes were vertically smoothed using the averaging kernels obtained in the retrieval procedure
Figure 7
Summary of FTIR-smoothed ECC comparison for the periods 1999–2004, 2005–May 2008, and June 2008–2018. Panels (a), (d), and (g) display the vertical profiles of the median () RD (FTIR-ECC, in %) for the three periods, respectively. Panels (b), (e), and (h) are the same as (a), (d), and (g) but for the standard deviation of RD distributions (, in %). Panels (c), (f), and (i) are the same as (a), (d), and (g) but for the Pearson correlation coefficient. The number of coincident FTIR-ECC measurements is 56, 49, and 167 for the periods 1999–2004, 2005–May 2008, and June 2008–2018, respectively.
[Figure omitted. See PDF]
Table 5The same as Table but for the FTIR-smoothed ECC comparison for the O partial columns computed between 2.37–13, 12–23, and 22–29 km. The number of coincident FTIR-ECC measurements is 56, 49, and 167 for the periods 1999–2004, 2005–May 2008, and June 2008–2018, respectively, and 272 for the whole dataset. The strategies showing the best performance (smallest and and largest ) are highlighted in bold for each period.
1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | 1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | ||
---|---|---|---|---|---|---|---|---|---|
Set-up | (%), (%), | (%), (%), | (%), (%), | (%), (%), | Set-up | (%), (%), | (%), (%), | (%), (%), | (%), (%), |
FTIR-ECC at 2.37–13 km | FTIR-ECC at 2.37–13 km | ||||||||
5A | 16.24, 7.09, 0.938 | 12.30, 6.64, 0.957 | 9.72, 5.12, 0.974 | 10.89, 6.56, 0.956 | 5AT | 16.93, 7.05, 0.940 | 12.76, 6.82, 0.956 | 10.11, 4.99, 0.976 | 11.23, 6.42, 0.958 |
5B | 16.01, 7.11, 0.938 | 11.67, 6.65, 0.957 | 9.33, 5.16, 0.974 | 10.52, 6.59, 0.956 | 5BT | 17.11, 7.01, 0.941 | 12.71, 6.84, 0.955 | 9.89, 5.00, 0.976 | 11.16, 6.46, 0.958 |
5C | 15.59, 7.06, 0.939 | 11.55, 6.65, 0.957 | 9.36, 5.14, 0.974 | 10.45, 6.52, 0.957 | 5CT | 16.86, 6.95, 0.942 | 12.43, 6.83, 0.955 | 9.90, 4.98, 0.976 | 11.09, 6.37, 0.959 |
5D | 16.76, 7.03, 0.941 | 12.15, 6.72, 0.956 | 9.54, 5.08, 0.975 | 10.77, 6.46, 0.957 | 5DT | 17.17, 7.01, 0.941 | 12.57, 6.81, 0.956 | 9.88, 5.02, 0.976 | 11.17, 6.42, 0.958 |
5E | 16.52, 6.97, 0.942 | 11.99, 6.71, 0.956 | 9.55,5.06, 0.975 | 10.65, 6.38, 0.958 | 5ET | 16.79, 6.94, 0.942 | 12.33, 6.80, 0.956 | 9.88, 5.00, 0.976 | 11.10, 6.34, 0.959 |
5F | 16.84, 6.96, 0.941 | 12.24, 6.70, 0.956 | 9.79, 5.07, 0.975 | 11.08, 6.39, 0.958 | 5FT | 17.18, 6.98, 0.941 | 12.52, 6.80, 0.956 | 9.92, 5.03, 0.976 | 11.25, 6.40, 0.958 |
5G | 16.65, 6.90, 0.942 | 12.09, 6.69, 0.956 | 9.77, 5.05, 0.975 | 10.93, 6.30, 0.959 | 5GT | 16.85, 6.92, 0.942 | 12.28, 6.79, 0.956 | 9.92, 5.01, 0.976 | 11.14, 6.31, 0.959 |
5G | 16.30, 6.84, 0.942 | 12.05, 6.70, 0.956 | 9.79, 5.10, 0.975 | 10.74, 6.44, 0.958 | 5GT | 16.70, 6.98, 0.942 | 12.39, 6.84, 0.955 | 9.86, 5.03, 0.976 | 11.06, 6.43, 0.958 |
5H | 15.86, 7.04, 0.939 | 12.17, 6.63, 0.957 | 9.63, 5.11, 0.975 | 10.81, 6.50, 0.957 | 5HT | 16.68, 6.99, 0.942 | 12.51, 6.81, 0.956 | 10.00, 4.98, 0.976 | 11.09, 6.36, 0.959 |
FTIR-ECC at 12–23 km | FTIR-ECC at 12–23 km | ||||||||
5A | 17.89, 5.43, 0.939 | 16.03, 4.94, 0.943 | 13.72, 4.81, 0.962 | 14.91, 5.21, 0.951 | 5AT | 17.48, 5.72, 0.935 | 16.91, 4.67, 0.948 | 15.07, 4.77, 0.963 | 16.02, 5.11, 0.953 |
5B | 17.85, 5.40, 0.939 | 16.55, 4.93, 0.943 | 13.68, 4.78, 0.963 | 14.92, 5.21, 0.951 | 5BT | 17.37, 5.81, 0.932 | 16.34, 4.65, 0.950 | 14.83, 4.80, 0.963 | 15.74, 5.13, 0.953 |
5C | 17.92, 5.44, 0.939 | 16.57, 4.93, 0.943 | 13.68, 4.79, 0.962 | 14.94, 5.23, 0.951 | 5CT | 17.27, 5.84, 0.932 | 16.25, 4.66, 0.949 | 14.81, 4.78, 0.963 | 15.72, 5.12, 0.953 |
5D | 17.76, 5.35, 0.941 | 16.58, 4.95, 0.942 | 13.72, 4.78, 0.963 | 14.89, 5.20, 0.951 | 5DT | 17.67, 5.37, 0.942 | 17.24, 4.70, 0.947 | 15.08, 4.78, 0.963 | 16.17, 5.03, 0.955 |
5E | 17.78, 5.38, 0.941 | 16.58, 4.95, 0.942 | 13.71, 4.78, 0.963 | 14.92, 5.22, 0.951 | 5ET | 17.61, 5.35, 0.942 | 17.24, 4.70, 0.947 | 15.07, 4.75, 0.963 | 16.18, 5.00, 0.955 |
5F | 17.66, 5.24, 0.942 | 16.82, 4.78, 0.945 | 14.95, 4.75, 0.963 | 16.01, 4.98, 0.955 | 5FT | 17.98, 5.31, 0.942 | 17.20, 4.69, 0.948 | 15.29, 4.71, 0.964 | 16.25, 4.96, 0.956 |
5G | 17.42, 5.26, 0.943 | 16.79, 4.78, 0.945 | 14.93, 4.75, 0.963 | 16.00, 4.98, 0.955 | 5GT | 17.81, 5.32, 0.942 | 17.20, 4.68, 0.948 | 15.26, 4.69, 0.964 | 16.13, 4.94, 0.956 |
5G | 17.27, 5.38, 0.941 | 16.78, 4.92, 0.942 | 14.89, 4.78, 0.962 | 16.00, 5.07, 0.953 | 5GT | 17.34, 5.47, 0.941 | 17.15, 4.61, 0.950 | 15.17, 4.63, 0.965 | 15.98, 4.94, 0.956 |
5H | 17.90, 5.46, 0.939 | 16.02, 4.95, 0.942 | 13.73, 4.81, 0.962 | 14.93, 5.22, 0.951 | 5HT | 17.40, 5.73, 0.935 | 16.90, 4.68, 0.948 | 15.05, 4.75, 0.963 | 16.02, 5.10, 0.953 |
FTIR-ECC at 22–29 km | FTIR-ECC at 22–29 km | ||||||||
5A | 16.00, 4.10, 0.783 | 16.43, 2.87, 0.892 | 14.06, 3.23, 0.777 | 14.79, 3.68, 0.776 | 5AT | 16.31, 3.67, 0.856 | 15.28, 3.53, 0.829 | 14.47, 3.12, 0.775 | 15.06, 3.50, 0.798 |
5B | 15.41, 4.17, 0.775 | 15.10, 2.83, 0.895 | 13.85, 3.27, 0.772 | 14.40, 3.59, 0.783 | 5BT | 16.55, 3.76, 0.852 | 16.28, 3.51, 0.817 | 14.87, 3.11, 0.775 | 15.42, 3.53, 0.795 |
5C | 15.50, 4.20, 0.772 | 15.15, 2.83, 0.895 | 13.86, 3.27, 0.772 | 14.41, 3.61, 0.781 | 5CT | 16.75, 3.82, 0.849 | 16.33, 3.50, 0.818 | 14.91, 3.12, 0.774 | 15.49, 3.55, 0.792 |
5D | 15.45, 4.14, 0.781 | 15.63, 2.86, 0.893 | 14.09, 3.24, 0.774 | 14.67, 3.70, 0.774 | 5DT | 15.41, 3.95, 0.802 | 15.34, 3.23, 0.857 | 14.67, 2.99, 0.788 | 14.82, 3.46, 0.787 |
5E | 15.64, 4.17, 0.779 | 15.70, 2.86, 0.893 | 14.10, 3.25, 0.773 | 14.72, 3.72, 0.772 | 5ET | 15.74, 3.97, 0.801 | 15.33, 3.21, 0.859 | 14.66, 3.00, 0.787 | 14.95, 3.48, 0.785 |
5F | 15.23, 3.72, 0.826 | 14.93, 2.72, 0.906 | 12.43, 3.10, 0.795 | 13.60, 3.54, 0.792 | 5FT | 15.49, 3.53, 0.844 | 15.09, 3.04, 0.877 | 13.45, 2.87, 0.806 | 14.05, 3.25, 0.812 |
5G | 15.85, 3.73, 0.825 | 14.97, 2.71, 0.907 | 12.53, 3.12, 0.792 | 13.62, 3.60, 0.785 | 5GT | 15.73, 3.60, 0.838 | 15.12, 3.03, 0.880 | 13.41, 2.89, 0.803 | 14.24, 3.32, 0.803 |
5G | 15.87, 3.96, 0.798 | 15.15, 2.84, 0.894 | 12.54, 3.20, 0.784 | 13.57, 3.80, 0.766 | 5GT | 16.10, 3.59, 0.847 | 15.18, 3.41, 0.840 | 13.45, 3.05, 0.783 | 14.33, 3.55, 0.788 |
5H | 16.10, 4.12, 0.782 | 16.43, 2.87, 0.892 | 14.14, 3.24, 0.777 | 14.83, 3.69, 0.774 | 5HT | 16.52, 3.70, 0.855 | 15.42, 3.52, 0.831 | 14.49, 3.12, 0.776 | 15.12, 3.52, 0.797 |
The impact of the instrumental characterization exhibits a clear vertical stratification, which depends on the FTIR instrument's status. Below the UTLS region the vertical performance of the different ILS characterization approaches is quite similar for the two FTIR spectrometers (slightly superior for the most refined set-ups), with scatter values of 6 %–8 % for the IFS 120M and for the 2005–2008 period of the IFS 120/5HR instrument ( 6 %–8 % for the PCs at 2.37–13 and 12–23 km, Table ). When the latter was aligned better (2008–2018 period), the dispersion between FTIR and smoothed ECC sondes is reduced by 5 %–6 % (3% for the PCs at 2.37–13 and 12–23 km). As expected, the effect of the temperature retrieval becomes noticeable beyond the UTLS region.
In the middle stratosphere (above 25 km), Fig. illustrates that for both IFS systems the best overall agreement with respect to ECC data is reached when the ILS, including the MEA parameter, and the temperature fits are simultaneously carried out with the O vertical retrieval (5DT, 5ET, 5FT, and 5GT). In fact, a substantial improvement is documented for the more sophisticated ILS retrievals (5FT and 5GT) as compared to the strategy using the cell-derived MEA values (5AT and 5HT). The scatter values at 29 km range from 4.5 % (5AT and 5HT) to 2.6 % (5FT and 5GT) and from 3.8 % (5AT and 5HT) to 2.9 % (5FT and 5GT) for the 2005–May 2008 and June 2008–2018 periods, respectively. Nonetheless, if only the temperature fit is applied (5AT, 5BT, 5CT, and 5HT), the cross-interference between the temperature retrieval and the instrumental performance (ILS uncertainties and spectral measurement noise) worsens the comparability to the ECC data. This is especially critical for the IFS 120M, for which the scatter values increase from 4 % up to 7 %–8 %, while for the well-aligned IFS 120/5HR period the dispersion increases by less than 0.2 % at 29 km. Refer to Table for the comparison statistics between PCs.
In relation to systematic differences, the altitude dependence of bias is found to consistently decline as the FTIR instrument gains stability and become better aligned. The IFS 120M is less sensitive to changes between the ILS set-ups and temperature treatments: bias is 15 %–18 % up to the UTLS region and drops by up to 9 %–12 % in the middle stratosphere for all configurations. Nonetheless, for the IFS 120/5HR, together with an overall reduction of the bias (especially up to the UTLS altitudes), the cross-interference between the ILS, temperature, and O concentrations becomes evident in the middle stratosphere. Considering the 2008–2018 period as reference (better instrumental alignment and more FTIR-ECC coincidences), above the middle stratosphere the most accurate O profiles are obtained with the most refined ILS set-ups with and without temperature retrieval (5F and 5FT, 5G and 5GT) with bias ranging between 6 %–8 % ( 13 %–14 % for the PCs 2.37–13 and 12–23 km, Table ). At lower altitudes, the differences between the strategies are not so significant and lie within the overall variance.
More remarkable discrepancies between the ILS approaches would be expected beyond the middle stratosphere, as suggested by the uncertainty analysis (see Fig. ). In addition, compensations between the ILS, temperature, measurement errors, and O vertical distribution might occur at these altitudes to be consistent with the findings from the O TC comparison (e.g. the subtle improvement due to the PE fit considering the cell-derived MEA values). Unfortunately, the ECC database does not allow these effects to be examined in detail given its limited altitude coverage. Other measurement techniques, such as ground-based lidar or microwave radiometers capable of probing the O profile at higher altitudes, could therefore provide an added value for the O vertical profile analysis.
6 Summary and conclusionsA precise instrumental characterization is indispensable to retrieve correct information from the measured solar absorption spectra by ground-based FTIR (Fourier transform infrared) spectrometers, since it directly affects the absorption line shapes of atmospheric gases. Currently, different approaches are used to deal with the FTIR instrumental response through the spectrometer's ILS (instrumental line shape) function. The most widely used options are to assume an ideal instrument, to retrieve the instrumental information simultaneously with the atmospheric gas concentrations from the measured solar absorption spectra, or to determine it from independent gas cell measurements. In this context, this paper has assessed the impact of these strategies on the quality of the FTIR ozone (O) products (total columns and vertical profiles) using the 20-year time series of FTIR measurements acquired at the subtropical Izaña Observatory (IZO) between 1999 and 2018.
The theoretical and experimental quality assessment addressed by this work has documented how critical the treatment of the ILS function can be. Assuming an ideal ILS can provide imprecise O concentrations, especially when the instrument is not properly aligned, while retrieving the ILS information in the inversion procedure has proven to lead to a misinterpretation of the actual O variations on a daily and seasonal scale (the more sophisticated the ILS fit, the more pronounced this artefact). The optimal approach to deal with the FTIR instrumental characterization is therefore the continuous monitoring of the ILS function by means of independent data, such as the low-pressure NO cell measurements used in this study. These measurements should be routinely taken about every 2 months for stable FTIR instruments, such as IFS 120/5HR, and even more frequently for IFS 120M spectrometers.
Nonetheless, if the simultaneous ILS retrieval is carried out due to technical necessities of the FTIR station (e.g. in absence of gas cell measurements), an atmospheric temperature profile retrieval is strongly recommended to improve the precision of the O retrievals and reduce the cross-interference between the atmospheric temperature and instrumental performance. However, this enhancement is only achieved provided the FTIR spectrometer is stable over time. For unstable instruments, the temperature retrieval exhibits a drastic negative impact on O products even though the ILS fit is simultaneously performed. In this sense, it is worth highlighting that retrieving only the ILS jointly to the O concentrations (without a simultaneous temperature retrieval) might worsen the precision of the FTIR O products, and it may be a better approach to assume an ideal ILS function.
Other tentative approaches to assess the FTIR instrumental performance from atmospheric trace gas retrievals have been also evaluated in this study (e.g. and HF). This analysis reveals that (in absence of independent ILS measurements) the most promising option is indeed the retrieval of the ILS parameters jointly with O concentrations (provided the ILS estimates are averaged annually to reduce seasonal effects). Nonetheless, it is important to be aware that the selection of the ILS characterization approach (and the instrumental degradation) can significantly affect the determination of the O vertical distributions. The impact is especially critical beyond the lower stratosphere, where most of the O recovery is currently taking place and instrumental artefacts might lead to imprecise observational estimates of its evolution. Given that the projected O trends are rather small, the use of the highest-quality O measurements is mandatory.
Appendix AMonitoring of the ILS function from NO-cell measurements: non-clamped phase error retrieval
Figure shows the ILS time series evaluated from the NO cell measurements assuming a non-clamped PE retrieval (i.e. free PE) at ZPD, as described by . Figure furthermore includes the comparison between the cell-derived PE time series at an OPD of 133 cm for the clamped and non-clamped approaches, as well as the retrieved PE values from the O atmospheric retrievals using the set-up 5H, which superimposes a simultaneous fit of a PE offset to the retrieved path-dependent cell-derived PE. A remarkable discrepancy between the approaches is clearly observed, especially for the IFS 120M instrument. While the clamped PE retrievals from cell and solar spectra are rather consistent, the cell-derived non-clamped retrieval provides larger and more erratic PE corrections, which are in turn reproduced by the retrieved PE values from atmospheric spectra (set-up 5H). This fact seems to point to the conclusion that NO cell measurements may not contain enough information to allow for a precise PE retrieval at ZPD. In addition, Table summarizes the comparison of the theoretical performance of the O retrieval strategies using both cell-derived ILS estimates in terms of the obtained DOFS and fitting residuals. The clamped ILS approach consistently offers the best performance independent of the instrument's status or period: it provides a better vertical sensitivity and a superior interpretation of the measured atmospheric spectra. These results further corroborate the idea that a clamped PE retrieval at ZPD when evaluating the NO cell measurements is a superior choice to characterize the instrumental performance of the NDACC FTIR spectrometers.
Figure A1
(a) Time series of the normalized MEA and PE values at four OPDs (38, 85, 133, and 180 ) between 1999 and 2018. Data points represent individual NO cell measurements, and solid lines depict the smoothed MEA and PE curves. The solid grey arrows indicate punctual interventions made regarding the IZO FTIR instruments: changes in the field stop instruments between 1999 and 2004, the switch from the IFS 120M to the IFS 120/5HR in January 2005, optical re-alignments in June 2008 and February 2013, and internal laser replacements in August 2016 and June 2017. (b) Time series of the PE values at an OPD of 133 cm (including the non-clamped and clamped PE retrieval at ZPD) and the time series of the retrieved PE values from the O retrievals using the 5H set-up (including the non-clamped and clamped PE retrievals as a priori information). Note that the retrieved PE from the 5H set-ups is constant throughout the whole OPD range.
[Figure omitted. See PDF]
Table A1The same as Table but for the O retrievals using the cell-derived ILS time series considering non-clamped PE retrievals at ZPD (Fig. ) (so-called 5). For a better comparison, the DOFS and fitting residuals obtained with the clamped cell-derived ILS time series have also been included (5A, 5AT, 5H, and 5HT).
DOFS | Residuals () | |||||||
---|---|---|---|---|---|---|---|---|
1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | 1999–2004 | 2005–2008 | 2008–2018 | 1999–2018 | |
Set-up | , | , | , | , | , | , | , | , |
5A | 4.19, 0.29 | 4.46, 0.14 | 4.30, 0.11 | 4.31, 0.16 | 3.95, 2.04 | 2.79, 0.92 | 3.29, 0.59 | 3.28, 0.97 |
5H | 4.35, 0.30 | 4.58, 0.15 | 4.51, 0.12 | 4.51, 0.17 | 3.34, 1.94 | 2.56, 0.88 | 2.72, 0.53 | 2.73, 0.90 |
5A | 4.29, 0.29 | 4.56, 0.15 | 4.52, 0.13 | 4.51, 0.18 | 3.51, 1.96 | 2.57, 0.86 | 2.70, 0.56 | 2.72, 0.93 |
5H | 4.35, 0.30 | 4.58, 0.15 | 4.52, 0.13 | 4.52, 0.17 | 3.37, 1.93 | 2.54, 0.85 | 2.70, 0.55 | 2.71, 0.90 |
5AT | 3.99, 0.34 | 4.32, 0.19 | 4.14, 0.13 | 4.15, 0.19 | 3.90, 2.04 | 2.78, 0.91 | 3.28, 0.58 | 3.27, 0.97 |
5HT | 4.13, 0.35 | 4.43, 0.21 | 4.35, 0.15 | 4.35, 0.21 | 3.27, 1.94 | 2.54, 0.87 | 2.70, 0.53 | 2.70, 0.89 |
5AT | 4.09, 0.34 | 4.42, 0.20 | 4.35, 0.15 | 4.35, 0.21 | 3.44, 1.96 | 2.55, 0.86 | 2.68, 0.56 | 2.70, 0.93 |
5HT | 4.14, 0.36 | 4.44, 0.20 | 4.36, 0.15 | 4.36, 0.21 | 3.28, 1.93 | 2.51, 0.85 | 2.67, 0.55 | 2.68, 0.90 |
Monitoring of the ILS function from CO and HF retrievals
An alternative approach to ensure the independence of the ILS and O retrievals in cases where ILS measurements are not available might be to retrieve the ILS information from atmospheric trace gas retrievals with well-known vertical distribution. For this purpose, the ILS parameters have also been evaluated from the measured absorption lines of very stable tropospheric and stratospheric gases, such as and HF, respectively. For the approach, the four isolated CO lines between 960–970 cm considered in the simultaneous temperature fit have been used, while the HF strategy retrieves the ILS parameters jointly with the HF concentrations from two micro-windows in the 4000.90–4001.05 and 4038.85–4039.08 cm spectral ranges. Figure displays the obtained ILS time series from the two approaches.
Figure B1
Time series of the MEA as a function of the OPD from 1999 to 2018 evaluated from simultaneous ILS retrievals (a) from and (c) from HF absorbing lines. Panels (b) and (d) are the same as (a) and (c) but for the MEA at an OPD of 133 cm and including the phase error (PE), which is constant throughout the whole OPD range. Note that the coloured scale for the MEA is different between (a) and (c).
[Figure omitted. See PDF]
Both strategies are found to produce unrealistic ILS estimates (especially the HF approach) and a strong artificial annual cycle in the retrieved MEA and PE values. The tropospheric CO absorbing lines seem to be too broad to allow us to accurately estimate the ILS, while the HF strategy is strongly affected by the mix between the ILS signatures and dynamical signals (seasonal shift of the tropopause height and stratosphere–troposphere exchange events). In light of these results, the most promising approach to characterize the instrumental performance might be to retrieve the ILS parameters jointly with O concentrations in the absence of independent ILS measurements (provided the ILS estimates are averaged annually to reduce seasonal effects).
Data availability
The FTIR and Brewer data are available by request to the corresponding authors, and the ozone sondes are available at the NDACC archive (
Author contributions
The manuscript was prepared by OEG and ES with contributions from all co-authors. FH developed the LINEFIT and PROFFIT retrieval codes. FH, MS, and TB discussed the results and participated in the retrieval analysis. OGE, MS, and ES have performed the routine FTIR measurements and regular cell measurements since 1999 and performed the maintenance and quality control of the FTIR instruments. AR, SFLL, and VC managed the Brewer spectrometers and elaborated on the ozone observations. Finally, CT and NP are in charge of the ozone sonde programme at IZO.
Competing interests
At least one of the (co-)authors is a member of the editorial board of Atmospheric Measurement Techniques. The peer-review process was guided by an independent editor, and the authors also have no other competing interests to declare.
Disclaimer
Publisher's note: Copernicus Publications remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Acknowledgements
The Izaña FTIR station has been supported by the German Bundesministerium für Wirtschaft und Energie (BMWi), the Helmholtz Association, the Karlsruhe Institute of Technology (KIT) and the State Meteorological Agency of Spain. The authors also acknowledge the support of the KIT and Izaña Observatory staff for maintaining and quality assuring all the instrumentation.
Financial support
The Izaña FTIR station has been supported by the German Bundesministerium für Wirtschaft und Energie (BMWi) via the DLR under grant no. 50EE1711A and by the Helmholtz Association via the research programme ATMO. In addition, this research was funded by the European Research Council under FP7/(2007–2013)/ERC grant agreement no. 256961 (project MUSICA), by the Deutsche Forschungsgemeinschaft for the project MOTIV (Geschäftszeichen SCHN 1126/2-1), the Ministerio de Economía y Competitividad of Spain through the projects CGL2012-37505 (project NOVIA) and CGL2016-80688-P (project INMENSE), and EUMETSAT under its fellowship programme (project VALIASI).
Review statement
This paper was edited by Mark Weber and reviewed by two anonymous referees.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2022. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Retrieving high-precision concentrations of atmospheric trace gases from FTIR (Fourier transform infrared) spectrometry requires a precise knowledge of the instrumental performance. In this context, this paper examines the impact on the ozone (O
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details





1 Izaña Atmospheric Research Centre (IARC), State Meteorological Agency of Spain (AEMet), Santa Cruz de Tenerife, Spain
2 Izaña Atmospheric Research Centre (IARC), State Meteorological Agency of Spain (AEMet), Santa Cruz de Tenerife, Spain; now at: Employment Observatory of the Canary Islands (OBECAN), Santa Cruz de Tenerife, Spain
3 Karlsruhe Institute of Technology (KIT), Institute of Meteorology and Climate Research (IMK-ASF), Karlsruhe, Germany
4 Izaña Atmospheric Research Centre (IARC), State Meteorological Agency of Spain (AEMet), Santa Cruz de Tenerife, Spain; now at: TRAGSATEC, Madrid, Spain